TUTTNAUER ELECTRONIC AUTOCLAVE, MODELS E, EA, EK, EKA, EP
K993856 · Tuttnauer U.S.A Co, Ltd. · FLE · Feb 9, 2001 · General Hospital
Device Facts
Record ID
K993856
Device Name
TUTTNAUER ELECTRONIC AUTOCLAVE, MODELS E, EA, EK, EKA, EP
Applicant
Tuttnauer U.S.A Co, Ltd.
Product Code
FLE · General Hospital
Decision Date
Feb 9, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tuttnauer electronic controlled autoclaves are intended to be used for sterilization of solid metal instruments, solid metal hinged instruments, dental handpieces and other materials that will withstand temperatures and pressure required for sterilization. The sterilizers have 3 cycles: unwrapped instruments (134°C at 4 minutes and 29 psi); wrapped instruments (134°C at 8 minutes and 29 psi); and liquids (121℃ at 30 minutes and 16 psi). These cycles are preprogrammed and are recommended in the labeling. Tuttnauer E Series Electronic Table-top Autoclaves are intended for use in ophthalmic, dental and medical clinics, first aid rooms, and small laboratories in conformity with 21 C.F.R. § 880.6880.
Device Story
Table-top steam sterilizer; utilizes electronic control, water reservoir, heat source, and sealed stainless steel chamber. Operates via three pre-programmed cycles: unwrapped instruments (134°C/4 min), wrapped instruments (134°C/8 min), and liquids (121°C/30 min). Used in dental, medical, and ophthalmic clinics; first aid rooms; small laboratories. Microprocessor-controlled; ensures sterilization of instruments, including dental handpieces. Healthcare providers load instruments into chamber; select cycle; device executes automated sterilization process. Benefits include effective sterilization of medical/dental instruments and liquids.
Clinical Evidence
Bench testing only. Validation studies performed using representative dental handpieces inoculated in difficult-to-sterilize areas. Three half-cycle tests conducted for both wrapped and unwrapped methods to confirm effective sterilization of up to 5 dental handpieces.
Indicated for sterilization of solid metal instruments, solid metal hinged instruments, dental handpieces, and other materials capable of withstanding required sterilization temperatures and pressures in ophthalmic, dental, and medical clinics, first aid rooms, and small laboratories.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
FEB - 9 2001
993856
## 510(K) SUMMARY
| Submitter: | Tuttnauer USA Co. Ltd.<br>33 Comac Loop, Equi-Park<br>Ronkonkoma, NY 11779 | | |
|----------------------|----------------------------------------------------------------------------|------------------------------------------------------------|--|
| | Phone: | 800-624-5836 | |
| | Fax: | 516-737-0720 | |
| Contact Name: | Robert R. Basile | | |
| Date Prepared: | October 25, 1999 | | |
| Common Name: | Electronic autoclave, table-top steam sterilizer | | |
| Trade Name: | Tuttnauer Electronic Autoclave Models (E, EA, EK, EP, EKA) | | |
| Classification Name: | | Steam Sterilizer<br>Class II Device - 21 C.F.R. § 880.6880 | |
## Substantial Equivalence:
The modified Tuttnaver E Series Electronic Table-top Autoclave is substantially equivalent to the following currently cleared and legally marketed autoclave:
| Company | Product Name | 510(k) Clearance Number |
|------------------------|-------------------------------|-------------------------|
| Tuttnauer USA Co. Ltd. | E Series Electronic Autoclave | K920478 |
With the exception of changes to the automated program cycle times and the labeling to allow sterilization of dental handpieces, the modified devices are identical to the Model E cleared under K920478.
#### General Description:
The Tuttnauer Model E series autoclave is a table-top steam sterilizer that includes as its main components an electronic control, water reservoir, heat source and sealed chamber. The E has three pre-programmed sterilization cycles. The three pre-programmed cycles are for unwrapped instruments (134°C at 4 minutes), wrapped instruments (134°C at 8 minutes) and liquids (121°C at 30 minutes). Dental handpieces can be sterilized as unwrapped or wrapped instruments without modification to cycle times or temperatures.
The Model E autoclave is produced in four sizes, the 1730 with a chamber size of 7" x 12"; the 2340 with a chamber size of 9" x 18.5"; the 2540 with a chamber size of 10" x 18.7"; and the 3870, which has a chamber size of 15" x 27."
{1}------------------------------------------------
Tuttnauer E Series Dental Handpiece Modification 510(k) Summary Page 2
### Design and Materials:
The pressure chamber, door and locking device are constructed from stainless steel. The pressure chamber, and components are designed, tested and stamped for conformance to ASME code.
A microprocessor controls the cycle parameters. The software provides the microprocessor with the necessary information to control the device within its desired parameters. The software has been validated to ensure compliance with all applicable requirements.
