Disposable Nitrile Medical Examination Glove (Tested for Use with Chemotherapy Drugs)

K211608 · Jiangsu Cureguard Glove Co., Ltd. · LZA · Aug 5, 2021 · General Hospital

Device Facts

Record IDK211608
Device NameDisposable Nitrile Medical Examination Glove (Tested for Use with Chemotherapy Drugs)
ApplicantJiangsu Cureguard Glove Co., Ltd.
Product CodeLZA · General Hospital
Decision DateAug 5, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

Disposable, non-sterile, ambidextrous, blue nitrile examination glove; intended for medical use to prevent cross-contamination between patient and examiner. Device features beaded cuff; available in XS, S, M, L, XL sizes. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Provides barrier protection; helps prevent exposure to patient contaminants and chemotherapy agents. Warning: Do not use with Carmustine.

Clinical Evidence

No clinical data. Bench testing only. Performance evaluated per ASTM D6319-19 (physical dimensions, tensile strength, elongation), ASTM D5151-19 (watertightness/holes), ASTM D6124-06 (powder content), and ASTM D6978-05 (chemotherapy drug permeation). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).

Technological Characteristics

Material: Nitrile. Form factor: Ambidextrous, beaded cuff, blue, powder-free. Dimensions: Length ≥220-230mm, thickness ≥0.05mm (finger/palm). Standards: ASTM D6319-19, ASTM D6978-05, ASTM D5151-19, ASTM D6124-06. Biocompatibility: ISO 10993-5, ISO 10993-10. Sterility: Non-sterile.

