N ANTISERA TO HUMAN TRANSFERRIN

K053075 · Dade Behring, Inc. · DDG · Mar 16, 2006 · Immunology

Device Facts

Record IDK053075
Device NameN ANTISERA TO HUMAN TRANSFERRIN
ApplicantDade Behring, Inc.
Product CodeDDG · Immunology
Decision DateMar 16, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5880
Device ClassClass 2

Indications for Use

In vitro diagnostic reagents for the quantitative determination of transferrin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometery on the BN™ Systems.

Device Story

N Antisera to Human Transferrin is an in vitro diagnostic reagent kit used on BN™ Systems. It detects transferrin in human serum, plasma, or urine samples. Principle of operation: immunonephelometry; proteins in sample form immune complexes with specific antibodies; complexes scatter light beam; scattered light intensity is proportional to protein concentration. Result is determined by comparing intensity against a standard of known concentration. Used in clinical laboratory settings by trained personnel. Output provides quantitative transferrin levels to assist clinicians in diagnosing malnutrition, inflammation, infection, and anemia.

Clinical Evidence

No clinical studies performed. Evidence consists of analytical performance data including matrix comparison studies and interference testing. Matrix comparison (N=40-99 per group) showed high correlation (R² 0.9851-0.9936) between serum and plasma samples. Interference testing confirmed no significant interference from bilirubin, hemoglobin, triglycerides, or heparin anticoagulants.

Technological Characteristics

Quantitative immunonephelometry assay. Reagents: polyclonal rabbit anti-human transferrin antiserum. Instrumentation: BN™ Systems (BNII, BN 100, BN Prospec). Detection principle: light scattering intensity proportional to protein concentration. Sample matrices: serum, heparinized plasma, EDTA plasma, urine.

Indications for Use

Indicated for quantitative determination of transferrin in human serum, heparinized/EDTA plasma, and urine to aid in diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders (e.g., iron deficiency anemia).

Regulatory Classification

Identification

A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

Predicate Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k053075 B. Purpose for Submission: Device modification. Addition of heparinized and EDTA plasmas as sample matrices for both human transferrin and haptoglobin (note: haptoglobin is Class II exempt) C. Measurand: Human transferring and human haptoglobin D. Type of Test: Quantitative immunonephelometry E. Applicant: Dade Behring, Inc. F. Proprietary and Established Names: N Antisera to Human Transferrin, Transferrin immunological test system N Antisera to Human Haptoglobin, Haptoglobin immunological test system G. Regulatory Information: 1. Regulation section: 21 CFR 866.5880, Transferrin immunological test system 21 CFR 866.5460, Haptoglobin immunological test system 2. Classification: Class II (Transferrin) Class II (Haptoglobin) 3. Product code: DDG, Transferrin, antigen, antiserum, control DAD, Haptoglobin, antigen, antiserum, control 4. Panel: Immunology 82 H. Intended Use: 1. Intended use(s): In vitro diagnostic reagents for the quantitative determination of transferrin and haptoglobin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometry on the BN™ systems. 2. Indication(s) for use: In vitro diagnostic reagents for the quantitative determination of transferrin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometry on the BN™ Systems. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: For use on the Dade Behring BNII, BN 100, and BN Prospec analyzers, {2} previously cleared under k860894. I. Device Description: The device consists of one vial containing 2 ml of N antiserum to human transferrin. J. Substantial Equivalence Information: 1. Predicate device name(s): N Antisera to Human Transferrin 2. Predicate 510(k) number(s): k972840 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | In vitro diagnostic reagents for the quantitative determination of transferrin and haptoglobin in human serum, heparinized and EDTA plasma, as well as transferrin in human urine by means of immunonephelometry on the BN™ systems. | In vitro diagnostic reagents for the quantitative determination of transferrin and haptoglobin in human serum, as well as transferrin in human urine by means of immunonephelometry on the BN™ systems. | | Antibody | Rabbit anti-Human transferrin (polyclonal) | Same | | Instrumentation | BN™ Systems | Same | | Assay Format | Quantitative nephelometry | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Sample | Serum, heparinized and EDTA plasma, urine | Serum and urine | K. Standard/Guidance Document Referenced (if applicable): None L. Test Principle: Proteins contained in human body fluids form immunochemical reaction with specific antibodies. These complexes scatter a beam o flight passed through the sample. The intensity of the scattered light is proportional to the concentration of the relevant protein in the sample. The result is evaluated by comparison with a standard of known concentration. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: No change. b. Linearity/assay reportable range: No change. {3} c. Traceability, Stability, Expected values (controls, calibrators, or methods): No change. d. Detection limit: No change. e. Analytical specificity: Interference by endogenous substances Transferrin: Normal serum samples (2.14-3.17 g/L) were spiked with increasing concentrations of bilirubin, hemoglobin, and triglycerides. The percent recovery was determined for each sample relative to a reference sample (± 20%). No interference was seen up to: 0.6 g/L bilirubin, 8.2 g/L triglycerides and 10 g/L hemoglobin. Normal serum samples (2.17-2.94 g/L) were compared to sera spiked with 5% of lithium, sodium, or ammonium heparin to determine potential interference by heparin anticoagulants for plasma samples. No interference was seen. Percent deviation (± 7%) between the mean recoveries of the heparin types was (-)1.06 to (+)0.528%. Haptoglobin: No data was provided for interference from bilirubin, hemoglobin, or triglycerides. Normal serum samples (0.6-2.11 g/L) were compared to sera spiked with 5% of lithium, sodium, or ammonium heparin to determine potential interference by heparin anticoagulants for plasma samples. No interference was seen. Percent deviation (± 7%) between the mean recoveries of the heparin types was (-)1.59 to (+)0.403%. f. Assay cut-off: No change. 2. Comparison studies: a. Method comparison with predicate device: Not applicable. b. Matrix comparison: Fresh and/or frozen serum and plasma samples, covering the reportable range (1:20 dilutions, Transferrin: 0.35-5.6 g/L; Haptoglobin: 0.26-8.3 g/L) were compared to determine if any significant bias existed between matrices. The heparin samples were a mixture of heparin types, however since the percent deviation between the heparin types (see Interference studies above) was low, this was deemed acceptable. | | N (pooled) | Regression equation | R² | 95% Confidence intervals (slope) | | --- | --- | --- | --- | --- | | Transferrin | | | | | | Heparin | 99 | y = 0.9768x - 0.0204 | 0.9851 | 0.9499, 1.003 | | EDTA | 40 | y = 1.0067x - 0.0508 | 0.9916 | 0.9767, 1.0425 | | Haptoglobin | | | | | | Heparin | 79 | y = 0.9777x - 0.0095 | 0.9932 | 0.9579, 1.0037 | | EDTA | 47 | y = 0.9963x - 0.0208 | 0.9936 | 0.9799, 1.0180 | 3. Clinical studies: a. Clinical Sensitivity: No change. {4} b. Clinical specificity: No change. 4. Clinical cut-off: No change. 5. Expected values/Reference range: No change. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...