Black Nitrile Powder Free Patient Examination Glove, Non Sterile
K222715 · Safeskin Retailing (Hk) Limited · LZA · Dec 3, 2022 · General Hospital
Device Facts
Record ID
K222715
Device Name
Black Nitrile Powder Free Patient Examination Glove, Non Sterile
Applicant
Safeskin Retailing (Hk) Limited
Product Code
LZA · General Hospital
Decision Date
Dec 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A nitrile patient examination glove is a disposable device made of nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
Device Story
Disposable, non-sterile, ambidextrous patient examination glove made of nitrile rubber. Designed for use by healthcare professionals in clinical settings to provide a barrier against contamination. Surface treatment ensures smooth donning without powder or silicone lubricants. Device is single-use. Performance verified through physical property testing (tensile strength, elongation), watertight integrity (ASTM D5151), and biocompatibility assessments (irritation, sensitization, systemic toxicity).
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D412 (tensile/elongation), ASTM D3767 (dimensions), ASTM D5151 (watertight/holes), and ASTM D6124 (residual powder). Biocompatibility testing per ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (acute systemic toxicity) confirmed safety.
Technological Characteristics
Material: Nitrile rubber. Form: Ambidextrous, powder-free, non-sterile. Dimensions/Physicals: Compliant with ASTM D6319. Thickness: 0.06mm (palm) to 0.10mm (finger). Biocompatibility: ISO 10993-10 and ISO 10993-11 compliant. Manufacturing: Surface treatment for smooth donning.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Disposable Powder Free Nitrile Examination Glove, Black Color (K190942)
Related Devices
K210755 — Powder Free Nitrile Patient Examination Gloves, Non-Sterile (Blue) · Meditech Gloves Sdn Bhd · May 7, 2021
K140418 — POWDER FREE BLUE NITRILE PATIENT EXAMINATION GLOVES, NON-STERILE · Wrp Rubber Products Sdn Bhd · Sep 3, 2014
K210366 — Blue Nitrile Powder Free Patient Examination Glove, Non Sterile · Onetexx Sdn Bhd · May 7, 2021
K212916 — Black Nitrile Powder Free Patient Examination Glove, Non Sterile · Rimba Glove Sdn Bhd · Mar 24, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 3, 2022
Safeskin Retailing (HK) Limited % Stephan Toupan President Dawa Medical LLC 7320 NW 12th Street Suite 103 Miami, Florida 33126
Re: K222715
Trade/Device Name: Black Nitrile Powder Free Patient Examination Glove, Non Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA Dated: September 5, 2022 Received: September 8, 2022
Dear Stephan Toupan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Bifeng Qian -S
Bifeng Oian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K222715
Device Name
BLACK NITRILE POWDER-FREE PATIENT EXAMINATION GLOVE, NON STERILE
Indications for Use (Describe)
A nitrile patient examination glove is a disposable device made of nitrile rubber intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K222715 As required by 21CFR§807.92(c)
| 1.0 | Submitter:<br>Name | : Alvin Ho |
|-----|--------------------------|-----------------------------------------------------------------------------------------------------------------|
| | Address | : Safeskin Retailing (HK) Limited<br>26th Floor, Beautiful Group Tower, 77 Connaught Road<br>Central, Hong Kong |
| | Phone No. | : +6012 826 5625 |
| | Date of Summary Prepared | : 11 November 2022 |
#### 2.0 Identification of the subject device:
| Trade Name: | Black Nitrile Powder-Free Patient Examination Glove, Non-<br>Sterile |
|-----------------------|----------------------------------------------------------------------|
| Common Name: | Patient Examination Gloves |
| Classification Name: | Patient Examination Gloves |
| Device Classification | 1 |
| Regulation Number: | 21 CFR 880.6250 |
| Product Code | LZA |
#### 3.0 Predicate Device:
## K190942
Trade Name : Disposable Powder Free Nitrile Examination Glove, Black Color Company: Ever Growth (Vietnam) Co., Ltd
#### Description of The Subject Device: 4.0
Black Nitrile Powder-Free Patient Examination Glove, Non-Sterile is manufactured from nitrile rubber. Innersurface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder or silicone on the glove surface. The glove is ambidextrous, i.e., can be worn on right or left hand.
