GLOVETEX LATEX EXAMINATION GLOVES, POWDERED, BLUE, NON STERILE
K030273 · Pt. Wrp Buana Multicorpora · LYY · Mar 26, 2003 · General Hospital
Device Facts
Record ID
K030273
Device Name
GLOVETEX LATEX EXAMINATION GLOVES, POWDERED, BLUE, NON STERILE
Applicant
Pt. Wrp Buana Multicorpora
Product Code
LYY · General Hospital
Decision Date
Mar 26, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
The Latex Examination Glove, Powdered, Blue, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Disposable natural rubber latex examination gloves; worn on examiner's hand or finger; intended to prevent cross-contamination between patient and examiner during medical examinations. Device is non-sterile and powdered. Performance verified against ASTM D3578-01ae2 standards for dimensions, physical properties, and pinhole freedom. Biocompatibility confirmed via primary skin irritation and dermal sensitization testing.
Clinical Evidence
No clinical data; bench testing only. Device performance verified against ASTM D3578-01ae2 standards for dimensions, physical properties, and pinhole freedom (FDA 21 CFR 800.20). Biocompatibility testing (primary skin irritation and dermal sensitization) passed.
Indicated for use as a disposable, natural rubber latex glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner during medical procedures.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K202377 — Palm Care Latex Examination Powder Free Gloves · Hi-Care Thai Gloves Co. , Ltd. · Dec 1, 2020
K051987 — FLEXISKIN POLYMER COATED LATEX EXAMINATION GLOVES, POWDER FREE, NON-STERILE · Pt. Mandiri Inti Buana · Oct 6, 2005
K221374 — Disposable Medical Rubber Examination Gloves (XS/S/M/L/XL) · Hebei Titans Hongsen Medical Technology Co., Ltd. · Oct 26, 2022
K221185 — TC Gloves · Thanh Cong Pharmaceutical and Trading Company Limited · Jun 22, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
**WRP Buana Multicorpora**
K030273
Jalan Jermal No. 20-B, Kelurahan Sei Mati, Medan-Labuhan Km 17, INDONESIA
MAR 2 6 2003
| Tel | +62-61-6942461 |
|-----|----------------|
| | +62-61-6942462 |
| Fax | +62-61-6942463 |
# 510 (K) SUMMARY
### 1.0 Submitter :
| Name | : PT. WRP Buana Multicorpora |
|-----------|---------------------------------------------------------------------------|
| Address | : Jl. Jermal No.20B Kelurahan Sei Mati<br>Medan Labuhan KM 17, INDONESIA. |
| Phone No. | : +61 6942461, 6942462 |
| Fax No. | : +61 6942463 |
Date of Summary Prepared : January 21, 2003
#### 2.0 Contact Person :
| Name | : Mr. Ng Poy Sin |
|-----------|------------------------|
| Phone No. | : +61 6942461, 6942462 |
| Fax No. | : +61 6942463 |
#### 3.0 Name of the device :
| Trade Name | 1). Glovetex, and<br>2). Multiple or Customers' Trade Name |
|---------------------|------------------------------------------------------------|
| Device Name | : Latex Examination Gloves, Powdered, Blue, Non-Sterile |
| Common Name | : Examination Gloves |
| Classification Name | : Patient Examination Gloves (Class I) |
## 4.0 ldentification of The Legally Marketed Device :
Class I patient examination gloves, 80LYY, powdered, that meets all the requirements of ASTM standard D 3578 – 01a 2 and FDA 1000 ml Water Leak Test.
#### 5.0 Description of The Device :
The Latex Examination Gloves, Powdered, Blue, Non Sterile meets all the requirements of ASTM standard D 3578 - 01a22 and FDA 1000 ml Water Leak Test.
#### 6.0 Intended Use of the Device :
The Latex Examination Glove, Powdered, Blue, Non Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a sequence of alphanumeric characters, specifically "KO30273". The characters are written in a bold, sans-serif font, and they appear to be handwritten. The image is in black and white, and the characters are clearly legible against the white background.
Jalan Jermal No. 20-B, Kelurahan Sei Mati, Medan-Labuhan Km 17, INDONESIA
+62-61-6942461 Tel +62-61-6942462 +62-61-6942463 Fax
# Summary of The Technological Characteristics of The Device : 7.0
RP Buana Multicorpora
The Latex Examination Gloves, Powdered, Blue, Non Sterile are summarized with the following technological characteristics compared to ASTM or equivalent standards.
| CHARACTERISTICS | STANDARDS | DEVICE<br>PERFORMANCE |
|-------------------------------|-------------------------------------|----------------------------------------|
| Dimensions | D 3578 – 01ae2 | Meets |
| Physical Properties | D 3578 - 01ae2 | Meets |
| Freedom from Pinholes | D 3578 - 01ae2<br>FDA 21 CFR 800.20 | Meets |
| Powder Amount | D 3578 - 01ae2<br>D 6124 - 01 | < 10 mg/dm² |
| Water Soluble Protein Content | D 3578 – 01ae2<br>D 5712 - 99 | < 200mg/dm² |
| Biocompatability | Primary Skin Irritation in Rabbits | Passes<br>(No primary skin irritation) |
| | Dermal Sensitization | Passes<br>(No contact sensitizer) |
#### 8.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.
# Substantial Equivalent Based on Assessment of Clinical Performance Data 9.0
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
#### 10.0 Conclusion
It can be concluded that the Latex Examination Gloves, Powdered, Blue, Non Sterile will perform according to the glove performance standards referenced in Section (7) above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 6 2003
Mr. Ng Poy Sin President Director PT. WRP Buana Multicorpora Jalan Jermal No. 20-B. Kelurahan Sei Mati, Medan-Labuhan Km 17, INDONESIA
Re: K030273
Trade/Device Name: Latex Examination Gloves, Powdered, Blue, Non-Sterile Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: January 22, 2003 Received: January 27, 2003
Dear Mr. Sin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Sin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters "WRP" in white, set against a black, rounded rectangle. The letters are bold and slightly distorted, giving them a textured appearance. The overall design is simple and eye-catching, with a clear contrast between the text and the background.
Jalan Jermal No. 20-B, Kelurahan Sei Mati, Medan-Labuhan Km 17, INDONESIA
Tel +62-61-6942461 +62-61-6942462 +62-61-6942463 Fax
# INDICATIONS FOR USE
: PT. WRP Buana Multicorpora Applicant K030273 510(k) Number (if known) : _ : LATEX EXAMINATION GLOVES, Device Name POWDERED, BLUE, NON STERILE
Indications For Use:
Latex Examination Gloves, Powdered, Blue, Non-Sterile is a disposable device and made of Natural Rubber Latex intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chin S. Lin
(Division Sian-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K030273
______________________________________________________________________________________________________________________________________________________________________________
Prescription Use (Per 21 CFR 801.109) Over-The-Counter_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.