POWDER FREE NITRILE PATIENT EXAMINATION GLOVE, BLUE, WHITE AND GREEN COLORED. NON-STERILE, LOW DERMATITIS POTENTIAL
K120066 · Kossan Latex Industries(M)Sdn Bhd · LZA · Nov 16, 2012 · General Hospital
Device Facts
Record ID
K120066
Device Name
POWDER FREE NITRILE PATIENT EXAMINATION GLOVE, BLUE, WHITE AND GREEN COLORED. NON-STERILE, LOW DERMATITIS POTENTIAL
Applicant
Kossan Latex Industries(M)Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Nov 16, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, powder-free nitrile patient examination gloves; available in blue, white, and green. Worn by healthcare examiners to provide a barrier against cross-contamination between patient and examiner. Manufactured from nitrile material; tested for low dermatitis potential using Modified Draize-95 testing to ensure absence of residual chemical additives capable of inducing Type IV allergies. Device meets ASTM D6319-10 standards for physical properties, dimensions, and freedom from pinholes. Used in clinical settings; provides protection for both patient and examiner during routine examinations.
Clinical Evidence
Clinical evidence consists of a Modified Draize-95 sensitization study conducted on 200 non-sensitized adult human subjects. The study evaluated the potential for dermal irritation and sensitization. Results were negative for all subjects, demonstrating no clinical evidence of residual chemical additives capable of inducing Type IV allergy in the general user population.
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Suitable for general patient examination use.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
<120066
## KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)
### FDA 510(k), Premarket Notification: 510(k) Summary of Safety and Effectiveness Information
#### 1.0 Submitter:
Kossan Latex Industries (M) Sdn Bhd Lot 6129, Jalan Haji Abdul Manan, Batu 5 1/4, Jalan Meru, Klang, 41050 Selangor, Malaysia
| Telephone No.: | +603 3392 3088 |
|----------------|----------------|
| Fax No.: | +603 3291 0584 |
#### 2.0 Contact Person:
| Contact: | Ms S F Cho |
|----------------|----------------|
| Telephone No.: | +603 3392 3088 |
| Fax No.: | +603 3291 0584 |
#### 3.0 09 November 2012 Preparation Date:
#### 4.0 Name of Device:
Trade Name: Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored. Non-Sterile, Low Dermatitis Potential Claim. Common Name: . Powder-Free Nitrile Patient Examination Glove Classification Name: Patient Examination Glove (21 CFR Part 880.6250) Regulatory Class: Class I Product Code: 80 LZA
#### 5.0 Identification of the Legally Marketed Device:
Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile, Low Dermatitis Potential Labeling Claim; Class I Patient Examination Gloves, Nitrile-80LZA, meets all of the requirements of ASTM D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application.
Predicate Device: K102790, Powder Free Nitrile Patient Examination Glove, Blue, Non-Sterile (Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim).
## NOV 1 6 2012
{1}------------------------------------------------
#### 6.0 Description of Device:
Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile, Low Dermatitis Potential Labeling Claim, meets all of the requirements of ASTM D 6319-10.
#### 7.0 Intended Use of the Device:
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
### Summary of the Technological Characteristics of the Device: 8.0
Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile, Low Dermatitis Potential Labeling Claim possesses the following technological characteristic (as compared to ASTM or equivalent standards):
| Characteristic | Standards | Device Performance |
|-----------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Dimensions | ASTM D 6319-10 | Meets |
| Physical Properties | ASTM D 6319-10 | Meets |
| Freedom from pin-<br>holes | ASTM D 5151-06<br>ASTM D 6319-10 | Meets |
| Powder Free Residue | ASTM D 6124-06<br>ASTM D 6319-10 | Meets |
| Biocompatibility | Dermal Sensitization<br>(as per ISO 10993-<br>10:2010) | Not a contact skin sensitizer |
| | Primary Skin<br>Irritation Test<br>(as per 16 CFR Part<br>1500.41) | Not a primary skin irritant |
| Low Dermatitis<br>Potential | Modified Draize-95<br>Test | No clinical evidence presence of<br>residual chemical additives that may<br>induce Type IV allergy in human<br>subjects. |
{2}------------------------------------------------
KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)
### 9.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data
Testing was performed per ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10:2010, and 16 CFR Part 1500.41. The gloves meet standard requirements referenced in Section 8.0 above. Biocompatibility test indicates the gloves are not a contact skin sensitizer and not a primary skin irritant.
#### 10.0 Substantial Equivalent Based on Assessment of Clinical Performance Data
Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile, Low Dermatitis Potential Labeling Claim, were tested in accordance with Modified Draize-95 Test, per FDA's guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510(k)] Submissions for Testing for Skin Sensitization to Chemicals in Natural Rubber Products, 1999".
