GEN 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for use with Chemotherapy Drugs Claims
K193121 · Eco Medi Glove Sdn. Bhd. · LZA · Mar 20, 2020 · General Hospital
Device Facts
Record ID
K193121
Device Name
GEN 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for use with Chemotherapy Drugs Claims
Applicant
Eco Medi Glove Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Mar 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Gen 2 Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner. In addition these gloves was tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards, Practice for assessment of medical glove to Permeation by Chemotherapy Drugs.
Device Story
Disposable blue nitrile examination glove; non-sterile; powder-free. Worn on hands/fingers by healthcare providers to prevent cross-contamination. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Low dermatitis potential claim supported by Modified Draize-95 clinical testing. Used in clinical settings; provides barrier protection; reduces risk of chemical exposure during chemotherapy handling.
Clinical Evidence
Clinical study using Modified Draize-95 test to evaluate skin sensitization/irritation potential. Conducted in two stages with 205 non-sensitized adult human subjects. Results showed no evidence of dermal irritation or sensitization; no clinical evidence of residual chemical additives capable of inducing Type IV allergy.
Indicated for use as a disposable patient examination glove by clinicians to prevent cross-contamination between patient and examiner. Suitable for use with chemotherapy drugs. No specific age or gender restrictions; intended for general population.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Examination (Blue) with Low Dermatitis Potential Claim with tested for use with Chemotherapy Drugs (K152542)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 20, 2020
ECO Medi Glove SDN BHD Suresh Kumar QA Manager Lot 23826, Jalan Tembaga Kuning, Kamunting Raya Industrial Estate, Perak, 34600 My
Re: K193121
Trade/Device Name: GEN 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for use with Chemotherapy Drugs Claims Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: December 14, 2020 Received: December 23, 2020
Dear Suresh Kumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elizabeth Claverie-Williams, MS Assistant Director, THT4B2: Disinfection, Reprocessing and Personal Protection DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020 |
|-------------------------------------------------------------------------|-----------------------------------------------------------------|
| Indications for Use | See PRA Statement below |
510(k) Number (if known)
K193121
Device Name
Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with Tested for use with Chemotherapy Drugs Claims
Indications for Use (Describe)
The Gen 2 Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim is a disposable device intended for Medical purpose that is worn on the examiner's hands or finger to prevent contamination patient and examiner. In addition these gloves was tested for use with Chemotherapy drugs in accordance with ASTM D6978-05 standards, Practice for assessment of medical glove to Permeation by Chemotherapy Drugs.
| Chemotherapy Drugs and Concentration | Minimum Breakthrough detection time in minutes(µg/cm²/minute) |
|---------------------------------------------------|---------------------------------------------------------------|
| 1)Carmustine (BCNU)(3.3 mg/ml)(3,300ppm) | 22 |
| 2)Cisplatin (1.0 mg/ml)(1,000ppm) | No breakthrough up to 240 min |
| 3)Cyclophosphamide (Cytoxan)(20 mg/ml)(20,000ppm) | No breakthrough up to 240 min |
| 4)Methotexale (25mg/ml)(25,000ppm) | No breakthrough up to 240 min |
| 5)Doxorubicin Hydrochloride (2.0 mg/ml)(2,000ppm) | No breakthrough up to 240 min |
| 6)Etoposide (Toposar)(20.0 mg/ml)(20,000ppm) | No breakthrough up to 240 min |
| 7)Fluorouracil (50.0 mg/ml)(50,000ppm) | No breakthrough up to 240 min |
| 8)Paclitaxel (Taxol)(6.0 mg/ml)(6,000ppm) | No breakthrough up to 240 min |
| 9)Thiotepa (10.0 mg/ml)(10,000ppm) | 47.4 |
The Maximum testing time is 240 minutes. Please note that the following drugs have low permeation time:
1) Carmustine (BCNU)(3.3mg/ml)with Permeation time of 22 minutes.
2) Thiotepa ( 10.0mg/ml)with Permeation time of 47.4 minutes.
