VARELISA SM ANTIBODIES, MODEL 18296

K042629 · Sweden Diagnostics (Germany) GmbH · LKP · Dec 8, 2004 · Immunology

Device Facts

Record IDK042629
Device NameVARELISA SM ANTIBODIES, MODEL 18296
ApplicantSweden Diagnostics (Germany) GmbH
Product CodeLKP · Immunology
Decision DateDec 8, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2

Intended Use

The Varelisa Sm Antibodies EIA kit is designed for the semiquantitative and qualitative determination of SmD antibodies in serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE). The device is for prescription use only.

Device Story

Varelisa Sm Antibodies is an enzyme-linked immunosorbent assay (ELISA) used in clinical laboratories to detect SmD antibodies in patient serum or plasma. The device utilizes microtiter plates coated with synthetic SmD peptide as the solid phase. Patient samples are added, allowing specific anti-SmD antibodies to bind to the immobilized antigen. A rabbit anti-human IgG horseradish peroxidase (HRP) conjugate is introduced, followed by a TMB substrate. The resulting color intensity, measured via a microplate reader at 450 nm and 620 nm, is proportional to the concentration of SmD antibodies in the sample. Results aid clinicians in the diagnosis of systemic lupus erythematosus (SLE).

Clinical Evidence

Clinical study evaluated 198 samples (100 SLE, 98 disease controls). Clinical sensitivity was 23.0% and specificity was 93.9%. Analytical performance included precision (within-run %CV 2.6-4.0%, between-run 1.8-3.9%), linearity (met specifications through 1:8 dilution), and analytical sensitivity (0.03 U/mL). Interference studies showed deviations within ±20% for bilirubin, hemoglobin, chyle, and rheumatoid factor. Heparin plasma showed potential interference (-53.7% to 29.9%).

Technological Characteristics

Indirect noncompetitive ELISA; microtiter plate solid phase coated with synthetic SmD peptide; rabbit anti-human IgG HRP conjugate; TMB substrate; requires microplate reader (450/620 nm).

Indications for Use

Indicated for the semiquantitative and qualitative determination of SmD antibodies in human serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE).

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(K) SUMMARY OF SAFETY AND 9. EFFECTIVENESS This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92. | Assigned 510(k) Number: | K042629 | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date of Summary Preparation: | September 21, 2004 | | Manufacturer: | Sweden Diagnostics (Germany) GmbH<br>Munzinger Strasse 7<br>D-79111 Freiburg, Germany | | Company Contact Person: | Sabine Klugbauer<br>Specialist, Regulatory Affairs<br>Sweden Diagnostics (Germany) GmbH<br>Munzinger Strasse 7<br>D-79111 Freiburg, Germany<br>+49-761-47805-419(Phone)<br>+49-761-47805-335 (Fax) | | Device Name: | Varelisa® Sm Antibodies | | Common Name: | Antinuclear antibody<br>immunological test system | # Classification | Product Name | Product Code | Class | CFR | |-------------------------|--------------|-------|----------| | Varelisa® Sm Antibodies | LKP | II | 866.5100 | ## Substantial Equivalence to Varelisa® Sm Antibodies (510(k) number: K000312) {1}------------------------------------------------ #### Intended Use Statement of the New Device #### Intended use/Indication for use The Varelisa Sm Antibodies EIA kit is designed for the semiquantitative and qualitative determination of SmD antibodies in serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE). Special condition for use statement The device is for prescription use only. Special instrument requirements A microplate reader capable of measuring OD at 450 mm and 620 nm is required. # General Description of the New Device The new device is an enzyme-linked immunosorbent assay (ELISA) using microtiter plates as the solid phase. The plate wells are coated with a synthetic SmD peptide as antigen, which allow anti-SmD antibodies to react with the immobilized antigen (sample). The conjugate is rabbit anti-human IgG horseradish peroxidase (HRP), which uses 3, 3'5, 5' tetramethylbenzidine dihydrochloride (TMB) as substrate. The kit contains a set of six calibrators, positive and negative controls. The kit also contains sample diluent, wash buffer concentrate and stop solution. # Test Principle of the New Device The new device is an indirect noncompetitive enzyme immunoassay for the semiquantitative and qualitative determination of SmD antibodies in human serum or plasma. The wells of a microplate are coated with a synthetic SmD peptide. Antibodies specific for SmD present in the patient sample bind to the antigen. In a second step the enzyme labeled second antibody (conjugate) binds to the antigen-antibody complex which leads to the formation of an enzyme labeled conjugate-antibody-antigen complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen is a function of the amount of conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample. #### Device Comparison The new device is developed as successor of the predicate device. Both assays share the same assay principle and indications for use. They are indirect noncompetitive enzyme immunoassays for the semiquantitative and qualitative determination of IgG antibodies against Sm in serum and plasma. Both assays recommend the same sample dilutions and use identical reagents (including the conjugate). The evaluation of the assays is identical. In accordance to the relevant scientific literature both assays state in the Intended Use, that the measuring of antibodies against Sm provides aid in the diagnosis of systemic lupus erythematosus (SLE). {2}------------------------------------------------ The new device is based on a synthetic peptide of the human SmD protein while the predicate device contains a native Sm protein isolated from calf thymus. Minor differences between the two devices pertain to the packaging size of the reagents and the leaving out of the prewashing step of the antigen strips. #### Laboratory equivalence The comparability of predicate device and new device is supported by a data set including - · results obtained within a comparison study analyzing positive, equivocal and negative sera. - · results obtained for clinically defined sera and for international reference sera. - results obtained for samples from apparently healthy subjects (normal population). In summary, all available data support that the new device is substantially equivalent to the predicate device and that the new device performs according to state-of-the-art expectations. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes that resemble birds in flight or flowing lines. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 DEC - 8 2004 Sabine Klugbauer, Ph.D. Specialist, Regulatory Affairs Sweden Diagnostics (Germany) GmbH Munzinger Strasse 7 D-79111 Freiburg Germany Re: k042629 Trade/Device Name: Varelisa® Sm Antibodies Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear Antibodies Immunological Test Systems Regulatory Class: Class II Product Code: LKP Dated: September 21, 2004 Received: September 27, 2004 Dear Dr. Klugbauer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Sabine Klugbauer, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html Sincerely yours. Robert L. Becker Jr. Robert L. Becker, Jr., M.D., Ph. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number: K042629 Device Name: Varelisa® Sm Antibodies ## Indications For Use: The Varelisa Sm Antibodies EIA kit is designed for the semiquantitative and qualitative determination of SmD antibodies in serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE). Maria Chan Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) *K042629* Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...