The study was conducted in two stages. In the first stage, a population of 50 human subjects was tested to evaluate the product for the potential to cause irritation or sensitization. The second stage was initiated on a further number of subjects to a total of a minimum of 200 individuals after the first stage has shown that the test product does not indicate a potential for inducing dermal irritation and does not shown sensitization capability.
The study completed on 200 non-sensitized adult human subjects, who reasonably reflect the general user population in the US, gave all negative results. There was no clinical evidence of the presence of residual chemical additives at the level that may induce Type IV allergy in the un-sensitized general user population in the tested article.
#### 11.0 Conclusion
It can be concluded that Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile, Low Dermatitis Potential Labeling Claim is safe and effective for use, and perform according to the glove performance standards referenced in Section 8.0 above, thereby meeting ASTM standards, FDA requirements, and the labeling claims for the product.
This device is substantially equivalent to currently marketed devices, per Substantial Equivalence Comparison Table below:-
{3}------------------------------------------------
KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K) K L 1
ubstantial Equivalence Comparison Tabl
| Feature | K102790, Powder Free Nitrile Patient Examination Glove, Blue, Non-Sterile (Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim) Predicate Device | Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile (Low Dermatitis Potential Labeling Claim) Proposed Device |
|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner | Identical |
| Indications for Use | This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. |
| | This glove has been tested for use with specific chemotherapy drugs listed below. | |
| | Chemotherapy Drug Breakthrough Detection Time in Minutes) | |
| | Dacarbazine (DTIC) (10.0 mg/mL) | |
| | Carmustine (BCNU) (3.3 mg/mL) | |
| | Cyclophosphamide (Cytoxan) (20.0 mg/mL) | |
| | Doxorubicin Hydrochloride (2.0 mg/mL) | |
| | 5-Fluorouracil (50.0 mg/mL) | |
| | Etoposide (20.0 mg/mL) | |
| | Paclitaxel (taxol) (6.0 mg/mL) | |
| | Thio-Tepa (10.0 mg/mL) | |
| Please note that Carmustine has extremely low permeation times of less than 30 minutes | | |
.
Page 4 of 5 (Rev 2)
{4}------------------------------------------------
KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)
K L 1 -
·
| Material | Color | Single Use | Sterilization | Dimensions | Physical Properties | Freedom from Pinhole | Residue Powder | Biocompatibility Test | Human Draize Test | Chemotherapy Drugs Permeation Test | Labeling Claim |
|-----------|-----------------------|------------|-----------------------------|------------------|---------------------|---------------------------------|---------------------------------|---------------------------------------------------------------------|-------------------|------------------------------------|---------------------------------------------------------------------------------------------------|
| Nitrile | Blue | Yes | Not Applicable; Non-Sterile | Meets ASTM D6319 | Meets ASTM D6319 | Meets ASTM D5151 and ASTM D6319 | Meets ASTM D6124 and ASTM D6319 | Pass Dermal Sensitization Test<br>Pass Primary Skin Irritation Test | Pass | Meets ASTM D6978-05 | Low Dermatitis and Chemotherapy Drugs Protection<br>Proposed device With Low Dermatitis Potential |
| Identical | Blue, White and Green | Identical | Identical | Identical | Identical | Identical | Identical | Identical | Identical | Not Applicable; Not Tested | |
·
. . . . . .
Page 5 of 5 (Rev 2)
・
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings outstretched.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
11
### November 16, 2012
Kossan Latex Industries (M) Sdn Bhd Ms. Cho Sow Fong Manager Regulatory Affairs Lot 6129, Jalan Haji Abdul Manan, Batu 5 ¼ Jalan Meru, Klang, 41050 Selangor, Malaysia
Re: K120066
Trade/Device Name: Powder Free Nitrile Patient Examination Gloves, Blue, White and Green Colored, Non-Sterile (Low Dermatitis Potential Labeling Claim)
Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: November 1, 2012
Received: November 2, 2012
Dear Ms. Fong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2 - Ms. Fong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
DN: c=US. o=U.S. Government, ou=HHS DA, ou=People, cn=Anthony U. Wa 342.19200300.100.1.1=1300092402 te: 2012.11.15 13:30:20 -05 00
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
<120066
Device Name:
Powder Free Nitrile Patient Examination Gloves. Blue, White and Green Colored. Non-Sterile (Low Dermatitis Potential Labeling Claim)
### Indications for Use:
A patient examination glove is a disposable device intended for medical purposes that is wom on the A patient exammation grove is a cispessor contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C) ×
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office Of Device Evaluation (ODE)
Page 1 of
Mary S. Runne : 2012, 11, 16 13:13:32 -05'00
(Division Sign-Off) (Division of Anesthesiology, General Hospital Division Control, Dental Devices
510(k) Number: K120066
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.