Type of Use (Select one or both, as applicable)
| <span style="display:inline-block; margin-right: 5px;">▢</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;">☒</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
{3}------------------------------------------------
## 510(K) Summary Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and Tested for use with Chemotherapy Drugs Claims
## 1.0 Submitter:
| Company Name: | ECO Medi Glove Sdn Bhd. |
|------------------|----------------------------------------------------------------------------------------------------------|
| Company Address: | Lot 23826,Jalan Tembaga Kuning,<br>Kamunting Raya Industrial Estate,<br>34600,Taiping<br>Perak,Malaysia. |
| Contact Person: | Mr Suresh Kumar |
| Telephone No: | 603-60283033 |
| Email: | qa1@riverstone.com.my |
## 2.0 Preparation Date: 16th March 2020
## 3.0 Device Identification:
Trade Name / Proprietary Name: Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and with tested for Use with Chemotherapy drugs Claims.
Device Name: Nitrile Patient Examination gloves
Device Classification Name: Patient Examination gloves (21 CFR 880.6250)
Device Class: Class I
Product Code: LZA, LZC
{4}------------------------------------------------
### 4.0 Predicate Device:
Class I patient Examination glove with claiming, this product contain Low Dermatitis Potential Claim and tested for use with Chemotherapy Drugs, Powder Free, LZC, which meets all the requirement of ASTM D 6319-10 and FDA 21 CFR 880.6250.
Predicate Device: K152542, Powder Free Nitrile Examination (Blue) with Low Dermatitis Potential Claim with tested for use with Chemotherapy Drugs
#### 5.0 Device Description:
The subject device in this 510(k) Notification is Blue Nitrile Examination gloves, with claiming, this product contains Low Dermatitis Potential Claim and tested for use with Chemotherapy drugs. The subject device is a patient examination glove made from nitrile compound, Blue color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
#### 6.0 Intended use of the Device:
Gen 2 patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 standards, Practice for assessment of Medical Glove to Permeation by chemotherapy Drugs. It is for over- the-counter use.
### 7.0 Specification for Nitrile gloves:
### 7.1 Dimension and Thickness of Gloves
| Dimension | Size S | Size M | Size L | Size XL |
|------------------------------|---------|---------|---------|---------|
| Overall Length (mm) | 230min | 230min | 230min | 230min |
| Width (± 5mm) | 85 | 95 | 105 | 115 |
| Thickness at Palm (mm) | 0.05min | 0.05min | 0.05min | 0.05min |
| Thickness at Finger Tip (mm) | 0.05min | 0.05min | 0.05min | 0.05min |
{5}------------------------------------------------
# 7.2 Gloves Physical Properties and Holes
| Measurement | Before Ageing | After Aging at 70℃ for<br>168 hrs @ 100℃ for 22 hrs |
|-------------------------|---------------------------------|-----------------------------------------------------|
| Tensile Strength (MPa) | 14min | 14 Min |
| Ultimate Elongation (%) | 500min | 400min |
| Pin-hole Level | AQL 2.5<br>Inspection Level G-1 | AQL 2.5<br>Inspection Level G-1 |
Gloves meet all the specification listed in ASTM D 6319-10
# 7.3 Technological Characteristics Comparison Table:
| Characteristics | Acceptance Criteria | Subject | Predicate | Comparison |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | | Gen 2 Nitrile Examination Gloves (Blue) with Low Dermatitis Potential Claim and with tested for use with chemotherapy drugs Claim, (K193121) | Powder Free Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim and with tested for used with Chemotherapy Drugs, (K152542) | |
| Product Code | LZA, LZC | LZA, LZC | LZA, LZC | same |
