K192147 · Owens & Minor (O&M) Halyard, Inc. · FRG · Oct 10, 2019 · General Hospital
Device Facts
Record ID
K192147
Device Name
Halyard ONE-STEP* Sterilization Wrap
Applicant
Owens & Minor (O&M) Halyard, Inc.
Product Code
FRG · General Hospital
Decision Date
Oct 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Halyard ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed devices by the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer (i.e., both the Lumen and Flexible Cycles). Additionally, the Halyard ONESTEP Sterilization Wrap was validated to allow effective aeration under the pre-programmed HC 80TT Sterilization Cycles. All models of the Halyard ONE-STEP Sterilization Wrap have been validated for use with the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer cycles as described below: Lumen Cycle: Reusable metal and nonmetal devices including devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors and up to fifteen (15) stainless steel lumens per load with the following dimensions: Single or dual channeled rigid and semi-rigid endoscopes, with stainless steel lumens that are: > 0.77 mm internal diameter (ID) and ≤ 410 mm long, or ≥ 1.33 mm ID and ≤ 430 mm long; and, Triple channeled rigid and semi-rigid endoscopes, with stainless steel lumens that are ≥ 1.00 mm ID and ≤ 310 mm long. (Refer to the HC 80TT User Manual for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e. 20.1 lb. per load). Flexible Cycle: Reusable rigid or semi-rigid non-lumen medical devices with metallic diffusion-restricted spaces such or mated surfaces such as the hinged portion of forceps or scissors; Single channel flexible endoscopes with flexible lumens that are ≥ 1.00 mm ID and ≤ 1280 mm long; and Dual channel flexible endoscopes with flexible lumens that are ≥ 0.80 mm ID and ≤ 1000 mm long. (Refer to the HC 80TT User Manual for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e. 25 lb. per load).
Device Story
Halyard ONE-STEP Sterilization Wrap consists of two ultrasonically seamed sheets of three-layer SMS (spunbound-meltblown-spunbound) polypropylene fabric; includes phthalocyanine blue and titanium dioxide pigments with antistatic treatment. Device functions as a protective barrier for medical instruments during sterilization in Sterilucent HC 80TT Hydrogen Peroxide Sterilizer. Healthcare providers wrap instruments in the fabric; the wrap allows sterilant penetration and subsequent aeration while maintaining sterility post-process. Used in clinical settings to prepare reusable devices (endoscopes, surgical instruments) for reuse. Benefits include aseptic presentation of sterile contents and maintenance of sterility until use. No software or electronic components.
Clinical Evidence
No clinical data. Bench testing only: Sterilant penetration/efficacy (no growth of challenge microorganisms in BI placement sites after worst-case half-cycles), maintenance of package integrity (30-day event-related shelf life), and simulated use testing (no growth after sterilization with soil challenge).
Technological Characteristics
Three-layer SMS (spunbound-meltblown-spunbound) polypropylene fabric; thermally bonded with round pin, hexagonal, triangle pattern. Dimensions vary by model (H100-H600). Single-use, non-sterile, OTC. No energy source, software, or connectivity.
Indications for Use
Indicated for use by healthcare providers to enclose reusable metal/nonmetal medical devices, including rigid/semi-rigid/flexible endoscopes and devices with diffusion-restricted spaces (e.g., hinged forceps/scissors), for sterilization in the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer (Lumen and Flexible cycles).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2019
Owens & Minor (O&M) Halyard, Inc % Peter Kalkbrenner Director of Engineering Sterilucent, Inc Minneapolis, Minnesota 55413
Re: K192147
Trade/Device Name: ONE-STEP Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: August 9, 2019 Received: August 13, 2019
Dear Peter Kalkbrenner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192147
### Device Name Halyard ONE-STEP* Sterilization Wrap
## Indications for Use (Describe)
The Halyard ONE-STEP* Sterilization Wraps are intended to allow sterilization of the enclosed devices by the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer (i.e., both the Lumen and Flexible Cycles). Additionally, the Halyard ONESTEP Sterilization Wrap was validated to allow effective aeration under the pre-programmed HC 80TT Sterilization Cycles. All models of the Halyard ONE-STEP Sterilization Wrap have been validated for use with the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer cycles as described below:
## Lumen Cycle:
Reusable metal and nonmetal devices including devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors and up to fifteen (15) stainless steel lumens per load with the following dimensions:
Single or dual channeled rigid and semi-rigid endoscopes, with stainless steel lumens that are: > 0.77 mm internal diameter (ID) and ≤ 410 mm long, or ≥ 1.33 mm ID and ≤ 430 mm long;
and, Triple channeled rigid and semi-rigid endoscopes, with stainless steel lumens that are ≥ 1.00 mm ID and ≤ 310 mm long. (Refer to the HC 80TT User Manual for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e. 20.1 lb. per load).
