HALYARD ONE-STEP Sterilization Wrap, HALYARD QUICK CHECK Sterilization Wrap

K214007 · Owens & Minor (O&M) Halyard, Inc. · FRG · Jul 27, 2022 · General Hospital

Device Facts

Record IDK214007
Device NameHALYARD ONE-STEP Sterilization Wrap, HALYARD QUICK CHECK Sterilization Wrap
ApplicantOwens & Minor (O&M) Halyard, Inc.
Product CodeFRG · General Hospital
Decision DateJul 27, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using the Stryker Sterizone VP4 Sterilizer Cycle 1. The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to allow sterilization of the enclosed devices by the Stryker Sterizone VP4 Sterilizer Cycle 1.

Device Story

Sterilization wrap system consisting of two sheets of SMS (spunbond-meltblown-spunbond) polypropylene fabric ultrasonically bonded on two edges; used by healthcare providers to wrap medical devices prior to sterilization; compatible with Stryker Sterizone VP4 Sterilizer Cycle 1 (vaporized hydrogen peroxide and ozone); allows sterilant penetration while maintaining package sterility for up to 12 months post-sterilization; facilitates aseptic opening of sterilized packages.

Clinical Evidence

Bench testing only. Evaluated sterilant penetration/efficacy (ANSI/AAMI ST79, ISO 11138-7), physical properties (ISO 11607-1, ISO 13938-2, ASTM D4966-12, CPSC 1610), package integrity (ANSI/AAMI/ISO 11607-1), and biocompatibility (ISO 10993 series). All tests passed, confirming 10^-6 sterility assurance level and 12-month shelf life.

Technological Characteristics

Three-layer SMS polypropylene fabric; antistatic treatment; ultrasonically bonded; dimensions vary by grade (H100-H600). Compatible with vaporized hydrogen peroxide and ozone sterilization. Complies with ANSI/AAMI/ISO 11607-1.

