ALOE VERA & VITAMIN E POWDERFREE LATEX EXAMINATION GLOVES
K023974 · Top Glove Sdn. Bhd. · LYY · Feb 13, 2003 · General Hospital
Device Facts
Record ID
K023974
Device Name
ALOE VERA & VITAMIN E POWDERFREE LATEX EXAMINATION GLOVES
Applicant
Top Glove Sdn. Bhd.
Product Code
LYY · General Hospital
Decision Date
Feb 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
The glove is disposable and intended for Medical purpose that is worn on the examiner's hand to prevent contamination between the patient and the examiner.
Device Story
Disposable latex examination glove; coated with aloe vera; features protein content labeling claim (50 micrograms or less). Used in clinical settings by healthcare professionals to provide barrier protection; prevents cross-contamination between patient and examiner. Device functions as a physical barrier; no electronic or mechanical components.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Material: Latex. Features: Powder-free, aloe vera coating, low protein content (≤ 50 micrograms). Form factor: Disposable examination glove. Regulatory class: I. Product code: LYY.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K013589 — CHLORINATED POWDER FREE LATEX MEDICAL EXAMINATION GLOVES WITH ALOE VERA · Supermax Glove Manufacturing Sdn Bhd · Jan 8, 2002
K013317 — TEKMEDIC POWDER-FREE WITH PROTEIN LABELING CLAIM (50 MICROGRAMS OR LESS) · Tekmedic (M) Sdn Bhd · Dec 17, 2001
K013953 — AURELIA CHLORINATED POWDER FREE LATEX MEDICAL EXAMINATION GLOVES WITH ALOE VERA AND GINSENG WITH PROTEIN LABELING CLAIM · Supermax Glove Manufacturing Sdn Bhd · Feb 26, 2002
K024121 — NON-STERILE, POWDER-FREE LATEX EXAMINATION GLOVES WITH POLYMER COATING, ALOE, VITAMIN E AND PROTEIN CLAIM · Siam Sempermed Corp., Ltd. · Dec 20, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines that resemble a ribbon or stream. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 3 2003
Dr. Lim Wee Chai Managind Director Top Glove Sdn. Bhd. Lot 4968 Jalan Teratai, Batu 6, Off Jalan Meru, 41050 Klang, Selangor, D.E., MALAYSIA
Re: K023974
Trade/Device Name: Powder Free Latex Examination Gloves with Aloe Vera and Protein Content Labeling Claim (50 Micrograms or Less) Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: November 26, 2002 Received: December 2, 2002
Dear Dr. Chai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Dr. Chai
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Susan Runner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
3.0 Indications for Use Statement: Include the following or equivalent Indications for Use page. The information, data and labeling claims in the entire the 510(k) submission nust support and agree with the Indications for Use statement.
## INDICATIONS FOR USE
| Applicant: | TOP GLOVE SDN. BHD. |
|---------------------------|------------------------------------------------------------------------------------------------------------------|
| 510(k) Number (If known): | K023974 |
| Device Name: | POWDER FREE LATEX EXAMINATION GLOVES WITH ALOE VERA AND - PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS) |
| Indications For Use: | |
The glove is disposable and intended for Medical purpose that is worn on the examiner's hand to prevent contamination between the patient and the examiner.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IN NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Clins S. lin
(Division Sign-Off) Division of Anesthesiology, General Ho Infection Control, Dental D
Prescription Use Per 21 CFR 801.109
OR
Over-The-Counter
(Optional Format 1-2-96)
* For a new submission, do NOT fill in the 510(k) number blank.
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.