## Flexible Cycle:
Reusable rigid or semi-rigid non-lumen medical devices with metallic diffusion-restricted spaces such or mated surfaces such as the hinged portion of forceps or scissors;
Single channel flexible endoscopes with flexible lumens that are ≥ 1.00 mm ID and ≤ 1280 mm long; and Dual channel flexible endoscopes with flexible lumens that are ≥ 0.80 mm ID and ≤ 1000 mm long.
(Refer to the HC 80TT User Manual for complete instructions on load(s) and cycle(s), including chamber loading instructions (i.e. 25 lb. per load).
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 201 Subpart D) |
| <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart G) |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff@fda.hhs.gov
{3}------------------------------------------------
#### K192147 510K Summary
| Submitted by: | O&M Halyard, Inc. |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Steven Dowdley<br>Associate Director of Regulatory Affairs<br>O&M Halyard Inc.<br>Steven.dowdley@hyh.com<br>678-451-8062 |
| Date of Summary: | October 8, 2019 |
| Device Trade Name: | Halyard ONE-STEP* Sterilization Wrap |
| Common or Usual Name: | Sterilization Wrap |
| Classification: | 21 CFR 880.6850 |
| Class: | Class II |
| Product Code: | FRG |
| Predicate Device(s): | Halyard ONE-STEP* Sterilization Wrap - K141712 |
| Device Description: | Halyard ONE-STEP* Sterilization Wrap is comprised of two sheets of Halyard<br>Sterilization Wrap that is ultrasonically seamed on two edges. This seamed<br>configuration allows for convenient wrapping of an article using two sheets<br>simultaneously. |
| | The sheets of sterilization wrap are square or rectangular fabric produced<br>using a three layer SMS (spunbound-meltblown-spunbound) process. The<br>wrap fabric is composed of polypropylene with phthalocyanine blue pigment,<br>titanium dioxide pigment, and antistatic treatment. The wrap allows a<br>sterilized package to be opened aseptically. |
| Indications for Use | The Halyard ONE-STEP Sterilization Wraps are intended to allow sterilization<br>of the enclosed devices by the Sterilucent HC 80TT Hydrogen Peroxide<br>Sterilizer (i.e., both the Lumen and Flexible Cycles). Additionally, the Halyard<br>ONE-STEP Sterilization Wrap was validated to allow effective aeration under<br>the pre-programmed HC 80TT Sterilization Cycles. All models of the Halyard<br>ONE-STEP Sterilization Wrap have been validated for use with the Sterilucent<br>HC 80TT Hydrogen Peroxide Sterilizer cycles as described below:<br>Lumen Cycle:<br>Reusable metal and nonmetal devices including devices with diffusion-<br>restricted spaces such as the hinged portion of forceps and scissors and up to<br>fifteen (15) stainless steel lumens per load with the following dimensions:<br>Single or dual channeled rigid and semi-rigid endoscopes, with stainless steel<br>lumens that are:<br>$\geq$ 0.77 mm internal diameter (ID) and $\leq$ 410 mm long, or $\geq$ 1.33 mm ID and $\leq$<br>430 mm long;<br>and, Triple channeled rigid and semi-rigid endoscopes, with stainless steel<br>lumens that are $\geq$ 1.00 mm ID and $\leq$ 310 mm long. (Refer to the HC 80TT<br>User Manual for complete instructions on load(s) and cycle(s), including<br>chamber loading instructions (i.e. 20.1 lb. per load).<br>Flexible Cycle<br>Reusable rigid or semi-rigid non-lumen medical devices including non-lumen<br>devices with metallic diffusion-restricted spaces such or mated surfaces such<br>as the hinged portion of forceps or scissors,<br>Single channel flexible endoscopes with flexible lumens that are $\geq$ 100 mm ID<br>and $\leq$ 1280 mm long, and Dual channel flexible endoscopes with flexible<br>lumens that are $\geq$ 0.80 mm ID and $\leq$ 1000 mm long<br>(Refer to the HC 80TT User Manual for complete instructions on load(s) and<br>cycle(s), including chamber loading instructions (ie: 25 lb per load) |
{4}------------------------------------------------
পিন্ধার করে না করে আর এক প্রকাশ করেন। এখানে পারে প
িত থাকায় করে অনুষ্ঠিত হয়েছে একটি পার্টির প্রায় প্রতিষ্ঠা
## Technological Characteristics Table:
| | Proposed | Predicate |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Halyard ONE-STEP*<br>Sterilization Wrap (K192147) | KIMGUARD ONE-STEP*<br>Sterilization Wrap (K141712) |
| Manufacturer | Halyard | Halyard |
| Device Model<br>numbers | H100<br>H200<br>H300<br>H400<br>H500<br>H600 | KC100<br>KC200<br>KC300<br>KC400<br>KC500<br>KC600<br>(Please note that the device model<br>number are the same The alphabetical<br>prefix on the product was changed when<br>the company's name changed from<br>Kimberly Clark to Halyard) |