#### Intended Use:
The Tuttnauer E Series Electronic Table-top Autoclave is intended to provide sterilization of medical and dental instruments (including dental handpieces) and to sterilize liquids for non-clinical applications.
#### Technology Considerations:
The modified E Series exhibits the same technological characteristics as the Tuttnauer E Series Electronic Table top Autoclave, its predicate device. All of the components that may be found in the unmodified Model E are identical to those in the modified version.
#### Non-Clinical Testing:
Tuttnauer conducted validation studies on the E Series Electronic Table-top Autoclave to ensure effective handpiece sterilization. Representative handpieces from various manufacturers were inoculated in their most difficult to sterilize areas for the testing. Worst-case handpiece loads were then determined for the sterilizer. Three half-cycle tests were performed for each of unwrapped and wrapped methods of sterilization. The testing shows that dental handpieces may be effectively sterilized in the E Series Electronic Table-top Autoclave.
#### Conclusion:
It is Tuttnauer USA Co. Ltd.'s conclusion that the modified Tuttnauer E Series Electronic Table-top Autoclave is substantially equivalent to its predicate device. Based upon test data submitted, Tuttnauer E Series Table-top autoclaves effectively sterilize up to 5 dental handpieces contained within them during normal autoclave operation.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of what appears to be an abstract caduceus symbol, featuring three wavy lines that resemble a stylized human figure.
#### Public Health Service
.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 9 2001
Tuttnauer USA Company Limited C/O Mr. Mark M. Yacura Buchanan Ingersoll, P.C. 1776 K Street, Northwest Suite 800 Washington, DC 20006-7910
Re: K993856 Tuttnauer Models 1730, 2340, 2540 and 3870 Trade Name: "E" Series Table-top Autoclave Regulatory Class: I Product Code: FLE Dated: January 26, 2001 Received: January 26, 2001
Dear Mr. Yacura:
We have reviewed your Section 510(k) notification of intent to we have reviewed your beoced above and we have determined the market the device islarguivalent (for the indications for use device is bubbeanceary eqo legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the marketed in theorbeate of the Medical Device Amendments, or to devices enatement dace of the neved in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, the reactar rood, brag, and, subject to the general controls chercrore, marked one act. The general controls provisions of the provisions or the nec. If annual registration, listing of Act include requirements re practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II il your device ID Class III (Premarket Approval), it may be (Spectal Controlo) or Srabbrontrols. Existing major regulations subject to back device can be found in the Code of Federal arrecting your dover that 800 to 895. A substantially Reguiralent determination assumes compliance with the Current Good equivalence decormination abiaments, as set forth in the Quality
Manufacturing Practice requirements, as set forth in the publicition rundracearing Fring (QS) for Medical Devices: General regulation System Regulacion (20) 102 through periodic QS inspections, the (Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further regulatory accron. In a your device in the Federal Register. announcements this response to your premarket notification ricuse note. Enro respect any obligation you might have under sections 531 through 542 of the Act for devices under the
{3}------------------------------------------------
Page 2 — Mr. Yacura
Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as mis icece wour 510 (k) premarket notification. The FDA described in your 510 m provilence of your device to a legally marketed predicate device results in a classification for your marketed predicate device robates in proceed to the market.
If you desire specific advice for your device on our labeling II you desire specific advice is additionally 809.00 for in regulation (21 crk rares), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on Compilance at (501) 32 resertising of your device, please contact che promocion and adversibling on July 2017 - 11so, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be Information on your responsible of Small Manufacturers Assistance obcarned from the Drviber. 01 638-2041 or (301) 443-6597 or at its internet address
its incerner addroby/cdrh/dsma/dsmamain.html".
Sincerely yours,
Susan Purpus
Timothy A.Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number:
Device Name:
Tuttnauer Models 1730, 2340, 2540 and 3870 "E" Series Table-top Autoclave
Tuttnauer electronic controlled autoclaves are intended to be used for Indications For Use: sterilization of solid metal instruments, solid metal hinged instruments, dental handpieces and other materials that will withstand temperatures and pressure required for sterilization.
> The sterilizers have 3 cycles: unwrapped instruments (134°C at 4 minutes and 29 psi); wrapped instruments (134°C at 8 minutes and 29 psi); and liquids (121℃ at 30 minutes and 16 psi). These cycles are preprogrammed and are recommended in the labeling.
Tuttnauer E Series Electronic Table-top Autoclaves are intended for use in ophthalmic, dental and medical clinics, first aid rooms, and small laboratories in conformity with 21 C.F.R. § 880.6880.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .............................................................................................................................................................................. .............................................................................................................................................................................. Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use __
Over -The-Counter Use
(Per 21 C.F.R. § 801.109)
(Optional Format 1-2-96)
Qun S. Lim
OR
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital D 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.