Powder Free Nitrile Examination Glove Black and Dual Color White-Blue
K201857 · Careplus (M) Sdn Bhd · LZA · Aug 20, 2021 · General Hospital
Device Facts
Record ID
K201857
Device Name
Powder Free Nitrile Examination Glove Black and Dual Color White-Blue
Applicant
Careplus (M) Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Aug 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable nitrile examination gloves; ambidextrous; available in black or dual-color white-blue. Manufactured via double-dipping for dual-color variant. Used in clinical settings by healthcare professionals to provide a barrier against contamination. Protects both patient and examiner during medical examinations. No electronic or software components.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified against ASTM D6319-19, ASTM D5151-06, and ASTM D6124-06 standards. Biocompatibility testing performed per ISO 10993-5, ISO 10993-10, and ISO 10993-11.
Technological Characteristics
Material: Nitrile with Titanium Dioxide pigment. Form factor: Ambidextrous, powder-free, various sizes (XS-XXXL). Standards: ASTM D6319-19 (physical properties, dimensions), ASTM D5151-06 (water tight test), ASTM D6124-06 (powder content). Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ISO 10993-11 (acute systemic toxicity).
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Examination Glove White (K142858)
Powder Free Nitrile Examination Glove Blue (K142862)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2021
Careplus (M) SDN BHD Bee Khoo Quality Manager Lot 120 & 121, Jalan Senawang 3, Senawang Industrial Estate Seremban. 70450 Mv
Re: K201857
Trade/Device Name: Powder Free Nitrile Examination Glove Black and Dual Color White-Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 8, 2021 Received: July 12, 2021
Dear Bee Khoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201857
Device Name
Powder Free Nitrile Examination Glove Black and Dual Color White-Blue
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K201857
## 510(K) SUMMARY
- 1.0 Applicant: CAREPLUS (M) SDN BHD
Address: Lot 120 & 121 . Jalan Senawang 3, Senawang Industrial Estate, 70450 Seremban, Negeri Sembilan Darul Khusus, Malaysia.
Phone No. 60-6-6772781 Fax No. 60-6-6772780
Contact Person Lim Kwee Shyan
- 2.0 Date of Summary 18th June, 2021
- 3.0 Device Information Device Name
## POWDER FREE NITRILE EXAMINATION GLOVE BLACK AND DUAL COLOR WHITE-BLUE
| Common Name | : Exam Gloves |
|----------------------|-----------------------------------------------------|
| Classification Name | : Patient Examination Gloves (21 CFR 880.6250, LZA) |
| Classification Panel | : Non-powdered patient examination glove |
| Class | : I |
#### 4.0 Predicate Device
Predicate device: Powder Free Nitrile Examination Glove White, 510(k) number K142858 and Powder Free Nitrile Examination Glove Blue, 510(k) number K142862, product code LZA.
#### 5.0 Device Description
Powder Free Nitrile Examination Glove, Black and dual color White-Blue are Class 1 Patient Examination Gloves. The dual color White-Blue is made dual color by double dipping. They are ambidextrous and come in different sizes, Extra Small, Small, Medium, Large and Extra Large, Double Extra-Large and Triple Extra-Large. Gloves meet the ASTM D6319-19 specification below:
| Size | XS | S | M | L | XL | XXL | XXXL |
|--------------------|----|-----|------|-----|-----|-----|------|
| Length, min. mm | | 220 | | 230 | | | NA |
| Thickness, min. mm | | | | | | | |
| finger | | | 0.05 | | | | NA |
| palm | | | 0.05 | | | | |
| Width, ± 10 mm | 70 | 80 | 95 | 110 | 120 | | NA |
#### 6.0 Indications for Use Statement
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
{4}------------------------------------------------
