PAL Sterilization Case

K243589 · Microaire Surgical Instruments, LLC · KCT · Jun 20, 2025 · General Hospital

Device Facts

Record IDK243589
Device NamePAL Sterilization Case
ApplicantMicroaire Surgical Instruments, LLC
Product CodeKCT · General Hospital
Decision DateJun 20, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The PAL Sterilization Case is a reusable sterilization case intended to provide storage for the PAL handpiece, the PAL cable, and 4 PAL cannulas during sterilization, storage, and transportation within the hospital environment. The PAL Sterilization Case is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with FDA-cleared sterilization wraps. The PAL Sterilization Case has only been evaluated for a non-stacked configuration. The validated steam sterilization cycle parameters are listed below: Cycle Type Exposure Temperature Exposure Time (minutes) Minimum Dry Time (minutes) Compatible Sterile Barriers Pre-Vacuum 132 °C (270 °F) 4 20 Wrap(s) Gravity 121 °C (250 °F) 30 20 Wrap(s)

Device Story

Reusable stainless-steel sterilization case; designed to house PAL handpiece, cable, and 4 cannulas. Features separable lid and base with latch mechanisms; internal medical-grade silicone mounts secure instruments. Used in hospital settings by sterile processing staff to organize and protect instruments during steam sterilization (gravity or pre-vacuum cycles). Perforated design allows steam penetration; double-wrapped with FDA-cleared wraps before autoclaving. Facilitates safe transport and storage of surgical instruments; ensures sterility maintenance post-autoclave.

Clinical Evidence

No clinical data. Bench testing only. Validation included steam sterilization efficacy (SAL 10^-6) for pre-vacuum (132°C/4 min) and gravity (121°C/30 min) cycles, cleaning validation per AAMI ST98, and weight verification.

Technological Characteristics

Materials: 304L and 304 stainless steel, medical-grade silicone inserts. Dimensions: 15.91" x 6.55" x 2.42". Sterilization: Steam (gravity and pre-vacuum). Standards: ANSI/AAMI ST77:2013/(R)2018, AAMI TIR12:2020, ISO 17665:2024, AAMI ST98:2022.

