K183007 · Phadia AB · LKP · Dec 24, 2018 · Immunology
Device Facts
Record ID
K183007
Device Name
EliA SmDP Immunoassay
Applicant
Phadia AB
Product Code
LKP · Immunology
Decision Date
Dec 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instrument Phadia 2500/5000.
Device Story
EliA SmDP Immunoassay is an in vitro diagnostic test for detecting IgG antibodies to Sm protein in human serum or EDTA-plasma. The device utilizes polystyrene wells coated with synthetic SmD3 peptide. Patient samples are incubated in these wells; if anti-Sm antibodies are present, they bind to the antigen. After washing, enzyme-labeled anti-IgG conjugate is added, followed by a development solution. The resulting fluorescence is measured to quantify antibody levels. The assay is performed on the Phadia 2500/5000 automated instrument system. The system uses disposable pipette tips to prevent carry-over and features an integrated barcode reader for automated reagent tracking. Results are calculated by comparing sample fluorescence to a calibration curve. The device is intended for use in clinical laboratories by trained personnel to assist in the diagnosis of systemic lupus erythematosus (SLE).
Clinical Evidence
No new clinical diagnostic studies were performed. Substantial equivalence was established via analytical performance testing (precision, linearity, LoD/LoQ) and a method comparison study. The instrument comparison study evaluated 105 samples (75 positive, 15 negative, 15 equivocal) across three Phadia 2500/5000 instruments versus the predicate. Passing-Bablok regression showed high correlation (R > 0.99). PPA and NPA were calculated, with TPA ranging from 98.1% to 100.0% (equivocal as positive). Reference range established using 400 healthy subjects.
Technological Characteristics
Solid-phase fluoroimmunoassay. Reagents: SmD3 peptide-coated wells, β-Galactosidase labeled anti-IgG conjugate, 4-Methylumbelliferyl-β-D-galactoside development solution. Instrument: Phadia 2500/5000 (automated, random access). Connectivity: Networked via IDM. Software: Onboard ISW. Sterilization: Not applicable (reagents are ready-to-use). Materials: High-density polyethylene (dilution wells), polyethylene (pipette tips).
Indications for Use
Indicated for the in vitro semi-quantitative measurement of IgG antibodies to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE).
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K183007
B. Purpose for Submission:
Addition of previously cleared assay on a new instrument platform (Phadia 2500/5000)
C. Measurand:
IgG autoantibodies specific to Sm protein (Sm) D component
D. Type of Test:
Automated semi-quantitative solid-phase fluoroimmunoassay
E. Applicant:
Phadia AB
F. Proprietary and Established Name:
EliA SmD⁸ Immunoassay
G. Regulatory Information:
1. Regulatory Section:
21 CFR 866.5100, Antinuclear antibody immunological test system
2. Classification:
Class II
3. Product Codes:
LKP: Anti-Sm Antibody, Antigen and Control
{1}
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use:
EliA SmD$^{\text{P}}$ is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmD$^{\text{P}}$ uses the EliA IgG method on the instrument Phadia 2500/5000.
2. Indication for use:
Same as Intended Use
3. Special conditions for use statement:
Prescription use only
4. Special instrument requirements:
For use on the Phadia 2500 and Phadia 5000 instruments
I. Device Description:
EliA uses a modular reagent system. The assay-specific, method-specific and general reagents are packaged and purchased as separate units. The reagents on Phadia 2500 and Phadia 5000 are identical; they are only filled in different containers.
(1) EliA Assay-specific reagent:
- EliA SmD$^{\text{P}}$ Wells are coated with a synthetic SmD3 peptide – four carriers (12 wells each), ready to use;
(2) EliA Method-specific reagent:
- EliA IgG Calibrator Strips: Human IgG (0, 4, 10, 20, 100, 600 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide in five strips, six single-use vials per strip, 0.3 mL each, ready to use;
- EliA IgG Curve Control Strips: Human IgG (20 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide in five strips, six single-use vials per strip, 0.3 mL each, ready to use;
- EliA Sample Diluent: PBS containing BSA, detergent and 0.095% (w/v) sodium azide in six bottles, 48 mL each, ready to use; or six bottles, 400 mL each, ready to use;
2
{2}
- EliA IgG Conjugate 50 or 200: β-Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide in six wedge-shaped bottles, 5 mL each, ready to use; or six wedge-shaped bottles, 19 mL each, ready to use;
- EliA IgG Calibrator Well: Coated with mouse monoclonal antibodies in four carriers (12 wells each), ready to use
(3) EliA General reagents:
- Development Solution (0.01% 4-Methylumbelliferyl-β-D-galactoside, <0.0010% preservative), ready for use;
- Stop Solution (4% Sodium Carbonate), ready for use;
- Dilution Wells (high density polyethylene wells), 50 carriers, ready to use;
- Pipette Tips in Racks (polyethylene tips), 24 racks × 160 tips, ready for use;
- Washing Solution (information in separate Washing Solution package insert)
Phadia 2500 and Phadia 5000 are identical instruments except for sample throughput. The Phadia 2500 consists of one process module (two process lines), whereas Phadia 5000 consists of two process modules (2 × 2 process lines). Instrument operation is handled by onboard Instrument Software (ISW). Data output is administered by Information Data Manager (IDM). All steps of an assay are performed within a single process line. Thus, study protocols used for Phadia 2500 are also valid for Phadia 5000.