#### 5.0 Indication for use:
A Nitrile patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### 6.0 Comparison of the Technological Characteristics of the Device:
The Black Nitrile Powder-Free Patient Examination Gloves, Non-Sterile are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1
{4}------------------------------------------------
## Table 1
| | | DEVICE PERFORMANCE | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| CHARACTERISTICS | STANDARDS | PREDICATE<br>BLACK | CURRENT<br>BLACK | COMPARISON<br>ANALYSIS |
| 510(k) Number | - | K190942 | K222715 | |
| Manufacturer(s) | - | Ever Growth (Vietnam) Co., Ltd | Safeskin Retailing (HK) Limited | |
| Material | ASTM D6319 | Nitrile | Nitrile | Same |
| Color | - | Black | Black | Same |
| Sterility | - | Non-Sterile | Non-Sterile | Same |
| Handedness | - | Ambidextrous | Ambidextrous | Same |
| Physical Properties<br><br>Before Aging<br>Tensile Strength:<br>Ultimate Elongation:<br><br>After Aging Tensile<br>Strength:<br>Ultimate Elongation: | ASTM D6319 | 14Mpa, min<br>500% min<br><br>14Mpa, min<br>400% min | 27.8 – 33.1 Mpa<br>530 -600%<br><br>31.5 – 35.3 Mpa<br>470- 550% | Different but<br>within the ASTM<br>standard<br><br>Different but<br>within the ASTM<br>standard |
| Thickness:<br>- Finger<br>- Palm<br><br>Powder Free | ASTM D6319<br>ASTM D6124 | 0.05mm min<br>0.05mm min<br><br>< 2mg per glove | Min 0.10mm for (XS, S, M, L, XL)<br>Min 0.06mm for (XS, S, M, L, XL)<br><br>Below 2mg of residual powder | Different but<br>within the ASTM<br>standard<br><br>Similar |
| | STANDARDS | DEVICE PERFORMANCE | | |
| CHARACTERISTICS | | PREDICATE<br>BLACK | CURRENT<br>BLACK | COMPARISON<br>ANALYSIS |
| | Primary Skin Irritation<br>- ISO 10993-10:2010<br>(E) & Consumer<br>Product Safety<br>Commission Title 16.<br>Chapter II, Part 1500 | Passes | Under conditions of this study, the<br>test material did not cause an<br>irritant response. The Primary<br>Irritant Response Category is<br>deemed 'Negligible' | Same |
| Biocompatibility | Dermal Sensitization-<br>ISO 10993-10: 2010<br>(E) & Consumer Product<br>Safety Commission,<br>Title 16,Chapter II, Part<br>1500.3 (c) (4) | Passes | Under conditions of this study,the<br>test material did not produce a skin<br>sensitization effect in the guinea<br>pigs. | Same |
| | Cytotoxicity - MEM<br>Elution, ISO 10993-5:<br>2009 | Passes | The Systemic Toxicity study was<br>conducted as the accepted<br>alternative to the cytotoxic test | Different - but<br>additional test of Acute<br>Systemic Toxicity is<br>conducted and passed |
| | Acute Systemic<br>Toxicity, ISO 10993-<br>11:2017 (E) | Not Tested | Under conditions of this study,the<br>test item did not induce any<br>systemic toxicity in Swiss albino<br>mice. | Different. The<br>subject glove was<br>tested using<br>systemic toxicity test<br>and passed, but the<br>Predicate did not<br>have the test<br>performed |
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON ANALYSIS |
| | | PREDICATE<br>BLACK | CURRENT<br>BLACK | |
| Watertight (1000ml) | ASTM D5151:2019 | In accordance with ASTM D6319-10<br>and ASTM D5141-06 (Reapproved<br>2011), G-1, AQL 2.5 | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10<br>(Reapproved 2015)<br><br>Tested in accordance with ASTM<br>D5151 (Reapproved 2015) with<br>acceptable results<br>At an AQL 1.5 | Different, but within<br>ASTM standard; AQL<br>1.5 is more stringent<br>than AQL 2.5 |
| Intended use | - | The Nitrile Powder Free Patient<br>examination glove is a non-sterile<br>disposable device intended for<br>medical purpose that is worn on<br>the examiners hands or finger to<br>prevent contamination between<br>patient and examiner. | A Nitrile patient examination glove<br>is a disposable device made of<br>nitrile rubber intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination between<br>patient and examiner. | Similar |
| Size | Medical Glove<br>Guidance Manual -<br>Labeling | Extra Small<br>Small<br>Medium<br>Large<br>X Large | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Same |
| Single use | Medical Glove<br>Guidance Manual -<br>Labeling | Single Use | Single Use | Same |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
#### Summary of Non-Clinical Testing 7.0
The performance test data of the non-clinical tests for this powder free nitrile examination glove is summarized as per below.
