Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid; Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid; Halyard STERLING NITRILE-XTRA * Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid

K231938 · O&M Halyard, Inc. · LZA · Mar 22, 2024 · General Hospital

Device Facts

Record IDK231938
Device NameHalyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid; Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid; Halyard STERLING NITRILE-XTRA * Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid
ApplicantO&M Halyard, Inc.
Product CodeLZA · General Hospital
Decision DateMar 22, 2024
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The Halyard STERLING NITRILE-XTRA* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Disposable, non-sterile, ambidextrous, chlorinated nitrile exam gloves; textured fingertips; available in 9.5-inch and 12-inch lengths. Used by healthcare professionals in clinical settings to provide a barrier against contamination between patient and examiner. Tested for permeation resistance against various chemotherapy drugs, fentanyl citrate, and gastric acid per ASTM D6978-05. Provides protection for clinicians handling hazardous drugs; however, specific breakthrough times are noted for certain agents (e.g., Carmustine, ThioTEPA), with warnings against use with these specific chemicals. Benefits include reduced risk of cross-contamination and potential for reduced Type IV allergic contact sensitization due to low dermatitis potential formulation.

Clinical Evidence

A 204-subject clinical study was conducted to evaluate Type IV allergic contact sensitization potential using the Jordan-King modification of the Draize test. Results showed the subject devices were non-irritating and demonstrated no clinical evidence of residual chemical additives capable of inducing Type IV allergy in human volunteers. Bench testing included ASTM D5151-06 (leakage), ASTM D6124-06 (residual powder), ASTM D6319-10 (physical properties), and ISO 10993 series (systemic toxicity, irritation, sensitization).

Technological Characteristics

Materials: Nitrile. Form factor: Ambidextrous, textured fingertips, chlorinated, powder-free. Dimensions: 9.5-inch or 12-inch length; 0.08-0.09 mm palm thickness. Sterility: Non-sterile. Standards: ASTM D5151-06, ASTM D6319-10, ASTM D6124-06, ASTM D6978-05, ISO 10993-10/11/23. Connectivity: None. Energy source: None.

Indications for Use

Indicated for use as a disposable medical exam glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs, fentanyl citrate, and gastric acid. Warning: Do not use with Carmustine or ThioTEPA.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 22, 2024 O&M Halyard, Inc. Caitlin Senter Director. Global Regulatory Affairs 1 Edison Drive Alpharetta, Georgia 30005 ## Re: K231938 Trade/Device Name: Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid; Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential. Tested for use with Chemotherapy Drugs. Fentanyl Citrate and Gastric Acid; Halyard STERLING NITRILE-XTRA * Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC, OPJ, QDO Dated: February 23, 2024 Received: February 23, 2024 Dear Caitlin Senter: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Allan Guan -S For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231938 #### Device Name Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid #### Indications for Use (Describe) The Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs. Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (Vidaza) (25 mg/mL) Bendamustine HCl (5 mg/mL) Bleomycin Sulfate (Blenoxane) (15 mg/mL) Bortezomib (Velcade) (1 mg/mL) Busulfan (6 mg/mL) Capecitabine (26 mg/mL) Carboplatin ( 10 mg/mL) Carfilzomib (2 mg/mL) Cetuximab (Erbitux) (2 mg/mL) Cisplatin (1 mg/mL) Cladribine (1 mg/mL) Cyclophosphamide (20 mg/mL) Cytarabine HCl (Cytosine) (100 mg/mL) Dacarbazine (DTIC) (10 mg/mL) Dactinomycin (0.5 mg/mL) Daunorubicin HCl (5 mg/mL) Decitabine ( 5 mg/mL) Docetaxel HCl (20 mg/mL) Doxorubicin HCl (2 mg/mL) Epirubicin HCl (Ellence) (2 mg/mL) Etoposide (20 mg/mL) Floxuridine (1 00mg/mL) 5-Fluorouracil (50 mg/mL) Gemcitabine HCl (38 mg/mL) Idarubicin HCl (1 mg/mL) lfosfamide (IFEX) (50 mg/mL) Irinotecan HCl (20 mg/mL) Lenvatinib (20mg/mL) Leuprolide Acetate Salt (5 mg/mL) Mechlorethamine HCl (1 mg/mL) Melphalan HCl (5 mg/mL) Methotrexate (25 mg/mL) Mitomycin C (0.5 mg/mL) Mitoxantrone HCl (2 mg/mL) Nelarabine (5 mg/mL) Oxaliplatin (5 mg/mL) Paclitaxel (Taxol) (6 mg/mL (6,000 ppm) Pemetrexed (25 mg/mL) FORM FDA 3881 (8/23) Page 1 of 2 {4}------------------------------------------------ Raltitrexed (0.5 mg/mL) Sorafenib Tosylate (200 mg/mL) Streptozocin (100mg/mL) Tamoxifen (2 mg/mL) Teniposide (10 mg/mL) Topotecan HCl (1 mg/mL) Trisenox (Arsenic Trioxide) (1 mg/mL) Vinblastine Sulfate (1 mg/mL) Vincristine Sulfate (Oncovin) (1 mg/mL) Vinorelbine (10 mg/mL) The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: Carmustine (3.3 mg/mL), breakthrough detected at 22.9 minutes The following chemotherapy drugs and concentration showed breakthrough detected in less than 40 minutes: ThioTEPA (10 mg/mL), breakthrough detected at 37.1 minutes Warning: Do not use with Carmustine or ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231938 #### Device Name Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid #### Indications for Use (Describe) The Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (Vidaza) (25 mg/mL) Bendamustine HCl (5 mg/mL) Bleomycin Sulfate (Blenoxane) (15 mg/mL) Bortezomib (Velcade) (1 mg/mL) Busulfan (6 mg/mL) Capecitabine (26 mg/mL) Carboplatin ( 10 mg/mL) Carfilzomib (2 mg/mL) Cetuximab (Erbitux) (2 mg/mL) Cisplatin (1 mg/mL) Cladribine (1 mg/mL) Cyclophosphamide (20 mg/mL) Cytarabine HCl (Cytosine) (100 mg/mL) Dacarbazine (DTIC) (10 mg/mL) Dactinomycin (0.5 mg/mL) Daunorubicin HCl (5 mg/mL) Decitabine ( 5 mg/mL) Docetaxel HCl (20 mg/mL) Doxorubicin HCl (2 mg/mL) Epirubicin HCl (Ellence) (2 mg/mL) Etoposide (20 mg/mL) Eribulin Mesylate (0.5 mg/mL) Fludarabine (25 mg/mL) 5-Fluorouracil (50 mg/mL) Gemcitabine HCl (38 mg/mL) Idarubicin HCl (1 mg/mL) lfosfamide (IFEX) (50 mg/mL) Irinotecan HCl (20 mg/mL) Lenvatinib (20 mg/mL) Leuprolide Acetate Salt (5 mg/mL) Mechlorethamine HCl (1 mg/mL) Melphalan HCl (5 mg/mL) Methotrexate (25 mg/mL) Mitomycin C (0.5 mg/mL) Mitoxantrone HCl (2 mg/mL) Oxaliplatin (5 mg/mL) Paclitaxel (Taxol) (6 mg/mL (6,000 ppm) Pemetrexed (25 mg/mL) FORM FDA 3881 (8/23) {6}------------------------------------------------ Raltitrexed (0.5 mg/mL) Ritux1mab ( 10 mg/mL) Sorafenib Tosylate (200 mg/mL) Tamoxifen (2 mg/mL) Topotecan HCl (1 mg/mL) Trisenox (Arsenic Trioxide) (1 mg/mL) Vinblastine Sulfate (1 mg/mL) Vincristine Sulfate (Oncovin) (1 mg/mL) Vinorelbine (10 mg/mL) The following chemotherapy drugs and concentration showed breakthrough