POWDER FREE NITRILE EXAMINATION GLOVES (BLUE, WIHTE)

K070862 · Purtech Cleanroom Products Co., Ltd. · LZA · May 30, 2007 · General Hospital

Device Facts

Record IDK070862
Device NamePOWDER FREE NITRILE EXAMINATION GLOVES (BLUE, WIHTE)
ApplicantPurtech Cleanroom Products Co., Ltd.
Product CodeLZA · General Hospital
Decision DateMay 30, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Disposable nitrile examination glove; worn by healthcare examiners on hands or fingers; acts as physical barrier to prevent cross-contamination between patient and examiner. Manufactured from nitrile material; powder-free. Used in clinical or medical settings. Provides protection during patient examinations.

Clinical Evidence

No clinical data. Bench testing only: physical dimensions, water leak test (AQL 2.5), biocompatibility (skin irritation/sensitization), and residual powder content (ASTM D6124-06).

Technological Characteristics

Nitrile material; powder-free (max 2 mg/glove). Complies with ASTM D6319-00a(2005). Dimensions and physical properties tested per ASTM standards. Biocompatible. Non-sterile.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Suitable for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAY 3 0 2007 EXHIBIT #1 Page 1 of 2 #### 510(K) SUMMARY This summary of 510(K) safety and cffectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(K) number is: _ 070 862 #### l. Submitter's Identification: Mr. Zhaolong Wang Purtech Cleanroom Products Co., Ltd. Fanjiazhuang Industrial Zonc Xinji City, Hebei Province, P.R. China 052360 Date Summary Prepared: March 22, 2007 ### 2. Name of the Device: Purtech Cleanroom Products Co., Ltd. Powder Free Nitrile Examination Gloves (Blue, White) ### 3. Predicate Device Information: Shanghai Poseidon Plastic Products Co., Ltd. Nitrile Exam Gloves, Powder-Free, Color Blue (K001128) #### 4. Device Description: Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power Free Nitrile Examination Gloves (Blue, White), 80 LZA, and meets all requirements of ASTM standard D 6319-00a(2005). #### 5. Intended Use: A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. {1}------------------------------------------------ ### 6. Comparison to Predicate Devices: Purtech Cleanroom Products Co., Ltd.'s Powder Free Nitrile Examination Gloves (Blue, White), is substantially equivalent in safety and effectiveness to the Shanghai Poseidon Plastic Products Co., Ltd.'s Nitrile Exam Gloves, Powder-Free, Color Blue, ### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows: The standards used for Purtech Cleanroom Products Co., Ltd.'s glove production are based on ASTM D-6319-00a(2005). All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, mecting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. A Residual Powder Test that based on ASTM D 6124-06 for Starch at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove). ### 8. Labeling: There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. ### 9. Discussion of Clinical Tests Performed: Not Applicable - There is no hypoallergenic Claim. ### 10. Conclusions: Purtech Cleanroom Products Co., Ltd.'s Powder Free Nitrile Examination Gloyes (Blue, White) conform fully to ASTM D6319-00a(2005) standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, representing health and medicine, with three figures intertwined, symbolizing the relationship between health, services, and people. The seal is in black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 3 0 2007 Purtech Cleanroom Products Company, Limited C/O Ms. Kathy Liu Official Correspondent Gloveco, Incorporated 3973 Schaefer Avenue Chino, California 91710 Re: K070862 Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue, White) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: May 15, 2007 Received: May 16, 2007 Dear Ms. Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the gencral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 -- Ms. Liu Please be advised that FDA's issuance of a substantial cquivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requircments as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Chiu-Ling, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ Page 1 of 1 ## INDICATION FOR USE 510(K) NUMBER APPLICANT: DEVICE NAME: (IF KNOWN): K070862 Purtech Cleanroom Products Co., Ltd. Powder Free Nitrile Examination Gloves (Blue, White) INDICATIONS FOR USE: A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter-Use V (Optional Format 1-2-96) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) | concurrent of CDRH, Office of Device Evaluation (ODE) | |-------------------------------------------------------| | <i>(Sign-Off)</i> | | Division of Anesthesiology, General Hospital, | | Location Control, Dental Devices | (k) Number ___________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...