Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low-Dermatitis Potential Claim and Tested For Use with Chemotherapy Drugs (White)

K180505 · Hartalega Sdn Bhd · LZA · Jun 14, 2018 · General Hospital

Device Facts

Record IDK180505
Device NameNitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low-Dermatitis Potential Claim and Tested For Use with Chemotherapy Drugs (White)
ApplicantHartalega Sdn Bhd
Product CodeLZA · General Hospital
Decision DateJun 14, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (White) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Is it also tested to be used against Chemotherapy Drugs.

Device Story

Nitrile powder-free examination glove; coated with colloidal oatmeal USP; intended for medical examination to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs. Used in clinical settings by healthcare professionals. Provides barrier protection; reduces potential for dermatitis. Not for use with Carmustine or Thiotepa.

Clinical Evidence

No clinical data. Bench testing only, including physical property testing and chemotherapy drug permeation testing.

Technological Characteristics

Nitrile examination glove; powder-free; colloidal oatmeal USP coating; non-sterile; disposable. Tested for chemotherapy drug resistance per ASTM standards.

Indications for Use

Indicated for use as a patient examination glove to prevent contamination between patient and examiner. Suitable for use with specific chemotherapy drugs. Contraindicated for use with Carmustine and Thiotepa.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{1}------------------------------------------------ Hartalega Sdn. Bhd. Nurul Kong Ouality Assurance Senior Manager No. 7, Kawasan Perusahaan Suria Bestari Jaya, 45600 My Re: K180505 Trade/Device Name: Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low-Dermatitis Potential Claim and Tested For Use with Chemotherapy Drugs (White) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: May 15, 2018 Received: May 17, 2018 Dear Nurul Kong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good {2}------------------------------------------------ manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray # lii III -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (ifknown) K180505 ### Device Name Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low Dermatial Claim and Tested for Use with Chemotherapy Drugs (White) #### Indications for Use (Describe) Nitrile Powder Free Examination Glove with Colloidal Oatmeal USP with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs (White) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Is it also tested to be used against Chemotherapy Drugs. The list of Chemotherapy Drugs tested (with breakthrough times) is as below; | Carmustine (3.3 mg/ml) | 17.2 minutes | |---------------------------------------|---------------| | Cisplatin (1.0 mg/ml) | >240 minutes | | Cyclophosphamide (20 mg/ml) | > 240 minutes | | Dacarbazine (10.0 mg/ml) | >240 minutes | | Doxorubicin Hydrochloride (2.0 mg/ml) | > 240 minutes | | Etoposide (20 mg/ml) | > 240 minutes | | Fluorouracil (50 mg/ml) | > 240 minutes | | Methotrexate (25 mg/ml) | > 240 minutes | | Mitomycin C (0.5 mg/ml) | > 240 minutes | | Paclitaxel (6.0 mg/ml) | > 240 minutes | | Thiotepa (10.0 mg/ml) | 36.1 minutes | | Vincristine Sulfate (1.0 mg/ml) | >240 minutes | Do not use with Carmustine. Do not use with Thiotepa. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) XOver-The-Counter Use(21CFR801 SubpartC) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.»
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...