Nitrile Powder-Free Exam Gloves 2.8, Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (light blue, dark blue)
K250193 · Basic Medical Technology, Inc. · LZA · Jul 9, 2025 · General Hospital
Device Facts
Record ID
K250193
Device Name
Nitrile Powder-Free Exam Gloves 2.8, Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (light blue, dark blue)
Applicant
Basic Medical Technology, Inc.
Product Code
LZA · General Hospital
Decision Date
Jul 9, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl and Gastric acid in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
Device Story
Nitrile powder-free examination gloves (light blue and dark blue variants) serve as a physical barrier between patient and clinician to prevent cross-contamination. Used in clinical settings by healthcare professionals. Device provides protection against specific chemotherapy agents, fentanyl citrate, and simulated gastric acid. Performance validated via permeation testing per ASTM D6978-05. Output is a protective barrier; clinician visually inspects glove integrity before use. Benefits include reduced exposure risk to hazardous substances during patient examination. Warning provided against use with Carmustine and Thiotepa due to rapid breakthrough times.
Clinical Evidence
Bench testing only. Permeation testing conducted per ASTM D6978-05 to determine breakthrough detection times for various chemotherapy drugs, fentanyl citrate, and simulated gastric acid. Results demonstrate >240 minutes for most tested substances, with specific exceptions for Carmustine and Thiotepa noted in labeling.
Technological Characteristics
Nitrile material; powder-free; disposable; form-fitting glove design. Tested for chemical permeation per ASTM D6978-05. No energy source or software components.
Indications for Use
Indicated for use as a disposable medical glove worn by examiners to prevent cross-contamination between patient and examiner. Suitable for use with specific chemotherapy drugs, opioid fentanyl citrate, and simulated gastric acid. Contraindicated for use with Carmustine and Thiotepa due to low permeation times.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K232266 — Powder Free Nitrile Patient Examination Gloves, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate · Better Care Plastic Technology Co., Ltd. · Nov 3, 2023
K250944 — Nitrile Powder Free Examination Glove, Blue Tested for Use with 60 Chemotherapy Drugs, Gastric Acid, Xylazine and Fentanyl · Gmp Medicare Sdn. Bhd. · Apr 23, 2025
K203236 — Rhino Health Non-Sterile, Powder-free Nitrile Exam Glove · Rhino Health, Inc. · Aug 16, 2021
K183562 — Powder-Free Nitrile Examination Glove (White Colored, Blue Colored, and Black Colored) Tested for Use with Chemotherapy Drugs · Encompass Industries Sdn. Bhd. · Oct 23, 2019
K200453 — NON-STERILE, POWDER FREE NITRILE EXAMINATION GLOVES, LOW DERMATITIS POTENTIAL, AND TESTED FOR USE WITH CHEMOTHERAPY DRUGS BLUE-AC · Yty Industry (Manjung) Sdn Bhd · May 1, 2020
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 9, 2025
Basic Medical Technology Inc.
John Zhao
General manager
5300 Concours Street
Ontario, California 91764
Re: K250193
Trade/Device Name: Nitrile Powder-Free Exam Gloves 2.8, Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (light blue, dark blue)
Regulation Number: 21 CFR 880.6250
Regulation Name: Non-powdered patient examination glove
Regulatory Class: Class I, reserved
Product Code: LZA, LZC, OPJ, QDO
Dated: June 20, 2025
Received: June 20, 2025
Dear John Zhao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250193 - John Zhao
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250193 - John Zhao
Page 3
Sincerely,
ALLAN GUAN -S
For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K250193
Device Name
Nitrile Powder-Free Exam Gloves 2.8, Tested for Use with Chemotherapy Drugs, Opioid Fentanyl Citrate, Simulated Gastric Acid (light blue, dark blue)
Indications for Use (Describe)
The glove is disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves were tested for use with Chemotherapy drugs, Fentanyl and Gastric acid in accordance with ASTM D6978-05 standards Practice for assessment of Medical Glove to Permeation by chemotherapy drugs.