detected in less than 20 minutes: Carmustine (BCNU) (3.3 mg/ml), breakthrough detected at 14.8 minutes The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: ThioTEPA (10 mg/ml), breakthrough detected at 23.9 minutes Warning: Do not use with Carmustine or ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes: Cyclosporin A (100 mg/mL) Cytovene (Ganciclovir) (10 mg/mL) Retrovir (Zidovudine) (10 mg/mL) Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {7}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231938 #### Device Name Halyard STERLING NITRILE-XTRA* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid #### Indications for Use (Describe) The Halyard STERLING NITRILE-XTRA* Powder-Free Exam Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs. Fentanyl Citrate and Gastric Acid are disposable devices intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes: Azacitidine (Vidaza) (25 mg/mL) Bendamustine HCl (5 mg/mL) Bleomycin Sulfate (Blenoxane) (15 mg/mL) Bortezomib (Velcade) (1 mg/mL) Busulfan (6 mg/mL) Capecitabine (26 mg/mL) Carboplatin ( 10 mg/mL) Carfilzomib (2 mg/mL) Cetuximab (Erbitux) (2 mg/mL) Cisplatin (1 mg/mL) Cladribine (1 mg/mL) Cyclophosphamide (20 mg/mL) Cytarabine HCl (Cytosine) (100 mg/mL) Dacarbazine (DTIC) (10 mg/mL) Dactinomycin (0.5 mg/mL) Daunorubicin HCl (5 mg/mL) Decitabine ( 5 mg/mL) Docetaxel HCl (20 mg/mL) Doxorubicin HCl (2 mg/mL) Epirubicin HCl (Ellence) (2 mg/mL) Etoposide (20 mg/mL) Floxuridine (100 mg/mL) Fludarabine (25 mg/mL) 5-Fluorouracil (50 mg/mL) Gemcitabine HCl (38 mg/mL) Idarubicin HCl (1 mg/mL) lfosfamide (IFEX) (50 mg/mL) Irinotecan HCl (20 mg/mL) Lenvatinib (20 mg/mL) Leuprolide Acetate Salt (5 mg/mL) Mechlorethamine HCl (1 mg/mL) Melphalan HCl (5 mg/mL) Methotrexate (25 mg/mL) Mitomycin C (0.5 mg/mL) Mitoxantrone HCl (2 mg/mL) Nelarabine (5 mg/mL) Oxaliplatin (5 mg/mL) Paclitaxel (Taxol) (6 mg/mL (6,000 ppm) {8}------------------------------------------------ Pemetrexed (25 mg/mL) Ralt1trexed (0.5 mg/mL) Rituximab ( 10 mg/mL) Sorafenib Tosylate (200 mg/mL) Streptozocin (100 mg/mL) Tamoxifen (2 mg/mL) Teniposide (10 mg/mL) Topotecan HCl (1 mg/mL) Trisenox (Arsenic Trioxide) (1 mg/mL) Vinblastine Sulfate (1 mg/mL) Vincristine Sulfate (Oncovin) (1 mg/mL) Vinorelbine (10 mg/mL) The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: Carmustine (3.3 mg/ml), breakthrough detected at 25.2 minutes The following chemotherapy drugs and concentration showed breakthrough detected in less than 40 minutes: ThioTEPA (10 mg/ml), breakthrough detected at 35.5 minutes Warning: Do not use with Carmustine or ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of a stylized "OM" in a dark red color, with the "O" being a circle with horizontal lines running through it. To the right of the "OM" is the text "Owens & Minor" in a gray color. The text is in a serif font and is slightly smaller than the "OM" symbol. # 510(k)Summary - K231938 | Date Summary<br>was Prepared | March 22, 2024 | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Submitter | O & M Halyard, Inc.