# K201857
## 7.0 Technological Characteristics
Following is a table summarized the technological characteristics compared to Predicate Device and equivalent standards:
| Parameter | Standard<br>Reference | Standard<br>Specification | Subject Glove<br>(Powder Free<br>Nitrile Examination<br>Glove, Black)<br>K201857 | Subject Glove<br>(Powder Free<br>Nitrile Examination<br>Glove, Dual Color<br>White-Blue)<br>K201857 | Predicate Device<br>(Nitrile Blue,<br>K142862<br>Nitrile White,<br>K142858) | Characteristics |
|------------------------------------------------------------------------------------|--------------------------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Product Code<br>Regulation number | - | - | LZA<br>21 CFR 880.6250 | LZA<br>21 CFR 880.6250 | LZA<br>21 CFR 880.6250 | Same |
| Class | - | - | I | I | I | Same |
| Material | - | - | Nitrile | Nitrile | Nitrile | Same |
| Color | - | - | Black | White-Blue | Blue, White | Similar |
| White Pigment | - | - | Titanium Dioxide | Titanium Dioxide | Titanium Dioxide | Same |
| Size | ASTM<br>D6319-19 | - | XS,S,M,L,XL, XXL,<br>XXXL | XS,S,M,L,XL, XXL,<br>XXXL | XS,S,M,L,XL | Similar |
| Water Tight Test,<br>1000ml | ASTM<br>D6319-19<br>ASTM<br>D5151-06 | G-I, AQL 2.5 | Meets specification | Meets specification | Meets specification | Same |
| Physical Properties<br>i) Before Ageing<br>Tensile Strength<br>Ultimate Elongation | ASTM<br>D6319-19 | Min. 14 Mpa<br>Min. 500 % | Meets specification | Meets specification | Meets specification | Same |
| ii) After Ageing<br>ii)Tensile Strength<br>Ultimate Elongation | ASTM<br>D6319-19 | Min. 14 Mpa<br>Min. 500 | Meets specification | Meets specification | Meets specification | Same |
| Powder Content | ASTM<br>D6319-19<br>ASTM<br>D6124-06 | Max.<br>2 mg/glove | Meets specification | Meets specification | Meets specification | Same |
| Biocompatability<br>Test<br>i) Primary Skin<br>Irritation Test | ISO 10993-<br>10 | | Conclusion: Under<br>the conditions of this<br>study, the test<br>material did not<br>cause and irritant<br>response. | Conclusion: Under<br>the conditions of this<br>study, the test<br>material did not<br>cause and irritant<br>response. | Conclusion: Under<br>the conditions of this<br>study, the test<br>material did not<br>cause and irritant<br>response. | Same |
| ii) Dermal<br>Sensitization Test | ISO 10993-<br>10 | | Conclusion: Under<br>the conditions of this<br>study, the test<br>material did not<br>produce a skin<br>sensitization effect. | Conclusion: Under<br>the conditions of this<br>study, the test<br>material did not<br>produce a skin<br>sensitization effect. | Conclusion: Under<br>the conditions of this<br>study, the test<br>material did not<br>produce a skin<br>sensitization effect. | Same |
| iii) Acute Systemic<br>Toxicity | ISO 10993-<br>11 | | Conclusion: Under<br>the conditions of this<br>study, the test<br>material both inner<br>and outer surface<br>did not reveal | Conclusion: Under<br>the conditions of this<br>study, the test<br>material both inner<br>and outer surface<br>did not reveal | Conclusion: Under<br>the conditions of this<br>study, the test<br>material both inner<br>and outer surface<br>did not reveal | Same |
{5}------------------------------------------------
# K201857
| iv) In vitro Cytotoxicity<br>Test | ISO 10993-5 | Conclusions: Under<br>the conditions of this<br>study, the test<br>material is<br>Cytotox(grade >2) at<br>undiluted, 1:2, and<br>Non-cyctotoxic,<br>grade ≤2 at dilutions<br>1:4, 1:8;1:16,1:32<br>and 1:64 | Conclusions: Under<br>the conditions of this<br>study, the test<br>material is<br>Cytotox(grade >2) at<br>undiluted, 1:2, 1:4,<br>1:8 and Non-<br>cyctotoxic, grade ≤2<br>at dilutions;1:16,1:32<br>and 1:64 | White Nitrile:<br>Conclusions: Under<br>the conditions of this<br>study, the test<br>material is<br>Cytotox(grade >2) at<br>undiluted, 1:2, 1:4<br>and Non-cyctotoxic,<br>grade ≤2 at<br>dilutions, 1:8 ,<br>1:16,1:32 and 1:64<br><br>Blue Nitrile:<br>Conclusions: Under<br>the conditions of this<br>study, the test<br>material is<br>Cytotox(grade >2) at<br>undiluted, 1:2, 1:4,<br>1:8, 1:16 and Non-<br>cyctotoxic, grade ≤2<br>at dilutions;1:32 and<br>1:64 | Similar |