J. Substantial Equivalence Information:
1. Predicate device name:
EliA SmDᵖ on Phadia 250 instrument
2. Predicate 510(k) number:
K132631
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Predicate Device Phadia 250 | Test Device Phadia 2500/5000 |
| Intended Use/Indications for Use | EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instrument Phadia 250. | EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instrument Phadia 2500/5000. |
| Analytical technology | Immuno-fluorescence measurement | Same |
| Reagents | Common Phadia EliA assay components | Same |
| Result calculation software | Phadia Information Data Manager (IDM) | Same |
| Assay cutoff / results interpretation | Negative: < 7 EliA U/mL Equivocal: 7–10 EliA U/mL Positive: > 10 EliA U/mL | Same |
| Traceability | IRP 67/86 of Human Serum IgA, G, M from World Health Organization | Same |
| Sample volume | 90 μL (20 μL of non-diluted sample) | Same |
| Incubation temperature | 37°C | Same |
| Conjugate volume | 90 μL | Same |
| Development solution volume | 90 μL | Same |
| Stop solution volume | 200 μL | Same |
| Assay set-up | Random access | Same |
| Reagent packaging size | Various/Common | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Predicate Device
Phadia 250 | Test Device
Phadia 2500/5000 |
| Onboard storage of reagents | Yes | Same |
| Time to 1st result | ~2 hours | Same |
| Differences | | |
| --- | --- | --- |
| Item | Predicate Device
Phadia 250 | Test Device
Phadia 2500/5000 |
| Sample matrix | Human serum or plasma
(heparin, EDTA, citrate) | Human serum or plasma (Li-heparin, EDTA) |
| Instrument Platform | Phadia 250 (ImmunoCAP 250) | Phadia 2500 / 5000 |
| Measuring Range | 0.8 – 480 U/mL | 1.6 – 480 U/mL |
| Daily throughput | ~250 tests | Phadia 2500: ~2500 tests
Phadia 5000: ~5000 tests |
| Sample dilution | Phadia 250 uses a steel pipette to dilute the samples in Dilution Plates | Phadia 2500/5000 uses disposable Pipette Tips in Racks for pipetting samples in Dilution Well |
| Risk for carry-over | The warning “DO NOT REUSE” in the Phadia 250 DFU for EliA Conjugates is due to the fact that a low risk of conjugate contamination by carry-over from samples was identified. In order to reduce the risk, the single use statement for the conjugate was included in the Phadia 250 DFU. | When running EliA tests on the Phadia 2500/5000 instruments, there is no need for this warning statement because these instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, and carry-over from samples to conjugate is impossible. |
| Loading of EliA carriers | The EliA carriers are loaded manually on the Loading Tray from where they can be processed directly or transferred to the cooled storage compartment. | The Phadia 2500/5000 instruments do not have such a Loading Tray. The EliA carriers are loaded into racks which are directly transferred to the cooled storage compartment. |
| Barcode reader | The Phadia 250 instrument has a built-in barcode reader at the front of the instrument, but the operator needs to scan the barcodes manually by showing the reagents to the barcode reader. Alternatively, the operator can also enter the characters below the barcode manually. | The Phadia 2500/5000 instruments dispose of a built-in barcode reader, and the reagents are on a moving belt which conveys them past the barcode reader. The lot-specific information will be read automatically by the instrument during loading. |
{5}
| Differences | | |
| --- | --- | --- |
| Item | Predicate Device
Phadia 250 | Test Device
Phadia 2500/5000 |
| Process time /
Time to patient
result | Phadia 250 needs one minute to
process one well and provides the
results at a 1-minute interval. | Phadia 2500/5000 processes two
wells in parallel in 48 seconds and
provides the results at a 24-second
interval. |
K. Standard/Guidance Documents Referenced:
- CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; September 2014
- CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; April 2003
- CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
L. Test Principle:
EliA tests are fluorescence immunoassays for the detection and measurement of human antibodies based on EliA solid-phase components, which contain specific antigens for the antibodies to be measured.
