IMMAGE IMMUNOCHEMISTRY SYSTEM TRANSFERRIN (TRF) REAGENT
K963427 · Beckman Instruments, Inc. · DDG · Oct 7, 1996 · Immunology
Device Facts
Record ID
K963427
Device Name
IMMAGE IMMUNOCHEMISTRY SYSTEM TRANSFERRIN (TRF) REAGENT
Applicant
Beckman Instruments, Inc.
Product Code
DDG · Immunology
Decision Date
Oct 7, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5880
Device Class
Class 2
Indications for Use
The IMMAGE Immunochemistry System Transferrin (TRF) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of transferrin concentrations in human serum samples by rate nephelometry.
Device Story
The IMMAGE Immunochemistry System Transferrin (TRF) Reagent is an in vitro diagnostic reagent used on the Beckman IMMAGE Immunochemistry System. It utilizes rate nephelometry to measure transferrin concentrations in human serum. The system processes samples using a reagent cartridge, evaporation caps, and a lot-specific bar code card. The assay involves an antibody-antigen reaction; the rate of light scattering is measured to determine analyte concentration. The device is intended for clinical laboratory use. Results are provided to healthcare providers to assist in clinical decision-making regarding patient iron metabolism and nutritional status.
Clinical Evidence
Bench testing only. Method comparison study between IMMAGE TRF and ARRAY system predicate showed a slope of 0.990, intercept of 0.692, and r=0.994. Within-run imprecision studies (n=80 per level) yielded %CVs of 3.1% (241 mg/dL), 3.9% (361 mg/dL), and 2.9% (653 mg/dL). Stability studies confirmed 24-month shelf-life and 14-day open-container/calibration stability.
Indicated for the quantitative determination of transferrin concentrations in human serum samples to aid in the diagnosis of transferrin-related conditions.
Regulatory Classification
Identification
A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
Predicate Devices
Beckman Immunochemistry Systems TRF Transferrin Reagent
Reference Devices
ARRAY® System
Related Devices
K971958 — SYNCHRON LX SYSTEMS TRANSFERRIN (TRFN) REAGENT · Beckman Instruments, Inc. · Aug 20, 1997
K991157 — N LATEX STFR REAGENT · Dade Behring, Inc. · Aug 2, 1999
K972840 — N ANTISERUM TO HUMAN TRANSFERRIN · Behring Diagnostics, Inc. · Oct 17, 1997
K053075 — N ANTISERA TO HUMAN TRANSFERRIN · Dade Behring, Inc. · Mar 16, 2006
Submission Summary (Full Text)
{0}
OCT 7 1996
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent
Summary of Safety & Effectiveness
K963427
# Summary of Safety & Effectiveness
## IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent
### 1.0 Submitted By:
Margie George
Project Manager, System Development
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-337
Brea, California 92822-8000
Telephone: (714) 961-3765
FAX: (714) 961-3759
### 2.0 Date Submitted:
30 August 1996
### 3.0 Device Name(s):
#### 3.1 Proprietary Names
IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent
#### 3.2 Classification Names
Transferrin immunological test system (21 CFR 866.5880)
### 4.0 Predicate Device(s):
Beckman Immunochemistry Systems TRF Transferrin Reagent
### 5.0 Description:
The IMMAGE Immunochemistry System Transferrin (TRF) Reagent is used in the determination of transferrin concentrations in human serum samples on Beckman's IMMAGE Immunochemistry System. The reagent kit contains one cartridge of reagent, evaporation caps and a lot specific bar code card. Calibrators and control materials are purchased separately.
file: trf510k.sse, 8/96
{1}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent
Summary of Safety & Effectiveness
## 6.0 Intended Use:
The IMMAGE Immunochemistry System Transferrin (TRF) reagent, in conjunction with Beckman Calibrator 1, is intended for use in the quantitative determination of transferrin concentrations in human serum samples by rate nephelometry.
## 7.0 Comparison to Predicate(s):
The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| SIMILARITIES | | |
| IMMAGE System (TRF) Reagent | Assay method - rate nephelometry | Same as Beckman's Immunochemistry System Transferrin TRF Reagent as run on the ARRAY® System |
| | Analytic Range 75 - 750 mg/dL
Extended Range 12.5 - 4,500 mg/dL | |
| | Sample and reagent ratios | |
| | Antibody | |
| | Shelf life of 24 months
(stored at 2 -8°C) | |
| DIFFERENCES | | |
| IMMAGE System (TRF) Reagent | Reaction Temperature | IMMAGE System TRF assays run at 37°C and the predicate runs at 26.7°C |
| | Reagent container | IMMAGE reagent is stored in a plastic cartridge where as the predicate is stored in glass. |
file: trf510k.sse, 8/96
{2}
Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Transferrin (TRF) Reagent
Summary of Safety & Effectiveness
## 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from Beckman's Immunochemistry System TRF Transferrin (on ARRAY) Reagent to the IMMAGE System TRF Reagent.
### Method Comparison Study Results
IMMAGE TRF Reagent vs. Beckman TRF Transferrin Reagent (on ARRAY)
| Analyte | Slope | Intercept | r | Predicate |
| --- | --- | --- | --- | --- |
| IMMAGE TRF | 0.990 | 0.692 | 0.994 | Beckman's TRF Transferrin Reagent on the ARRAY® Systems |
### Stability Study Results
| Reagent | Product Claim |
| --- | --- |
| IMMAGE TRF Reagent System | 24 month shelf-life
14 day open container stability
14 day calibration stability |
### Estimated Within-Run Imprecision
| MATERIAL | MEAN (mg/dL) | SD (mg/dL) | %CV | Number of Results |
| --- | --- | --- | --- | --- |
| TRF Reagent | | | | |
| Level 1 | 241 | 7.4 | 3.1 | 80 |
| Level 2 | 361 | 14.0 | 3.9 | 80 |
| Level 3 | 653 | 18.9 | 2.9 | 80 |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
file: trf510k.sse, 8/96
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.