GRAND WORK PLASTICS PRODUCTS CO., LTD., SYNTHETIC VINYL PATIENT EXAMINATION GLOVES - POWDERED

K041188 · Grand Work Plastic Products Co., Ltd. · LYZ · Jun 1, 2004 · General Hospital

Device Facts

Record IDK041188
Device NameGRAND WORK PLASTICS PRODUCTS CO., LTD., SYNTHETIC VINYL PATIENT EXAMINATION GLOVES - POWDERED
ApplicantGrand Work Plastic Products Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateJun 1, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Story

Synthetic vinyl patient examination glove; powdered; disposable. Worn by healthcare personnel on hands/fingers during patient examinations. Acts as physical barrier to prevent cross-contamination between examiner and patient body, fluids, waste, or environment. Manufactured to meet ASTM D5250-00E4 standards for physical dimensions and integrity. Used in clinical settings to protect both patient and provider.

Clinical Evidence

No clinical data. Bench testing only. Testing included physical dimensions, water fill leak testing (ASTM D-5151-99, AQL 2.5), residual powder testing (ASTM D6124-01), and biocompatibility (primary skin irritation and sensitization).

Technological Characteristics

Synthetic vinyl material. Powdered. Meets ASTM D5250-00E4 standards. Physical testing conducted at Inspection Level S-2, AQL 4.0. Water fill testing per ASTM D-5151-99 at AQL 2.5. Residual starch powder content ≤ 10 mg/dm2 per ASTM D6124-01.

Indications for Use

Indicated for use by healthcare and similar personnel as a disposable barrier worn on the hand or finger to prevent contamination between the patient and the examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JUN - 1 2004 ## 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: _| (04 1188 #### 1. Submitter's Identification: Mr. Shuangzu Ma Grand Work Plastics Products Co., Ltd. Donggao Industrial Zone Zanhuang, Hebei Province P.R. China Date Summary Prepared: March 01, 2004 #### 2. Name of the Device: Grand Work Plastics Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves—Powdered ## 3. Predicate Device Information: Shijiazhuang Great Vision Plastic Products Co., Ltd. Synthetic Powdered Vinyl Patient Examination Gloves (K992554) #### 4. Device Description: Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powdered Vinyl Patient Examination Glove, 80L YZ, and meets all requirements of ASTM Standard D5250-00E4. #### 5. Intended Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment. {1}------------------------------------------------ ### 6. Comparison to Predicate Devices: Grand Work Plastics Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves-Powdered, is substantially equivalent in safety and effectiveness to the Shijiazhuang Vision Plastic Products Co., Ltd. Synthetic Powdered Vinyl Patient Examination Gloves. ## 7. Discussion of Non-Clinical tests Performed for Determination of Substantial Equivalence are as follows: The standards used for Grand Work Plastics Products Co., Ltd. glove production are based on ASTM-D-5250-00E4. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test based on ASTM D-5151-99 was also conducted with samplings of AQL 2.5, Inspection Level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. A Residual Powder Test that based on ASTM D6124-01 for Starch at finished inspection is conducted to insure that our gloves meet our "powdered' claims (contain no more thon 10 mg/dm2). ### 8. Discussion of Clinical Tests Performed: Not Applicable - There is no hypoallergenic Claim. ### 9. Conclusions: Grand Work Plastics Co., Ltd. Synthetic Vinyl Patient Examination gloves-Powdered conform fully to ASTM-D-5250-00E4 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 1 2004 Grand Work Plastics Products Company Limited C/O Mr. James Chu Official Correspondent Gloveco, Incorporated 590 W. Central Avenue, #D Brea, California 92821 Re: K041188 Trade/Device Name: Grand Work Plastics Products Company Limited Synthetic Vinyl Patient Examination Glove--Powdered Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: May 5, 2004 Received: May 6, 2004 Dear Mr. Chu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Chu Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Attachment A Page 1 of 1 (IF KNOWN) : _ K 0 4 ) | 1 8 8 = 510(k) NUMBER Grand Work Plastics Products Co., Ltd. DEVICE NAME: INDICATIONS FOR USE: Synthetic Vinyl Patient Examination Gloves--Powdered A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ke Muley (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K041188
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...