POWDER FREE VINYL SYNTHETIC EXAMINATION GLOVES W/ALOE VERA, GREEN COLOR
Applicant
Ever Light Plastic Products Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
Jul 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamingtion between healthcare personnel and the patient's body, fluids, waste or environment.
Device Story
Disposable vinyl synthetic examination glove; green color; contains aloe vera. Used by healthcare personnel to prevent cross-contamination between examiner and patient/environment. Device acts as physical barrier. No electronic, software, or algorithmic components. Bench testing confirms compliance with ASTM D-5250-00E4 standards for physical dimensions, water-tightness (pinhole testing), and residual powder limits.
Clinical Evidence
No clinical data. Bench testing only, including ASTM D-5250-00E4 physical/dimensional testing, FDA 1000 ml water fill test for pinholes, and residual powder testing per ASTM D-6124-01. Biocompatibility testing confirmed no primary skin irritation or sensitization.
Technological Characteristics
Vinyl synthetic material; green color; aloe vera additive. Complies with ASTM D-5250-00E4. Powder-free (residual powder < 2 mg/glove). Non-sterile. No electronic components or software.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Shijiazhuang Grand Work Plastic Products Co., Ltd Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color (K042830)
Related Devices
K042830 — POWDER FREE VINYL SYNTHETIC EXAMINATION GLOVES · Grand Work Plastic Products Co., Ltd. · Nov 17, 2004
K092302 — POWDER FREE VINYL SYNTHETIC EXAMINATION GLOVES WITH ALOE VERA, GREEN COLOR · Bettercare Plastic Products Co., Ltd. · May 5, 2010
K081539 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR · Rich Mountain Medical Products, Inc. · Sep 26, 2008
K023728 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR · Canopus Medical Supply Co., Ltd. · Jan 7, 2003
K072314 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES, WITH ALOE VERA, GREEN COLOR · Xuzhou Full Sun Medical Products , Ltd. · Dec 12, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(K) number is: _______________________________________________________________________________________________________________________________________________
#### 1. Submitter's Identification:
JUL - 7 2006
Mr. Yang Ying Ever Light Products Co., Ltd. Donggao Industrial Zone Zanhuang, Hebei, China 050000
Date Summary Prepared: March 14, 2006
### 2. Name of the Device:
Ever Light Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color
# 3. Predicate Device Information:
Shijiazhuang Grand Work Plastic Products Co., Ltd Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color (K042830)
#### 4. Device Description:
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder Free Vinyl Synthetic Examination Gloves with Aloe Vert Creen
Color, 80 LYZ, and meets all requirements of ASTM standard D-5250-00E4.
## 5. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamingtion between healthcare personnel and the patient's body, fluids, waste or environment.
{1}------------------------------------------------
11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
# 6. Comparison to Predicate Devices:
Ever Light Plastic Products Co., Ltd.'s Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color is substantially equivalent in safety and effectiveness to the Shijiazhuang Grand Work Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color.
# 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:
The standards used for Ever Light Plastic Products Co., Ltd.'s glove production are based on ASTM-D-5250-00E4. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.
Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
A Residual Powder Test that based on ASTM D-6124-01 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).
### 8. Labeling:
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
# 9. Discussion of Clinical Tests Performed:
Not Applicable -- There is no hypoallergenic Claim.
## 10. Conclusions:
Ever Light Plastic Products Co., Ltd.'s Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color conform fully to ASTM D-5250-00E4 stadard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and wings at the top.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 7 2006
Ever Light Plastic Products Company, Limited C/O Ms. Kathy Liu Gloveco, Incorporated 3973 Schaefer Avenue Chino, California 91710
Re: K060785
Trade/Device Name: Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 16, 2006 Received: June 19, 2006
Dear Ms. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Liu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Sujit J. Michael, M.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
# INDICATION FOR USE
K060185 510 (k) NUMBER (IF KNOW): APPLICANT: Even Light Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe DEVICE NAME: Vera, Green Color
INDICATIONS FOR USE:
A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Per 21 CFR 801.109) OR
い Over-The-Counter-Use (Optional Format 1-2-96)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
| Concurrent of CDRH, Office of Device Evaluation (ODE) | |
|-------------------------------------------------------|--|
|-------------------------------------------------------|--|
Shula H. Murphy, R.D. 7/16/04
Sign-Off
ion of Anesthesiology, General Hospital,
ion Control, Dental Devices
Number K060785
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.