<br>1 Edison Drive<br>Alpharetta, GA 30005 | | Primary Contact for<br>this 510(k) Submission | Caitlin Senter, MS, RAC<br>Tel: 678-221-7330<br>Email: caitlin.senter@hyh.com | | Marketed Common Name | Nitrile Powder-Free Exam Gloves | | Device Submission Trade name<br>and Description | Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis<br>Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and<br>Gastric Acid,<br><br>Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves, Low<br>Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl<br>Citrate and Gastric Acid<br><br>Halyard STERLING NITRILE-XTRA * Powder-Free Exam Gloves, Low Dermatitis<br>Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and<br>Gastric Acid | | Device Common Name | Medical Exam Gloves | | Device Product Code<br>and Classification Name | LZA Class I, 21 CFR §880.6250 Patient Examination Glove<br>LZC Class I, 21 CFR §880.6250 Patient Examination Glove, Specialty;<br>OPJ Class I, 21 CFR §880.6250 Medical Gloves With Chemotherapy Labeling<br>Claims - Test For Use With Chemotherapy Drugs<br>QDO Class I, 21 CFR §880.6250 Fentanyl and other opioid protection glove | | Predicate Device | Halyard Sterling* Nitrile Powder-Free Exam Gloves, Halyard Sterling SG*<br>Nitrile Sensi-Guard Powder-Free Exam Gloves (K191230) | | Reference Device | Halyard Purple Xtra and Purple Sterile, Low Dermatitis Potential, Powder-Free<br>Exam Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate<br>(K192241) | | Subject Device Description | The subject devices are disposable, grey-colored, chlorinated, nitrile,<br>powder-free, textured fingertip, ambidextrous, non-sterile patient<br>examination gloves that are packed in a cardboard dispenser box. | | | The devices follow consensus standards: | | | ASTM D5151-06 Standard Test Method for Detection of Holes in Medical<br>Gloves | | | ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for<br>Medical Applications | | | ASTM D6124-06 Standard Test Method for Residual Powder on Medical<br>Gloves | | | ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical<br>Gloves to Permeation by Chemotherapy Drugs | | | ISO 10993-11: 2017 Biological evaluation of medical devices – Tests for<br>systemic toxicity. | | | ISO 10993-23: 2021 Biological Evaluation of Medical Devices - Part 23: Tests<br>for Irritation | | | ISO 10993-10: 2010 Biological evaluation of medical devices – Tests for skin<br>sensitization | | Indications for Use for Halyard<br>STERLING* Nitrile Powder-Free<br>Exam Gloves, Low Dermatitis<br>Potential, Tested for use with<br>Chemotherapy Drugs, Fentanyl<br>Citrate and Gastric Acid | The Halyard STERLING* Nitrile Powder-Free Exam Gloves, Low Dermatitis<br>Potential, Tested for use with Chemotherapy Drugs, Fentanyl Citrate and<br>Gastric Acid are disposable devices intended for medical purposes that is<br>worn on the examiner's hand to prevent contamination between patient and<br>examiner.<br><br>The following chemotherapy drugs and concentration had NO breakthrough<br>detected up to 240 minutes:<br>Azacitidine (Vidaza) (25 mg/mL)<br>Bendamustine HCl (5 mg/mL)<br>Bleomycin Sulfate (Blenoxane) (15 mg/mL)<br>Bortezomib (Velcade) (1 mg/mL)<br>Busulfan (6 mg/mL)<br>Capecitabine (26 mg/mL)<br>Carboplatin ( 10 mg/mL)<br>Carfilzomib (2 mg/mL)<br>Cetuximab (Erbitux) (2 mg/mL)<br>Cisplatin (1 mg/mL)<br>Cladribine (1 mg/mL)<br>Cyclophosphamide (20 mg/mL)<br>Cytarabine HCl (Cytosine) (100 mg/mL)<br>Dacarbazine (DTIC) (10 mg/mL)<br>Dactinomycin (0.5 mg/mL)<br>Daunorubicin HCl (5 mg/mL)<br>Decitabine ( 5 mg/mL)<br>Docetaxel HCl (20 mg/mL)<br>Doxorubicin HCl (2 mg/mL)<br>Epirubicin HCl (Ellence) (2 mg/mL)<br>Etoposide (20 mg/mL)<br>Floxuridine (100mg/mL)<br>5-Fluorouracil (50 mg/mL)<br>Gemcitabine HCl (38 mg/mL)<br>Idarubicin