DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR
K081539 · Rich Mountain Medical Products, Inc. · LYZ · Sep 26, 2008 · General Hospital
Device Facts
Record ID
K081539
Device Name
DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR
Applicant
Rich Mountain Medical Products, Inc.
Product Code
LYZ · General Hospital
Decision Date
Sep 26, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable vinyl synthetic examination glove; green color; infused with aloe vera. Worn by healthcare examiners on hands or fingers. Acts as a physical barrier to prevent cross-contamination between patient and examiner during medical procedures. Used in clinical settings. Device is non-sterile and powder-free.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including physical dimensions, watertightness (ASTM D5151-06), residual powder (ASTM D6124-06), and biocompatibility (skin irritation and sensitization).
Technological Characteristics
Material: Vinyl synthetic. Standards: ASTM D5250-06 (physical/dimensions), ASTM D5151-06 (watertightness), ASTM D6124-06 (residual powder). Form factor: Disposable, powder-free, green color, aloe vera infused. No electronic or software components.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hand or fingers to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Disposable Powder Free Vinyl Synthetic Examination Gloves With Aloe Vera, Green Color (K023728)
Related Devices
K024081 — DISPOSBLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR · Safehealth Medical Supply Corp. · Jan 28, 2003
K024026 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR · Shanghai PM Plastics Enterise Co., Ltd. · Dec 18, 2002
K032670 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA GREEN COLOR · Kingswell International, Inc. · Nov 14, 2003
K022087 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR · Shanghai PT Plastics Enterprise Company, Ltd. · Jul 11, 2002
K023728 — DISPOSABLE POWDER FREE VINYL SYNTHETIC EXAM GLOVES WITH ALOE VERA, GREEN COLOR · Canopus Medical Supply Co., Ltd. · Jan 7, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K081539
#### SEP 2 6 2008
# REVISED 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(K) safety and effectiveness information is being submitted in Accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(K) number is:
## 1. Submitter's Identification:
Michelle Shih Rich Mountain Medical Products Inc. No. 8, Jiang Yin Road, Yitang Town Pizhou County, Jiangsu Province China
Date Summary : April 15, 2008
# 2. Name of the Device:
Rich Mountain Medical Products Inc. Disposable Powder Free Vinyl Synthetic Exam Gloves With Aloe Vera, Green Color
#### 3. Predicate Device Information:
Canopus Medical Supply Company Limited. Disposable Powder Free Vinyl Synthetic Examination Gloves With Aloe Vera, Green Color (K023728)
#### 4. Device description:
Classified by FDA's General and Plastic Surgery Device panel as Class 1, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-06.
# 5. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
{1}------------------------------------------------
### 6.Comparison to Predicate Devices:
Rich Mountain Medical Products Inc.'s
Disposable Powder Free Vinyl Synthetic Exam Gloves With Aloe Vera, Green Color is substantially equivalent to the device manufactured by Canopus Medical Supply Company Limited. (K023728).
# 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows:
The standards used for Rich Mountain Medical Products Inc.'s glove production are Based on ASTM D5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection level S-2, AQL 4.0.
The FDA 1000 ml Watertight Test based on ASTM D-5151-06 was also conducted with samplings of AQL 2.5, Inspection level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
A residual Powder test that based on ASTM D6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per gloves)
### 8. Discussion of Clinic Tests Performed:
Not applicable - There is no Hypoallergenic Claim.
#### 9. Conclusions:
Rich Mountain Medical Products Inc's Disposable Powder Free Vinyl Synthetic Exam Gloves With Aloe Vera, Green Color conform fully to ASTM D-5250-06 standard as well as applicable 21CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a thick, black line.
Public Health Service
SEP 2 6 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Michelle Shih Coordinator Rich Mountain Medical Products, Incorporated No. 8, Jiang Yin Road, Yi Tang Town Pizhou County, Jiangsu CHINA
Re: K081539
Trade/Device Name: Disposable Powder Free Vinyl Synthetic Exam Glove, With Aloe Vera, Green Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: September 15, 2008 Received: September 18, 2008
Dear Ms. Shih:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Shih
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Manuel Lender for II
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
# Applicant: RICH MOUNTAIN MEDICAL PRODUCTS INC.
510(k) Number (if known): APPLIED
Device Name: Disposable Powder Free Vinyl Synthetic Exam Gloves, With
Aloe Vera, Green Color
Indications for Use:
A PATIENT EXAMINATION GLOVES IS A DISPOSABLE DEVICE INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINER'S HAND OR FINGERS TO PREVENT CONTAMINATION BETWEEN PATIENT AND EXAMINER.
> (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K081534
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.