K042680 · Corgenix, Inc. · DHN · Nov 19, 2004 · Immunology
Device Facts
Record ID
K042680
Device Name
RHIGENE HEP-ANA TEST SYSTEM
Applicant
Corgenix, Inc.
Product Code
DHN · Immunology
Decision Date
Nov 19, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Intended Use
The RhiGene HEP-ANA Test System is an indirect fluorescent antibody assay using Hep-2 cells as a substrate for the qualitative and/or semi-quantitative determination of antinuclear antibodies in human serum. The RhiGene HEP-ANA Test System is intended for use as an aid in the diagnosis of certain autoimmune diseases.
Device Story
The HEP-ANA Test System is an indirect fluorescent antibody (IFA) assay for detecting antinuclear antibodies (ANA) in human serum. The device uses HEp-2 tissue culture cells as a substrate on slides. Patient serum is incubated on the substrate, allowing ANA to bind to the cells. After washing, FITC-labeled anti-human immunoglobulin is added to form fluorescent complexes. Slides are examined via fluorescence microscopy. The assay requires 40 minutes of incubation at room temperature. The test is performed by laboratory personnel in hospital or reference labs. Results assist clinicians in diagnosing autoimmune diseases by identifying the presence of ANA.
Clinical Evidence
In-house studies compared the HEP-ANA Test System to the predicate device. Clinical specificity was 82% for the subject device and 78% for the predicate in a healthy donor population. Sensitivity was 75% for both devices in an autoimmune disease population. Performance characteristics are comparable and consistent with published literature.
Indicated for qualitative/semi-quantitative determination of antinuclear antibodies (ANA) in human serum as an aid in diagnosing autoimmune diseases in clinical laboratory settings.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
K120889 — AESKUSLIDES ANA HEP-2 · Aesku Diagnostics · Oct 15, 2012
K070763 — EUROIMMUN ANA IFA: HEP-20-10 · Euroimmun Us, LLC · May 22, 2007
K081104 — AESKULISA ANA HEP-2, REF 30-7115US · Aesku Diagnostics · May 2, 2008
K993294 — DIAMEDIX IS-ANA ELISA SCREEN TEST SYSTEM · Diamedix Corp. · Oct 25, 1999
K131791 — IFA 40: HEP-20-10 · Euroimmun US · Feb 26, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
**RhiGene** an MBL Compa
RhiGene Inc 55 State Street Suite 104 es Plaines, IL 60016 TEL : 847-375-9091 800-447-0781 FAX : 847-375-9093
K042680
NOV 19 2004
## SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS HEP-ANAN Test System September 10, 2004
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The HEP-ANA Test system is compared to a legally marketed predicate device and a substantial equivalence claim made. The predicate device is RhiGene Titer-Fluor ANA Test System (manufactured by BION ENTERPRISES, Des Plaines, IL, K872845), currently marketed by RhiGene. Inc. Des Plaines, IL,
The RhiGene HEP-ANA Test System is an indirect fluorescent antibody assay utilizing HEp-2 tissue culture cells as a substrate, similar to the predicate device. Diluted serum samples are incubated on substrate slides coated with HEP-2 (human epithelial) cells. Incubation allows the anti-nuclear antibodies (ANA) present in the samples to react with the antigen. After the removal of unbound serum proteins by washing, antibodies specific for human immunoglobulins, labeled with fluorescein isothiocyanate (FITC), are added forming complexes with the nuclear bound antibodies. Following another washing step, coverslips are mounted and then the slides are examined with a fluorescence microscope. The total incubation time (at room temperature in a moist chamber) of the assay is 40 minutes.
The intended use of the device is a qualitative and/or semi-quantitative indirect fluorescent antibody assay for the determination of antinuclear antibodies in human serum. The HEP-ANA Test System is intended for in vitro diagnostic use as an aid in the determination of certain autoimmune diseases.
Performance indicates that HEP-ANA Test system and the RhiGene Titer-Fluor ANA Test System are equivalent. In-house studies indicate a clinical specificity of 82% and 78% for ANA in a healthy donor serum population respectively. Additional studies resulted a sensitivity of 75% with an autoimmune diseases population on both assav. In general, the performance characteristics are comparable between the two methods. These results are also in compliance with those in published literature for ANA detection. The clinical studies performed demonstrate that the HEP-ANA test system is safe and effective.
Yuush Role
9/15/2004
Date
Yusuke Kobe Vice President, Sales and Marketing Department
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure embracing or protecting another, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the emblem. The emblem is black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Corgenix, Inc. c/o Mr. Yusuke Kobe Vice President Rhigene. Inc. 455 State Street Des Plaines, IL 60016
## NOV 1 9 2004
一
Re: k042680
Trade/Device Name: RhiGene Hep-ANA Test System Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear Antibody, Immunological Test System Regulatory Class: Class II Product Code: DHN Dated: September 14, 2004 Received: September 29, 2004
Dear Mr Kobe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 -- Mr. Yusuke Kobe
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Robert Zeday
Robert L. Becker, Jr., M.D. PhD Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known); K 042680
RhiGene HEP-ANA Test Syotem Device Namé:
Indications For Use:
The HEP-ANA Test System is an indirect fluorescent antibody assay utilizing HEp-2 tissue culture cells as a substrate for the qualitative and/or semi-quantitative determination of antinuclear antibodies in human serum. The HEP-ANA Test System is intended for use as an aid in the diagnosis of certain autoimument diseases.
The HEP-ANA Test System is intended to be used by clinical (hospital and reference) laboratories.
Prescription Use _ X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mana Chan
**Division Sign-Off**
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of ________
510(k) K042680
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.