HCl (1 mg/mL)<br>Ifosfamide (IFEX) (50 mg/mL)<br>Irinotecan HCl (20 mg/mL)<br>Lenvatinib (20mg/mL)<br>Leuprolide Acetate Salt (5 mg/mL)<br>Mechlorethamine HCl (1 mg/mL)<br>Melphalan HCl (5 mg/mL)<br>Methotrexate (25 mg/mL)<br>Mitomycin C (0.5 mg/mL)<br>Mitoxantrone HCl (2 mg/mL)<br>Nelarabine (5 mg/mL)<br>Oxaliplatin (5 mg/mL)<br>Paclitaxel (Taxol) (6 mg/mL (6,000 ppm)<br>Pemetrexed (25 mg/mL)<br>Raltitrexed (0.5 mg/mL)<br>Sorafenib Tosylate (200 mg/mL)<br>Streptozocin (100mg/mL)<br>Tamoxifen (2 mg/mL)<br>Teniposide (10 mg/mL) | {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of two parts: a stylized, maroon-colored emblem on the left and the company name "Owens & Minor" in gray on the right. The emblem is made up of the letters "OM" with horizontal lines running through them. The text is in a serif font and is stacked vertically. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of the letters "OM" in a dark red color, with horizontal lines running through the letters. To the right of the letters, the words "Owens & Minor" are written in a light gray color. The logo is simple and professional, and it is likely used to represent the company in its marketing and branding materials. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of a stylized "OM" in maroon color on the left, with horizontal lines running through the letters. To the right of the letters is the company name "Owens & Minor" in gray color. The font is serif and the ampersand is stylized. | Topotecan HCl (1 mg/mL) | |-----------------------------------------| | Trisenox (Arsenic Trioxide) (1 mg/mL) | | Vinblastine Sulfate (1 mg/mL) | | Vincristine Sulfate (Oncovin) (1 mg/mL) | | Vinorelbine (10 mg/mL) | The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: Carmustine (3.3 mg/mL), breakthrough detected at 22.9 minutes The following chemotherapy drugs and concentration showed breakthrough detected in less than 40 minutes: ThioTEPA (10 mg/mL), breakthrough detected at 37.1 minutes Warning: Do not use with Carmustine or ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: Fentanyl Citrate Injection (100 mcg/2 mL) Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of two parts: a red symbol on the left and the company name "Owens & Minor" in gray on the right. The red symbol is a stylized representation of the letters "OM" with horizontal lines running through them. The company name is written in a serif font, with the "&" symbol connecting the two names. | Indications for Use for Halyard<br>STERLING SG* SENSI-GUARD<br>Powder-Free Nitrile Exam<br>Gloves, Low Dermatitis Potential,<br>Tested for use with<br>Chemotherapy Drugs, Fentanyl<br>Citrate and Gastric Acid | The Halyard STERLING SG* SENSI-GUARD Powder-Free Nitrile Exam Gloves,<br>Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl<br>Citrate and Gastric Acid are disposable devices intended for medical purposes<br>that is worn on the examiner's hand to prevent contamination between<br>patient and examiner.