K982332 · STERIS Corporation · FLE · Jan 20, 1999 · General Hospital
Device Facts
Record ID
K982332
Device Name
STERIS TEN SIXTEEN STEAM STERILIZER
Applicant
STERIS Corporation
Product Code
FLE · General Hospital
Decision Date
Jan 20, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Indications for Use
The STERIS Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture- stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.
Device Story
Tabletop steam sterilizer; uses saturated steam under pressure to achieve sterilization. Operates via factory-programmed cycles (Unwrapped, Wrapped, Liquid, Packed) with specific temperature, time, and drying parameters. Used in hospitals, clinics, and dental/veterinary offices; operated by trained healthcare personnel. Device monitors process conditions; includes built-in safeguards to abort cycles and provide alerts if parameters are not met. Healthcare providers use biological and chemical indicators to verify sterility assurance. Benefits include effective decontamination of reusable medical devices and reduction of cross-contamination risks.
Clinical Evidence
No clinical data. Efficacy demonstrated via bench testing using Bacillus stearothermophilus spores as the Most Resistant Organism (MRO) to validate sterilization cycles. Testing confirms sterility assurance level of less than 10^-6. Performance validated against AAMI-ST55 and AAMI-ST37 standards.
Technological Characteristics
Tabletop steam sterilizer; 10" x 16" chamber. Materials and construction meet ASME Section VII, Division 1 standards. Energy source: electrical/steam. Safety certifications: UL 544, CSA C22.2 No. 125. Features automated cycle control, failure mode monitoring, and alarm systems for process deviations.
Indications for Use
Indicated for sterilization of heat and moisture-stabile materials in hospital, dental, or veterinary environments. Loads include unwrapped instruments (no lumens/dental handpieces), wrapped instrument packs, paper-peel pouches, vented borosilicate glass containers (≤500ml), and fabric packs. Contraindicated for items sensitive to high temperature, pressure, or moist heat.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
Related Devices
K993856 — TUTTNAUER ELECTRONIC AUTOCLAVE, MODELS E, EA, EK, EKA, EP · Tuttnauer U.S.A Co, Ltd. · Feb 9, 2001
K021223 — BELIMED STEAM STERILIZER, MODEL TOP 5000 · Sauter AG · Jun 26, 2002
K021504 — TUTTNAUER EHS SERIES TABLE-TOP AUTOCLAVE, MODEL 2540 · Tuttnauer U.S.A Co, Ltd. · Jul 25, 2002
K111736 — TUTTNAUER EZ PLUS SERIES ELECTRONIC TABLETOP AUTOCLAVES · Tuttnauer Co. , Ltd. · Dec 8, 2011
{0}------------------------------------------------
JAN 20 1999
K 982332
## A 510(k) SUMMARY PERTAINING TO THE SAFETY AND EFFECTIVENESS OF THE STERIS TEN SIXTEEN STEAM STERILIZER
## Submitter Information.
Raymond Ursick Senior Director, Regulatory Affairs and Quality Systems 5960 Heisley Road Mentor, Ohio 44060 (440) 354-2600 Phone Date Summary Prepared: June 26, 1998
Introduction. Moist heat in the form of saturated steam under pressure is the oldest, safest and most dependable form of sterilization. Steam sterilization is universally recognized as the method of choice, except when high temperatures, pressure and/or moist heat would damage or degrade items to be sterilized. Steam sterilizers in a wide variety of sizes are used in hospitals, clinics, doctors' and dentists' offices, clinical and research laboratories and industry. Uses include sterilization of reusable medical devices such as surgical instruments, gowns and drapes; sterilization of single-use medical devices such as syringes and catheters; decontamination of potentially infectious waste generated during surgery or research; sterilization of other materials, instruments and goods used in biotechnology and pharmaceutical houses, and in other applications whenever biological contamination and infection control are essential.
The STERIS Ten Sixteen Steam Sterilizer is a Class II medical device as defined by 21 CFR §880.6880. The Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture-stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.
STERIS's manufacturing facility for the STERIS Ten Sixteen Steam Sterilizer is located in Erie, Pennsylvania and is registered with and regulated by the FDA. STERIS's Erie facility is inspected by the FDA to assure that the requirements of Quality System Regulation are met. The Erie manufacturing facility is certified to the ISO 9001 Quality System Standard and the EN Medical Device Directive.
{1}------------------------------------------------
The STERIS Ten Sixteen Steam Sterilizer cycle descriptions are as follows:
| CYCLES | RECOMMENDED<br>LOAD | STERILIZE<br>TEMP. | STERILIZE<br>TIME | DRY<br>TIME |
|-----------|------------------------------------------------------------------------------------------------------|--------------------|-------------------|---------------------------------|
| UNWRAPPED | Unwrapped Instrument<br>tray with a single<br>instrument (No lumens<br>and no dental<br>handpieces). | 270°F (132°C) | 3 minutes | 0 minute |
| WRAPPED | Wrapped instrument<br>pack. Paper-peel<br>Pouches. (No lumens<br>and no dental<br>handpieces). | 270°F (132°C) | 10 minutes | 45<br>minutes |
| LIQUID | Vented borosilicate glass<br>containers (500ml or<br>smaller). | 250°F (121°C) | 40 minutes | 10<br>minutes<br>(Cool<br>Time) |
| PACKED | Fabric packs. | 250°F (121°C) | 30 minutes | 30<br>minutes |
Effectiveness. Under controlled conditions of temperature and pressure, steam is capable of killing all forms of sporulating and vegetative bacteria, viruses, fungi, yeast and bacterial and fungal endospores. It has been well established that Bacillus stearothermophilus in the form of spores is the Most Resistant Organism (MRO) to steam. Thus, during the development, qualification and validation of STERIS steam sterilizers, B. stearothermophilus spores are used in order to ensure a worst-case test of the steam sterilizer's efficacy in killing all forms of microbial and viral contamination.