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hands to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 5, 2021 Jiangsu Cureguard Glove Co., Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801, No.161 Lujiazui East Rd., Pudong Shanghai. 200120 China Re: K211608 Trade/Device Name: Disposable Nitrile Medical Examination Glove (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, LZC Dated: May 5, 2021 Received: May 25, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K211608 #### Device Name Nitrile Exam Gloves, Powder Free,Blue (Tested for use with Chemotherapy Drugs) Indications for Use (Describe) A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. | Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes | |----------------------------|------------------------|----------------------------------------| | Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | 23.6 Minutes | | Cisplatin | 1.0 mg/ml(1,000 ppm) | > 240 Minutes | | Cyclophosphamide (Cytoxan) | 20.0 mg/ml(20,000 ppm) | > 240 Minutes | | Dacarbazine (DTIC) | 10.0 mg/ml(10,000 ppm) | > 240 Minutes | | Doxorubicin HCl | 2.0 mg/ml(2,000 ppm) | > 240 Minutes | | Etoposide | 20.0 mg/ml(20,000 ppm) | > 240 Minutes | | Fluorouracil | 50.0 mg/ml(50,000 ppm) | > 240 Minutes | | Methotrexate | 25 mg/ml(25,000 ppm) | > 240 Minutes | | Mitomycin C | 0.5 mg/ml(500 ppm) | > 240 Minutes | | Paclitaxel | 6.0 mg/ml(6,000 ppm) | > 240 Minutes | | Thio Tepa | 10.0 mg/ml(10,000 ppm) | 57.4 Minutes | | Vincristine Sulfate | 1.0 mg/ml(1,000 ppm) | > 240 Minutes | Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 23.6 Minutes Thio-Tepa 10.0 mg/ml 57.4 Minutes Caution: Testing showed an average breakthrough time of 57.4 minutes with Thio-Tepa WARNING: Do not use with Carmustine Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary (K211608) This summary of 510(k) is being submitted in accordance with 21 CFR 807.92. #### 1.0 Submitter's Information Name: Jiangsu Cureguard Glove Co., Ltd. Address: No.65 Shenzhen Road, The Economic Development Zone, Sugian , Jiangsu, 223800 China. Phone Number: +86-13485097856 Contact: Guo Hua Date of Preparation: Aug.5,2021 #### Designated Submission Correspondent Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn #### 2.0 Device Information Trade name: Disposable Nitrile Medical Examination Glove(Tested for Use with Chemotherapy Drugs) Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): XS,S, M, L, XL #### 3.0 Classification Production code: LZA,LZC Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital #### 4.0 Predicate Device Information Manufacturer: Medline Industries, Inc. Medline Powder-Free Light Blue Nitrile Exam Glove (Tested Device: for Use with Chemotherapy Drugs) {5}------------------------------------------------ 510(k) number: K201390 ### 5.0 Device Description The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in five sizes: extra-small, small, medium, large, and extra-large. #### 6.0 Indication for Use A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs, per ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. | Chemotherapy Drug | Concentration | Breakthrough Detection<br>Time in Minutes | |-------------------------------|------------------------|-------------------------------------------| | Carmustine (BCNU) | 3.3 mg/ml (3,300 ppm) | 22.9 | | Cisplatin | 1.0 mg/ml(1,000 ppm) | > 240 | | Cyclophosphamide<br>(Cytoxan) | 20.0 mg/ml(20,000 ppm) | > 240 | | Dacarbazine (DTIC) | 10.0 mg/ml(10,000 ppm) | > 240 | | Doxorubicin HCI | 2.0 mg/ml(2,000 ppm) | > 240 | | Etoposide | 20.0 mg/ml(20,000 ppm) | > 240 | | Fluorouracil | 50.0 mg/ml(50,000 ppm) | > 240 | | Methotrexate | 25 mg/ml(25,000 ppm) | > 240 | | Mitomycin C | 0.5 mg/ml(500 ppm) | > 240 | | Paclitaxel | 6.0 mg/ml(6,000 ppm) | > 240 | | Thio Tepa | 10.0 mg/ml(10,000 ppm) | 44.2 | | Vincristine Sulfate | 1.0 mg/ml(1,000 ppm) | > 240 | Please note that the following drugs have low permeation times: Carmustine (BCNU) 3.3 mg/ml 22.9 Minutes; Thio