Indications for Use

Indicated for healthcare providers to enclose medical devices for sterilization using the Stryker Sterizone VP4 Sterilizer Cycle 1. Applicable to general instruments, rigid/semi-rigid channeled instruments, and flexible endoscopes.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 27, 2022 Owens & Minor (O&M) Halyard, Inc. Anureet Singh Regulatory Affairs Specialist II 9120 Lockwood Blvd Mechanicsville, Virginia 23116 # Re: K214007 Trade/Device Name: HALYARD ONE-STEP Sterilization Wrap, HALYARD QUICK CHECK Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: June 29, 2022 Received: June 30, 2022 Dear Anureet Singh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K214007 #### Device Name HALYARD* ONE-STEP* Sterilization Wrap and HALYARD* QUICK CHECK* Sterilization Wrap ## Indications for Use (Describe) HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to enclose another medical device that is to be sterilized by a healthcare provider using the Stryker Sterizone VP4 Sterilizer Cycle 1. The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended to allow sterilization of the enclosed devices by the Stryker Sterizone VP4 Sterilizer Cycle 1. All models of the Halyard ONE-STEP and QUICK CHECK Sterilization Wrap have been validated for use with the Stryker Sterilizer Cycle 1 as described below: Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ | Type of Device | HALYARD ONE-STEP* and QUICK CHECK* Validation Loads Description<br>in the Stryker Sterizone VP4 Sterilizer Cycle 1 | Validated<br>Sterilization<br>Wrap GradesB | |-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | General<br>instruments | Load consisted of three (3) trays each containing three (3) lb of general medical<br>instruments. The load included:<br>General devices representing the following geometries:<br>Box-lock hinge<br>Pivot hinge<br>Luer-lock<br>Testing was performed with trays wrapped in H100 sterilization wrap.<br>Load #7A - Consisted of three (3) trays each containing 25 lb of general medical<br>instruments, for a total of 75 lb per load (excluding the loading rack). The load included:<br>General devices representing the following geometries:<br>Box-lock hinge<br>Pivot hinge<br>Luer-lock<br>Testing was performed with trays wrapped in H600 sterilization wrap. | H100C<br>H200<br>H300<br>H400<br>H500<br>H600C | | Rigid channel<br>instruments | Load #4A - Consisted of 15 lumens from rigid and semi-rigid channeled devices. The load<br>included:<br>Three (3) double channel (six (6) lumens) semi-rigid endoscopes (ureteroscope - 0.7 mm × 500 mm and 1.1 mm × 500 mm) were packaged in three (3)<br>sterilization trays including appropriate silicone brackets. Additional rigid channel instruments (nine (9) lumens) were added. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C | | Single and double<br>channel flexible<br>endoscopes | Load #8A - Consisted of five (5) lumens from single and double channel flexible<br>endoscopes. The load included:<br>Two (2) double channel flexible endoscopes (ureteroscope) with inside<br>diameter of 1 mm and lengths of 850 mm and 989 mm; One (1) single channel flexible endoscope (ureteroscope) with inside diameter<br>of 1 mm and length of 850 mm; The endoscopes were individually packaged in sterilization trays, including<br>appropriate silicone brackets. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C | | Multi-channel<br>flexible<br>endoscopes | Load #9A - Consisted of one (1) multichannel flexible endoscope with four (4) channels.<br>The load included:<br>One (1) multi flexible endoscope (colonoscope) with no more than four (4)<br>channels having inside diameter of 1.2 mm and lengths of 1955 mm or inside<br>diameter of 1.45 mm and lengths of 3500 mm, packaged individually in a<br>sterilization tray. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C | A: Representative Sterizone® VP4 Sterilizer Cycle 1 Valifand* ONE-STEP and QUICK CHECK* Sterliization Wrap are comprised of two sheets of Halyard Sequential Sterilization Wrap ultrasonically bonded together on two sides. Therefore, these grades are applicable to HALYARD* ONE-STEP*, QUICK CHECK* and Sequential Sterilization Wrap. C: Indicates the bracketed grades for validation testing. Type of Use (Select one or both, as applicable) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary – K214007 | Submitter: | O&M Halyard, Inc.<br>9120 Lockwood Boulevard<br>Mechanicsville, VA 23116<br>Phone: 804-723-7000/800-488-8850<br>Fax: 804-723-7100 | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulatory Contact: | Anureet Singh<br>Regulatory Affairs Specialist II<br>O&M Halyard, Inc.<br>1 Edison Drive<br>Alpharetta, GA 30005 | | Date of Summary: | 25 July 2022 | | Device Trade Name: | HALYARD* ONE-STEP* Sterilization Wrap and HALYARD* QUICK CHECK*<br>Sterilization Wrap | | Common Name: | Sterilization Wrap | | Classification Name: | Sterilization wrap (21 CFR 880.6850, Product Code FRG) | | Predicate Device: | Halyard ONE-STEP Sterilization Wrap, K192147 | | Device Description: | Halyard Sterilization Wrap is supplied to the customer as bulk packages of single<br>sheets, where in accordance with standard hospital practices, two sheets are then<br>used to wrap a medical device or a collection of medical devices for sterilization.