Nitrile Examination Gloves, Powder Free [Purple] Tested for Use with Chemotherapy Drugs
Applicant
Brightway Holdings Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Jul 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; purple color; textured fingers. Provides single-use barrier protection for examiner and patient. Tested for permeation resistance against various chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Output is physical barrier protection; prevents cross-contamination. Does not provide protection against all chemotherapy agents; specifically contraindicated for use with Carmustine and ThioTEPA due to low breakthrough times.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ASTM D6319-10 (physical properties, dimensions, tensile strength, elongation, freedom from holes, residual powder) and ASTM D6978-05 (chemotherapy drug permeation). Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (systemic toxicity).
Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with specific chemotherapy drugs per ASTM D6978-05.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim (K992162)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three human profiles incorporated into the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2017
Brightway Holdings Sdn. Bhd. % Dr. Wava Truscott Consultant Truscott Medsci Associates, LLC 180 Burkemeade Ct Roswell. Georgia 30075
Re: K170686
Trade/Device Name: Nitrile Examination Gloves, Powder Free (Purple) Tested For Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: June 25, 2017 Received: June 26, 2017
Dear Dr. Truscott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tara A. Ryan -S
for
Lori Wiggins Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
## 510(k) Number (if known)
K170686
### Device Name
#### BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUG5
#### Indications for Use (Describe)
### BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.
The tested chemotherapy drugs and their breakthrough detection times are as follows :
| Test Chemotherapy Drug Name and Concentration | Minimum Breakthrough Detection Time |
|---------------------------------------------------|-------------------------------------|
| · Blenoxane (15mg/ml),(15,000 ppm) | >240 Minutes |
| · Busulfan(6mg/ml),(6,000 ppm) | >240 Minutes |
| · Carmustine (BCNU) (3.3mg/ml),(3,300 ppm) | 3.6 Minutes |
| · Cisplatin(1.0mg/ml), (1.000 ppm) | >240 Minutes |
| · Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm) | >240 Minutes |
| · Cytarabine(100mg/ml), (100,000 ppm) | >240 Minutes |
| · Dacarbazine(DTIC) 10mg/ml, (10,000 ppm) | >240 Minutes |
| · Daunorubicin(5mg/ml), (5,000 ppm) | >240 Minutes |
| · Docetaxel(10mg/ml), (10,000 ppm) | >240 Minutes |
| · Doxorubicin HCL(2mg/ml),(2,000 ppm) | >240 Minutes |
| · Ellence (2mg/ml), (2,000 ppm) | >240 Minutes |
| · Etoposide /Toposar (20mg/ml), (20,000 ppm) | >240 Minutes |
| · Fludarabine (25mg/ml), (25,000 ppm) | >240 Minutes |
| · Fluorouracil(50mg/ml), (50,000 ppm) | >240 Minutes |
| · Gemcitabine (38mg/ml), (38,000 ppm) | >240 Minutes |
| · Idarubicin(1.0mg/ml), (1,000 ppm) | >240 Minutes |
| · Ifosfamide (50mg/ml), (50,000 ppm) | >240 Minutes |
| · Irinotecan(20mg/ml), (20,000 ppm) | >240 Minutes |
| · Mechlorethamine HCI(1.0mg/ml), (1,000 ppm) | >240 Minutes |
| · Melphalan(5mg/ml), (5,000 ppm) | >240 Minutes |
| · Methotrexate(25mg/ml),(25,000 ppm) | >240 Minutes |
| · Mitomycin C(0.5 mg/ml), (500 ppm) | >240 Minutes |
| · Mitoxantrone(2mg/ml),(2,000 ppm) | >240 Minutes |
| · Paclitaxel(6.0mg/ml),(6,000 ppm) | >240 Minutes |
| · Paraplatin (10mg/ml),(10,000 ppm) | >240 Minutes |
| · Rituximab(10mg/ml),(10,000 ppm) | >240 Minutes |
| · Thiotepa (10mg/ml),(10,000 ppm) | 15.9 Minutes |
| · Trisenox(0.1mg/ml), (100 ppm) | >240 Minutes |
| · Vincristine Sulfate(1.0mg/ml),(1,000 ppm) | >240 Minutes |
Please note that Carmustine and Thiotepa have extremely low permeation times of 3,6 minutes, respectively
#### WARNING: Do Not Use With : Carmustine, ThioTEPA
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
FORM FDA 3881 (8/14)
Page 1 of 2
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### K170686 510(K) SUMMARY