| Intended use | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the-counter use. | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over-the- counter use. | A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device is for over- the-counter use. | same |
| Material use | Nitrile compound | Nitrile compound | Nitrile compound | same |
| Colour | Blue | Blue | Blue | same |
| Sterility | Non sterile | Non sterile | Non sterile | same |
| Single used | Single used | Single used | Single used | same |
| Non Sterile | Non Sterile | Non Sterile | Non Sterile | same |
{6}------------------------------------------------
| Dimensions | Overall Length (mm)<br>Min 230mm Width<br>(±5mm)<br>Size S= 85mm<br>Size M =95mm<br>Size L= 105mm Size<br>XL = 115mm<br>Thickness at Palm (mm) Min; 0.05<br>mm Thickness at Finger Tip (mm)<br>Min 0.05 mm | Meets ASTM D6319- 10 | Meets ASTM D6319-10 | same |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------|
| Physical properties | Before Ageing Tensile Strength<br>(MPa)<br>= 14min Ultimate<br>Elongation (%)<br>= 500min After Aging at 70°C for<br>168 hrs @ 100°C for 22 hrs<br>Tensile Strength (MPa)<br>= 14min Ultimate Elongation<br>(%)<br>= 400min | Meets ASTM<br>D6319-10 | Meets ASTM<br>D6319-10 | same |
| Freedom from pinholes | AQL 2.5<br>Inspection Level G-1 | Meets ASTM D5151-06 | Meets ASTM<br>D5151-06 | same |
| Residual Powder | ≤ 2.0 mg/glove | Meets ASTM<br>D6124-06 | Meets ASTM<br>D6124-06 | same |
| Biological Evaluation on<br>Medical Device - Part 10<br>-Primary Skin Irritation<br>Test | Under the conditions of this<br>study, the test article was a<br>non- irritant. | Under the conditions of this<br>study, the test<br>article was a non- irritant. | same | |
| Biological Evaluation on<br>Medical Device - Part 10<br>-Dermal Sensitization<br>Assay | Under the conditions of this<br>study, the test article was a<br>non- sensitizer. | Under the conditions of this<br>study, the test article was a non-<br>sensitizer. | same | |
| Biological Evaluation on<br>Medical Device – Part 5<br>-Tests for in vitro<br>Cytotoxicity | Under the conditions of this<br>study, the test article was not<br>cytotoxic. | NA | Different | |
| Resistance against<br>Chemotherapy Drugs | 1) Carmustine (3.3mg/ml<br>or 3000ppm),<br>Breakthrough : 22 min. | 1) Carmustine (3.3mg/ml or<br>3000ppm), Breakthrough : 20.1<br>min. | similar | |
| | 2) Cyclophosphamide<br>(20mg/ml or 20,000ppm),<br>Breakthrough time :<br>>240 min. | 2)Cyclophosphamide (20mg/ml or<br>20,000ppm),Breakthrough time :<br>>240 min. | | |
| | 3) Cisplatin (1mg/ml or<br>1000ppm), Breakthrough<br>time :<br>> 240 min | 3) Cisplatin (1mg/ml or<br>1000ppm), Breakthrough time :<br>>240 min | | |
| | 4)Doxorubicin Hydrochloride<br>(2.0mg/ml or 2000ppm),<br>Breakthrough time :<br>>240 min. | 4)Doxorubicin Hydrochloride<br>(2.0mg/ml or 2000ppm),<br>Breakthrough time :<br>>240 min. | | |
| | 5) Etoposide (20mg/ml or<br>20,000ppm), Breakthrough<br>time :<br>>240 min. | 5) Etoposide (20mg/ml or<br>20,000ppm), Breakthrough time :<br>>240 min. | | |
| | 6) Flourouracil (50mg/ml or<br>50,000), Breakthrough time :<br>>240 min. | 6) Flourouracil (50mg/ml or<br>50,000), Breakthrough time :<br>>240 min. | | |
| | | | | |
| | | 7) Methotexate (25mg/ml or 25,000ppm), Breakthrough time:<br>> 240 min. | 7) Methorexate (25mg/ml or 25,000ppm), Breakthrough time :<br>> 240 min. | |
| | | 8) Paclitaxel (6mg/ml or 6,000ppm), Breakthrough time:<br>>240 min. | 8) Paclitaxel (6mg/ml or 6,000ppm), Breakthrough time:<br>>240 min | |
| | | 9) Thiotepa (10mg/ml or 10,000ppm), Breakthrough time:<br>47.4 min. | 9) Thiotepa (10mg/ml or 10,000ppm), Breakthrough time:<br>50.6 min. | |
| Low Dermatitis<br>Potential Claim | 1)Modified<br>Draize 95 test | No Clinical evidence<br>presence of residual<br>chemical additives<br>that may induce Type<br>IV allergy in human<br>subject | No Clinical<br>evidence presence<br>of residual<br>chemical additives<br>that may induce<br>Type IV allergy in<br>human subject | similar |
