K213019 · Hebei Astro Medical Supply Co., Ltd. · LYZ · Jan 5, 2022 · General Hospital
Device Facts
Record ID
K213019
Device Name
Vinyl Exam Gloves
Applicant
Hebei Astro Medical Supply Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
Jan 5, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Vinyl Exam Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Vinyl Exam Gloves are disposable, non-sterile, powder-free medical gloves made of vinyl. They are worn on the examiner's hand or finger to serve as a barrier to prevent contamination between the patient and the examiner. The device is available in five sizes (XS, XL) and is transparent. It is intended for use in clinical or medical settings by healthcare providers or other users. The gloves function as a physical barrier; they do not involve electronic processing, software, or active components. The device benefits the patient and examiner by reducing the risk of cross-contamination during medical examinations.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM standards for physical dimensions, tensile strength, elongation, residual powder, and freedom from holes. Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).
Technological Characteristics
Material: Vinyl. Form factor: Disposable, non-sterile, powder-free glove. Sizes: XS, S, M, L, XL. Standards: ASTM D5250-19 (Polyvinyl chloride gloves), ASTM D5151-19 (Hole detection), ASTM D3767-03 (Dimensions), ASTM D412-16 (Tension), ASTM D6124-06 (Residual powder). Biocompatibility: ISO 10993-5, ISO 10993-10. No energy source, software, or connectivity.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Suitable for use by healthcare professionals or in over-the-counter settings.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 5, 2022
Hebei Astro Medical Supply Co., Ltd Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai. 200120 China
Re: K213019
Trade/Device Name: Vinyl Exam Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: November 26, 2021 Received: December 8, 2021
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K213019
Device Name Vinyl Exam Gloves
Indications for Use (Describe)
Vinyl Exam Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K213019
#### 1. Date of Preparation: 01/03/2022
#### 2. Sponsor Identification
## Hebei Astro Medical Supply Co., Ltd
East of Xiaoxixian, West of Jingsan Street, South of Weiwu Road, North of Weiqi Road, Jinzhou Economic Development Zone, Hebei Province, P.R.China, 052260
Establishment Registration Number: 3015537296
Contact Person: Ning Zheng Position: General Manager Tel: +86-18617965639 Fax: +86-311-85125615 Email: erin@wallyplastic.net
#### 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
# Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-2281-5850, Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Vinyl Exam Gloves Common Name: Examination Vinyl Gloves
# Regulatory Information
Classification Name: Vinyl Patient Examination Glove Classification: I; Product Code: LYZ; Regulation Number: 21CFR 880.6250 Review Panel: General Hospital;
### Indication for Use:
Vinyl Exam Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
# Device Description
The proposed device is a powder free medical glove. The device is available in transparent. The device meets the requirements of ASTM D5250-19: Standard specification for Poly (vinyl chloride) Gloves for Medical Application. The proposed gloves are available in five sizes, which are XS, S, M, L, XL, it could be selected by the user depended on size of hand. The different between each size is just in the dimension. The proposed device is provided in non-sterile.
#### ડ. Identification of Predicate Device
510(k) Number: K210799 Product Name: Powder Free Vinyl Patient Examination Glove
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ASTM D5250-19 Standard specification for Poly (vinyl chloride) Gloves for Medical Application
- > ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves
- > ASTM D3767-03 (2020) Standard Practice for Rubber-Measurement of Dimensions
- > ASTM D412-16 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension
{5}------------------------------------------------
- > ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
- > ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization;
- > ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for In Vitro Cytotoxicity ;
- Summary of Technological characteristics 7.