13
## BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS
- 1. Submitter :
| Company Name : | BRIGHTWAY HOLDINGS SDN. BHD. |
|------------------|--------------------------------------------|
| Street Address : | Lot 1559, Jalan Istimewa, |
| | Batu Belah, 42100 Klang |
| | Selangor Darul Ehsan. |
| Country : | Malaysia |
| Phone No. : | 603-3343 1007 & 603-3343 1094. |
| Fax No. : | 603-3341 4800 |
| E-mail Address : | brightway@brightway919.com |
| Contact Person : | Mr. G. Baskaran (Group Managing Director) |
| | baskar@brightway919.com |
| | Mr. Felix Darrel (Group Marketing Manager) |
| | felix.marketing@brightway919.com |
2. Preparation Date : July 13, 2017
3. Name of the Device :
Device trade or proprietary name: BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE[PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS
{4}------------------------------------------------
Device Classification Name: Polymer Patient Examination Glove (21 CFR 88-6250)
Device common or usual name: Powder-Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs
FDA Device Class : Class 1
Product Code : LZC, LZA
- 4. ldentification of the Device :
This Class 1 medical device; product code LZA, LZC with the Trade Name: BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE (Purple) TESTED FOR USE WITH CHEMOTHERAPY DRUGS, was tested for Chemotherapy drug penetration resistance per ASTM D6978-05 and meets all the requirements of ASTM D6319-10 and FDA 21 CFR 880.6250.
#### Predicate device:
Legally marked device to which substantial equivalence is claimed:
- 1) K992162 Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim
- 5. Device Description :
The subject device in this 510(k) Notification is a Purple Nitrile Examination glove tested for use with Chemotherapy drugs.
The subject device is a patient examination glove made from nitrile compound, Purple color, powder free and non-sterile (as per 21 CFR 880.6250, class I).
The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
This device also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05.
{5}------------------------------------------------
#### 6. Intended use of the Device / Indication of Use
ੇ ਮੁੱਤਰ
BRIGHTWAY BRAND EXAMINATION GLOVES, POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is a disposable device intended for medical purpose worn on the examiners hand or finger to prevent contamination between patient and examiner.
This device has been tested with chemotherapy drugs and their breakthrough detection times are as follows:
Test Chemotherapy Drug Name and Concentration
| | Test Chemotherapy Drug Name and Concentration | Minimum Breakthrough Detection Time |
|---|-------------------------------------------------|-------------------------------------|
| • | Blenoxane (15mg/ml), (15,000 ppm) | >240 Minutes |
| • | Busulfan(6mg/ml),(6,000 ppm) | >240 Minutes |
| • | Carmustine (BCNU) (3.3mg/ml),(3,300 ppm) | 3.6 Minutes |
| • | Cisplatin(1.0mg/ml), (1,000 ppm) | >240 Minutes |
| • | Cyclophosphamide/Cytoxan(20mg/ml), (20,000 ppm) | >240 Minutes |
| • | Cytarabine(100mg/ml), (100,000 ppm) | >240 Minutes |
| • | Dacarbazine (DTIC) 10mg/ml, (10,000 ppm) | >240 Minutes |
| • | Daunorubicin(5mg/ml), (5,000 ppm) | >240 Minutes |
| • | Docetaxel(10mg/ml), (10,000 ppm) | >240 Minutes |
| • | Doxorubicin HCL(2mg/ml),(2,000 ppm) | >240 Minutes |
| • | Ellence (2mg/ml), (2,000 ppm) | >240 Minutes |
| • | Etoposide/Toposar (20mg/ml), (20,000 ppm) | >240 Minutes |
| • | Fludarabine (25mg/ml), (25,000 ppm) | >240 Minutes |
| • | Fluorouracil(50mg/ml), (50,000 ppm) | >240 Minutes |
| • | Gemcitabine (38mg/ml), (38,000 ppm) | >240 Minutes |
| • | Idarubicin (1.0mg/ml), (1,000 ppm) | >240 Minutes |
| • | Ifosfamide (50mg/ml), (50,000 ppm) | >240 Minutes |
| • | Irinotecan(20mg/ml), (20,000 ppm) | >240 Minutes |
| • | Mechlorethamine HCI(1.0mg/ml), (1,000 ppm) | >240 Minutes |
| • | Melphalan(5mg/ml), (5,000 ppm) | >240 Minutes |
| • | Methotrexate(25mg/ml), (25,000 ppm) | >240 Minutes |
| • | Mitomycin C(0.5 mg/ml), (500 ppm) | >240 Minutes |
| • | Mitoxantrone(2mg/ml), (2,000 ppm) | >240 Minutes |