<br><br>The following chemotherapy drugs and concentration had NO breakthrough<br>detected up to 240 minutes:<br>Azacitidine (Vidaza) (25 mg/mL)<br>Bendamustine HCl (5 mg/mL)<br>Bleomycin Sulfate (Blenoxane) (15 mg/mL)<br>Bortezomib (Velcade) (1 mg/mL)<br>Busulfan (6 mg/mL)<br>Capecitabine (26 mg/mL)<br>Carboplatin ( 10 mg/mL)<br>Carfilzomib (2 mg/mL)<br>Cetuximab (Erbitux) (2 mg/mL)<br>Cisplatin (1 mg/mL)<br>Cladribine (1 mg/mL)<br>Cyclophosphamide (20 mg/mL)<br>Cytarabine HCl (Cytosine) (100 mg/mL)<br>Dacarbazine (DTIC) (10 mg/mL)<br>Dactinomycin (0.5 mg/mL)<br>Daunorubicin HCl (5 mg/mL)<br>Decitabine ( 5 mg/mL)<br>Docetaxel HCl (20 mg/mL)<br>Doxorubicin HCl (2 mg/mL)<br>Epirubicin HCl (Ellence) (2 mg/mL)<br>Etoposide (20 mg/mL)<br>Eribulin Mesylate (0.5 mg/mL)<br>Fludarabine (25 mg/mL)<br>5-Fluorouracil (50 mg/mL)<br>Gemcitabine HCl (38 mg/mL)<br>Idarubicin HCl (1 mg/mL)<br>Ifosfamide (IFEX) (50 mg/mL)<br>Irinotecan HCl (20 mg/mL)<br>Lenvatinib (20 mg/mL)<br>Leuprolide Acetate Salt (5 mg/mL)<br>Mechlorethamine HCl (1 mg/mL)<br>Melphalan HCl (5 mg/mL)<br>Methotrexate (25 mg/mL)<br>Mitomycin C (0.5 mg/mL)<br>Mitoxantrone HCl (2 mg/mL)<br>Oxaliplatin (5 mg/mL)<br>Paclitaxel (Taxol) (6 mg/mL (6,000 ppm)<br>Pemetrexed (25 mg/mL)<br>Raltitrexed (0.5 mg/mL)<br>Ritux1mab ( 10 mg/mL)<br>Sorafenib Tosylate (200 mg/mL)<br>Tamoxifen (2 mg/mL)<br>Topotecan HCl (1 mg/mL) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of two parts: a red symbol on the left and the company name in gray on the right. The red symbol is a stylized "OM" with horizontal lines running through it, while the company name "Owens & Minor" is written in a serif font. | Trisenox (Arsenic Trioxide) (1 mg/mL) | |-----------------------------------------| | Vinblastine Sulfate (1 mg/mL) | | Vincristine Sulfate (Oncovin) (1 mg/mL) | | Vinorelbine (10 mg/mL) | The following chemotherapy drugs and concentration showed breakthrough detected in less than 20 minutes: | Carmustine (BCNU) (3.3 mg/ml), breakthrough detected at 14.8 minutes | |----------------------------------------------------------------------| |----------------------------------------------------------------------| The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: | ThioTEPA (10 mg/ml), breakthrough detected at 23.9 minutes | |------------------------------------------------------------| |------------------------------------------------------------| Warning: Do not use with Carmustine or ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: | Fentanyl Citrate Injection (100 mcg/2 mL) | |----------------------------------------------------------------------------| | Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution | The following hazardous drugs and concentration had NO breakthrough detected up to 240 minutes: | Cyclosporin A (100 mg/mL) | |-----------------------------------| | Cytovene (Ganciclovir) (10 mg/mL) | | Retrovir (Zidovudine) (10 mg/mL) | {15}------------------------------------------------ Image /page/15/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of two parts: a maroon-colored symbol on the left and the company name in gray on the right. The symbol is a stylized "OM" with horizontal lines running through it. The company name, "Owens & Minor," is written in a serif font, with the "&" symbol connecting the two names. | Indications for Use for Halyard<br>STERLING NITRILE-XTRA*<br>Powder-Free Exam Gloves, Low<br>Dermatitis Potential, Tested for<br>use with Chemotherapy Drugs,<br>Fentanyl Citrate and Gastric Acid | The Halyard STERLING NITRILE-XTRA* Powder-Free Exam Gloves, Low<br>Dermatitis Potential, Tested for use with Chemotherapy Drugs, Fentanyl<br>Citrate and Gastric Acid are disposable devices Intended for medical purposes<br>that is worn on the examiner's hand to prevent contamination between<br>patient and examiner.