K211904 · Jiangsu Dihong Industry and Trade Co., Ltd. · LYZ · Sep 17, 2021 · General Hospital
Device Facts
Record ID
K211904
Device Name
Vinyl Glove
Applicant
Jiangsu Dihong Industry and Trade Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
Sep 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Vinyl Glove is disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Device Story
Vinyl Glove is a non-sterile, powder-free, ambidextrous patient examination glove. It acts as a physical barrier worn by clinicians or users to prevent cross-contamination between patient and examiner. The device is manufactured from vinyl material and is available in sizes S, M, L, and XL. It is intended for single-use applications in medical environments. The device provides a protective barrier; it does not involve electronic inputs, software, or automated processing. Clinical benefit is derived from the reduction of contamination risk during patient examinations.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ASTM D6124 (residual powder), ASTM D5151 (water tightness/holes), ASTM D5250 (dimensions), and ASTM D412 (tensile strength/elongation) was demonstrated. Biocompatibility testing per ISO 10993-5 and ISO 10993-10 confirmed safety.
Indicated for use as a disposable medical glove worn on the examiner's hands to prevent contamination between patient and examiner. Suitable for use by clinicians and in over-the-counter settings.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
September 17, 2021
Jiangsu Dihong Industry and Trade Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K211904
Trade/Device Name: Vinyl Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: April 19, 2021 Received: June 21, 2021
Dear Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Liqun Zhao -S
For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211904
Device Name Vinyl Glove
Indications for Use (Describe)
The Vinyl Glove is disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary- K211904
This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.
#### 1.0 Submitter's Information
Jiangsu Dihong Industry and Trade Co.,Ltd. Name: Address: Donqxiaodian Industrial Park, Shuyang City, Jiangsu, China Phone Number: +86-18761104047 Contact: Sue Chen Date of Preparation: 09/09/2021
#### Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
#### 2.0 Device Information
Trade name: Vinyl Glove Common name: Vinyl Patient Examination Glove Classification name: Non-powdered Patient Examination Glove Model(s): S, M, L, XL
#### 3.0 Classification
Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital
#### 4.0 Predicate Device Information
Manufacturer: Hebei Hongtai Plastic Products Company Limited Vinyl Patient Examination Gloves (White, Blue, Yellow) Device: 510(k) number: K163168
#### 5.0 Indication for Use
{4}------------------------------------------------
The Vinyl Glove is disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
#### 6.0 Device Description
The subject device is powder free vinyl patient examination gloves. The subject device is clear. The design of subject device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D5250. The subject device is non-sterile.
#### 7.0 Technological Characteristic Comparison Table
| Item | Subject device | Predicate device | Comparison |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) number | K211904 | K163168 | - |
| Product Code | LYZ | LYZ | Same |
| Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same |
| Class | I | I | Same |
| Intended Use | The Vinyl Glove is disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | The Vinyl Examination Glove (White, Blue, or Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same |
| Powdered or<br>Powered free | Powdered free | Powdered free | Same |
| Design Feature | Ambidextrous | Ambidextrous | Same |
| Labeling<br>Information | Single use, powder free, device color, device name, glove size and quantity, Vinyl Glove, Non-Sterile | Single use, powder free, device color, device name, glove size and quantity, Vinyl Examination Gloves, Non-Sterile | Similar |
#### Table1-General Comparison
#### Table2 Device Dimensions Comparison
| Predicate<br>Device<br>(K163168) | Designation | Size | | | | | Tolerance |
|----------------------------------|----------------|------|-----|-----|-----|-----|-----------|
| | Length, mm | XS | S | M | L | XL | |
| | | 230 | 230 | 235 | 245 | 245 | min |
| | Width, mm | 80 | 85 | 95 | 105 | 115 | $\pm 5$ |
| | Thickness, mm: | | | | | | |
| | Finger | 0.05 | | | | | min |
| | Palm | 0.08 | | | | | min |
| Subject Device | Designation | Size | | | | | Tolerance |
{5}------------------------------------------------
| (K211904) | | S | M | L | XL | |
|-----------|----------------|------|-----|-----|-----|-----|
| | Length, mm | 230 | 230 | 230 | 230 | min |
| | Width, mm | 85 | 95 | 105 | 115 | ±5 |
| | Thickness, mm: | | | | | |
| | Finger | 0.08 | | | | min |
| | Palm | 0.08 | | | | min |
| Remark | Similar | | | | | |
Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250,so the differences do not raise any new safety or performance questions.
| Item | Subject device<br>(K211904) | Predicate device<br>(K163168) | Comparison | |
|------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|------------|------------|
| Colorant | Clear | White, Blue, Yellow | Analysis 1 | |
| Physical<br>Properties | Before Aging<br>Tensile Strength | 11MPa, min | 15MPa, min | Analysis 2 |
| | Ultimate Elongation | 300%min | 380%min | Analysis 2 |
| | After Aging<br>Tensile Strength | 11MPa, min | 15MPa, min | Analysis 2 |
| | Ultimate Elongation | 300%min | 380%min | Analysis 2 |
| Comply with ASTM D5250 | Comply with ASTM D5250 | Comply with ASTM D5250 | Same | |
| Freedom from Holes | Be free from holes when tested in accordance with ASTM D5151 AQL=2.5 | Be free from holes when tested in accordance with ASTM D5151 AQL=2.5 | Same | |
| Powder Content | < 0.04 mg per glove, Meet the requirements of ASTM D6124 | Meet the requirements of ASTM D6124 | Similar | |
#### Table3 Performance Comparison
Analysis 1: The colorant of the subject device is different with that of the predicate device, the subject device was evaluated according to ISO 10993-1 standards, and there were no risks identified.
