K990841 · Terang Nusa Sdn Bhd · LZA · May 17, 1999 · General Hospital
Device Facts
Record ID
K990841
Device Name
NUGARD NITRIL NITRILE EXAMINATION GLOVE
Applicant
Terang Nusa Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
May 17, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
These nitrile examination gloves are to be worn by healthcare workers or similar personnel during work to prevent cross contamination between the user and the patient.
Device Story
Nugard Nitril is a blue, prepowdered nitrile examination glove; intended for use by healthcare workers to prevent cross-contamination between user and patient. Device is disposable, single-use, and ambidextrous. Manufactured from nitrile latex; does not contain natural rubber. Used in clinical settings; provides a physical barrier. No complex processing or electronic components involved.
Clinical Evidence
No clinical data; bench testing only. Testing conducted per ASTM D3578 and ASTM D512; biocompatibility tests performed.
Indicated for use by healthcare workers or similar personnel to prevent cross-contamination between the user and the patient.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K990840 — NUGARD NITRIL POWDERFREE NITRILE EXAMINATION GLOVE · Terang Nusa Sdn Bhd · May 17, 1999
K210145 — Nitrile Glove Powder Free Blue · Changzhou Universal Medical Equipment Co., Ltd. · Apr 21, 2021
K222813 — Nitrile Examination Gloves · Ecoglove Medical Manufacturing Company Limited · Mar 16, 2023
K142588 — Nitrile Powder Free Examination Glove (Blue) · Hartalega Sdn Bhd · Dec 18, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
K990841//A'
Image /page/0/Picture/1 description: The image shows the text "MAY 17" on the left side. On the right side of the image, there is a compass rose graphic. The compass rose has eight points and is black and white.
## TERANG NUSA Sdn Bhd
510(k) Summary for NUGARD Nitrile Examination Glove
## 510(k) Summary
| Submitter Name | Terang Nusa Sdn Bhd |
|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address | E4(4), Jalan 8<br>Pengkalan Chepa 2 Industrial Zone<br>16100 Kota Bharu,<br>Kelantan, Malaysia. |
| Submitter Telephone | +60 9 7735133 |
| Submitter Fax | +60 9 7737755 |
| Contact Person | LOW, Chin Guan |
| Date of preparation | 8 Mar 99 |
| Trade Name | Nugard Nitril |
| Common Name | Examination Glove |
| Classification | Patient Examination Glove |
| Legally marketed device to which<br>substantial equivalence is being<br>claimed. | The Nugard Nitril examination glove described in<br>this 510(k) is substantially equivalent to the nitrile<br>examination glove currently being marketed. |
| Description of device | Nugard Nitril meet the requirement for<br>examination glove described by the American<br>Standard for Testing and Material ASTM D 3578<br>with the exception of ultimate elongation before<br>aging. It is blue in color and prepowdered. Sizes<br>available is from XS - XL |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a compass rose with eight points. The compass rose is black and white, and it is oriented with north pointing upwards. There are some dotted lines around the compass rose, which may be part of a design or logo. The image is simple and clear, and it would be easily recognizable as a compass rose.
的一个人的一个人的一个人
## TERANG NUSA Sdn Bhd
e produktion programment of the provinsion of the second of the second of the second of the second of the second of the seconding the seconding the section of the section of
510(k) Summary for NUGARD Nitrile Examination Glove
... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ..
| Intended Use of the device | These nitrile examination gloves are to be worn by healthcare workers or similar personnel during work to prevent cross contamination between the user and the patient. |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of technological characteristics compared to predicate device | This notification describes the similarities to the approved device described. |
| Brief description of non-clinical tests | Test conducted per ASTM D3578, ASTM D512 indicates that the product meet the requirements.<br>Biocompatibility tests are carried out. |
| Brief description of clinical tests | Not carried out |
| Conclusion drawn from clinical and non clinical tests | Not applicable |
| Additional information deemed necessary by the FDA | None |
{2}------------------------------------------------
Appendix A
Dispenser Box Design
Image /page/2/Figure/2 description: The image shows two views, top and side, of a box of "NUGARD" brand blue nitrile examination gloves. The top view indicates the gloves are disposable, for single use only, prepowdered with absorbable dusting powder, USP, and ambidextrous. The side view includes a caution about residual powder and states the gloves are made from nitrile latex and do not contain natural rubber.
+
*
*
*
*
*
*
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is black.
MAY 17 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Chin-Guan Low Managing Director TERANG NUSA Sdn. Bhd. 1 Jalan 8, Pengkalan Chepa 2 Industrial Zone 16100 Kota Bharu, Kelantan, MALAYSIA
K990841 Re: Nugard Nitril® Nitrile Examination Glove Trade Name: Requlatory Class: I Product Code: LZA Dated: March 30, 1999 Received: April 6, 1999
Dear Mr. Chin-Guan Low:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in Please note: this response to your the Federal Reqister. premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Chin-Guan Low
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda/gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timoťhy) A Ulatowski Directdr Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a compass rose with four points oriented to the cardinal directions. The compass rose is black and white, with the points being black and the background being white. There is a faint outline of a square around the compass rose. The compass rose is a symbol of navigation and direction.
TERANG NUSA Sdn Bhd
510(k) Submission for Nitrile Examination Glove
## 3. Indication for use Statement
Applicant 510(k) Number Device Name Trade Name
Terang Nusa Sdn Bhd Not available K990841 Patient Examination Glove NUGARD Nitril ® Prepow dered, USP, BLUE
Indication for use
The examination gloves are to be worn by healthcare workers or similar personnel during work to prevent cross contamination between the user and the patient.
Concurrence of CDHR Office of Device Evaluation (ODE)
Burell & Chin
(Division Sign Off) (Division of Dental, Infection Control, and General Hospital Devices 510(k) Number KGA D841
Prescription Use Per 21 CFR 801.109 OR Over the counter
CANV Mitcil C. C., -- 1-- 4: - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.