The EliA wells are coated with a synthetic $\mathrm{SmD}_3$ peptide. If present in the patient's specimen, antibodies to $\mathrm{SmD}_3$ peptide bind to the specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away, and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the higher the amount of antibody bound and detected in the sample tested. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.
M. Performance Characteristics:
1. Analytical performance:
All results presented below met the manufacturer’s pre-determined acceptance criteria for all analytical performance studies.
{6}
a. Precision/Reproducibility:
To determine the precision of the assay, the variability was assessed in a study with a total of 21 runs (three instruments total and seven runs per instrument). The study was performed with a single run/day over a period of seven days. Each sample was tested in four replicates/run giving in total 84 replicates per sample (3 instruments × 7 runs × 4 replicates = 84). The data were calculated against the calibration curve from Day 1. Five native patient serum samples were tested: one negative, two in the equivocal range, and two positive specimens. All samples were tested on all instruments. The results are summarized in the tables below:
| Sample | Mean (EliA U/mL) | Within-Run | | Between-Run | | Between-instrument | | Total Imprecision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 2.2 | 0.2 | 8.7 | 0.2 | 11.0 | 0.4 | 20.5 | 0.5 | 24.8 |
| 2 | 7.1 | 0.3 | 3.6 | 0.2 | 2.8 | 0.3 | 4.7 | 0.5 | 6.5 |
| 3 | 10.0 | 0.4 | 3.7 | 0.2 | 1.9 | 0.8 | 7.7 | 0.9 | 8.8 |
| 4 | 146.6 | 5.1 | 3.5 | 3.7 | 2.5 | 7.2 | 4.9 | 9.5 | 6.5 |
| 5 | 408.1 | 15.5 | 3.8 | 9.8 | 2.4 | 31.1 | 7.6 | 36.1 | 8.8 |
b. Linearity/assay reportable range:
Four patient serum samples were diluted in EliA Sample Diluent and tested in triplicate with one lot of EliA SmD⁰ immunoassay and one set of system reagents on Phadia 2500/5000. The results of the dilutions were compared with their expected values. The ratio observed/expected (O/E) was calculated. Mean observed value was used in the calculation. The results of the regression analysis are summarized in the tables below:
| Sample | Dilution Range (EliA U/mL) | Slope (95% CI) | Intercept (95% CI) | R² | %CV Range |
| --- | --- | --- | --- | --- | --- |
| 1 | 6.5–285.5 | 1.04 (0.95–1.12) | 4.21 (-5.70–4.21) | 0.99 | 2.7–15.4 |
| 2 | 8.7–392.8 | 0.98 (0.94–1.02) | -0.56 (-7.29 to -0.56) | 1.00 | 0.8–10.3 |
| 3 | 27.1–537.4 | 1.01 (0.97–1.05) | -10.58 (-21.14 to -0.03) | 1.00 | 2.9–8.6 |
| 4 | 0.9–11.4 | 1.01 (0.99–1.02) | -0.09 (-0.19–0.01) | 1.00 | 1.6–7.2 |
{7}
The reportable range (Limit of Detection to upper limit) for EliA SmD$^{\mathrm{P}}$ is from 0.8 to 480 EliA U/mL. The measuring range (Limit of Quantitation to upper limit) is from 1.6 to 480 EliA U/mL.
The following statements are included in the package inserts: “Please note that concentration values between LoD and LoQ may show a higher uncertainty.” and “Please note that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range.”
The hook effect was previously reviewed in K132631.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Same as in the predicate devices. Refer to K132631.
d. Detection limit:
The limit of blank (LoB) and limit of detection (LoD) studies were performed on a Phadia 2500 instrument. LoB was measured in six assay runs with each run including a blank sample in 12 replicates, giving a total of 72 replicates. LoD was measured in six assay runs by testing five low level serum samples in 12 replicates per sample in a run, giving a total of 360 replicates. The limit of quantitation (LoQ) was based on 360 determinations and a precision target of 20%. The results are summarized in the table below:
| EliA SmD$^{\mathrm{P}}$ Assau (EliA U/mL) | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| Phadia 2500/5000 | 0.3 | 0.8 | 1.6 |
e. Analytical specificity:
i) Interference:
Interference by endogenous substances was previously reviewed in K132631.
ii) Cross-reactivity:
Cross-reactivity was previously reviewed in K132631.
iii) Carry-over:
The Phadia 2500/5000 instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, therefore carry-over from samples to conjugate was not evaluated.