Tepa 10.0 mg/ml 44.2 Minutes. Caution: Testing showed an average breakthrough time of 44.2 minutes with Thio-Tepa WARNING: Do not use with Carmustine {6}------------------------------------------------ # 7.0 Technological Characteristic Comparison Table | Table1-General Comparison | | | | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------| | Item | Subject Device<br>(K211608) | Predicated Device<br>(K201390) | Remark | | Product Code | LZA,LZC | LZA,LZC | Same | | Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same | | Class | I | I | Same | | Intended Use | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05(2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05(2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. | Same | | Powdered or<br>Powered free | Powdered free | Powdered free | Same | | Design Feature | Ambidextrous | Ambidextrous | Same | | Sterility | Non-Sterile | Non-Sterile | Same | | Labeling Information | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile, a statement of standard ASTM D6978-05 compliance and a summary of the testing results. | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile, a statement of standard ASTM D6978-05 compliance and a summary of the testing results. | Same | #### Table1-General Comparison #### Table2 Device Dimensions Comparison | Table2 Device Dimensions Comparison | | | | | | | | |-------------------------------------|-------------|------|-----|-----|-----|-----|-----------| | | | Size | | | | | | | Predicate | Designation | XS | S | M | L | XL | Tolerance | | Predicate<br>Device(K201390) | Length, mm | NA | 240 | 240 | 240 | 240 | min | | | Width, mm | NA | 85 | 95 | 105 | 115 | ±10 | {7}------------------------------------------------ | | Thickness, mm: | | | | | | | |-----------------------------|----------------|----------------|-----|-----|-----|-----|-----------| | | Finger | 0.16 | | min | | | | | | Palm | 0.14 | | min | | | | | | Designation | Size | | | | | | | Subject Device<br>(K211608) | Length, mm | XS | S | M | L | XL | Tolerance | | | Width, mm | 220 | 220 | 230 | 230 | 230 | min | | | | 70 | 80 | 95 | 110 | 120 | ±10 | | | | Thickness, mm: | | | | | | | | Finger | 0.05 | | min | | | | | | Palm | 0.05 | | min | | | | | Remark | Different | | | | | | | Analysis: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D6319-19, so the differences do not raise any new safety or performance questions. | Item | Subject device<br>(K211608) | Predicated device<br>(K201390) | Remark | | | |------------------------|-----------------------------|---------------------------------------------------------------------------|------------------------|---------------------------------------------------------------------------|-----------| | Colorant | Blue | Light Blue | Different | | | | Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14Mpa, min | 17Mpa, min | Different | | | | Ultimate<br>Elongation | 500% min | 500% min | Same | | | After<br>Aging | Tensile<br>Strength | 14Mpa, min | 14Mpa, min | Same | | | | Ultimate<br>Elongation | 400%min | 400%min | Same | | | Comply with ASTM D6319 | | Comply with ASTM D6319 | Same | | | Freedom from Holes | | Be free from holes when tested<br>in accordance with<br>ASTMD5151 AQL=2.5 | | Be free from holes when tested<br>in accordance with<br>ASTMD5151 AQL=2.5 | Same | | Powder Content | | 0.10 mg per glove, Meet the<br>requirements of ASTM D6124 | | Meet the requirements<br>of ASTM D6124 | Same | | | | Carmustine (BCNU) 3.3 mg/ml:<br>22.9 Minutes | | Carmustine (BCNU)<br>3.3 mg/ml:<br>25.3 Minutes | Different | | | | Cisplatin 1.0 mg/ml: > 240<br>Minutes | | Cisplatin 1.0 mg/ml:<br>$\ge$ 240 Minutes | Same | | | | Cyclophosphamide (Cytoxan)<br>20.0 mg/ml: > 240 Minutes | | Cyclophosphamide<br>(Cytoxan) | Same | #### Table3 Performance Comparison {8}------------------------------------------------ | Chemotherapy Drugs<br>Tested with Minimum<br>Breakthrough<br>Detection Time as<br>Tested per ASTM D<br>6978 | | 