Powder Free Blue Nitrile Examination Gloves Tested for use with 32 Chemotherapy Drugs and Fentanyl Citrate
K183354 · Gx Corporation Sdn Bhd · LZA · Jun 20, 2019 · General Hospital
Device Facts
Record ID
K183354
Device Name
Powder Free Blue Nitrile Examination Gloves Tested for use with 32 Chemotherapy Drugs and Fentanyl Citrate
Applicant
Gx Corporation Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Jun 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.
Device Story
Disposable blue nitrile examination gloves; worn by healthcare personnel to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for permeation resistance against 32 chemotherapy drugs and Fentanyl Citrate. Breakthrough detection times provided for specific agents (e.g., Carmustine, Thio-Tepa). Used in clinical settings; protects clinicians during handling of hazardous drugs.
Clinical Evidence
Bench testing only. Permeation testing performed to determine breakthrough detection times for 32 chemotherapy drugs and Fentanyl Citrate per ASTM standards.
Technological Characteristics
Material: Nitrile. Form factor: Powder-free blue examination glove. Principle: Physical barrier protection. Sterilization: Not specified. Connectivity: None.
Indications for Use
Indicated for use as a disposable patient examination glove worn by examiners to prevent cross-contamination between patient and examiner. Tested for permeation resistance against 32 specific chemotherapy drugs and Fentanyl Citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K140816 — POWDER FREE NITRILE EXAMINATION GLOVE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS (BLUE) · Better Care Plastic Technology Co., Ltd. · Jul 8, 2014
K222129 — Nitrile Examination Powder Free Glove, Blue Tested For Use With 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim, Nitrile Examination Powder Free Glove, Black Tested For Use With 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim · Top Glove Medical (Thailand) Co., Ltd. · Feb 10, 2023
K222815 — Powder Free Nitrile Examination Gloves (Blue, White and Black), Tested For Use With Chemotherapy Drugs and Fentanyl · Zibo Lanhua Medical Packaging Material Co., Ltd. · Dec 4, 2022
K223375 — Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-sterile, Low Dermatitis Potential. Tested for Use with Chemotherapy Drugs and Fentanyl Citrate. · Kossan International Sdn Bhd3 · Mar 15, 2023
K211810 — Shamrock Powder Free Blue Nitrile Examination Gloves (Tested for use with Chemotherapy Drugs and Fentanyl) · Pt. Shamrock Manufacturing Corpora · May 23, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 20, 2019
GX Corporation SDN BHD Foo Khon Pu Managing Director LOT 6491, Batu 5 3/4, Sementa, Jalan Kapar Klang, 42100 My
Re: K183354
Trade/Device Name: Powder Free Blue Nitrile Examination Gloves - Tested for use with 32 Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZC, LZA, ODO Dated: May 13, 2019 Received: May 20, 2019
Dear Foo Khon Pu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie-Williams, M.S. Assistant Director THT4B2: Disinfection, Reprocessing and Personal Protection Acting Assistant Director,THT4B1: Sterility Devices DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183354
#### Device Name
Powder Free Blue Nitrile Examination Gloves - Tested for Use with 32 Chemotherapy Drugs and Fentanyl Citrate
### Indications for Use (Describe)
A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs as listed below.
Chemotherapy Drug permeation and concentration ( Breakthrough Detection Time) in Minutes. The following chemicals have been tested with these gloves.
| Chemotherapy Drugs: | Breakthrough Detection Time in Minutes |
|--------------------------------------------------------|----------------------------------------|
| 1. Carmustine (BCNU)3.3 mg/ml(3,300 ppm) | 54.8 min |
| 2. Cisplatin 1.0 mg/ml (1,000ppm) | >240 min |
| 3. Cyclophosphamide (Cytoxan) 20mg/ml(20,000ppm) | >240 min |
| 4. Dacarbazine (DTIC)10.0 mg/ml(10,000ppm) | >240 min |
| 5. Doxorubicin Hydrochloride 2.0 mg/ml (2,000ppm) | >240 min |
| 6. Etoposide (Toposar) 20.0 mg/ml (20,000ppm) | >240 min |
| 7. Fluorouracil 50.0 mg/ml (50,000ppm) | >240 min |
| 8. Ifosfamide 50.0 mg/ml (50,000ppm) | >240 min |
| 9. Methotrexate 25 mg/ml (25,000ppm) | >240 min |
| 10.Mitomycin C 0.5 mg/ml (500ppm) | >240 min |
| 11. Mitoxantrone 2.0 mg/ml (2,000ppm) | >240 min |
| 12. Paclitaxel (Taxol) 6.0 mg/ml (6,000ppm) | >240 min |
| 13. Thio-Tepa 10.0 mg/ml (10,000ppm) | 46.9 min |
| 14.Vincristine Sulfate 1.0 mg/ml (1,000ppm) | >240 min |
| 15. Amethopterin (Methotrexate) 25 mg/ml, (25,000 ppm) | >240 min |
| 16. Bleomycin 15 mg/ml (15,000 ppm) | >240 min |
| 17. Busulfan 6 mg/ml (6,000 ppm) | >240 min |
| 18. Carboplatin 10 mg/ml (10,000 ppm) | >240 min |
| 19. Cytarabine 100 mg/ml (100.000 ppm) | >240 min |
| 20. Daunorubicin 5 mg/ml (5,000 ppm) | >240 min |
| 21. Doxetacel 10.0 mg/ml (10,000 ppm) | >240 min |
| 22. Ellence 2 mg/ml (2,000 ppm) | >240 min |
| 23. Fludarabine 25 mg/ml (25,000 ppm) | >240 min |
| 24. Gemcitabine (Gemzar) 38 mg/ml (38,000 ppm) | >240 min |
| 25. Idarubicin 1 mg/ml (1,000 ppm) | >240 min |
| 26. Irinotecan 20.0 mg/ml (20,000 ppm) | >240 min |
| 27. Mechloretamine HCL 1.0 mg/ml (1,000 ppm) | >240 min |
| 28. Melphalan 5 mg/ml (5,000 ppm) | >240 min |
| 29. Oxaliplatin 2.0 mg/ml (2,000 ppm) | >240 min |
| 30. Rituximab 10 mg/ml (10,000 ppm) | >240 min |
| 31. Trisonex 0.1 mg/ml (100 ppm) | >240 min |
| 32. Vinorelbin 10 mg/ml (10,000 ppm) | >240 min |
| plus Other Drugs | |
| 1. Fentanyl Citrate Injection 100 mcg/2ml | >240min |
| FORM FDA 3881 (7/17) | Page 1 of 2 |
{3}------------------------------------------------
Please Note that the following Drugs have extremely low permeation times : Carmustine : 54,8 minutes and Thio-Tepa: 46.9 minutes.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Z Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.