Dermoaroma Disposable Nitrile Examination Glove

K212716 · Chuzhou Harmony Gloves Medical Technology Co.,Ltd · LZA · Dec 27, 2021 · General Hospital

Device Facts

Record IDK212716
Device NameDermoaroma Disposable Nitrile Examination Glove
ApplicantChuzhou Harmony Gloves Medical Technology Co.,Ltd
Product CodeLZA · General Hospital
Decision DateDec 27, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Dermoaroma Disposable Nitrile Examination Glove are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

Device Story

Dermoaroma Disposable Nitrile Examination Glove; blue, powder-free, non-sterile, ambidextrous nitrile glove. Used by healthcare examiners to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. No electronic, software, or algorithmic components. Performance verified via bench testing against ASTM standards for dimensions, physical properties (tensile strength, elongation), and freedom from holes. Biocompatibility assessed per ISO 10993 standards.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-5, -10, -11), residual powder (ASTM D6124), freedom from holes (ASTM D5151), and physical properties/dimensions (ASTM D6319).

Technological Characteristics

Material: Nitrile. Form factor: Ambidextrous, powder-free, non-sterile. Dimensions and physical properties comply with ASTM D6319. Biocompatibility testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity). No software, electronics, or energy source.

Indications for Use

Indicated for use as a disposable, non-sterile, ambidextrous, powder-free nitrile glove worn on the examiner's hands to prevent cross-contamination between patient and examiner in medical settings.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 27, 2021 ChuZhou Harmony Gloves Medical Technology Co.,Ltd Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room 608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China Re: K212716 Trade/Device Name: Dermoaroma Disposable Nitrile Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: November 19, 2021 Received: November 29, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K212716 #### Device Name Dermoaroma Disposable Nitrile Examination Glove Indications for Use (Describe) The Dermoaroma Disposable Nitrile Examination Gloveare disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92(a)(1). # 1.0 Submitter's information Name: ChuZhou Harmony Gloves Medical Technology Co.,Ltd Address: Technology Road 6, branching stream economic development zone, LaiAn Town,Chuzhou City ,AnHui Province,China Phone Number: +86-18689321491 Contact: WeiHong Ou Date of Preparation: December 7, 2021 # Designated Submission Correspondent Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn # 2.0 Device information Trade name: Dermoaroma Disposable Nitrile Examination Glove Patient Examination Gloves Common name: Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL ### 3.0 Classification Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital ### 4.0 Predicate device information Manufacturer: Jiangxi Surefine Medical Co., Ltd. Blue Nitrile Exam Gloves Device: 510(k) number: K211341 {4}------------------------------------------------ #### 5.0 Indications for Use The Dermoaroma Disposable Nitrile Examination Glove are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. #### 6.0 Device description The proposed device is Powder Free Dermoaroma Disposable Nitrile Examination Glove. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile. #### 7.0 Summary comparing technological characteristics with predicate device | Item | Proposed device | Predicated device | Remark | |--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | 510(k) number | K212716 | K211341 | | | Product Code | LZA | LZA | Same | | Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same | | Class | I | I | Same | | Indications for Use | The Dermoaroma Disposable Nitrile Examination Glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | The Blue Nitrile Exam Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same | | Powdered or Powered free | Powdered free | Powdered free | Same | | Design Feature | ambidextrous | ambidextrous | Same | | Labeling Information | Single-use indication, powder free, device color, device name, glove size and quantity, Dermoaroma Disposable Nitrile Examination Glove, Non-Sterile | Single-use indication, powder free, device color, device name, glove size and quantity, Blue Nitrile Exam Gloves, Non-Sterile | Same | #### Table1-General Comparison #### Table2 Device Dimensions Comparison | Predicate | Designation | Size | | | | Tolerance | |-----------------|-------------|------|---|---|----|-----------| | Device(K211341) | | S | M | L | XL | | {5}------------------------------------------------ | | Length, mm | 220 | 230 | 230 | 230 | min | |-----------------|----------------|------|-----|-----|-----|-----------| | | Width, mm | 80 | 95 | 110 | 120 | ±10 | | | Thickness, mm: | | | | | | | | Finger | 0.05 | | | | min | | | Palm | 0.05 | | | | min | | Proposed Device | Designation | Size | | | | Tolerance | | | | S | M | L | XL | | | | Length, mm | 220 | 230 | 230 | 230 | min | | | Width, mm | 80 | 95 | 110 | 120 | ±10 | | | Thickness, mm: | | | | | | | | Finger | 0.05 | | | | min | | | Palm | 0.05 | | | | min | | Remark | Same | | | | | | Analysis1: The proposed device has same sizes to the predicate device, and all proposed devices are conducted the properties test, the test results shown that the sizes comply with the requirements of standard ASTM D6319-19, Standard Specification For Dermoaroma Disposable Nitrile Examination Glove For Medical Application. | Item | | | Proposed device | Predicated device | Remark | | |------------------------|------------------------|------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------|----------------------|------| | Colorant | | | | blue | Blue | SAME | | Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | SAME | | | | | Ultimate<br>Elongation | 500%min | 500%min | SAME | | | | After<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | SAME | | | | | Ultimate<br>Elongation | 400%min | 400%min | SAME | | | | Comply with ASTM D6319 | | Comply with ASTM D6319 | Comply with ASTM D6319 | SAME | | | Freedom from Holes | | | Be free from holes<br>when tested in<br>accordance with<br>ASTMD5151<br>AQL=2.5 | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5 | SAME | | | | | | | | | | | | | | | | | | | | | | | | | | | Powder Content | | | 0.13mg -0.16 mg | 0.03 mg | SIMILAR<br>Analysis2 | | #### Table3 Performance Comparison Analysis2: The proposed device has different powder content to the predicate device, but the proposed device is conducted the biocompatibility and performance tests, the test results shown that the difference does not affect the safety and efficacy of proposed device. {6}------------------------------------------------ | Item | Proposed device | Predicated device | Remark | | |------------------|-------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------|---------| | Material | Nitrile | Nitrile | SAME | | | Biocompatibility | Irritation | Under the conditions of the study, not an irritant | Comply with ISO10993-10 | SAME | | | Sensitization | Under conditions of the study, not a sensitizer. | | | | | Cytotoxicity | Under the conditions of the study, the device is potentially cytotoxic | Comply with ISO10993-5 | SIMILAR | | | Systemic toxicity | Under the conditions of the study, the device does not elicit a systemic toxicity response in the model animal. | | | #### Biocompatibility Testing Comparison Table4 #### 8.0 Summary of Non-Clinical Performance Testing The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard. | No. | Name of the Test<br>Methodology / Standard | Purpose | Acceptance Criteria | Results | |-----|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | ISO 10993-10:2010<br>Biological Evaluation Of<br>Medical Devices - Part<br>10: Tests For Irritation<br>And Skin Sensitization. | This part of ISO<br>10993 assesses<br>possible contact<br>hazards from<br>chemicals<br>released from<br>medical devices, | Skin Sensitization<br>Test:<br>provided grades less than 1,<br>otherwise<br>sensitization. | All grades are 0.<br><br>All animals were survived and no<br>abnormal signs were observed<br>during the study. | | 2 | | which may<br>produce skin and<br>mucosal irritation,<br>eye irritation or<br>skin sensitization. | Skin Irritation Test:<br>If the primary irritation<br>index is 0-0,4, the<br>response category is<br>Negligible.<br>0,5-1,9 means slight<br>2-4,9 means moderate<br>5-8 means severe | The primary irritation index is 0.<br>The response of the proposed<br>device was categorized as<br>negligible under the test<br>condition | | | | | | | | 3 | ISO 10993-5:2009<br>Biological Evaluation Of<br>Medical Devices - Part<br>5: Tests For In Vitro<br>Cytotoxicity | This part of ISO<br>10993 describes<br>test methods to<br>assess the in vitro<br>cytotoxicity of<br>medical devices. | The viab.% of the<br>100% extract of the<br>test article is the final<br>result, and if viability is<br>reduced to <70% of<br>the blank, it has<br>cytotoxic potential. | Viab.