<br>HALYARD ONE-STEP and HALYARD QUICK CHECK Sterilization Wraps<br>are comprised of two sheets of HALYARD* Sequential Sterilization Wrap<br>ultrasonically seamed on two edges. This allows for convenient wrapping with<br>two sheets simultaneously.<br><br>Sterilization wrap is a square or rectangular sheet made of three-layer SMS<br>(spunbond-meltblown-spunbond) polypropylene fabric treated with an antistatic<br>treatment. The wrap allows a sterilized package to be opened aseptically.<br><br>The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are<br>intended to enclose another medical device that is to be sterilized by a healthcare<br>provider using the Stryker Sterizone VP4 Sterilizer Cycle 1 which is a dual-<br>sterilant, low temperature sterilizer that uses vaporized hydrogen peroxide (H2O2)<br>and ozone (O3) in a single cycle. All models of the Halyard ONE-STEP and<br>QUICK CHECK Sterilization Wrap have been validated for use with the Stryker<br>Sterizone VP4 Sterilizer Cycle 1 as described below. | | Indication for Use: | HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are intended<br>to enclose another medical device that is to be sterilized by a healthcare provider<br>using the single cycle Stryker Sterizone VP4 Sterilizer Cycle 1.<br><br>The HALYARD ONE-STEP and QUICK CHECK Sterilization Wraps are<br>intended to allow sterilization of the enclosed devices by the Stryker Sterizone | {5}------------------------------------------------ VP4 Sterilizer Cycle 1. All models of the Halyard ONE-STEP and QUICK CHECK Sterilization Wrap have been validated for use with the Stryker Sterizone VP4 Sterilizer Cycle 1 as described below: | Type of Device | HALYARD ONE-STEP* and QUICK CHECK* Validation Loads Description<br>in the Stryker Sterizone VP4 Sterilizer Cycle 1 | Validated<br>Sterilization<br>Wrap GradesB | |-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | General<br>instruments | Load consisted of three (3) trays each containing three (3) lb of general medical<br>instruments. The load included:<br>General devices representing the following geometries: Box-lock hinge Pivot hinge Luer-lock Testing was performed with trays wrapped in H100 sterilization wrap.<br>Load #7A - Consisted of three (3) trays each containing 25 lb of general medical<br>instruments, for a total of 75 lb per load (excluding the loading rack). The load included: General devices representing the following geometries: Box-lock hinge Pivot hinge Luer-lock Testing was performed with trays wrapped in H600 sterilization wrap. | H100C<br>H200<br>H300<br>H400<br>H500<br>H600C | | Rigid channel<br>instruments | Load #4A - Consisted of 15 lumens from rigid and semi-rigid channeled devices. The load<br>included: Three (3) double channel (six (6) lumens) semi-rigid endoscopes (ureteroscope<br>- 0.7 mm × 500 mm and 1.1 mm × 500 mm) were packaged in three (3)<br>sterilization trays including appropriate silicone brackets. Additional rigid channel instruments (nine (9) lumens) were added. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C | | Single and double<br>channel flexible<br>endoscopes | Load #8A - Consisted of five (5) lumens from single and double channel flexible<br>endoscopes. The load included: Two (2) double channel flexible endoscopes (ureteroscope) with inside<br>diameter of 1 mm and lengths of 850 mm and 989 mm; One (1) single channel flexible endoscope (ureteroscope) with inside diameter<br>of 1 mm and length of 850 mm; The endoscopes were individually packaged in sterilization trays, including<br>appropriate silicone brackets. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C | | Multi-channel<br>flexible<br>endoscopes | Load #9A - Consisted of one (1) multichannel flexible endoscope with four (4) channels.<br>The load included: One (1) multi flexible endoscope (colonoscope) with no more than four (4)<br>channels having inside diameter of 1.2 mm and lengths of 1955 mm or inside<br>diameter of 1.45 mm and lengths of 3500 mm, packaged individually in a<br>sterilization tray. Testing was performed with trays wrapped in H400 and H600 sterilization wrap. | H400C<br>H500<br>H600C | A: Representative Sterizone® VP4 Sterilizer Cycle 1 Validation Load from K172191. B'HALYARD* ONE-STEP* and QUICK CHECK* Sterilization Wrap are comprised of two sheets of Halyard Sequential Sterilization Wrap ultrasonically bonded together on two sides are applicable to HALYARD* ONE-STEP*, QUICK CHECK* and Sequential Sterilization Wrap. C: Indicates the bracketed grades for validation testing. {6}------------------------------------------------ | | Proposed<br>HALYARD ONE-STEP and<br>QUICK CHECK Sterilization Wrap<br>(K214007) | Predicate<br>HALYARD ONE-STEP<br>Sterilization Wrap<br>(K192147) | Same<br>Similar, or<br>Different | |-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------| | Manufacturer | O&M Halyard, Inc. | O&M Halyard, Inc. | Same | | Device Model<br>Numbers | H100<br>H200<br>H300<br>H400<br>H500<br>H600 | H100<br>H200<br>H300<br>H400<br>H500<br>H600 | Same | | Common or<br>Usual Name | Sterilization Wrap | Sterilization Wrap | Same | | Classification | 21 CFR 880.6850 | 21 CFR 880.6850 | Same | | Class | II | II | Same | | Product Code | FRG | FRG | Same | | Indication for<br>Use | HALYARD ONE-STEP and QUICK<br>CHECK Sterilization Wraps are<br>intended to enclose another medical<br>device that is to be sterilized by a<br>healthcare provider using the Stryker<br>Sterizone VP4 Sterilizer Cycle 1.