COMFIT POWDER FREE BLUE EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN)
K962188 · Wembley Rubber Products (M) Sdn Bhd · LYY · Jul 26, 1996 · General Hospital
Device Facts
Record ID
K962188
Device Name
COMFIT POWDER FREE BLUE EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN)
Applicant
Wembley Rubber Products (M) Sdn Bhd
Product Code
LYY · General Hospital
Decision Date
Jul 26, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
The gloves are intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient’s body, fluids, waste, or environment.
Device Story
Device is a powder-free, low-protein, hypoallergenic latex patient examination glove. Used by healthcare and operating room personnel as a protective barrier to prevent contamination between the wearer and the patient's body, fluids, waste, or environment. Device functions as a physical barrier; no electronic or mechanical components.
Clinical Evidence
Clinical study conducted on 200 human subjects using the Modified Draize test. Results demonstrated no potential for eliciting dermal irritation or sensitization, supporting the 'hypoallergenic' labeling claim.
Technological Characteristics
Latex patient examination glove; powder-free; low protein (<50 µg/g). Meets ASTM D3578-91 standards for dimensions, tensile strength, and ultimate elongation. Watertight integrity verified via 1000 ml test (AQL 2.5).
Indications for Use
Indicated for use by healthcare and operating room personnel as a protective barrier to prevent cross-contamination between personnel and patients.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
K953590 — COMFIT · Wembley Rubber Products (M) Sdn Bhd · Jun 27, 1996
K032940 — BRIGHTWAY BRAND LATEX EXAMINATION GLOVE-(POWDER FREE, STERILE) CONTAINING 50 UGM OR LESS OF WATER EXTRACTABLE PROTEIN PE · Brightway Holdings Sdn. Bhd. · Nov 20, 2003
K962189 — COMFIT POWDER FREE GREEN EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN) · Wembley Rubber Products (M) Sdn Bhd · Jul 26, 1996
K974832 — POWDERFREE LATEX EXAMINATION GLOVES · Servicom Services Sdn Bhd · Feb 11, 1998
Submission Summary (Full Text)
{0}
WKP
CONTACT PERSON : MR. Y.W. CHOW
WEMBLEY RUBBER PRODUCTS
(M) SDN BHD
Lot 1, Jalan 3
Kawasan Perusahaan Bandar Baru
Salak Tinggi, 43900 Sepang,
Selangor, Malaysia.
Corporate Office Tel: 8461486
8461867
Fax: 8461485
8461557
Factory Office: 8461495
8461496
Fax: 8461488
8461608
K962188
JUL 26 1996
# 510 (K) SUMMARY
1. **Trade Name** : COMFIT POWDER FREE BLUE EXAMINATION GLOVES (HYPOALLERGENIC, LOW PROTEIN)
2. **Common Name** : Examination Gloves
3. **Classification Name** : Patient Examination Glove
4. **Substantial Equivalence** :
Class I latex patient examination glove 80 LYY, powder free, hypoallergenic, low protein, that meets all of the requirements of ASTM standard D3578-91.
5. **Description of device** :
Class I latex patient examination glove 80 LYY, powder free, hypoallergenic, low protein, that meets all of the requirements of ASTM standard D3578-91.
6. **Intended use of device** :
The gloves are intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient’s body, fluids, waste, or environment.
Page 1 of 3
BS EN ISO 9002 : 1994
BRITISH STANDARDS INSTITUTION
REGISTRATION NO: FM 13934
GOOD MANUFACTURING PRACTICE
UK DEPARTMENT OF HEALTH
REGISTRATION NO: R0589/SP
REGISTERED
MALAYSIA
QUALITY SYSTEM MS ISO 9002 : 1991
REGISTRATION NO: AR 0027
{1}
Performance data of gloves to ASTM D 3578-91 and FDA 1000 ml watertight test.
| TEST | ASTM D 3578-91 | COMFIT Powder Free Blue Examination Gloves (Hypoallergenic, Low Protein) |
| --- | --- | --- |
| 1. Watertight (1000 ml) | S-4, AQL 4.0 | Pass based on AQL of 2.5 |
| 2. Length (mm) | | |
| Size | XS | - |
| | S | min 230 |
| | M | min 230 |
| | L | min 230 |
| | XL | - |
| 3. Palm width (mm) | | |
| Size | XS | - |
| | S | 80 ± 10 |
| | M | 95 ± 10 |
| | L | 111 ± 10 |
| | XL | - |
| 4. Thickness (mm) | | |
| Finger | min 0.08 | 0.40 |
| Palm | min 0.08 | 0.30 |
| 5. Physical Properties | | |
| Before Ageing : | | |
| Tensile Strength (MPa) | min 21 | 31.23 |
| Ultimate Elongation (%) | min 700 | 895 |
| After Ageing : | | |
| Tensile Strength (MPa) | min 16 | 26.64 |
| Ultimate Elongation (%) | min 500 | 946 |
| 6. Powder Content | - | below 2 mg / glove |
| 7. Protein Content | - | below 50 microgram / gram |
Page 2 of 3
{2}
Page 3 of 3
## 8. Substantial Equivalence based on assessment of Non-Clinical performance data
The performance test data of device as shown above indicate that this glove meets requirements of ASTM D 3578-91.
Protein content tested on recently manufactured and accelerated ageing gloves using ASTM D5712 is below 50 microgram / gram.
## 9. Assessment of Clinical Performance data
Clinical data of 200 Human subjects on Modified Draize test done on this glove demonstrated no potential for eliciting either dermal irritation or sensitization and support the labeling claim for “hypoallergenic”.
## 10. Conclusion
This glove exceeds the ASTM D 3578-91 requirements, meet pinhole FDA requirements, meet Hypoallergenic labeling claim and below 50 microgram / gram protein content labeling claim.
Date Summary Prepared : May 27, 1996
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.