| Common or<br>Usual Name: | Sterilization Wrap | Sterilization Wrap |
| Classification: | 21 CFR 880.6850 | 21 CFR 880.6850 |
| Class: | II | II |
| Product Code | FRG | FRG |
| Manufacturer | Halyard | Halyard |
| Indication for<br>Use | The Halyard ONE-STEP Sterilization<br>Wraps are intended to allow sterilization<br>of the enclosed devices by the<br>Sterilucent HC 80TT Hydrogen Peroxide<br>Sterilizer (ie. both the Lumen and<br>Flexible Cycles) Additionally, the<br>Halyard ONESTEP Sterilization Wrap<br>was validated to allow effective aeration<br>under the pre-programmed HC 80TT<br>Sterilization Cycles. All models of the<br>Halyard ONE-STEP Sterilization Wrap<br>have been validated for use with the<br>Sterilucent HC 80TT Hydrogen Peroxide<br>Sterilizer cycles as described below | KIMGUARD ONE-STEP* Sterilization<br>Wraps are intended to enclose another<br>medical device that is to be sterilized by<br>a healthcare provider using<br>Sterilucent PSD-85 Hydrogen Peroxide<br>Sterilizer that include<br>• Lumen Cycle and<br>• Non-Lumen Cycle<br><br>KIMGUARD ONE-STEP* Sterilization<br>Wrap (KC100, KC200, KC300, KC400,<br>KC500 and KC600) are intended to allow<br>sterilization of the enclosed medical<br>device(s) and also maintain sterility of the<br>enclosed device(s) until used<br><br>Test results validated that KIMGUARD |
{5}------------------------------------------------
### Lumen Cycle
Reusable metal and nonmetal devices including devices with diffusion- restricted spaces such as the hinged portion of forceps and scissors and up to fifteen (15) stainless steel lumens per load with the following dimensions:
Single or dual channeled rigid and semirigid endoscopes. with stainless steel lumens that are ≥ 0.77 mm internal diameter (ID) and ≤ 410 mm long, or ≥ 1.33 mm ID and ≤ 430 mm long, and, Triple channeled rigid and semi- rigid endoscopes, with stainless steel lumens that are ≥ 1.00 mm ID and ≤ 310 mm long (Refer to the HC 80TT User Manual for complete instructions on load(s) and cycle(s), including chamber loading instructions (ie. 20.1 Ib per load)
### Flexible Cycle
Reusable rigid or semi-rigid non-lumen medical devices including non-lumen devices with metallic diffusion-restricted spaces such or mated surfaces such as the hinged portion of forceps or scissors. Single channel flexible endoscopes with flexible lumens that are ≥ 1.00 mm ID and ≤1280 mm long, and Dual channel flexible endoscopes with flexible lumens that are ≥ 0.80 mm ID and ≤1000 mm long (Refer to the HC 80TT User Manual for complete instructions on load(s) and cycle(s), including chamber loading instructions (ie. 25 Ib per load)
ONE-STEP* Sterilization Wraps (KC100, KC200, KC300, KC400, KC500,and KC600) allowed sterilization of the enclosed devices by the Sterilucent PSD- 85 Hydrogen Peroxide Sterilizer (ie. both the Lumen and Non-Lumen Cycles) Additionally, the KIMGUARD ONE-STEP* Sterilization Wrap was validated to allow effective aeration under the pre- programmed PSD-85 Sterilization Cycles
The PSD-85 Lumen Cycle has been validated to sterilize a load of up to ten (10) pounds (combined pouch and wrapped tray load) containing a maximum of ten (10) single channel stainless steel lumens per load with the following dimensions
- . An inside diameter of 1 mm or larger and a length of 60 mm or shorter,
- An inside diameter of 2 mm or . larger and a length of 250 mm or shorter,
- . An inside diameter of 3 mm or larger and a length of 350 mm or shorter
The PSD-85 Non-Lumen Cycle has been validated to sterilize a load of up to 25 pounds (combined pouch and wrapped tray load)
All models of the KIMGUARD ONE- STEP* Sterilization Wrap (KC100, KC200, KC300, KC400. KC500. and KC600) have been validated for use with the Sterilucent PSD-85 Hydrogen Peroxide Sterilizer cycles listed below
### Lumen
Reusable metal and nonmetal devices including devices with diffusion-restricted spaces such as the hinged portion of forceps and scissors and up to 10 single channel stainless steel lumened devices of the following dimensions per chamber load
o An inside diameter of 1 mm or larger and a length of 60 mm or shorter
o An inside diameter of 2 mm or larger and a length of 250 mm or shorter
o An inside diameter of 3 mm or larger and a length of 350 mm or shorter
{6}------------------------------------------------