| Test<br>Method | Standard | Purpose of Testing | Acceptance Criteria | | Results | | Status | |
|------------------------|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------|-----------------|-----------------|------------------------------------------------------------|------------------------------------------------------------|------|
| | | | Before aging | After aging | Before aging | After aging | | |
| Physical<br>Properties | ASTM D412<br>(Standard Test Method<br>for Vulcanized Rubber<br>and Thermoplastic<br>Elastomers-Tension) | To evaluate the<br>tensile (tension)<br>properties of glove. | Tensile<br>strength | Min 14.0<br>MPa | Min 14.0<br>MPa | XS – 28.9<br>S – 27.8<br>M – 28.7<br>L – 28.3<br>XL – 28.0 | XS – 31.7<br>S – 32.1<br>M – 32.4<br>L – 31.9<br>XL – 31.5 | Pass |
| | | | Ultimate<br>elongation | Min<br>500% | Min<br>400% | XS – 540<br>S – 530<br>M – 560<br>L – 540<br>XL – 540 | XS – 480<br>S – 470<br>M – 490<br>L – 480<br>XL – 490 | Pass |
{8}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Glove<br>Size | Acceptance Criteria | Results | Status |
|----------------|----------------------------------------------------------------------------------|----------------------------------------------------------|------------------------------------------|-------------------------------------------------------|---------------------------------|--------|
| Dimension | ASTM D3767<br>Standard<br>Practice for<br>Rubber—<br>Measurementof<br>Dimensions | To measure the<br>length, width and<br>thickness ofglove | X-Small | Length<br>Min 240 mm | Length<br>244 mm | Pass |
| | | | | Width<br>$70 \pm 10$ mm | Width<br>74.0 mm | Pass |
| | | | | Thickness<br>Finger – min 0.05mm<br>Palm – min 0.05mm | Thickness<br>0.10 mm<br>0.06 mm | Pass |
| | | | Small | Length<br>Min 240 mm | Length<br>245 mm | Pass |
| | | | | Width<br>$80 \pm 10$ mm | Width<br>85.0 mm | Pass |
| | | | | Thickness<br>Finger – min 0.05mm<br>Palm – min 0.05mm | Thickness<br>0.10 mm<br>0.06 mm | Pass |
| | | | Medium | Length<br>Min 240 mm | Length<br>244 mm | Pass |
| | | | | Width<br>$95 \pm 10$ mm | Width<br>95.0 mm | Pass |
| | | | | Thickness<br>Finger – min 0.05mm<br>Palm – min 0.05mm | Thickness<br>0.10 mm<br>0.06 mm | Pass |
| | | | Large | Length<br>Min 240 mm | Length<br>243 mm | Pass |
| | | | | Width<br>$110 \pm 10$ mm | Width<br>106 mm | Pass |
| | | Thickness | Finger – min 0.05mm<br>Palm - min 0.05mm | Thickness | 0.10 mm<br>0.06 mm | Pass |
| | | Length | Min 240 mm | Length | 242 mm | Pass |
| | X-Large | Width | $120 \pm 10$ mm | Width | 117 mm | Pass |
| | | Thickness | Finger - min 0.05mm<br>Palm - min 0.05mm | Thickness | 0.10 mm<br>0.06 mm | Pass |
{9}------------------------------------------------
{10}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|----------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Watertight | ASTM D5151<br>(Standard Test<br>Method for<br>Detection of Holesin<br>Medical Gloves) | To detect holes<br>that leak water and<br>thereby<br>compromise the<br>usefulness of the<br>glove. | Sample size: 315 pcs<br>Inspection level: G1<br>AQL: 1.5, Acceptance No. 10 | The batch size for this<br>sampling is 150,001 to<br>500,000. Hence,<br>according to the single<br>sampling plan GI, the<br>sample to be drawn is<br>under code M equivalent<br>to 315 pieces with accept<br>10 and reject 11 to be<br>accepted under AQL<br>1.5.<br>For Size XS, during the<br>test, 1piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size S, during the<br>test, 3piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size M. during the<br>test, 2piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size L, during the<br>test, 3piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria.<br>For Size XL, during the<br>test, 1piece was found<br>with leaks. Hence it falls<br>within the acceptance<br>criteria. | Pass |
{11}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | | Status |
|--------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------|
| Residual<br>Powder | ASTM D6124<br>(Standard Test Method<br>for Residual Powder on<br>Medical Gloves) | To determine the<br>amount of residual<br>and non-powder<br>solids found on<br>gloves | Less than 2 mg per glove<br>Requirement : <2mg/glove | Sample size<br>Result XS<br>Result S<br>Result M<br>Result L<br>Result XL | : 5 pcs<br>:0.26mg/glove<br>:0.32mg/glove<br>:0.28mg/glove<br>:0.34mg/glove<br>:0.38mg/glove | Pass |
{12}------------------------------------------------
#### 8.0 Non-clinical performance testing methods full titles:
- ASTM D412 Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— O Tension
- ASTM D573 Test Method for Rubber—Deterioration in an Air Oven x ASTM D3578 o Specification for Rubber Examination Gloves
- ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical o Application
- ASTM D5151 Test Method for Detection of Holes in Medical Gloves o
- ASTM D6124 Test Method for Residual Powder on Medical Gloves O
- ISO 2859 Sampling Procedures and Tables for Inspection by Attributes Test results show O that under the conditions of the testing, there is no difference inphysical attributes between the proposed device and the predicate device.
- ISO 10993 Part 10: Tests for Irritation and Sensitization. Both Skin Irritation and Dermal O Magnuson/Kligman Sensitization performed.
- o ISO 10993 - Part 11: Tests for assessment of Systemic Toxicity
#### 9.0 Summary of Clinical Testing
No clinical testing is included in this submission
#### 10.0 Conclusion
The conclusion drawn from the non-clinical tests demonstrates that the subject "Black Nitrile Powder-Free Patient Examination Glove, Non-Sterile" is as safe, as effective, and performs as well as, or better than the legally marketed predicate device predicate K190942.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.