{7}------------------------------------------------
{8}------------------------------------------------
## Resistance against Chemotherapy Drugs
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection time<br>(Specimen 1/2/3)(Minutes) |
|----------------------------------------------------|------------------------------------------------------------------|
| 1)Carmustine (BCNU) (3.3mg/ml)(3,300 ppm) | 22 |
| 2) Cisplatin (1.0mg/ml)(1,000 ppm) | No breakthrough up to 240 min |
| 3)Cyclophosphamide (Cytoxan)(20mg/ml)(20,000 ppm) | No breakthrough up to 240 min |
| 4)Methotexate(25.0mg/ml)(25,000 ppm) | No breakthrough up to 240 min |
| 5)Doxorubicin Hydrochloride ( 2.0mg/ml)(2,000 ppm) | No breakthrough up to 240 min |
| 6)Etoposide (Toposar)(20.0mg/ml)(20,000 ppm) | No breakthrough up to 240 min |
| 7)Fluorouracil (50.0mg/ml)(50,000 ppm) | No breakthrough up to 240 min |
| 8)Paclitaxel (Taxol)(6.0mg/ml)(6,000 ppm) | No breakthrough up to 240 min |
| 9)Thiotepa(10.0mg/ml)(10,000 ppm) | 47.4 |
The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time. Carmustine (BCNU) (3.3mg/ml) Thiotepa (10mg/ml)
{9}------------------------------------------------
### 8.0) Summary of Non-Clinical Testing:
| Test Method | Purpose | Acceptance Criteria | Results |
|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| ASTM D5151-06 | Freedom from<br>Holes | AQL 2.5<br>Inspection Level G-1 | Meets ASTM D6319-<br>10 |
| ASTM D6319-10 | Dimensions,<br>Physical Properties | ASTM D6319- 10 | Meets ASTM D6319-<br>10 |
| ASTM D6124-06 | Powder-free<br>residue | < 2.0 mg/glove | Meets ASTM D6319-<br>10 |
| Primary skin<br>irritation - ISO<br>10993-10<br>Skin Sensitization<br>- ISO 10993-10<br>In vitro<br>Cytotoxicity –<br>ISO 10993-5 | Bio-compatibility | 1) The test article was a non-<br>irritant.<br>2) The test article was a non-<br>sensitizer<br>3) The test article was not<br>cytotoxic | Test article is non<br>irritant and Non<br>sensitizer and not<br>cytotoxic |
| ASTMD 6978-05 | Chemotherapy Drug<br>test | As per test report | As labeling claim |
| Modify Draize 95<br>Test | Low Dermatitis<br>Potential Claim | Induce Type<br>IV allergy in human | Test article do not<br>induce type IV<br>allergy to human |
### 9.0) Summary of Clinical Testing:
Gen 2 Nitrile Examination Glove (Blue) With Low Dermatitis Potential Claim and tested for Use with Chemotherapy drugs were tested in accordance with Modified Draize -95 test, per FDA's guidance document "Guidance for Industry and FDA Reviewer/Staffs: Premarket Notification [510k] Submissions for testing for skin sensitization to chemical in natural Rubber Products".
The study was conducted in two stages. In the first, a population of 30 human subjects was tested to evaluate the product for the potential to cause irritation or sensitization. The second stage was initiated on a further number of subjects to a total of a minimum 205 individuals after the first stage has shown that the test product does not indicate a potential for inducing dermal irritation and does not shown sensitization capability
The study completed on 205 non sensitized adult human subjects, who reasonably reflect the general user population in the US, gave all negative results. There was no clinical evidence of the presence of residual chemical additives at the level that may induce type IV allergy in the un-sensitized general user population in the tested articles
## 10.0 Conclusion:
The conclusion drawn from the nonclinical and clinical test demonstrate that the subject device in 510(k) K193121, Gen 2 Nitrile Examination Glove (Blue) with Low Dermatitis Potential Claim and tested for Use with Chemotherapy drugs is as safe, as effective and performs as well as or better than the legally marketed predicate device (K15254)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.