{8}
f. Assay cut-off:
The assay cut-offs are the same as in the predicate device and are summarized in the tables below:
| < 7 EliA U/mL | Negative |
| --- | --- |
| 7–10 EliA U/mL | Equivocal |
| > 10 EliA U/mL | Positive |
2. Comparison studies:
a. Method comparison with predicate device:
See 2c. Instrument Comparison below.
b. Matrix comparison:
Matrix comparison between serum and plasma was reviewed in K132631.
c. Instrument comparison:
The purpose of this study was to evaluate conformance and show comparability of EliA SmD$^{\mathrm{P}}$ on the Phadia 250 instrument versus the Phadia 2500/5000 instrument. A total of 105 samples (75 positive, 15 negative and 15 equivocal) were run in singlicate on one Phadia 250 instrument and three Phadia 2500/5000 instruments for comparison. Only samples inside the measuring range were included in the calculations. Results were analyzed by Passing-Bablok regression and summarized below:
| Instrument | Range Tested (EliA U/mL) | Intercept (95% CI) | Slope (95% CI) | *R |
| --- | --- | --- | --- | --- |
| PH2500/5000 A | 1.3–473.7 | 0.07
(−0.38–0.53) | 1.00
(0.98–1.04) | 0.995 |
| PH2500/5000 B | 1.0–467.6 | 0.06
(−0.39–0.24) | 1.03
(1.01–1.05) | 0.995 |
| PH2500/5000 C | 0.9–479.9 | −0.34
(−0.70 to −0.08) | 1.06
(1.03–1.09) | 0.997 |
| *Pearson correlation coefficient | | | | |
{9}
Positive percent agreement (PPA), negative percent agreement (NPA), and total percent agreement (TPA) were evaluated for each instrument pair comparison with equivocal results considered positive in the table below:
| | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
| --- | --- | --- | --- |
| PPA
(95% CI) | 98.9%
(94.0%-100.0%) | 98.9%
(94.0%-100.0%) | 100.0%
(96.0%-100.0%) |
| NPA
(95% CI) | 93.3%
(68.1%-99.8%) | 100.0%
(73.5%-100.0%) | 100.0%
(78.2%-100.0%) |
| TPA
(95% CI) | 98.1%
(93.3%-99.8%) | 99.0%
(94.7%-100.0%) | 100.0%
(96.5%-100.0%) |
Results when equivocal results are considered negative are presented in the table below:
| | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
| --- | --- | --- | --- |
| PPA
(95% CI) | 98.7%
(92.8%-100.0%) | 100.0%
(95.2%-100.0%) | 100.0%
(95.2%-100.0%) |
| NPA
(95% CI) | 86.7%
(69.3%-96.2%) | 88.9%
(70.8%-97.6%) | 93.3%
(77.9%-99.2%) |
| TPA
(95% CI) | 95.2%
(89.2%-98.4%) | 97.1%
(91.6%-99.4%) | 98.1%
(93.3%-99.8%) |
No sample switched from negative to positive. One sample in the negative range was tested equivocal. One sample in the equivocal range was tested negative and four samples in the equivocal range were tested positive. One sample in the positive range was tested equivocal.
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
Clinical performance values were reviewed in K132631.
b. Other clinical supportive data (when a. is not applicable):
Not Applicable
4. Clinical cut-off:
Same as assay cut-off.
{10}
11
5. Expected values/Reference range:
The frequency distribution for $\mathrm{SmD}^{\mathrm{P}}$ antibodies was investigated in a group of apparently healthy subjects (N=400) equally distributed by age and gender, using sera from a Caucasian population obtained from a blood bank. None of test samples was within the equivocal or the positive range in the three EliA assays. The test results are given in the table below and all measurements are shown in EliA U/mL:
| Test | N | Mean | Median | Range | 95^{th} Percentile | 99^{th} Percentile |
| --- | --- | --- | --- | --- | --- | --- |
| EliA SmD^{P} on Phadia 2500/5000 | 400 | 1.2 | 1.1 | 0.0–5.4 | 2.2 | 4.2 |
N. Proposed Labelling:
The labelling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.