Efficacy of sterilizer function and exposure time recommendations are ultimately demonstrated by showing complete kill of biological indicators and documenting an appropriate safety factor or sterility assurance level of less than 10* (probability of less than one chance out of one million of a non-sterile indicator). STERIS qualifies its recommended sterilization cycles through the application of rigorous performance standards and recommended practices developed by various independent organizations that have specified worst-case loads. The STERIS Ten Sixteen Steam Sterilizer meet the applicable requirements of the AAMI-ST55 "Tabletop Steam Sterilizers" performance standard. The STERIS Ten Sixteen Unwrapped Cycle meets the applicable requirements of AAMI-ST37 standard for flash sterilization. Biological indicators used in qualification studies were in compliance with the Association for the Advancement of Medical Instrumentation (AAMI) standard for biological indicators (AAMI-ST19).
{2}------------------------------------------------
Safety. STERIS sterilizers have been designed, constructed and tested to meet the safety and performance requirements of various national codes and standards. Before release to market, STERIS sterilizers are tested and certified by independent testing agents to the following requirements:
- Underwriters Laboratory (UL) 544 as certified by ETL Testing Laboratories, Inc.; 1.
- American Society of Mechanical Engineers (ASME), Section VII, Division 1; and 2.
- Canadian Standards Association (CSA) C22.2 No. 125 as certified by ETL Testing 3. Laboratories, Inc.
A Failure Modes Effect Criticality Analysis was also conducted on the STERIS Ten Sixteen Steam Sterilizer's electrical, mechanical and piping system.
Hazards-Failure of Performance. Failure of the sterilization process can lead to incidents of cross contamination, the transmission of potentially infectious organisms from an infected person to another who was not otherwise infected prior to the incident.
To avoid failure, the operator must ensure the following:
- That the materials, instruments and devices to be sterilized are thoroughly 1. cleaned and decontaminated;
- That the manufacturer's instructions for use are followed; 2.
- 3. That the operator must select the validated cycle applicable for the load;
- That the sterilizer has been maintained in accordance with the manufacturer's 4. recommended maintenance schedule and is operating properly; and
- That each sterilizer load is monitored with available biological and chemical 5. sterilization process indicators.
Today there are thousands of steam sterilizers in daily use in hospitals and other healthcare environments throughout the United States running several million cycles per year. Despite its widespread use, the incidence of sterilizer malfunction or sterilization process failure is relatively rare. Further, there are no known reports in the literature of patient infections that have resulted from steam sterilizer failure. Although the ability to prove that nosocomial infections result from failure of the sterilization process or a malfunction of the sterilizer is at best difficult, the technology designed in today's STERIS sterilizers provides numerous builtin safeguards that abort the cycle and give appropriate signals, alerts and warnings when required conditions have not been met or when a malfunction occurs.
User Information. STERIS conducts in-house customer training and has developed a series of user training videos that provide helpful information about the appropriate use of steam sterilizers. Through its careful attention to detail in instructions for use and other labeling, STERIS further provides information to the user that is intended to ensure safe and effective use of steam sterilization. Numerous recommended practice documents have been developed by AAMI and AORN to give guidance in techniques that are intended to reduce the risks of cross-contamination. The use and periodic review of, as well as adherence to these materials, give further assistance of the safe and effective use of steam sterilization equipment in healthcare environments.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three profiles, representing health, human services, and the USA.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 20 1999
Mr. Raymond Ursick Senior Director Regulatory Affairs and Quality Systems Steris® Corporation 5960 Heisley Road Mentor, Ohio 44060
K982332 Re : STERIS Ten Sixteen Steam Sterilizer Trade Name: Requlatory Class: II FLE Product Code: October 21, 1998 Dated: Received: October 22, 1998
Dear Mr. Ursick:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਕੇ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Ursick
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Tim Director
Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## STERIS TEN SIXTEEN TABLETOP STEAM STERILIZER
## INDICATIONS FOR USE:
The STERIS Ten Sixteen is a tabletop steam sterilizer designed to sterilize heat and moisture- stabile materials used in hospital, dental or veterinary environments. The STERIS Ten Sixteen Tabletop Steam Sterilizer is available in a 10" x 16" inch chamber configuration.
The STERIS Ten Sixteen Tabletop Steam Sterilizer is equipped with the following factoryprogrammed set sterilization cycles and cycle values:
| CYCLES | RECOMMENDED LOAD | STERILIZE<br>TEMP. | STERILIZE<br>TIME | DRY TIME |
|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|-------------------|---------------------------|
| UNWRAPPED | Unwrapped Instrument tray with<br>a single instrument (No lumens<br>and no dental handpieces). | 270°F (132°C) | 3 minutes | 0 minute |
| WRAPPED | Wrapped instrument pack: one<br>pack, maximum weight 1300 g<br>(2.87 lbs). Paper-peel Pouches:<br>maximum 5 pouches, one<br>instrument per pack, total<br>maximum weight 500 g (1.1 lb).<br>(No lumens and no dental<br>handpieces). | 270°F (132°C) | 10 minutes | 45 minutes |
| LIQUID | Vented Borosilicate glass<br>containers (500ml or smaller). | 250°F (121°C) | 40 minutes | 10 minutes<br>(Cool Time) |
| PACKED | Fabric pack(s): maximum weight<br>1050 g (2.32 lbs). | 250°F (121°C) | 30 minutes | 30 minutes |
(Please Do Not Write Below This Line - Continue on Another Page If Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Qam Silm
(Optional Format 1-2-96)
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.