<br>The HALYARD ONE-STEP and<br>QUICK CHECK Sterilization Wraps<br>are intended to allow sterilization of<br>the enclosed devices by the Stryker<br>Sterizone VP4 Sterilizer Cycle 1. All<br>models of the Halyard ONE-STEP and<br>QUICK CHECK Sterilization Wrap<br>have been validated for use with the<br>Stryker Sterizone VP4 Sterilizer Cycle<br>1 as described below:<br><br>Type of Device: General instruments<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load consisted of three (3) trays each<br>containing three (3) lb of general medical<br>instruments. The load included general<br>devices representing the following<br>geometries: Box-lock hinge, Pivot hinge,<br>Luer-lock.<br>Testing was performed with trays wrapped<br>in H100 sterilization wrap.<br>Load #7A - Consisted of three (3) trays<br>each containing 25 lb of general medical<br>instruments, for a total of 75 lb per load<br>(excluding the loading rack). The load<br>included general devices representing the<br>following geometries: Box-lock hinge,<br>Pivot hinge, Luer-lock.<br>Testing was performed with trays wrapped<br>in H600 sterilization wrap. | The Halyard ONE-STEP Sterilization<br>Wraps are intended to allow<br>sterilization of the enclosed devices by<br>the Sterilucent HC 80TT Hydrogen<br>Peroxide Sterilizer (ie, both the Lumen<br>and Flexible Cycles) Additionally, the<br>Halyard ONE-STEP Sterilization<br>Wrap was validated to allow effective<br>aeration under the pre-programmed<br>HC 80TT Sterilization Cycles. All<br>models of the Halyard ONE-STEP<br>Sterilization Wrap have been validated<br>for use with the Sterilucent HC 80TT<br>Hydrogen Peroxide Sterilizer cycles as<br>described below.<br>Lumen Cycle<br>Reusable metal and nonmetal devices<br>including devices with diffusion-<br>restricted spaces such as the hinged<br>portion of forceps and scissors and up<br>to fifteen (15) stainless steel lumens<br>per load with the following<br>dimensions:<br>Single or dual channeled rigid and<br>semi-rigid endoscopes, with stainless<br>steel lumens that are<br>$\ge$ 0.77 mm internal diameter (ID) and<br>$\le$ 410 mm long, or $\ge$ 1.33 mm ID and $\le$<br>430 mm long;<br>and, Triple channeled rigid and semi-<br>rigid endoscopes, with stainless steel<br>lumens that are $\ge$ 1.00 mm ID and $\le$<br>310 mm long (Refer to the HC 80TT<br>User Manual for complete instructions<br>on load(s) and cycle(s), including<br>chamber loading instructions (i.e. 20.1<br>lb per load)<br>Flexible Cycle | Different | Technological Characteristics Comparison Table: {7}------------------------------------------------ | Validated Sterilization Wrap GradesB:<br>H100C, H200, H300, H400, H500, H600C<br><br>Type of Device: Rigid channel<br>instruments<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load #4A - Consisted of 15 lumens from<br>rigid and semi-rigid channeled devices.<br>The load included:<br>Three (3) double channel (six (6) lumens)<br>semi-rigid endoscopes (ureteroscope - 0.7<br>mm × 500 mm and 1.1 mm × 500 mm)<br>were packaged in three (3) sterilization<br>trays including appropriate silicone<br>brackets.<br>Additional rigid channel instruments (nine<br>(9) lumens) were added.<br>Testing was performed with trays wrapped<br>in H400 and H600 sterilization wrap.<br>Validated Sterilization Wrap GradesB:<br>H400C, H500, H600C | Reusable rigid or semi-rigid non-<br>lumen medical devices including non-<br>lumen devices with metallic diffusion-<br>restricted spaces such or mated<br>surfaces such as the hinged portion of<br>forceps or scissors;<br>Single channel flexible endoscopes<br>with flexible lumens that are ≥ 1.00<br>mm ID and ≤ 1280 mm long; and Dual<br>channel flexible endoscopes with<br>flexible lumens that are ≥ 0.80 mm ID<br>and ≤ 1000 mm long.<br>(Refer to the HC 80TT User Manual<br>for complete instructions on load(s)<br>and cycle(s), including chamber<br>loading instructions (i.e. 25 lb per<br>load). | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Type of Device: Single and double<br>channel flexible endoscopes<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load #8A - Consisted of five (5) lumens<br>from single and double channel flexible<br>endoscopes. The load included: Two (2)<br>double channel flexible endoscopes<br>(ureteroscope) with inside diameter of 1<br>mm and lengths of 850 mm and 989 mm;<br>One (1) single channel flexible endoscope<br>(ureteroscope) with inside diameter of 1<br>mm and length of 850 mm; The<br>endoscopes were individually packaged in<br>sterilization trays, including appropriate<br>silicone brackets.<br>Testing was performed with trays wrapped<br>in H400 and H600 sterilization wrap.