CHLORINATED POWDER FREE LATEX MEDICAL EXAMINATION GLOVES WITH ALOE VERA
K013589 · Supermax Glove Manufacturing Sdn Bhd · LYY · Jan 8, 2002 · General Hospital
Device Facts
Record ID
K013589
Device Name
CHLORINATED POWDER FREE LATEX MEDICAL EXAMINATION GLOVES WITH ALOE VERA
Applicant
Supermax Glove Manufacturing Sdn Bhd
Product Code
LYY · General Hospital
Decision Date
Jan 8, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A medical examination glove is worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient.
Device Story
Chlorinated powder-free latex medical examination glove; features aloe vera coating and low protein content (50 micrograms or less). Used by healthcare personnel in clinical settings as a protective barrier to prevent cross-contamination between patient and provider. Device functions as a physical barrier; no electronic or algorithmic components.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Material: Latex. Type: Chlorinated, powder-free. Features: Aloe vera coating, protein content labeling claim (≤ 50 micrograms). Form factor: Medical examination glove. Sterilization: Not specified.
Indications for Use
Indicated for use by health care and similar personnel as a protective barrier to prevent cross-contamination between the wearer and the patient.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K013953 — AURELIA CHLORINATED POWDER FREE LATEX MEDICAL EXAMINATION GLOVES WITH ALOE VERA AND GINSENG WITH PROTEIN LABELING CLAIM · Supermax Glove Manufacturing Sdn Bhd · Feb 26, 2002
K023974 — ALOE VERA & VITAMIN E POWDERFREE LATEX EXAMINATION GLOVES · Top Glove Sdn. Bhd. · Feb 13, 2003
K013317 — TEKMEDIC POWDER-FREE WITH PROTEIN LABELING CLAIM (50 MICROGRAMS OR LESS) · Tekmedic (M) Sdn Bhd · Dec 17, 2001
K032940 — BRIGHTWAY BRAND LATEX EXAMINATION GLOVE-(POWDER FREE, STERILE) CONTAINING 50 UGM OR LESS OF WATER EXTRACTABLE PROTEIN PE · Brightway Holdings Sdn. Bhd. · Nov 20, 2003
K024121 — NON-STERILE, POWDER-FREE LATEX EXAMINATION GLOVES WITH POLYMER COATING, ALOE, VITAMIN E AND PROTEIN CLAIM · Siam Sempermed Corp., Ltd. · Dec 20, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or feathers.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## 8 2002 JAN
Ms. Priscilla Wee Supermax Glove Manfacturing Sdn. Bhd. Lot 6070 Jalan Haji Abdul Manan 6th Miles Off Jalan Meru Klang Selangor, MALAYSIA
Re: K013589
Trade/Device Name: Chlorinated Powder Free Latex Medical Examination Gloves with Aloe Vera and Protein Content Labeling Claim (50 Micrograms or Less) Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: November 19, 2001 Received: November 27, 2001
Dear Ms. Wee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## INDICATIONS FOR USE
Applicant : Supermax Glove Manufacturing Sdn. Bhd.
510 (k) Number (if known) : K 013589
Chlorinated Powder Free Latex Medical Examination Gloves with Aloe Device Name : Vera and Protein Content labeling Claim (50 m ( Rogerms or less
Indications For Use :
A medical examination glove is worn on the hand of health care and similar personnel to prevent contamination between health care personnel and the patient.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Prescription Use Per 21 CFR 801.109) OR
Over-The-Counter
Qian S. Lin
ું તેમ છે.
C on of Dental, Infection Control. neral Hospital Devic Number
(Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.