K210145 · Changzhou Universal Medical Equipment Co., Ltd. · LZA · Apr 21, 2021 · General Hospital
Device Facts
Record ID
K210145
Device Name
Nitrile Glove Powder Free Blue
Applicant
Changzhou Universal Medical Equipment Co., Ltd.
Product Code
LZA · General Hospital
Decision Date
Apr 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Nitrile Glove Powder Free Blue is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
Device Story
Nitrile Glove Powder Free Blue is a non-sterile, powder-free, disposable examination glove. Worn by healthcare examiners to provide a protective barrier against potentially infectious materials and contaminants. Ambidextrous design. Used in clinical settings. Provides protection by preventing cross-contamination between patient and examiner. No electronic or mechanical components.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -10, -11) and physical performance testing (ASTM D6319, EN 455-1, EN 455-2).
Technological Characteristics
Nitrile rubber construction. Ambidextrous, powder-free. Dimensions: 230 mm length, 85-115 mm width, 0.05 mm thickness. Complies with ASTM D6319 for physical dimensions, tensile strength (≥ 14 MPa), and elongation (≥ 400%). Biocompatibility verified per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity). Non-sterile.
Indications for Use
Indicated for use as a disposable barrier on the examiner's hands to prevent contamination between patient and examiner in a medical setting.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
K142588 — Nitrile Powder Free Examination Glove (Blue) · Hartalega Sdn Bhd · Dec 18, 2014
K221192 — Disposable Nitrile Powder-Free Examination Gloves · Shandong Maida Medical Technology Co., Ltd. · Jun 22, 2022
K102593 — SYNTHETIC, POWDER-FREE NITRILE EXAMINATION · Zibo Yinghao Medical Products Co., Ltd. · Dec 20, 2010
K213859 — Stronghold Group Nitrile Examination Glove, Powder Free · Stronghold Group, LLC · Jan 13, 2022
K210057 — Blue Nitrile Exam Glove, Powder Free · Real Star Medical Technology Co., Ltd. · Mar 31, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 21, 2021
Changzhou Universal Medical Equipment CO. LTD. % Roxanne Dubois Regulatory Consultant R. Dubois Consulting, LLC 1399 Robnick Ct. Campbell, California 95008
Re: K210145
Trade/Device Name: Nitrile Glove Powder Free Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: March 31, 2021 Received: April 5, 2021
Dear Roxanne Dubois:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210145
Device Name
Nitrile Glove Powder Free Blue
#### Indications for Use (Describe)
The Nittile Glove Powder Free Blue is a disposable device intended purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Changzhou Universal Medical Equipment Co., Ltd. The logo consists of a blue circular emblem with stars around the perimeter and Chinese characters inside. To the right of the emblem is the company name in both Chinese and English. The English translation reads "Changzhou Universal Medical Equipment Co., Ltd."
## 510(k) Summary - K210145
### 1.0 Submitter's Information:
| Name: | Changzhou Universal Medical Equipment Co. Ltd |
|----------------------|-----------------------------------------------|
| Address: | No 6, Xinxi Road, Xinbei District |
| | Changzhou, Jiangsu Province, 213000 China |
| Tel: | +86-519-85483888 |
| Date of Preparation: | January 12, 2021 |
| Contact Person: | Roxanne Dubois |
| Address: | R. Dubois Consulting, LLC |
| | 1399 Robnick Ct., Campbell, CA 95008 |
| Tel: | 408-828-5019 |
| Email: | rduboisconsulting@gmail.com |
#### 2.0 Device Information
| Trade name: | Nitrile Glove Powder Free Blue |
|------------------|----------------------------------------|
| Common name: | Patient Examination Gloves |
| Regulation name: | Non-powdered patient examination glove |
#### 3.0 Classification
| Production code: | LZA |
|--------------------|------------------|
| Regulation number: | 21 CFR 880.6250 |
| Classification: | Class I |
| Panel: | General Hospital |
## 4.0 Predicate Device Information
| Manufacturer: | Haining Medical Products Co. Ltd |
|----------------|----------------------------------|
| Device: | Nitrile Glove Powder Free White |
| 510(k) number: | K183068 |
#### 5.0 Indications for Use
The Nitrile Glove Powder Free Blue is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
#### 6.0 Device Description
The proposed device is powder free, non-sterile nitrile examination gloves.
