Palm Care Latex Examination Powder Free Gloves

K202377 · Hi-Care Thai Gloves Co. , Ltd. · LYY · Dec 1, 2020 · General Hospital

Device Facts

Record IDK202377
Device NamePalm Care Latex Examination Powder Free Gloves
ApplicantHi-Care Thai Gloves Co. , Ltd.
Product CodeLYY · General Hospital
Decision DateDec 1, 2020
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Disposable natural rubber latex examination gloves; powder-free; non-sterile. Worn by clinicians to provide barrier protection against cross-contamination between patient and examiner. Manufactured to meet ASTM D3578-05 standards. Device is single-use; intended for medical examination purposes.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including physical property testing (tensile strength, elongation), barrier integrity (ASTM D5151), residual powder content (ASTM D6124), protein content (ASTM D5712), and biocompatibility (ISO 10993).

Technological Characteristics

Material: Natural Rubber Latex. Form factor: Powder-free, finger-textured, non-sterile. Dimensions and physical properties comply with ASTM D3578-05 (Reapproved 2015). Biocompatibility verified per ISO 10993-5, -10, and -11. Residual powder < 2 mg/glove. Extractable protein < 50 µg/dm².

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner. Suitable for use by healthcare professionals and in general medical settings.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration." December 1, 2020 Hi-Care Thai Gloves Co. Ltd. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr. STE 114 Aurora, Illinois 60504 Re: K202377 Trade/Device Name: Latex Examination Powder Free Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LYY Dated: October 31, 2020 Received: November 3, 2020 Dear Manoj Zacharias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For: CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202377 Device Name Latex Examination Powder Free Gloves #### Indications for Use (Describe) Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.. | Type of Use (Select one or both, as applicable) | |---------------------------------------------------------------------------------| | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | Description 303 (Part 21, SFAR 36, Subpart B) |X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510K SUMMARY ## K202377 # As required by 21CFR§807.92(c) #### A. APPLICANT INFORMATION | Applicant | Hi-Care Thai Gloves Co. Ltd. | |----------------|----------------------------------------------------------| | Address | 199Moo.11, T.Banpru, A.Hatyai, Songkhla 90250, Thailand. | | Phone | +66-92 225 8472 | | Fax | +66-74-291044 | | E-mail | daniel@hicarethai.com | | Contact Person | Mr. Daniel John | | Designation | International Operations Manager, | | Contact Number | +66-92 225 8472 | | Contact Email | daniel@hicarethai.com | | Date Submitted | August 5th, 2020 | # B. DEVICE IDENTIFICATION | Name of the device | Latex Examination Powder Free Gloves | |---------------------------|------------------------------------------------| | Proprietary or trade name | Palm Care Latex Examination Powder Free Gloves | | Common or usual name | Latex Examination Powder Free Gloves | | Classification name | Patient Examination Glove | | Device Classification | Class I | | Product Code | LYY | | Regulation Number | 21 CFR 880.6250 | | Review Panel | General Hospital | ### C. PREDICATE DEVICE | Predicate device | JR Medic Powder Free Latex Examination Gloves | |------------------|-----------------------------------------------| | Predicate 510(K) | K192329 | | Regulatory Class | Class I | | Product code | LYY | #### D. DESCRIPTION OF THE DEVICE Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D 3578-05 (Reapproved 2015), Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (no color is added) and are powder free. {4}------------------------------------------------ ### E. INDICATIONS FOR USE/INTENDED USE OF THE DEVICE: Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner. ### F. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS BETWEEN PREDICATE AND SUBJECT DEVICES | CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON | | |---------------------------------------------|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|------------| | | | PREDICATE | CURRENT | | | | 510(k) Number | | K192329 | K202377 | | | | Name of device | --- | JR MEDIC Blue<br>Latex<br>Examination<br>Powder Free<br>Gloves | Latex Examination Powder Free<br>Gloves | --- | | | Dimensions- Length | ASTMD3578-05<br>(Reapproved<br>2015) | Length > 230 mm | Length > 230 mm | Similar | | | | | | Size<br>X-Small<br>Small<br>Medium<br>Large | | | | | | | Average<br>303<br>304<br>304<br>305 | | | | Dimensions- Width | ASTMD3578-05<br>(Reapproved<br>2015) | Width Min 95+/-<br>10 mm (for<br>medium size) | Width Min 95+/-10 mm<br>(for medium size) | Similar | | | | | | Size<br>X-Small<br>Small<br>Medium<br>Large | | | | | | | Average<br>75<br>84<br>94<br>105 | | | | Physical Properties-<br>Tensile Strength | ASTMD3578-05<br>(Reapproved<br>2015) | Before Ageing<br>Tensile Strength<br>> 18 Mpa_<br>After Ageing<br>Tensile Strength<br>> 14 Mpa | Before Ageing<br>Tensile Strength > 18 Mpa | Similar | | | | | | Size<br>X-Small<br>Small<br>Medium<br>Large | | | | | | | Actual value<br>22.13<br>22.20<br>22.25<br>22.28 | | | | | | | After Ageing<br>Tensile Strength > 14 Mpa | Similar | | | | | | Size<br>X-Small<br>Small<br>Medium<br>Large | | | | | | | Actual value<br>18.57<br>18.64<br>18.70<br>18.74 | | | | CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | | COMPARISON | | | | PREDICATE | | CURRENT | | | 510(K) Number | | K192329 | | K202377 | | | Physical Properties-<br>Ultimate Elongation | ASTMD3578-05<br>(Reapproved 2015) | Before Ageing<br>Ultimate Elongation<br>> 650%<br>After Ageing<br>Ultimate Elongation<br>>500% | | Before Ageing<br>Ultimate Elongation > 650% | Similar | | | | | Size | Actual value | | | | | | X-Small | 867 | | | | | | Small | 871 | | | | | | Medium | 874 | | | | | | Large | 877 | | | | | | | After Ageing<br>Ultimate Elongation > 500% | | | | | | Size | Actual value | | | | | | X-Small | 845 | | | | | | Small | 848 | | | | | | Medium | 854 | | | | | | Large | 860 | | | Thickness | ASTMD3578-05<br>(Reapproved 2015) | Palm > 0.08 mm<br>Finger > 0.08 mm | | Palm > 0.08 mm<br>Finger > 0.08 mm | Similar | | | | | Size | Palm<br>(Actual value)<br>Finger<br>(Actual value) | | | | | | X-Small | 0.16<br>0.21 | | | | | | Small | 0.16<br>0.21 | | | | | | Medium | 0.16<br>0.21 | | | | | | Large | 0.16<br>0.21 | | | Powder Free Residue | ASTMDD 3578-10<br>(Reapproved2015) | $≤$ 2 mg/glove | $≤$ 2 mg/glove | | Similar | | | | | Size | Residual<br>powder content<br>(mg/glove) | | | | | | X-Small | 0.21 | | | | | | Small | 0.21 | | | | | | Medium | 0.22 | | | | | | Large | 0.22 | | | Biocompatibility | Primary Skin<br>Irritation-ISO<br>10993-10:2010(E) | Under the condition<br>of study, not an<br>irritant | Under the condition of study, not<br>an irritant | | Same | | | Dermal Sensitization-<br>ISO 10993-<br>10:2010(E) | Under the conditions<br>of the study, not a<br>sensitizer | Under the conditions of the<br>study, not a sensitizer | | Same | | | In vitro<br>cytotoxicity<br>ISO10993-5<br>:2009(E) | Under the conditions<br>of the study,<br>non-cytotoxic | Under the conditions of the<br>study, non-cytotoxic | | Same | | | Material mediated<br>Pyrogenicity<br>ISO 10993-<br>11:2017(E) / USP<br>41<151> | Under the<br>conditions of the<br>study non pyrogenic | Under the conditions of the<br>study, non-pyrogenic | | Same | | CHARACTERSTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON | | | | | PREDICATE | CURRENT | | | | 510(K) Number | | K192329 | K202377 | --- | | | Biocompatibility | Bacterial Endotoxin<br>test<br>USP 42<85> | No data available | <20EU/pair of gloves | ---- | | | | Acute Systemic<br>Toxicity Test ISO<br>10993-11:2017(E) | Under the condition of<br>study, not systemic<br>toxic | Under the conditions<br>of study, the device<br>extracts do not pose a<br>systemic toxicity concern | same | | | Water Tight (1000<br>ml) | ASTM D5151-06<br>(Reapproved 2015) | Passes AQL-1.5 | Passes AQL-1.5 | Same | | | Intended use | | JR MEDIC Blue<br>Latex Examination<br>Powder Free Gloves<br>are disposable devices<br>intended for medical<br>purpose that are won<br>on the examiner's<br>hand to prevent<br>contamination<br>between patient and<br>examiner. | Latex Examination Powder<br>Free Gloves are disposable<br>devices intended for<br>medical purpose that are<br>won on the examiner's<br>hand to prevent<br>contamination between<br>patient and examiner. | Same | | | Material | - | Natural Latex | Natural Latex | Identical | | | Color | - | Blue | Natural (No color is added) | different | | | Texture | - | Finger Texture | Finger texture | Identical | | | Size | ASTMD 3578-5<br>(Reapproved 2015) | Small, Medium,<br>Large & X Large | X Small, Small,<br>Medium, Large, | Similar | | | Single Use | Medical Glove<br>Guidance Manual<br>- Labeling | Single Use | Single Use | Same | | | Sterile/non sterile | - | Non sterile | Non sterile | Same | | | Powder/Powder free | - | Powder free | Powder free | Same | | | Label and Labeling | FDA Label<br>requirements | Meets FDA's label<br>and labeling<br>requirements | Meets FDA's label and<br>labeling requirements | Same | | | Manufacturer(s) | - | JR Engineering &<br>Medical Technologies<br>(M) SDN.BHD.<br>Malaysia. | Hi-Care Thai Gloves Co.<br>Ltd. | --- | | {5}------------------------------------------------ {6}------------------------------------------------ # G. SUMMARY OF NON-CLINICAL PERFORMANCEDATA | TEST METHOD | PURPOSE | ACCEPTANCE<br>CRITERIA | RESULT | |--------------------------------------------------------------------------------------------|---------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------| | ASTM D3578-05 (Reapproved 2015)<br>Standard Specification for Rubber<br>Examination Gloves | To determine<br>the length of the<br>gloves | Min 230 mm for all sizes | X-Small:-303 mm<br>Small:-304mm<br>Medium:-304mm<br>Large:-305mm | | ASTM D3578-05 (Reapproved 2015)<br>Standard Specification for Rubber<br>Examination Gloves | To determine the<br>width of the gloves | X-Small:-70+/-10 mm<br>Small:-80+/-10mm<br>Medium:-95+/-10 mm<br>Large:-111+/-10 mm | X-Small:-75 mm<br>Small:-84 mm<br>Medium:-94mm<br>Large:-105mm | {7}------------------------------------------------ | TEST METHOD | PURPOSE | ACCEPTANCE CRITERIA | RESULT | | | |-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------| | ASTM D3578-05 (Reapproved 2015) Standard Specification For Rubber Examination Gloves | To determine the thickness of the gloves | Palm 0.08 mm min<br>Finger 0.08 mm min for all sizes | Size<br>X-Small<br>Small<br>Medium<br>Large | Palm<br>0.16mm<br>0.16mm<br>0.16mm<br>0.16mm | Finger<br>0.21mm<br>0.21mm<br>0.21mm<br>0.21mm | | | To Determine the physical properties-Tensile strength | Before Ageing Tensile Strength 18Mpa Min for all sizes<br>After Ageing Tensile Strength 14Mpa Min for all sizes | Size<br>X-Small<br>Small<br>Medium<br>Large | Before ageing<br>22.13Mpa<br>22.20Mpa<br>22.25Mpa<br>22.28Mpa | After ageing<br>18.57Mpa<br>18.64Mpa<br>18.70Mpa<br>18.74Mpa | | ASTM D3578-05 (Reapproved 2015) Standard Specification for Rubber Examination Gloves | To Determine the physical properties-Ultimate Elongation | Before Ageing Ultimate elongation 650% Min for all sizes After Ageing Ultimate Elongation 500% Min for all sizes | Size<br>X-Small<br>Small<br>Medium<br>Large | Before ageing<br>867%<br>871%<br>874%<br>877% | After ageing<br>845%<br>848%<br>854%<br>860% | | | To Determine the physical properties-stress at 500% Elongation | Before Ageing 5.5 Mpa Max for all sizes | Size<br>X-Small<br>Small<br>Medium<br>Large | Before ageing<br>5.1 Mpa<br>5.1Mpa<br>5.2 Mpa<br>5.2Mpa | NA | | ASTM D5151-06 (Reapproved 2015) Standard Test Method for Detection of Holes in Medical Gloves | To determine the holes in the gloves | AQL 2.5 | Gloves Passes AQL 1.5 | | | | ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves | To determine the residual powder in the gloves | 2 Mg/Glove Max | Size<br>X-Small<br>Small<br>Medium<br>Large | Residual Powder Content<br>0.21 mg/glove<br>0.21 mg/glove<br>0.22 mg/glove<br>0.22 mg/glove | | | ASTM D 5712-95 ( Re approved 2010) Standard Test Method for the Analysis of Protein in Natural Rubber | To determine the extractable protein in the gloves. | 200 µg/ dm² Max for all sizes | Size<br>X-Small<br>Small<br>Medium | Extractable Protein content<br>50 µg/ dm²<br>50 µg/ dm²<br>50 µg/ dm² | | ## H. SUMMARY OF CLINICAL PERFORMANCEDATA Not applicable. ### I. CONCLUSION The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission Latex Examination Powder Free Gloves (K202377) is as safe, as effective, and performs as well as or better than the legally marketed predicate device K192329.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...