| • | Paclitaxel(6.0mg/ml),(6,000 ppm) | >240 Minutes |
| • | Paraplatin (10mg/ml), (10,000 ppm) | >240 Minutes |
| • | Rituximab(10mg/ml), (10,000 ppm) | >240 Minutes |
| • | Thiotepa (10mg/ml), (10,000 ppm) | 15.9 Minutes |
| • | Trisenox(0.1mg/ml), (100 ppm) | >240 Minutes |
| • | Vincristine Sulfate(1.0mg/ml), (1,000 ppm) | >240 Minutes |
Please Note: Carmustine and ThioTEPA have extremely low permeation times with breakthrough detected in less than 30 minutes: Carmustine (3.3mg/mL): 3.6 minutes; ThioTEPA (10mg/mL): 15.9 minutes WARNING: Not for Use with: Carmustine, ThioTEPA
{6}------------------------------------------------
## 7. Substantial Equivalence Based on Assessment of Non-Clinical Performance Data
a.) Testing was performed per ASTM D6319: Standard Specification for Nitrile Examination Gloves for Medical Application, utilizing the currently re-approved versions of the test methods there-in referenced:
- ASTM D412 Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— ● Tension
- ASTM D573 Test Method for Rubber—Deterioration in an Air Oven
- ASTM D3578 Specification for Rubber Examination Gloves
- ASTM D3767 Practice for Rubber—Measurement of Dimensions
- o ASTM D5151 Test Method for Detection of Holes in Medical Gloves
- ASTM D6124 Test Method for Residual Powder on Medical Gloves
- ISO 2859 Sampling Procedures and Tables for Inspection by Attributes
Test results show that under the conditions of the testing, there is no difference in physical attributes between the proposed device and the predicate device.
b.) Both the proposed device and the predicate were tested for use with chemotherapy drugs per:
- o ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
c.) Biocompatibility testing was performed utilizing:
- . ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for Irritation and Sensitization. Both dermal irritation and sensitization (Magnusson & Kingman) were performed.
- . ISO 10993-11: Biological evaluation of medical devices – Part 11: Tests for Systemic toxicity were conducted as had been performed on the predicate device.
Both the proposed device and the predicate device demonstrated that they were nonirritating and non-sensitizing, and that they did not demonstrate systemic toxicity under conditions of the studies performed.
All testing performed demonstrated that under the conditions of the test or study conducted, that the proposed medical device and the predicate device performed the same or similarly.
## 8. A summary, side by side Comparison Table is presented:
{7}------------------------------------------------
# 510(k) Summary Table
# Comparison of Proposed Device to Predicate Device (pg. 1 of 2)
| Device<br>Characteristics | Test<br>Standard | Proposed Device | Predicate Device<br>K992162 | Comparison<br>Analysis |
|-------------------------------------------------|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Trade Name | N/A | Brightway Brand Nitrile<br>Examination Gloves, Powder<br>Free, [Purple] Tested for Use<br>with Chemotherapy Drugs | Safeskin Purple Powder-Free<br>Nitrile Examination Gloves Tested<br>for use with Chemotherapy Drugs | Similar |
| 510(k)<br>Reference | N/A | | K992162 | N/A |
| Common Name | N/A | Patient Examination Glove | Patient Examination Glove | Same |
| Product Code | N/A | LZA, LZC | LZA, LZC | Same |
| Intended use | N/A | A powder-free patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiners hand or finger<br>to prevent contamination<br>between patient and examiner.<br>Additionally, the gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM D6978-<br>05 Standard Practice for<br>Assessment of Medical Glove to<br>Permeation by chemotherapy<br>Drugs. | A powder-free patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiners hand or finger<br>to prevent contamination<br>between patient and examiner.<br>Additionally, the gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM D6978-<br>05 Standard Practice for<br>Assessment of Medical Glove to<br>Permeation by chemotherapy<br>Drugs. | Same |