| Sterilization<br>Parameters | Stenlucent HC 80TT Hydrogen<br>Peroxide Sterilizer that includes<br>• Lumen Cycle<br>• Flexible Cycle | Sterilucent PSD-85 Hydrogen Peroxide<br>Sterilizer that includes<br>• Lumen Cycle<br>• Non-Lumen Cycle<br>(Refer to the PSD-85 User Manual for<br>complete instructions on load(s) and<br>cycle(s), including chamber loading<br>instructions (ie. 10 lbs per load))<br>Non-Lumen<br>Non-lumened reusable metal and nonmetal<br>devices including devices with stainless<br>steel diffusion-restricted spaces such as<br>the hinged portion of forceps and scissors<br>(Refer to the PSD-85 User Manual for<br>complete instructions on load(s) and<br>cycle(s), including chamber loading<br>instructions (ie. 25 lbs per load)) |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology | Tortuous sheet material used to enclose<br>medical devices that are to be sterilized<br>by a healthcare provider to allow<br>sterilization of the enclosed medical<br>device(s) and maintain sterility of the<br>enclosed device(s) | Tortuous sheet material used to enclose<br>medical devices that are to be sterilized<br>by a healthcare provider to allow<br>sterilization of the enclosed medical<br>device(s) and maintain sterility of the<br>enclosed device(s) |
| Device Design | Two sheets of nonwoven polypropylene<br>fabric Each sheet is composed of three<br>thermally- bonded layers consisting of a<br>meltblown polypropylene layer<br>surrounded by spunbound polypropylene<br>layers (SMS) | Two sheets of nonwoven polypropylene<br>fabric Each sheet is composed of three<br>thermally- bonded layers consisting of a<br>meltblown polypropylene layer surrounded<br>by spunbound polypropylene layers<br>(SMS) |
| Method for<br>bonding SMS layers | Thermal bonding with round pin,<br>hexagonal, triangle bond pattern ("daisy"<br>pattern) | Thermal bonding with round pin<br>hexagonal, triangle bond pattern ("daisy"<br>pattern) |
| Materials | Polypropylene with blue and white<br>pigments | Polypropylene with blue and white<br>pigments |
| Distribution | Non-Sterile and Over-the-Counter | Non-Sterile and Over-the-Counter |
| Single Use<br>Device | Yes | Yes |
{7}------------------------------------------------
## Performance Testing Summary
Performance Testing (Bench):
Performance of Halyard ONE-STEP* Sterilization Wrap (H100, H200, H300, H400, H500, H600) has been evaluated to show that all results of testing met acceptance criteria, demonstrating that the Halyard ONE-STEP* Sterilization Wrap allows sterilization of contents by the Sterilucent HC 80TT Hydrogen Peroxide Sterilizer and maintains sterility of contents for the designated test period (30 Days).
| est | Description | Result |
|----------------------------------|---------------------------------------------------------------------|--------|
| Sterilant Penetration / Efficacy | Demonstration of no growth of challenge microorganisms in | Passed |
| | inoculation BI placement sites of claimed challenge loads | |
| | after exposure to worst case half-cycles sterilization cycle | |
| | conditions for both the lumen and flexible sterilization cycles. | |
| Maintenance of Package Integrity | Demonstration of no growth of microorganisms after 30 Day Passed | |
| | event-related shelf life testing using simulated worst case | |
| | handling conditions. | |
| Simulated use testing | Demonstration no growth of challenge microorganisms, after Passed | |
| | sterilization, using an inoculum in the presence of a defined | |
| | soil challenge | |
| Performance Testing<br>(Clinical): | Clinical evaluations were not required and therefore are not submitted with<br>this 510(k) |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Discussion: | The Halyard ONE-STEP* Sterilization Wrap and the predicate device are<br>both intended to enclose another medical device that is to be sterilized by a<br>healthcare provider, to allow sterilization of the enclosed medical device(s)<br>and to maintain sterility of the enclosed device(s). Both devices have the<br>same intended use, design, materials, specifications, and composition, and<br>are manufactured using the exact same production methods |
| Overall Performance<br>Conclusions | Based on the nonclinical tests performed the subject device, the Halyard<br>ONE-STEP* Sterilization Wrap for use with the Sterilucent HC 80TT<br>Hydrogen Peroxide Sterilizer, is as safe, as effective and performs as well or<br>better than the legally marketed predicate device, the Halyard ONE-STEP*<br>Sterilization Wrap for use with the PSD-85 Hydrogen Peroxide Sterilizer |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.