Analysis 2: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250,so the differences do not raise any new safety or performance questions.
#### Table4 Safety Comparison
| Table: Safety Comparison | | | |
|--------------------------|----------------|-------------------|------------|
| Item | Subject device | Predicated device | Comparison |
{6}------------------------------------------------
| Material | | Vinyl | Vinyl | Same |
|--------------------|---------------|------------------------------------------------------------------------------------------------------------------------|----------------------------|-----------|
| Biocompatibility | Irritation | Comply with<br>ISO10993-10.<br>Under the conditions<br>of the study, not an<br>irritant | | Same |
| | Sensitization | Comply with<br>ISO10993-10.<br>Under conditions of<br>the study, not a<br>sensitizer. | Comply with<br>ISO10993-10 | |
| | Cytotoxicity | Comply with<br>ISO10993-10.<br>Under conditions of<br>the study, did not<br>show potential<br>toxicity to L-929 cells. | Not provided | Different |
| Label and Labeling | | Meet FDA's<br>Requirement | Meet FDA's<br>Requirement | Same |
#### 8.0 Summary of Non-clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards as shown in Table 5 Biocompatibility and Table 6 Non-Clinical Testing below:
| Test Method | Purpose | Acceptance Criteria | Results |
|---------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------------------------|---------|
| ISO 10993-10:2010<br>Tests For Irritation And<br>Skin Sensitization | To determine if device<br>is a skin irritant | The device must be a<br>non-irritant | Pass |
| ISO 10993-10:2010<br>Tests For Irritation And<br>Skin Sensitization | To determine if device<br>is a skin sensitizer | The device must be a<br>non- sensitizer | Pass |
| ISO 10993-5:2009<br>Tests For In Vitro<br>Cytotoxicity | To determine if the<br>device is potential<br>toxicity to L-929<br>cells. | The device must be a<br>non toxicity. | Pass |
#### Table 5 Biocompatibility Testing
#### Table 6 Non-Clinical Testing
| Test Method | Purpose | Acceptance Criteria | Results |
|------------------------------------------------------------|---------------------------------|---------------------|---------|
| ASTM D6124-06<br>(Reapproved 2017)<br>Standard Test Method | To determine residual<br>powder | ≤ 2 mg/glove | Pass |
{7}------------------------------------------------
| for Residual Powder on<br>Medical Gloves | | | |
|-----------------------------------------------------------------------------------------------------------------|-------------------------------------|---------------------------------------------------|------------------------------------|
| ASTMD5151-19<br>Standard Test Method<br>for Detection of Holes<br>in Medical Gloves | To determine water<br>tightness | Meet the requirements<br>of ASTM D5151 AQL<br>2.5 | 0/125 leaks<br>Pass |
| ASTM D5250-19<br>Standard Specification<br>for Poly (vinyl chloride)<br>Gloves for Medical<br>Application | To determine physical<br>dimensions | Length(mm):≥230. | Length(mm):>230 |
| | | Width(mm): | Width(mm): |
| | | S: 85±5; | S: 84-86 |
| | | M: 95±5; | M: 95-97 |
| | | L: 105±5; | L: 105-107 |
| | | XL: 115±5; | XL: 115-117 |
| | | | Pass |
| ASTM D412-06a-2013<br>Standard Test Methods<br>for Vulcanized Rubber<br>and Thermoplastic<br>Elastomers—Tension | To determine physical<br>properties | Thickness (mm): | Finger: 0.09-0.11 |
| | | Finger: ≥0.08 | Palm: 0.08-0.11 |
| | | Palm: ≥0.08 | Pass |
| | | Before Aging: | Before Aging: |
| | | Tensile<br>Strength≥11MPa | Tensile Strength:<br>15.1~17.4 MPa |
| | | Ultimate<br>Elongation≥300% | Ultimate Elongation:<br>552%~560% |
| | | After Aging: | After Aging: |
| | | Tensile<br>Strength≥11MPa | Tensile Strength:<br>15.7~17.4 MPa |
| | | Ultimate<br>Elongation≥300% | Ultimate Elongation:<br>527%~550% |
| | | | Pass |
#### 9.0 Summary of Clinical Testing
Clinical testing is not needed for this device.
#### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device in 510(k) submission K211904, Vinyl Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K163168.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.