Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs; Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs
K170397 · Kossan International Sdn Bhd · LZA · Jul 10, 2017 · General Hospital
Device Facts
Record ID
K170397
Device Name
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs; Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs
Applicant
Kossan International Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Jul 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable nitrile patient examination gloves; non-sterile; polymer coated; blue or white colored. Used by healthcare professionals in clinical settings to provide a protective barrier against contamination between patient and examiner. Tested for permeation resistance against various chemotherapy drugs per ASTM D6978-05. Provides protection during handling of hazardous chemotherapy agents, though specific breakthrough times vary by drug type.
Clinical Evidence
Bench testing only. Permeation testing performed per ASTM D6978-05 (Reapproved 2013) to determine breakthrough detection times for 15 chemotherapy drugs. Results indicate breakthrough times >240 minutes for most drugs, with exceptions for Carmustine (BCNU) and Thiotepa.
Technological Characteristics
Material: Nitrile rubber; polymer coated. Form factor: Disposable, non-sterile examination glove. Testing standard: ASTM D6978-05 (Reapproved 2013) for chemotherapy drug permeation.
Indications for Use
Indicated for use as a disposable medical glove worn by examiners to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Submission Summary (Full Text)
{0}------------------------------------------------
July 10, 2017
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Kossan International Sdn. Bhd. Cho Fong RA Manager Wisma Kossan, Lot 782, Jalan Sungai Putus, Off Batu 3 3/34, Jalan Kapar. Klang, 42100 MY
Re: K170397
Trade/Device Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: June 8, 2017 Received: June 12, 2017
Dear Cho Fong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark S. Fellman -S
for
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170397
#### Device Name
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotheray Drugs
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with cherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
|-----------------------------------------|------------------------------------------------|
| Carmustine (BCNU) (3.3 mg/ml) | 10.2 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | >240 |
| Cytarabine(100 mg/ml) | >240 |
| Dacarbazine (DTIC) (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| Etoposide (20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Ifosfamide (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Mitoxantrone (2.0 mg/ml) | >240 |
| Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 30.4 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
Please note that Carmustine (BCNU) and Thiotepa has extremely low permeation times of 10.2 minutes and 30.4 minutes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170397
## Device Name
Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Polymer Coated, Tested for Use with Chemotheray Drugs
#### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
|-----------------------------------------|------------------------------------------------|
| Carmustine (BCNU) (3.3 mg/ml) | 10.1 |
| Cisplatin (1.0 mg/ml) | >240 |
| Cyclophosphamide (Cytoxan) (20.0 mg/ml) | >240 |
| Cytarabine(100 mg/ml) | >240 |
| Dacarbazine (DTIC) (10.0 mg/ml) | >240 |
| Doxorubicin Hydrochloride (2.0 mg/ml) | >240 |
| Etoposide (20.0 mg/ml) | >240 |
| Fluorouracil (50.0 mg/ml) | >240 |
| Ifosfamide (50.0 mg/ml) | >240 |
| Methotrexate (25.0 mg/ml) | >240 |
| Mitomycin C (0.5 mg/ml) | >240 |
| Mitoxantrone (2.0 mg/ml) | >240 |
| Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| Thiotepa (10.0 mg/ml) | 30.4 |
| Vincristine Sulfate (1.0 mg/ml) | >240 |
Please note that Carmustine (BCNU) and Thiotepa has extremely low permeation times of 10.1 minutes and 30.4 minutes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.