20.0 mg/ml: ≥240<br>Minutes | | |-------------------------------------------------------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------------------|-----------| | | Dacarbazine (DTIC) 10.0<br>mg/ml: > 240 Minutes | Dacarbazine (DTIC)<br>10.0 mg/ml:<br>≥240 Minutes | Same | | | Doxorubicin HCl 2.0 mg/ml: ><br>240 Minutes | Doxorubicin<br>Hydrochloride 2.0<br>mg/ml: ≥240 Minutes | Same | | | Etoposide 20.0 mg/ml: > 240<br>Minutes | Etoposide (Toposar)<br>20.0<br>mg/ml: ≥240 Minutes | Same | | | Fluorouracil 50.0 mg/ml: >240<br>Minutes | Fluorouracil 50.0<br>mg/ml: ≥240<br>Minutes | Same | | | Methotrexate 25 mg/ml: >240<br>Minutes | Methotrexate 25<br>mg/ml: ≥240<br>Minutes | Same | | | Mitomycin C 0.5 mg/ml: ><br>240 Minutes | Mitomycin C 0.5<br>mg/ml: ≥ 240<br>Minutes | Same | | | Paclitaxel 6.0 mg/ml: > 240<br>Minutes | Paclitaxel (Taxol) 6.0<br>mg/ml:<br>≥240 Minutes | Same | | | Thio Tepa 10.0 mg/ml: 44.2<br>Minutes | Thio-Tepa 10.0 mg/ml:<br>43.7<br>Minutes | Different | | | Vincristine Sulfate 1.0 mg/ml:<br>> 240 Minutes | Vincristine Sulfate<br>(Oncovin) 1.0<br>mg/ml: ≥240 Minutes | Same | ## Table4 Safety Comparison | Item | Subject device<br>(K211608) | Predicated device<br>(K201390) | Remark | | |------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|-------------------------|------| | Material | Nitrile | Nitrile | Same | | | Biocompatibility | Irritation (ISO 10993-10:2010 Biological Evaluation of Medical Devices – Part 10: Tests For Irritation And Skin Sensitization)<br>Sensitization (ISO 10993-10:2010 | Under the conditions of the study, not an irritant | Comply with ISO10993-10 | Same | {9}------------------------------------------------ | Biological Evaluation of Medical Devices –<br>Part 10: Tests For Irritation And Skin<br>Sensitization) | the study, not a sensitizer. | | | |---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------|-----------| | Cytotoxicity (ISO<br>10993-5:2009<br>Biological Evaluation of Medical Devices –<br>Part 5: Tests For In Vitro Cytotoxicity) | Under conditions of the study, device extract is not cytotoxic | Under conditions of the study, device extract is cytotoxic | Different | | Acute Systemic Toxicity(ISO 10993-<br>11:2017,Biological evaluation of medical<br>devices - Part 11: Tests for systemic<br>toxicity.) | N/A | Under conditions of the study, device extract is non-toxic | Different | ### 8.0 Summary of Non-Clinical Testing The biocompatibility evaluation for Disposable Nitrile Medical Examination Glove(Tested for Use with Chemotherapy Drugs) was conducted in accordance with the following standards: ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization. ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity Performance Testing (Bench) Physical performance qualities of the proposed device were evaluated per ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application. Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. In addition, the proposed device was tested according to ASTM D6978-05 (Reapproved 2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs, in which minimum breakthrough times were determined for a wide range of chemotherapy drugs. {10}------------------------------------------------ In summary, the performance testing of the subject device was conducted to adequately demonstrate the effectiveness of the device in accordance with the relevant test methods cited below: - ASTM D6124-06 (Reapproved 2017), Standard Test Method for -Residual Powder on Medical Gloves - ASTM D5151-19, Standard Test Method for Detection of Holes in -Medical Gloves. - ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. - ASTM D 6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. | Test Method | Purpose | Acceptance Criteria | Results | |-------------|--------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ASTM D6319 | Physical<br>Dimensions<br>Test | Length(mm):<br>XS/S:≥220;<br>M/L/XL: ≥230; | Length: >230 | | | | Width(mm):<br>XS: 70±10;<br>S: 