| Labeling | N/A | There are no special labeling<br>claims. Does not claim<br>hypoallergenic on label | There are no special labeling<br>claims. Does not claim<br>hypoallergenic on label | Same |
| Regulation<br>Number | N/A | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Materials | N/A | Nitrile | Nitrile | Same |
| Color | N/A | Purple | Purple | Same |
| Textures fingers | N/A | Yes | Yes | Same |
| Design<br>Configurations | ASTM<br>D6319-10 | Extra-Small<br>Small<br>Medium<br>Large<br>Extra-Large | Extra-Small<br>Small<br>Medium<br>Large<br>Extra-Large | Same |
| Dimensions<br>Length | ASTM<br>D6319-10 | All sizes comply with length<br>dimensions | All sizes comply with length<br>dimensions | Same |
| Dimensions<br>Width | ASTM<br>D6319-10 | All sizes comply with width<br>dimensions | All sizes comply with width<br>dimensions | Same |
| Dimensions<br>Thickness | ASTM<br>D6319-10 | All sizes comply with thickness<br>dimensions | All sizes comply with thickness<br>dimensions | Same |
| Device<br>Characteristics | Test<br>Standard | Proposed Device | Predicate Device<br>K992162 | Comparison<br>Analysis |
| Physical<br>Properties -<br>Tensile Strength | ASTM<br>D6319-10 | Complies with ASTM<br>D6319-10 both before and<br>after accelerated aging | Complies with ASTM<br>D6319-10 both before and<br>after accelerated aging | Same |
| Physical<br>Properties -<br>Elongation | ASTM<br>D6319-10 | Complies with ASTM<br>D6319-10 both before and<br>after accelerated aging | Complies with ASTM<br>D6319-10 both before and<br>after accelerated aging | Same |
| Freedom from<br>Holes | ASTM<br>D6319-10<br>by D5151-<br>06 | Pass | Pass | Same |
| Powder Free<br>Designation | ASTM<br>D6319-10<br>by<br>D6124-06 | Less than 2.0 mg per glove;<br>Pass | Less than 2.0 mg per glove;<br>Pass | Same |
| Biocompatibility | ISO 10993-<br>10:2010<br>Reapproved<br>2014 | Under conditions of the test,<br>Not a skin irritant | Under conditions of the test<br>Not a skin irritant | Same |
| | ISO 10993-<br>10:2010<br>Reapproved<br>2014 | Under conditions of the test,<br>Not a skin sensitizer | Under conditions of the test,<br>Not a skin sensitizer | Same |
| | ISO 10993-<br>11:2006<br>Reapproved<br>2010 | Under conditions of the test<br>Does not cause<br>systemic toxicity | Under conditions of the test<br>Does not cause<br>systemic toxicity | Same |
| Prescription vs.<br>ОТС | N/A | ОТС | ОТС | Same |
| Sterile vs.<br>Non-sterile | N/A | Non-Sterile | Non-Sterile | Same |
| Single Use | N/A | Yes | Yes | Same |
| Tested for use<br>with<br>Chemotherapy<br>Drugs | ASTM<br>D6978-05 | Please note: Carmustine and<br>ThioTEPA have extremely low<br>breakthrough times of less<br>than 30 minutes | Please note: Carmustine and<br>ThioTEPA have extremely low<br>breakthrough times of less<br>than 30 minutes | Same |
| Warning | N/A | WARNING: Not for Use with:<br>Carmustine, ThioTEPA | WARNING: Not for Use with:<br>Carmustine, ThioTEPA | Same |
{8}------------------------------------------------
## Part 10. Executive Summary
# Comparison of Proposed Device to Predicate Device (pg. 2 of 2)
{9}------------------------------------------------
## 9.0 Conclusion
Based on intended uses, technological characteristics and non-clinical performance data, the subject device BRIGHTWAY BRAND NITRILE EXAMINATION GLOVES, POWDER FREE, [PURPLE] TESTED FOR USE WITH CHEMOTHERAPY DRUGS is substantially equivalent to the predicate device K992162 - Safeskin Purple Powder-Free Nitrile Examination Gloves for Use with Chemotherapeutic Drugs Labeling Claim.
The subject device meets the requirements of ASTM D 6319- 10 standards as well as applicable 21 CFR and meets FDA recognized, physical properties requirements, pinhole requirements, powder free requirements, resistance to permeation by chemotherapy drugs testing, biocompatibility studies and labeling requirements presented in this document.
There are no safety or efficacy issues or new claims different from the "substantially equivalent" predicate.
Based on the complete list of non-clinical tests, biological safety studies, labeling, intended use, materials, and processes of manufacture, the subject device herein proposed is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.