Following FDA Code LZA, a "nitrile (or polymer) patient examination glove is a disposable device made of nitrile rubber or synthetic polymers that may or may not bear a trace amount of residual powder, and is intended to be worn on the hand for medical purposes to provide a barrier against potentially infectious materials and other contaminants."
{4}------------------------------------------------
# 7.0 Technological Characteristic Comparison Table
| Table 1 General Comparison |
|----------------------------|
|----------------------------|
| Topic | Proposed Device | Predicate Device<br>(K183068) | Comparison |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code | LZA | LZA | Same |
| Regulation No | 21 CFR 880.6250<br>Non-powdered patient<br>examination glove | 21 CFR 880.6250<br>Non-powdered patient<br>examination glove | Same |
| Identification | A non-powdered patient<br>examination glove is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between patient<br>and examiner. | A non-powdered patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or finger<br>to prevent<br>contamination between<br>patient and examiner. | Same |
| Class | Class I reserved medical<br>devices that require a 510(k)<br>premarket notification | Class I reserved medical devices<br>that require a 510(k) premarket<br>notification. | Same |
| Indications For Use | The Nitrile Glove Powder<br>Free Blue is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hands to prevent<br>contamination<br>between patient and examiner. | The Nitrile Glove Powder Free<br>Blue is a disposable device<br>intended for medical purposes<br>that is worn on the examiner's<br>hands to prevent contamination<br>between patient and examiner. | Same |
| Powdered or Powder<br>Free | Non-powdered | Non-powdered | Same |
| Design Feature | ambidextrous | ambidextrous | Same |
| Labeling Information | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Nitrile Glove<br>Powder-Free, Non Sterile | Single-use indication, powder<br>free, device color, device<br>name, glove size and quantity,<br>Nitrile Glove Powder-Free,<br>Non Sterile | Same |
| Dimensions | Length:<br>230 mm<br>Width:<br>85-115 mm<br>Thickness:<br>0.05 mm | Length:<br>230 mm<br>Width:<br>85-115 mm<br>Thickness: 0.05 mm | Same |
| Colorant | Blue | White | Different |
# 8.0 Discussion of Non-clinical and Performance Testing
Non-clinical tests were conducted to verify that the proposed device meets all design specifications, the FDA-recognized consensus standard, proposed labeling claims and pinhole acceptable quality level (AQL). The relevant standards and test results are provided below.
{5}------------------------------------------------
Table 2 Relevant Standards and Test Results
| Test Methodology | Purpose | Acceptance Criteria | Results |
|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| ISO 10993-5 Biological<br>evaluation of medical<br>devices Test for <i>in vitro</i><br>cytotoxicity | To determine the<br>potential<br>cytotoxicity. | Non-cytotoxic | PASS |
| ISO 10993-11 Biological<br>evaluation of medical<br>devices Part 11: Tests for<br>Systemic Toxicity | To determine the<br>potential systemic<br>toxicity. | Does not cause systemic toxicity | PASS |
| ISO 10993-10 Biological<br>evaluation on medical<br>device Part 10: Test for<br>Irritation and Skin<br>Irritation | To determine the<br>potential for<br>irritation and skin<br>irritation. | Non-irritant, and<br>Non skin irritant | PASS |
| ISO 10993-10 Biological<br>evaluation on medical<br>device Part 10: Test Skin<br>Sensitization | To determine the<br>potential skin<br>sensitization. | Does not cause skin sensitization | PASS |
| ASTM D6319, Standard<br>Specification for Nitrile<br>Examination Gloves for<br>Medical Application | To test for:<br>(1) freedom from<br>holes<br>(2) physical<br>dimensions<br>(3) Aging | (1) Shall not leak<br>(2) For size M (mm):<br>Width: 110 +/- 10<br>Length: $≥$ 230<br>Finger and palm thickness:<br>median value $≥$ 0.05<br>(3) After Aging:<br>Tensile Strength: $≥$ 14 MPa<br>Ultimate elongation: $≥$ 400% | PASS |
| EN 455-1, EN 455-2,<br>Medical Gloves For<br>Single Use | To test for:<br>(1) freedom from<br>holes<br>(2) dimensions and<br>strength | (1) Shall not leak<br>(2) For size M (mm):<br>Width: 110<br>Length: $≥$ 240<br>Strength:<br>Force at Break: $≥$ 6N<br>Force at break after challenge<br>testing $≥$ 6N at 7 days | PASS |
# 9.0 Discussion of Clinical Testing
Clinical testing was not required to support this device.
#### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.