SYNTHETIC POWDERED VINYL PATIENT EXAMINATION GLOVES

K992554 · Shijiazhuang Great Vision Plastic Products Co., LT · LYZ · Mar 29, 2000 · General Hospital

Device Facts

Record IDK992554
Device NameSYNTHETIC POWDERED VINYL PATIENT EXAMINATION GLOVES
ApplicantShijiazhuang Great Vision Plastic Products Co., LT
Product CodeLYZ · General Hospital
Decision DateMar 29, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Story

Synthetic powdered vinyl patient examination gloves; disposable; worn by healthcare personnel on hands/fingers. Function: barrier protection against contamination between patient and examiner (fluids, waste, environment). Used in clinical settings. Device conforms to ASTM D5250-92 standards for physical dimensions and integrity.

Clinical Evidence

No clinical data. Bench testing only: physical/dimensional testing (ASTM D5250-92, AQL 4.0), water fill test (AQL 2.5), and biocompatibility (primary skin irritation and sensitization).

Technological Characteristics

Material: Synthetic vinyl. Standard: ASTM D5250-92. Form factor: Disposable powdered patient examination glove. Sterilization: Not specified. Energy source: None. Connectivity: None.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: _K 9925334 ### 1. Submitter's Identification: Ms. Fang Yi Liu Shijiazhuang Great Vision Plastics Products Co., Ltd. Cangshi Road (S) Jinzhou City, Hebei Province P.R. China Date Summary Prepared: July 14, 1999 ### 2. Name of the Device: Shijiazhuang Great Vision Plastics Products Co., Ltd. Synthetic Powdered Vinyl Patient Examination Gloves ### 3. Predicate Device Information: Sunmax Enterprise Shanghai Co., Ltd. Powdered Vinyl Patient Examination Gloves, K#960746 Shanghai Super Gloves Company, Ltd. Powdered Vinyl Patient Examination Gloves, K#974151 ### 4. Device Description: Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 800.6250. Powdered Vinyl Patient Examination Glove, 80LYZ, Powdered with Absorbable Dusting Powder, USP, Class III and meets all requirements of ASTM Standard D5250-92. ### 5. Intended Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment. ### 6. Comparison to Predicate Devices: {1}------------------------------------------------ Shiijazhuang Great VisionPlastics Products Co., Ltd. Synthetic Powdered Vinyl Patient Examination Gloves, is substantially equivalent in safety and effectiveness to the Sunmax Enterprise Shanghai Co., Ltd. Powdered Vinyl Patient Examination Gloves and the Shanghai Super Gloves Company, Ltd. Powder Vinyl Patient Examination Gloves. #### 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows: The standards used for Shijiazhuang Great Vision Plastics Products Co., Ltd. glove production are based on ASTM-D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5. Inspection Level S-4, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. #### Discussion of Clinical Tests Performed: 8. Not Applicable - There is no hypoallergenic Claim. #### 9. Conclusions: Shijiazhuang Great Vision Plastics Co., Ltd. Synthetic Vinyl Patient Examination gloves (Powdered) conform fully to ASTM-D-5250-92 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image. The logo is simple and recognizable, representing the department's role in health and human services. MAR 2 9 2000 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Shijiazhuang Great Vision Plastics Products Co., Ltd. C/O Ms. Karen Abell Basic Medical Industries, Inc. 13945 Magnolia Avenue Chino, California 91710 Re : K992554 Synthetic Powdered Vinyl Patient Examination Trade Name: Gloves Requlatory Class: I Product Code: LYZ February 28, 2000 Dated: February 28, 2000 Received: Dear Ms. Abell: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your {3}------------------------------------------------ Page 2 -Ms. Abell premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labelinq regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4690. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html". Sincerely yours Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Attachment A Page 1 of 1 510(k) NUMBER (IF KNOWN) K ag 25524 Shijiazhuang Great Vision Plastics Products Co., Ltd. DEVICE NAME: INDICATIONS FOR USE: Synthetic Powdered Vinyl Patient Examination Gloves-Powdered A patient examination glove is disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) | | Concurrent of CDRH, Office of Device Evaluation (ODE) | |--|-------------------------------------------------------| |--|-------------------------------------------------------| | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter-Use<br>(Optional Format 1-2-96) | <div style="text-align:right;">✓</div> | |------------------------------------------|----|--------------------------------------------------|----------------------------------------| |------------------------------------------|----|--------------------------------------------------|----------------------------------------| | | <i>Barito for Clin</i><br>(Division Sign-Off) | |---------------|------------------------------------------------------------------------| | | Division of Dental, Infection Control,<br>and General Hospital Devices | | 510(k) Number | K99 2554 |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...