POWDER FREE NITRILE EXAMINATION GLOVES (WHITE, BLUE, AND PURPLE COLOR)

K051318 · Platinum Glove Industries Sdn. Bhd. · LZA · Jun 17, 2005 · General Hospital

Device Facts

Record IDK051318
Device NamePOWDER FREE NITRILE EXAMINATION GLOVES (WHITE, BLUE, AND PURPLE COLOR)
ApplicantPlatinum Glove Industries Sdn. Bhd.
Product CodeLZA · General Hospital
Decision DateJun 17, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

This glove is disposable and intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Story

Powder-free nitrile examination gloves; disposable; worn by healthcare examiners; prevents cross-contamination between patient and examiner; provides barrier protection.

Clinical Evidence

Bench testing only.

Technological Characteristics

Nitrile material; powder-free; disposable; examination glove form factor; non-sterile.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines representing movement or progress. The seal is simple and monochromatic, with a focus on the department's name and the human-centered symbol. JUN 17 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. So Soon Cheaw General Manager Platinum Glove Industries Sdn. Bhd. 4015A, 4 Floor, President House, Jalan Sultan Ismail, 50250 Kuala Lumpur, MALAYSIA Re: K051318 Trade/Device Name: Powder Free Nitrile Examination Gloves (White, Blue and Purple Color) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: May 17, 2005 Received: May 23, 2005 Dear Mr. Cheaw: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Mr. Cheaw Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chi-Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ుండి ప్రాథమ ## 3.0 Indication for Use Statement: ## INDICATIONS FOR USE 510(k) Number (if known) : K 0513 18 Device Name: Powder Free Nitrile Examination Gloves (White, Blue and Purple Color) Indication For Use: This glove is disposable and intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. Prescription Use (Part 21 CFR 801 Subpart D) NEEDED) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF Concurrence of CDRH, Office Of Device Evaluation (ODE) Shuley H. Murphy, M.D. (Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices Page 1 of ___________ 510(k) Number: K051318
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...