The gloves are intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient’s body, fluids, waste, or environment.
Device Story
Comfit Powder Free Latex Examination Gloves are disposable, beaded, hypoallergenic latex gloves. Used by healthcare and operating room personnel to provide a protective barrier against contamination. Device functions as a physical barrier between the wearer and the patient or environment. Performance verified through physical property testing and biocompatibility assessment.
Clinical Evidence
Clinical performance supported by a Modified Draize test conducted on 200 human subjects. Results demonstrated no potential for dermal irritation or sensitization, supporting the 'hypoallergenic' labeling claim.
Indicated for use by healthcare and operating room personnel as a protective barrier to prevent cross-contamination between the user and the patient's body, fluids, waste, or environment.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K962189 — COMFIT POWDER FREE GREEN EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN) · Wembley Rubber Products (M) Sdn Bhd · Jul 26, 1996
K962188 — COMFIT POWDER FREE BLUE EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN) · Wembley Rubber Products (M) Sdn Bhd · Jul 26, 1996
K953590 — COMFIT · Wembley Rubber Products (M) Sdn Bhd · Jun 27, 1996
K980618 — LOVYTEX SDN BHD POWDER-FREE LATEX EXAMINATION GLOVES · Lovytex Sdn Bhd · Mar 13, 1998
K021855 — TOP GLOVE POWDERFREE LATEX EXAMINATION GLOVE WITH PROTEIN LABELLING CLAIMS (50 MICROGRAMS OR LESS) · Top Glove Sdn. Bhd. · Aug 5, 2002
Submission Summary (Full Text)
{0}
OCT 11 1996
WRP
K953213
# CONTACT PERSON: MR. Y. W. CHOW
## Wembley Rubber Products (M) Sdn Bhd
### Marketing / Factory
Lot 1, Jalan 3
Kawasan Perusahaan Bandar Baru
Salak Tinggi, 43900 SEPANG
Selangor, Malaysia
Telex: MA 39599 WRP
Tel: +603-846 1486
Fax: +603-846 1557
## APPENDIX 4
### 510 (K) SUMMARY
1. **Trade Name** : COMFIT POWDER FREE LATEX EXAMINATION GLOVES (HYPOALLERGENIC, BEADED, PROTEIN LABELING CONTENT)
2. **Common Name** : Examination Gloves
3. **Classification Name** : Patient Examination Glove
4. **Substantial Equivalence** :
Class I latex patient examination glove 80 LYY, powder free, hypoallergenic, protein labeling content, that meets all of the requirements of ASTM standard D3578-91.
5. **Description of device** :
Class I latex patient examination glove 80 LYY, powder free, hypoallergenic, protein labeling content, that meets all of the requirements of ASTM standard D3578-91.
6. **Intended use of device** :
The gloves are intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient’s body, fluids, waste, or environment.
Page 1 of 3
REGISTERED
{1}
Page 2 of 3
# 7. Summary of Performance data :
Performance data of gloves to ASTM D 3578-91 and FDA 1000 ml watertight test.
| TEST | ASTM D 3578-91 | COMPETIT Powder Free Latex Examination Gloves (Hypo, Beaded, Protein Labeling Content) |
| --- | --- | --- |
| 1. Watertight (1000 ml) | S-4, AQL 4.0 | Pass based on AQL of 2.5 |
| 2. Length (mm) | | |
| Size
XS
S
M
L
XL | -
min 230°
min 230
min 230
- | 240
242
238
239
240 |
| 3. Palm width (mm) | | |
| Size
XS
S
M
L
XL | -
80 ± 10
95 ± 10
111 ± 10
- | 79
85
96
105
110 |
| 4. Thickness (mm) | | |
| Finger
Palm | min 0.08
min 0.08 | 0.25
0.19 |
| 5. Physical Properties | | |
| Before Ageing : | | |
| Tensile Strength (MPa)
Ultimate Elongation (%) | min 21
min 700 | 28.4
873 |
| After Ageing : | | |
| Tensile Strength (MPa)
Ultimate Elongation (%) | min 16
min 500 | 25.8
854 |
| 6. Powder Content | - | below 2 mg / glove |
| 7. Protein Content | - | below 50 microgram / gram |
{2}
Page 3 of 3
## 8. Substantial Equivalence based on assessment of Non-Clinical performance data
The performance test data of device as shown above indicate that this glove meets requirements of ASTM D 3578-91.
Protein content tested on recently manufactured and accelerated ageing gloves using ASTM D5712 is below 50 microgram / gram.
## 9. Assessment of Clinical Performance data
Clinical data of 200 Human subjects on Modified Draize test done on this glove demonstrated no potential for eliciting either dermal irritation or sensitization and support the labeling claim for “hypoallergenic”.
## 10. Conclusion
This glove exceeds the ASTM D 3578-91 requirements, meet pinhole FDA requirements, meet Hypoallergenic labeling claim and below 50 microgram / gram protein content labeling claim.
Date Summary Prepared: June 05, 1996
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.