Indications for Use

Indicated for storage and steam sterilization of MicroAire PAL handpiece, cable, and up to 4 cannulas (≤30 cm) in hospital environments. For use with FDA-cleared sterilization wraps in non-stacked configurations.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION June 20, 2025 MicroAire Surgical Instruments, LLC Ashley Sanguinetti Senior Regulatory Affairs Specialist 3590 Grand Forks Blvd. Charlottesville, Virginia 22911 Re: K243589 Trade/Device Name: PAL Sterilization Case Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: May 21, 2025 Received: May 21, 2025 Dear Ashley Sanguinetti: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243589 - Ashley Sanguinetti Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K243589 - Ashley Sanguinetti Page 3 Sincerely, Stephen A. Anisko -S Digitally signed by Stephen A. Anisko -S Date: 2025.06.20 17:34:35 -04'00' for: Christopher Dugard Director DHT4C: Division of Infection Control OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K243589 Device Name PAL Sterilization Case Indications for Use (Describe) The PAL Sterilization Case is a reusable sterilization case intended to provide storage for the PAL handpiece, the PAL cable, and 4 PAL cannulas during sterilization, storage, and transportation within the hospital environment. The PAL Sterilization Case is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with FDA-cleared sterilization wraps. The PAL Sterilization Case has only been evaluated for a non-stacked configuration. The validated steam sterilization parameters are listed below: | Cycle Type | Exposure Temperature | Exposure Time (minutes) | Minimum Dry Time (minutes) | Compatible Sterile Barriers | | --- | --- | --- | --- | --- | | Pre-Vacuum | 132 °C (270 °F) | 4 minutes | 20 minutes | Wrap(s) | | Gravity | 121 °C (250 °F) | 30 minutes | 20 minutes | Wrap(s) | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) P3C Publishing Services (501) 443-6740 {4} Page 1 of 8 K243589 510(k) Summary MicroAire PAL Sterilization Case In accordance with 21 CFR 807.92 the following summary of information is provided: 1. Submitter MicroAire Surgical Instruments, LLC 3590 Grand Forks Boulevard Charlottesville, Virginia 22911 USA 2. Primary Contact Person Ashley Sanguinetti Senior Regulatory Affairs Specialist MicroAire Surgical Instruments, LLC Phone: (434) 975-8344 Fax: (434) 975-4144 Email: ashley.sanguinetti@microaire.com 3. Date Prepared June 18, 2025 4. Device Identification Trade/Proprietary Name: PAL Sterilization Case Common/Usual Name: Sterilization Case Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Containers, Trays, Cassettes &amp; Other Accessories Regulatory Class: Class 2 Product Code: KCT Classification Panel: General Hospital 5. Legally Marketed Predicate Device(s) Device name: Sonicision™ reusable sterilization tray 510(k) number: K180915 Manufacturer: Covidien, LLC No reference devices were used in this submission. {5} Page 2 of 8 K243589 # 6. Indications for Use Statement The PAL® Sterilization Case is a reusable sterilization case intended to provide storage for the PAL® handpiece, the PAL® cable, and 4 PAL® cannulas during sterilization, storage, and transportation within the hospital environment. The PAL® Sterilization Case is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with FDA-cleared sterilization wraps. The PAL® Sterilization Case has only been evaluated for a non-stacked configuration. The validated steam sterilization cycle parameters are listed below: | Cycle Type | Exposure Temperature | Exposure Time (minutes) | Minimum Dry Time (minutes) | Compatible Sterile Barriers | | --- | --- | --- | --- | --- | | Pre-Vacuum | 132 °C (270 °F) | 4 | 20 | Wrap(s) | | Gravity | 121 °C (250 °F) | 30 | 20 | Wrap(s) | # 7. Device Description The MicroAire PAL Sterilization Case is a non-surgical, non-patient contacting, accessory to the already existing PAL (Power-Assisted Liposuction) System (cleared for market under K212024, K171286 and K220318), designed to contain reusable surgical instruments during steam sterilization. The case is designed to contain four (4) PAL Multi-Use Cannulas, one (1) PAL Electric Handpiece, and one (1) PAL Handpiece Cable. The subject product scope only includes the sterilization case, constructed of a separable base and lid, as illustrated in Figure 11-A. ![img-0.jpeg](img-0.jpeg) Figure 11-A PAL Sterilization Case (PAL-CASE) {6} Page 3 of 8 K243589 The PAL Sterilization Case (subject device) is a stainless-steel case comprised of a separable lid and base. These two components latch together via the latch mechanisms on either side of the base which mate with the flanges on either side of the lid. Inside the case there are silicone mounts for a PAL handpiece, cable, and 4 cannulas (30 cm or less in length). The sterilization case is used to steam sterilize its contents via standard gravity or dynamic air removal autoclave processes. The subject device is a non-surgical, non-patient contacting, accessory device to the MicroAire PAL System. The subject device allows a user to organize and transport PAL reusable surgical instruments before, during, and after