<br><br>The following chemotherapy drugs and concentration had NO breakthrough<br>detected up to 240 minutes:<br>Azacitidine (Vidaza) (25 mg/mL)<br>Bendamustine HCl (5 mg/mL)<br>Bleomycin Sulfate (Blenoxane) (15 mg/mL)<br>Bortezomib (Velcade) (1 mg/mL)<br>Busulfan (6 mg/mL)<br>Capecitabine (26 mg/mL)<br>Carboplatin ( 10 mg/mL)<br>Carfilzomib (2 mg/mL)<br>Cetuximab (Erbitux) (2 mg/mL)<br>Cisplatin (1 mg/mL)<br>Cladribine (1 mg/mL)<br>Cyclophosphamide (20 mg/mL)<br>Cytarabine HCl (Cytosine) (100 mg/mL)<br>Dacarbazine (DTIC) (10 mg/mL)<br>Dactinomycin (0.5 mg/mL)<br>Daunorubicin HCl (5 mg/mL)<br>Decitabine ( 5 mg/mL)<br>Docetaxel HCl (20 mg/mL)<br>Doxorubicin HCl (2 mg/mL)<br>Epirubicin HCl (Ellence) (2 mg/mL)<br>Etoposide (20 mg/mL)<br>Floxuridine (100 mg/mL)<br>Fludarabine (25 mg/mL)<br>5-Fluorouracil (50 mg/mL)<br>Gemcitabine HCl (38 mg/mL)<br>Idarubicin HCl (1 mg/mL)<br>Ifosfamide (IFEX) (50 mg/mL)<br>Irinotecan HCl (20 mg/mL)<br>Lenvatinib (20 mg/mL)<br>Leuprolide Acetate Salt (5 mg/mL)<br>Mechlorethamine HCl (1 mg/mL)<br>Melphalan HCl (5 mg/mL)<br>Methotrexate (25 mg/mL)<br>Mitomycin C (0.5 mg/mL)<br>Mitoxantrone HCl (2 mg/mL)<br>Nelarabine (5 mg/mL)<br>Oxaliplatin (5 mg/mL)<br>Paclitaxel (Taxol) (6 mg/mL (6,000 ppm)<br>Pemetrexed (25 mg/mL)<br>Ralt1trexed (0.5 mg/mL)<br>Rituximab ( 10 mg/mL)<br>Sorafenib Tosylate (200 mg/mL) | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| {16}------------------------------------------------ Image /page/16/Picture/0 description: The image shows the logo for Owens & Minor. The logo consists of the letters "OM" in a stylized, maroon-colored font on the left side. To the right of the letters is the company name "Owens & Minor" in a gray, sans-serif font. The ampersand symbol is placed between the two words. | Tamoxifen (2 mg/mL) | |-----------------------------------------| | Teniposide (10 mg/mL) | | Topotecan HCI (1 mg/mL) | | Trisenox (Arsenic Trioxide) (1 mg/mL) | | Vinblastine Sulfate (1 mg/mL) | | Vincristine Sulfate (Oncovin) (1 mg/mL) | | Vinorelbine (10 mg/mL) | The following chemotherapy drugs and concentration showed breakthrough detected in less than 30 minutes: | Carmustine (3.3 mg/ml), breakthrough detected at 25.2 minutes | |---------------------------------------------------------------| |---------------------------------------------------------------| The following chemotherapy drugs and concentration showed breakthrough detected in less than 40 minutes: | ThioTEPA (10 mg/ml), breakthrough detected at 35.5 minutes | |------------------------------------------------------------| |------------------------------------------------------------| Warning: Do not use with Carmustine or ThioTEPA The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes: | Fentanyl Citrate Injection (100 mcg/2 mL) | |----------------------------------------------------------------------------| | Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution | | Technological Characteristics Comparison Table | | | | | | | | |------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--| | | Subject Device<br>Halyard<br>STERLING*<br>Nitrile Powder-<br>Free Exam | Subject Device<br>Halyard STERLING<br>SG* NITRILE Sensi-<br>Guard Powder-<br>Free Exam Gloves | Subject Device<br>Halyard<br>STERLING*<br>NITRILE-XTRA*<br>Powder-Free | Predicate Device<br>K191230 | Reference<br>Device<br>K192241…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...