| Table 1 Comparison of Technology Characteristics | | | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|---------------|
| ITEM | Proposed Device<br>K213019 | Predicate Device<br>K210799 | Remark | |
| Product Code | LYZ | LYZ | Same | |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same | |
| Class | I | I | Same | |
| Indication for use | Vinyl Exam Gloves is a disposable<br>device intended for medical purposes<br>that is worn on the examiner's hand<br>or finger to prevent contamination<br>between patient and examiner. | The Powder Free Patient<br>Examination Glove is a disposable<br>device intended for medical purposes<br>that is worn on the examiner's hand<br>or finger to prevent contamination<br>between patient and examiner. | Same | |
| Material | Vinyl | Vinyl | Same | |
| Color | Clear | Clear | Same | |
| Sterility | Non-sterile | Non-sterile | Same | |
| Single-use | Yes | Yes | Same | |
| Size | XS, S, M, L, XL | XS, S, M, L, XL | Same | |
| Dimensions<br>(ASTM D5250-19) | Width | | | |
| | XS | $75 \pm 5mm$ | XS | $75 \pm 5mm$ |
| | S | $85 \pm 5mm$ | S | $85 \pm 5mm$ |
| | M | $95 \pm 5mm$ | M | $95 \pm 5mm$ |
| | L | $105 \pm 5mm$ | L | $105 \pm 5mm$ |
| | Length | | | Same |
| | XS | 230mm min | XS | 230mm min |
| | S | 230mm min | S | 230mm min |
| | M | 230mm min | M | 230mm min |
| | L | 230mm min | L | 230mm min |
| | XL | 230mm min | XL | 230mm min |
| Thickness | | | | |
Table 1 Comparison of Technology Characteristics
{6}------------------------------------------------
| | Palm | 0.08mm min | Palm | 0.08mm min | | |
|---------------------------------------------------------------|--------------------------|-----------------------------------------------------------------|--------------------------|-----------------------------------------------------------------|------|------|
| | Finger | 0.08mm min | Finger | 0.08mm min | | |
| | Before Aging | | | | | |
| Physical Properties<br>(ASTM D5250-19<br>and ASTM<br>D412-16) | Tensile Strength | 11MPa min | Tensile Strength | 11MPa min | Same | |
| | Ultimate Elongation | 300% min | Ultimate Elongation | 300% min | | |
| | After Aging | | | | | |
| | Tensile Strength | 11MPa min | Tensile Strength | 11MPa min | | |
| | Ultimate Elongation | 300% min | Ultimate Elongation | 300% min | | |
| Power free residue<br>(ASTM D6124-06) | Less than 2mg per glove | | Less than 2mg per glove | | Same | |
| Freedom from<br>Holes (ASTM<br>D5151-19) | No water leakage occurs. | | No water leakage occurs. | | Same | |
| Biocompatibility | | | | | | |
| | Skin Irritation | Under the conditions of the study, not an Irritant | | Under the conditions of the study, not an Irritant | | |
| | Sensitization | Under the conditions of the study, not a Sensitization | | Under the conditions of the study, not a Sensitization | | Same |
| | Cytotoxicity | Under the conditions of the study, the device is non-cytotoxic. | | Under the conditions of the study, the device is non-cytotoxic. | | |
#### 8. Summary of Non-Clinical Performance Testing
The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard.
| No. | Name of the Test Methodology / Standard | Purpose | Acceptance Criteria | Test Results |
|-----|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| 1 | ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization | Evaluated for the potential to cause delayed dermal contact sensitization and skin irritation | Magnusson and Kligman grade shall be less than control group<br>No significant reaction than the control group | No skin sensitization<br>No skin irritation |
| 2 | ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity | Evaluated for the potential cytotoxicity | The viability shall be not reduced to less than 70% | No cytotoxic |
| 3 | ASTM D6124-06 (Reapproved | Evaluate the | Less than 2.0mg | Less than 2.0mg |
Table 2. Summary of Non-Clinical Performance Testing
{7}------------------------------------------------
| | 2017), Standard Test Method for<br>Residual Powder on Medical<br>Gloves | residue powder | | |
|---|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 4 | ASTM D5151-06(Reapproved<br>2015), Standard Test Method for<br>Detection of Holes in Medical<br>Gloves. | Detection the<br>holes that allow<br>water leakage | Do not show droplet,<br>stream or other type of<br>water leakage | No leakage |
| 5 | ASTM D6319-10 (Reapproved<br>2015), Standard Specification for<br>Nitrile Examination Gloves for<br>Medical Application. | Evaluate the<br>glove physical<br>dimension | Length: > 230 mm<br>Width (±5mm)<br>XS = 75mm<br>S = 85mm<br>M = 95mm<br>L = 105mm<br>XL = 115mm<br>Thickness at Finger<br>(mm)<br>All Sizes ≥ 0.08 mm<br>Thickness at Palm<br>All Sizes ≥ 0.08 mm | Length<br>Larger than 230mm<br>Width<br>XS: within 75±5mm<br>S: within 85±5mm<br>M: within 95±5mm<br>L: within 105±5mm<br>XL: within 115±5mm<br>Thickness<br>Larger than 0.08mm |
| 6 | ASTM 412-16 Standard Test<br>Methods for Vulcanized Rubber<br>and Thermoplastic<br>Elastomers-Tension | Evaluate the<br>physical<br>requirement | Tensile strength: 11Mpa<br>Ultimate elongation:<br>300% | Larger than 11Mpa and<br>300% |
#### 9. Clinical Test Conclusion
No clinical study is included in this submission.
# 10. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as the legally marketed predicate device K210799.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.