Light Blue glove:
| Chemotherapy Drugs | Minimum Breakthrough Detection Time, minutes |
| --- | --- |
| 1 Bendamustine HCl, 5 mg/ml | >240 min |
| 2 Bortezomib (Velcade), 1 mg/ml | >240 min |
| 3 Busulfan, 6 mg/ml | >240 min |
| 4 Carboplatin, 10 mg/ml | >240 min |
| 5 Carmustine, 3.3 mg/ml | 12.6 min. |
| 6 Carfilzomib, 2 mg/ml | >240 min. |
| 7 Cetuximab (Erbitux), 2 mg/ml | >240 min. |
| 8 Chloroquine, 50mg/ml | >240 min. |
| 9 Cisplatin, 1 mg/ml | >240 min. |
| 10 Cladribine, 1.0 mg/ml | >240 min. |
| 11 Cyclophosphamide, 20 mg/ml | >240 min. |
| 12 Cyclosporin A, 100mg/ml | >240 min. |
| 13 Cytarabine HCl, 100 mg/ml | >240 min. |
| 14 Cytovene, 10 mg/ml | >240 min. |
| 15 Dacarbazine, 10 mg/ml | >240 min. |
| 16 Daunorubicin HCl, 5 mg/ml | >240 min. |
| 17 Decitabine, 5 mg/ml | >240 min. |
| 18 Docetaxel, 10 mg/ml | >240 min. |
| 19 Doxorubicin HCl, 2 mg/ml | >240 min. |
| 20 Epirubicin HCl, 2 mg/ml | >240 min. |
| 21 Etoposide, 20 mg/ml | >240 min. |
| 22 Fludarabine Phosphate, 25 mg/ml | >240 min. |
| 23 5-Fluorouracil, 50 mg/ml | >240 min. |
| 24 Gemcitabine HCl, 38 mg/ml | >240 min. |
| 25 Idarubicin HCl, 1 mg/ml | >240 min. |
| 26 Ifosfamide, 50 mg/ml | >240 min. |
| 27 Irinotecan HCl, 20 mg/ml | >240 min. |
| 28 Mechlorethamine HCl, 1 mg/ml | >240 min. |
| 29 Melphalan HCl, 5 mg/ml | >240 min. |
| 31 Methotrexate, 25 mg/ml | >240 min. |
| 30 Mitomycin-C, 0.5 mg/ml | >240 min. |
| 32 Mitoxantrone HCl, 2 mg/ml | >240 min. |
| 33 Oxaliplatin, 2 mg/ml | >240 min. |
| 34 Pacilitaxel, 6 mg/ml | >240 min. |
| 35 Retrovir, 10 mg/ml | >240 min. |
FORM FDA 3881 (8/23)
Page 1 of 3
PSC Publishing Services (301) 443-6740
{4}
36 ThioTEPA, 10 mg/ml
27.1 min.
37 Trisonex, 1 mg/ml
>240 min.
38 Vidaza(Azacitidine),25 mg/ml
>240 min.
39 Vincristine Sulfate,1 mg/ml
>240 min.
Fentanyl, Gastric Acid and Xylazine
Minimum Breakthrough Detection Time
1 Fentanyl Citrate Injection,50mcg/ml
>240 min.
2 Gastric Acid
>240 min.
3 Fentanyl Citrate: Xylazine HCl, 50:50
>240 min.
*Please note that the following drugs have extremely low permeation times:
Carmustine: 12.6 minutes, Thio Tepa:27.1 minutes
Warning: Do not use with Carmustine and Thiotepa.
Dark Blue Glove:
Chemotherapy Drugs
Minimum Breakthrough Detection Time, minutes
1 Bendamustine HCl, 5 mg/ml
>240 min
2 Bortezomib (Velcade), 1 mg/ml
>240 min
3 Busulfan, 6 mg/ml
>240 min
4 Carboplatin, 10 mg/ml
>240 min
5 Carmustine, 3.3 mg/ml
11.0 min.
6 Carfilzomib, 2 mg/ml
>240 min.
7 Cetuximab (Erbitux), 2 mg/ml
>240 min.
8 Chloroquine, 50mg/ml
>240 min.
9 Cisplatin, 1 mg/ml
>240 min.
10 Cladribine, 1.0 mg/ml
>240 min.
11 Cyclophosphamide, 20 mg/ml
>240 min.
12 Cyclosporin A,100mg/ml
>240 min.
13 Cytarabine HCl, 100 mg/ml
>240 min.
14 Cytovene, 10 mg/ml
>240 min.
15 Dacarbazine, 10 mg/ml
>240 min.
16 Daunorubicin HCl, 5 mg/ml
>240 min.
17 Decitabine, 5 mg/ml
>240 min.
18 Docetaxel, 10 mg/ml
>240 min.
19 Doxorubicin HCl, 2 mg/ml
>240 min.
20 Epirubicin HCl, 2 mg/ml
>240 min.
21 Etoposide, 20 mg/ml
>240 min.
22 Fludarabine Phosphate, 25 mg/ml
>240 min.
23 5-Fluorouracil, 50 mg/ml
>240 min.
24 Gemcitabine HCl, 38 mg/ml
>240 min.
25 Idarubicin HCl, 1 mg/ml
>240 min.
26 Ifosfamide, 50 mg/ml
>240 min.
27 Irinotecan HCl, 20 mg/ml
>240 min.
28 Mechlorethamine HCl, 1 mg/ml
>240 min.
29 Melphalan HCl, 5 mg/ml
>240 min.
31 Methotrexate, 25 mg/ml
>240 min.
30 Mitomycin-C, 0.5 mg/ml
>240 min.
32 Mitoxantrone HCl, 2 mg/ml
>240 min.
33 Oxaliplatin, 2 mg/ml
>240 min.
34 Pacilitaxel, 6 mg/ml
>240 min.
35 Retrovir, 10 mg/ml
>240 min.
36 ThioTEPA, 10 mg/ml
26.0 min.
37 Trisonex, 1 mg/ml
>240 min.
FORM FDA 3881 (8/23)
Page 2 of 3
PSC Publishing Services (301) 443-6740
{5}
38 Vidaza(Azacitidine),25 mg/ml
>240 min.
39 Vincristine Sulfate,1 mg/ml
>240 min.
Fentanyl, Gastric Acid and Xylazine
Minimum Breakthrough Detection Time
1 Fentanyl Citrate Injection,50mcg/ml
>240 min.
2 Gastric Acid
>240 min.
3 Fentanyl Citrate: Xylazine HCl, 50:50
>240 min.
*Please note that the following drugs have extremely low permeation times:
Carmustine: 11.0 minutes, Thio Tepa:26.0 minutes
Warning: Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
PSC Publishing Services (301) 443-6740
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.