<br>Validated Sterilization Wrap GradesB:<br>H400C, H500, H600C | | | Type of Device: Multi-channel flexible<br>endoscopes<br>HALYARD ONE-STEP* and QUICK<br>CHECK* Validation Loads Description in<br>the Stryker Sterizone VP4 Sterilizer Cycle<br>1:<br>Load #9A - Consisted of one (1)<br>multichannel flexible endoscope with four<br>(4) channels. The load included: One (1)<br>multi flexible endoscope (colonoscope)<br>with no more than four (4) channels having<br>inside diameter of 1.2 mm and lengths of<br>1955 mm or inside diameter of 1.45 mm | | {8}------------------------------------------------ | | and lengths of 3500 mm, packaged<br>individually in a sterilization tray.<br>Testing was performed with trays wrapped<br>in H400 and H600 sterilization wrap.<br>Validated Sterilization Wrap GradesB:<br>H400C, H500, H600C<br><br>A: Representative Sterizone® VP4<br>Sterilizer Cycle 1 Validation Load from<br>K172191. B: HALYARD* ONE-STEP*<br>and QUICK CHECK* Sterilization Wrap<br>are comprised of two sheets of Halyard<br>Sequential Sterilization Wrap<br>ultrasonically bonded together on two<br>sides. Therefore, these grades are<br>applicable to HALYARD* ONE-STEP*,<br>QUICK CHECK* and Sequential<br>Sterilization Wrap. C: Indicates the<br>bracketed grades for validation testing. | | | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Sterilization<br>Parameters | Stryker Sterizone VP4 Sterilizer Cycle<br>1 which is a dual-sterilant, low<br>temperature sterilizer that uses<br>vaporized hydrogen peroxide (H2O2)<br>and ozone (O3) in a single cycle | Sterilucent PSD-85 Hydrogen<br>Peroxide Sterilizer that includes<br>● Lumen Cycle<br>● Flexible Cycle | Different | | Technology | Tortuous sheet material used to<br>enclose medical devices that are to be<br>sterilized by a healthcare provider to<br>allow sterilization of the enclosed<br>medical device(s) and maintain<br>sterility of the enclosed device(s) until<br>used | Tortuous sheet material used to<br>enclose medical devices that are to be<br>sterilized by a healthcare provider to<br>allow sterilization of the enclosed<br>medical device(s) and maintain<br>sterility of the enclosed device(s) until<br>used | Same | | Device Design | Two sheets of nonwoven<br>polypropylene fabric. Each sheet is<br>composed of three thermally-bonded<br>layers consisting of a meltblown<br>polypropylene layer surrounded by<br>spunbond polypropylene layers (SMS) | Two sheets of nonwoven<br>polypropylene fabric. Each sheet is<br>composed of three thermally-bonded<br>layers consisting of a meltblown<br>polypropylene layer surrounded by<br>spunbond polypropylene layers (SMS) | Same | | Method for<br>Bonding SMS<br>Layers | Thermal bonding with round pin,<br>hexagonal, triangle bond pattern<br>( | | | ## Summary of Performance Testing Performance Testing (Bench): Performance testing of HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap was evaluated and the results showed that acceptance criteria were met demonstrating that the HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap allows sterilization of its contents using the Stryker Sterizone VP4 Sterilizer Cycle 1 and that sterility is maintained for the testing period of 12 months. {9}------------------------------------------------ | Purpose | Test | Acceptance Criteria | Results | |-----------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------| | Sterilant<br>Penetration/Efficacy | ANSI/AAMI ST79<br>ANSI/AAMI/ISO 11138-7 | Achieving a 10-6 sterility assurance<br>level following processing in a worst-case half-cycle | Passed | | Performance<br>Testing (Non-sterile<br>and Sterile) | ANSI/AAMI/ISO 11607-1 Annex B<br>ISO 13938-2<br>ASTM D4966-12<br>CPSC 1610 | Complies with the selected physical<br>properties | Passed | | Maintenance of<br>Package Integrity | ANSI/AAMI/ISO 11607-1<br>ANSI/AAMI ST79 | Maintain sterility for up to 12 months | Passed | | Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-7 | Non-cytotoxic<br>Non-irritating<br>H2O2 ≤ 0.56 μg/cm2 | Passed | ## Summary of Non-Clinical Testing Performed ## Performance Testing (Clinical): Clinical evaluations were not required and therefore are not submitted with this 510(k). Discussion: The HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap in this submission and the predicate device submission are intended to enclose another medical device that is to be sterilized by a healthcare provider, to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s). The ONE-STEP and QUICK CHECK Sterilization Wrap in this submission and the predicate device submission have identical intended use, design, materials, specifications, and composition, and are manufactured using identical production methods. The different technological characteristics, that is, the Indication for Use and the Sterilization Parameters, do not affect the safety and effectiveness of the device as evidenced by the results of the nonclinical testing. Overall Performance Conclusions: The conclusions drawn from the nonclinical tests demonstrate that the HALYARD ONE-STEP and QUICK CHECK Sterilization Wrap (K214007) is as safe, as effective, and performs as well as or better than the legally marketed HALYARD ONE-STEP Sterilization Wrap (K192147).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...