% of 100% test article<br>extract is 20.8%<br>It means the proposed device<br>have potential toxicity to L-929 in<br>the MTT method | | 4 | ISO 10993-11: 2017<br>Biological evaluation of<br>medical devices — Part<br>11: Tests for systemic<br>toxicity | To evaluate the<br>potential for<br>medical device<br>materials to cause<br>adverse systemic<br>reactions. | Within the monitoring<br>period (72 h), if the<br>toxicosis response of<br>testing group is not<br>greater than that of<br>control group, the<br>testing sample is<br>regarded as<br>acceptable. | No toxicosis response in testing<br>group.<br>It means the test<br>article has no potential acute<br>system toxicity on ICR mice in<br>the extraction method. | | 5 | ASTM D6124-06<br>(Reapproved 2017),<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves | This standard is<br>designed to<br>determine the<br>amount ofresidual<br>powder (or<br>filter-retained<br>mass) found<br>on medical gloves | powder residue limit of<br>2.0 mg | 0.13mg -0.16 mg /glove | | 6 | ASTM<br>D5151-06(Reapproved2<br>015), Standard Test<br>Method for Detection of<br>Holes in Medical Gloves. | This test method<br>covers the<br>detection of holes<br>in<br>medical gloves. | Samples number: 125<br>gloves<br>AQL: 2.5 (ISO 2859)<br>Criterion ≤7 gloves<br>for water leakage | no glove water leakage found | | 7 | ASTM<br>D6319-10(Reapproved<br>2015),Standard<br>Specification For Nitrile<br>Examination Gloves For<br>Medical Application. | This specification<br>covers certain<br>requirements for<br>nitrile rubber<br>gloves used in<br>conducting<br>medical<br>examinations and<br>diagnostic and<br>therapeutic<br>procedures. | Sterility: no need<br>Freedom from holes:<br>pl. Refer to No. 5 in<br>table 5<br>Dimensions:<br>S: width 80±10mm<br>Length ≥220 mm<br>M: width 95±10mm<br>Length ≥230 mm<br>L: width 110±10mm<br>Length ≥230 mm | N.A.<br>Please refer to No. 5 in table 5<br>Lot no.:202100505<br>Dimensions:<br>S: width: 80-87 mm<br>Length 240-246 mm<br>M: width 93-98 mm<br>Length 240-245 mm<br>L: width 103-110 mm<br>Length 249-252 mm<br>XL: width 118-123 mm | #### Table 5 Summary of Non-Clinical Performance Testing {7}------------------------------------------------ {8}------------------------------------------------ | | XL: width 120±10mm | Length 265-270 mm | | |--|----------------------------------------------------|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | | Length ≥230 mm | Thickness: | | | | Thickness: | Finger 0.10-0.12 mm | | | | Finger ≥0.05 mm | Palm 0.07-0.10 mm | | | | Palm ≥0.05 mm | | | | | | | | | | Physical properties: | Physical properties: | | | | Before aging | Before aging | | | | Tensile strength ≥ 14MPa | Tensile strength 15.3-19.8 MPa | | | | Ultimate Elongation ≥ 500% | Ultimate Elongation 508.480% - 669.787% | | | | After Accelerated Aging | After Accelerated Aging | | | | Tensile strength ≥ 14MPa | Tensile strength 14.3-15.2MPa | | | | Ultimate Elongation ≥ 400% | Ultimate Elongation 435.176% - 542.968% | | | | | Powder-free Residue: pl. Refer to No. 4 in table 5 | | | | Powder-free Residue: pl. Refer to No. 4 in table 5 | Lot no.:202100520 | | | | | Dimensions: | | | | | S: width: 80-85 mm<br>Length 245-251 mm | | | | | M: width 89-96 mm<br>Length 240-245 mm | | | | | L: width 105-112 mm<br>Length 249-253 mm | | | | | XL: width 118-125 mm<br>Length 259-264 mm | | | | | Thickness: | | | | | Finger 0.10-0.12 mm | | | | | Palm 0.08-0.10 mm | | | | | | | | | | Physical properties: | | | | | Before aging | | | | | Tensile strength 14.9-19.8 MPa | | | | | Ultimate Elongation 508.480% - 669.787% | | | | | After Accelerated Aging | | | | | Tensile strength 14.3-15.2MPa | | | | | Ultimate Elongation 503.493% - 644.640% | | | | | | | | | | | Powder-free Residue:<br>pl. Refer to No. 4 in table 5 | | | | | Lot no.:202100601 | | | | | Dimensions: | | | | | S: width: 82-90 mm<br>Length 241-245 mm | | | | | M: width 95-103 mm<br>Length 240-245 mm | | | | | L: width 103-112 mm<br>Length 248-255 mm | | | | | XL: width 114-120 mm<br>Length 258-265 mm | | | | | Thickness:<br>Finger 0.09-0.12 mm<br>Palm 0.08-0.09 mm | | | | | Physical properties:<br>Before aging<br>Tensile strength 15.1-19.9 MPa<br>Ultimate Elongation 521.280% -<br>665.453% | | | | | After Accelerated Aging<br>Tensile strength 14.1-15.9MPa<br>Ultimate Elongation 430.518% -<br>513.693% | | | | | Powder-free Residue:<br>pl. Refer to No. 5 in table 5 | {9}------------------------------------------------ ### 9. Summary of Clinical Performance Test No clinical study is included in this submission. ### 10.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...