80±10;<br>M: 95±10;<br>L: 110±10;<br>XL: 120±10 | Width:<br>Lot1:<br>XS: 77-81;<br>S: 85-89<br>M: 97-100<br>L: 108-114<br>XL: 117-120<br>Lot2:<br>XS: 78-80<br>S: 85-90<br>M: 97-100<br>L: 109-114<br>XL: 117-1120<br>Lot3:<br>XS: 77-81<br>S: 85-89<br>M: 97-100<br>L: 109-113<br>XL: 117-121<br>Pass<br>Lot1:<br>XS:<br>Finger: 0.08-0.10<br>Palm: 0.08-0.10<br>S: | | | | | | | | | Finger: 0.07-0.11<br>Palm: 0.08-0.09 | | | | | M:<br>Finger: 0.08-0.10<br>Palm: 0.08-0.10 | | | | | L:<br>Finger: 0.08-0.12<br>Palm: 0.08-0.12 | | | | | XL:<br>Finger: 0.08-0.13<br>Palm: 0.08-0.11 | | | | | Lot2:<br>XS:<br>Finger: 0.08-0.10<br>Palm: 0.08-0.11 | | | | | S:<br>Finger: 0.08-0.10<br>Palm: 0.08-0.12 | | | | Thickness (mm) :<br>Finger: ≥0.05<br>Palm: ≥0.05 | M:<br>Finger: 0.08-0.12<br>Palm: 0.08-0.11 | | | | | L:<br>Finger: 0.08-0.12<br>Palm: 0.08-0.12 | | | | | XL:<br>Finger: 0.08-0.13<br>Palm: 0.08-0.12 | | | | | Lot3:<br>XS:<br>Finger: 0.08-0.10<br>Palm: 0.08-0.10 | | | | | S:<br>Finger: 0.08-0.11<br>Palm: 0.08-0.10 | | | | | M:<br>Finger: 0.08-0.12<br>Palm: 0.08-0.12 | | | | | L:<br>Finger: 0.08-0.11 | | #### Table 5 Summary of Non-Clinical Testing {11}------------------------------------------------ {12}------------------------------------------------ | | | | | | Palm: 0.08-0.12<br>XL:<br>Finger: 0.08-0.13<br>Palm: 0.08-0.15<br>Pass | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|---------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------|---------------------------------------------------------------------------------------------| | ASTM D5151 | Watertightness<br>Test<br>for<br>Detection<br>of<br>Holes | Meet the requirements of ASTM D5151 AQL 2.5 | | | XS:0/125 leaks<br>S:1/125 leaks<br>M:0/125 leaks<br>L:0/125 leaks<br>XL:0/125 leaks<br>Pass | | ASTM D6124 | Powder Content | Meet the requirements of ASTM D6124 < 2.0mg | | | 0.10 mg/Pass | | ASTM D412 | Physical<br>properties | Before<br>Aging | Tensile<br>Strength | ≥14MPa | Lot 1:15.4-17.1<br>Lot2:15.6-17.3<br>Lot3:15.7-16.6<br>Pass | | | | | Ultimate<br>Elongation | ≥500% | Lot1:542-580<br>Lot 2:542-570<br>Lot3:545-562<br>Pass | | | | After<br>Aging | Tensile<br>Strength | ≥14MPa | Lot 1:15.6-17.0<br>Lot2:15.8-17.0<br>Lot3:15.4-19.2<br>Pass | | | | | Ultimate<br>Elongation | ≥400% | Lot1:546-577<br>Lot 2:547-572<br>Lot3:536-571<br>Pass | | | | ASTM<br>D6978 | Chemotherapy<br>Drugs<br>Tested with<br>Minimum<br>Breakthrough<br>Detection Time | Carmustine (BCNU) 3.3 mg/ml: 22.9 Minutes | | | Cisplatin 1.0 mg/ml: > 240 Minutes | | | | | | | Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes | | | | | | | Dacarbazine (DTIC) 10.0 mg/ml: > 240 Minutes<br>Doxorubicin HCI 2.0 mg/ml: > 240 Minutes<br>Etoposide 20.0 mg/ml: > 240 Minutes<br>Fluorouracil 50.0 mg/ml: > 240 Minutes<br>Methotrexate 25 mg/ml: > 240 Minutes<br>Mitomycin C 0.5 mg/ml: > 240 Minutes | | | | | | {13}------------------------------------------------ | | | Paclitaxel 6.0 mg/ml: > 240 Minutes | | |--------------|---------------|----------------------------------------------|----------------------------------------------------------------------------------------| | | | Thio Tepa 10.0 mg/ml: 44.2 Minutes | | | | | Vincristine Sulfate 1.0 mg/ml: > 240 Minutes | | | ISO 10993-5 | Cytotoxicity | Non-cytotoxic | Under conditions of the study, did not show potential toxicity to L-929 cells.<br>Pass | | ISO 10993-10 | Irritation | Non-irritating | Under the conditions of the study, not an irritant.<br>Pass | | ISO 10993-10 | Sensitization | Non-sensitizing | Under conditions of the study, not a sensitizer.<br>Pass | # 9.0 _Summary of Clinical Testing Clinical testing is not needed for this device. #### 10.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device, Disposable Nitrile Medical Examination Glove(Tested for Use with Chemotherapy Drugs) is as safe, as effective, and performs as well as or better than the legally marketed predicated device under K201390.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...