AESKULISA ANA HEP-2, REF 30-7115US

K081104 · Aesku Diagnostics · LKJ · May 2, 2008 · Immunology

Device Facts

Record IDK081104
Device NameAESKULISA ANA HEP-2, REF 30-7115US
ApplicantAesku Diagnostics
Product CodeLKJ · Immunology
Decision DateMay 2, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

AESKULISA ANA-HEp2 is a solid phase enzyme immunoassay for the combined qualitative detection of IgG artibodies against HEp2 cells in human serum. Each well is coated with lysed HEp2 cells and spectic antigens. The test collectively detects, in one well, total ANAs against double stranded DNA (dsDNA), histones, SS-A (Ro), SS-B (La), Sm, snRNP/Sm, Scl-70, Jo-1 and centromeric antigens along with sera positive for HEp2 immunofluorescence test (IFT). The assay is a tool in the diagnosis of certain systemic theumatic diseases and should be used in conjunction with other serological tests and clinical findings.

Device Story

AESKULISA ANA-HEp2 is a solid-phase enzyme immunoassay (EIA) for detecting IgG antibodies against HEp2 cells and specific nuclear antigens in human serum. The device utilizes wells coated with lysed HEp2 cells and specific antigens to capture total antinuclear antibodies (ANAs). It is intended for use in clinical laboratories by trained personnel as an aid in diagnosing systemic rheumatic diseases. Results are interpreted in conjunction with other serological tests and clinical findings. The assay provides a qualitative result, assisting clinicians in identifying patients requiring further evaluation for autoimmune conditions.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Solid-phase enzyme immunoassay (EIA). Wells coated with lysed HEp2 cells and specific antigens (dsDNA, histones, SS-A, SS-B, Sm, snRNP/Sm, Scl-70, Jo-1, centromeric). Qualitative detection of IgG antibodies.

Indications for Use

Indicated for qualitative detection of IgG antibodies against HEp2 cells and specific antigens (dsDNA, histones, SS-A, SS-B, Sm, snRNP/Sm, Scl-70, Jo-1, centromeric) in human serum to aid in diagnosis of systemic rheumatic diseases. For prescription use only.

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Submission Summary (Full Text)

{0} Special 510(k): Device Modification OIVD Review Memorandum To: The File Re: k081104 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II device requiring 510(k). The submission was reviewed by a Third Party Review entity. The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. AESKU.DIAGNOSTICS AESKULISA ANA HEp-2, K040953 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). There are no changes in Intended Use or Indications for Use. (Section 4, page 2 of 6 and Section 8, page 4-18) 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for: a. Removal of mercury containing stabilizers (thimerosal) from reagents b. Change in incubation times from 30-15-15 minutes to 30-30-30 minutes 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and Similarities: | Feature | Modified device | | --- | --- | | Indicated use | No Change | | Fundamental Scientific Technology | No Change | | Coated antigen | No Change | | Shelf life | No Change | | Production Process | No Change | {1} | Feature | Modified device | | --- | --- | | Standardization | No Change | | Quality Control Criteria | No Change | | Cut-off Range | No Change | ## Differences: - The preservative thimerosal which was a non-reactive compound of very low concentration in some components of the product (Sample buffer 5x, Wash buffer 50x, Controls and Conjugate) was removed for environmental issues (Thimerosal contains 50% mercury). The other preservative content (e.g. NaN3) in the respective components still ensures the stability of the product. Thimerosal does not play any role in the reactivity, sensitivity and specificity of the test. The role as a preservative is to keep the test stable for the claimed shelf life (18 Months) and it is per definition and requirement a non-reactive compound. - The different incubation steps of the device, which were initially performed in 30 minutes, 15 minutes and 15 minutes, were changed. The modification is a prolongation of the 2nd and 3rd step from 15 minutes to 30 minutes. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis: The risk analysis method used to assess the impact of the modification was a Failure Mode and Effects Analysis (FMEA) – (Section 8, page 4-19) b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied: As a result of the risk analysis, real time stability testing for low, medium and high levels of ANA autoantibodies, over the 18 month shelf life and sera distributed over the range of the assay tested on predicate and modified device were tested. The test methods used are the same as those used in the original submission. {2} c) A declaration of conformity with design controls. The declaration of conformity included: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met (Section 8, page 4-20), and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review (Section 8, page 4-20). 6. A Truthful and Accurate Statement (Section 7, page 4-25), a 510(k) Statement (cover letter and Section 8, page 4-24) and the Indications for Use Enclosure (cover letter and Section 8, page 6-53) The labeling for this modified subject device was reviewed to verify that the intended use for the device was unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter provided the design control information as specified in The New 510(k) Paradigm and, on this basis, I recommend the device be determined substantially equivalent to the previously cleared device. This submission was reviewed by Intertek Testing Services NA, Inc. a certified Third Party Review entity. They submitted a signed Authorization to Submit letter from AESKU.DIAGNOSTICS and signed Third Party Certification statements.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...