|-----------------------------------|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Single Use | - | Single Use | Single Use | Single Use | Same |
| Labeling | - | Single Use | Single Use | Single Use | Same |
| Intended Use | - | A patient<br>examination gloves<br>is a disposal device<br>intended for medical<br>purposes that is<br>worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | A patient<br>examination gloves<br>is a disposal device<br>intended for medical<br>purposes that is<br>worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | A patient<br>examination gloves<br>is a disposal device<br>intended for medical<br>purposes that is<br>worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | Same |
{6}------------------------------------------------
| Parameter | Standard References | Standard Specification | Device Performance<br>Black | Device Performance<br>White-Blue | Predicate Device |
|------------------------------------|---------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------|
| Freedom from pin-holes | ASTM D6319-19 | AQL 2.5 | 1.5 and below | 1.5 and below | 1.5 and below |
| Width, mm | ASTM D6319-19 | XS - 70 ± 10<br>S - 80 ± 10<br>M - 95 ± 10<br>L - 110 ± 10<br>XL- 120 ± 10<br>XXL- NA<br>XXXL- NA | 78- 80<br>86- 88<br>96- 98<br>106- 108<br>116- 118<br>120- 125<br>126- 129 | 78- 80<br>86- 88<br>95- 98<br>107- 108<br>115- 117<br>122- 123<br>129- 132 | Meet Specification |
| Length, mm | ASTM D6319-19 | XS - 220 min<br>S - 220 min<br>M - 230 min<br>L - 230 min<br>XL- 230 min<br>XXL- NA<br>XXXL- NA | min. 293<br>min. 285<br>min. 287<br>min. 281<br>min. 282<br>min. 292<br>min. 290 | min. 297<br>min. 285<br>min. 285<br>min. 286<br>min. 301<br>min. 289<br>min. 300 | Meet Specification |
| Thickness Palm, mm | ASTM D6319-19 | XS – 0.05 min<br>S - 0.05 min<br>M - 0.05 min<br>L - 0.05 min<br>XL- 0.05 min<br>XXL- NA<br>XXXL- NA | min. 0.13<br>min. 0.15<br>min. 0.15<br>min. 0.16<br>min. 0.15<br>min. 0.12<br>min. 0.13 | min. 0.15<br>min. 0.16<br>min. 0.15<br>min. 0.15<br>min. 0.14<br>min. 0.16<br>min. 0.14 | Meet Specification |
| Thickness Finger, mm | ASTM D6319-19 | XS – 0.05 min<br>S - 0.05 min<br>M - 0.05 min<br>L - 0.05 min<br>XL- 0.05 min<br>XXL- NA<br>XXXL- NA | min. 0.18<br>min. 0.21<br>min. 0.21<br>min. 0.18<br>min. 0.21<br>min. 0.16<br>min. 0.20 | min. 0.19<br>min. 0.20<br>min. 0.19<br>min. 0.20<br>min. 0.20<br>min. 0.20<br>min. 0.19 | Meet Specification |
| Tensile Strength Before aging, Mpa | ASTM D6319-19 | 14 min | min. 25 | min. 25 | Meet Specification |
| Tensile Strength After aging, Mpa | ASTM D6319-19 | 14 min | min. 29 | min. 26 | Meet Specification |
| Ultimate Elongation Before aging,% | ASTM D6319-19 | 500 min | min. 605 | min. 606 | Meet Specification |
| Ultimate Elongation After aging,% | ASTM D6319-19 | 400 min | min. 559 | min. 576 | Meet Specification |
| Powder Free | ASTM D6319-19 | Max. 2 | Max. 1.02 | Max. 0.90 | Meet Specification |
#### Summary of Non-Clinical Testing 8.0
The performance test data from the non-clinical test met the specifications from ASTM testing methodology.
#### Summary of Clinical Testing 9.0
Clinical data is not needed for medical gloves or of most devices cleared by the 510(k) process.
### 10.0 Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.