steam sterilization. The case is provided non-sterile and contains silicon mounting brackets to keep the devices in place, and a perforated stainless-steel enclosure to allow penetration of steam. The case has an overall footprint of approximately 16" x 6" x 2". Its construction of 1mm thick stainless steel and silicone make it lightweight and easy to transport when fully loaded with instruments. The lid also has two carrying handles on its lid for easy transport. ## 8. Technological Characteristics Comparison The following table compares the subject device MicroAire PAL Sterilization Case with the predicate device Covidien IIc, Sonicision™ reusable sterilization tray (K180915) with respect to indications for use, mechanisms of operation, technological characteristics, materials, and performance testing. {7} Page 4 of 8 K243589 Table 11-A Comparison to Predicate | Characteristic | Predicate Device Sonicision™ Reusable Sterilization Tray, K180915 | | | | | Subject Device MicroAire, PAL® Sterilization Case, K243589 | | | | | Device Comparison | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | CLASSIFICATION & INDICATIONS | | | | | | | | | | | | | FDA Classification | Class II | | | | | Class II | | | | | Same | | FDA Product Code | KCT | | | | | KCT | | | | | Same | | FDA Classification Name | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories | | | | | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories | | | | | Same | | Indications for Use | The Sonicision™ reusable sterilization tray is intended to provide storage for the compatible reusable generator and reusable battery insertion guides during sterilization, storage, and transportation within the hospital environment. The Sonicision™ reusable sterilization tray is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with an FDA-cleared sterilization wrap or rigid container (validated with Aesculap™* rigid container model JN741/JK789). The sterilization tray has only been evaluated for a non-stacked configuration. The validated sterilization cycle parameters are as follows: Steam Sterilization Cycles | | | | | The PAL Sterilization Case is a reusable sterilization case intended to provide storage for the PAL handpiece, the PAL cable, and 4 PAL cannulas during sterilization, storage, and transportation within the hospital environment. The PAL Sterilization Case is only intended to maintain sterility of the enclosed components during autoclave sterilization when used in conjunction with FDA-cleared sterilization wraps. The PAL Sterilization Case has only been evaluated for a non-stacked configuration. The validated sterilization cycle parameters are as follows: Steam Sterilization Cycles | | | | | Similar | | | Cycle Type | Exposure Temperature | Exposure Time (minutes) | Minimum Dry Time¹ (minutes) | Compatible Sterile Barriers | Cycle Type | Exposure Temperature | Exposure Time (minutes) | Minimum Dry Time (minutes) | Compatible Sterile Barriers | | | | Pre- Vacuum | 132 °C (270 °F) | 4 | 40 | Wrap(s) and rigid cotainers² | Pre- Vacuum | 132 °C (270 °F) | 4 | 20 | Wrap(s) | | | | Pre- Vacuum | 135 °C (275 °F) | 3 | 40 | Wrap(s) and rigid cotainers² | Gravity | 121 °C (250 °F) | 30 | 20 | Wrap(s) | | | | Gravity | 132 °C (270 °F) | 15 | 40 | Wrap(s) | | | | | | | | | Gravity | 135 °C (275 °F) | 10 | 30 | Wrap(s) | | | | | | | | ¹ Due to the variations in autoclave sterilization equipment, actual dry times may vary from the minimum dry times validated for this product. ² Validated with Aesculap™* rigid container model JN741/JK789. | | | | | | | | | | | | {8} Page 5 of 8 K243589 | Characteristic | Predicate Device Sonicision™ Reusable Sterilization Tray, K180915 | Subject Device MicroAire, PAL® Sterilization Case, K243589 | Device Comparison | | --- | --- | --- | --- | | **SUMMARY OF PHYSICAL CHARACTERISTICS** | | | | | Image | | | Similar | | Material Specifications | • Base: Polyphenylsulfone • Lid: Stainless Steel | • Base and Lid: 304L and 304 stainless steels • Perforation inserts: Medical Grade Gray Silicone | Same/ Different | | Dimensions | LxWxH: 8.5” x 11.0” x 3.0” | LxWxH: 15.91” x 6.55” x 2.42” | Similar | | **SUMMARY OF TECHNOLOGICAL CHARACTERISTICS** | | | | | Case Type | Sterilization Case | Sterilization Case | Same | | Enclosed devices for sterilization | The tray can contain at a maximum: One (1) Sonicision™ Reusable Generator, Two (2) Sonicision™ Reusable Battery Insertion Guides. | The tray can contain at a maximum: One (1) PAL Handpiece One (1) PAL Cable Up to four (4) PAL Cannulas (30 cm or less in length). The sterilization case is an accessory to the approved PAL System, designed to contain reusable surgical instruments during steam sterilization. The PAL Sterilization Case is a stainless-steel case comprised of a separable lid and base. These two components latch together via the latch mechanisms on either side of the base which mate with the flanges on either side of the lid. Inside the case there are silicone mounts for a PAL handpiece, cable, and 4 cannulas | Different | {9} Page 6 of 8 K243589 | Characteristic | Predicate Device Sonicision™ Reusable Sterilization Tray, K180915 | Subject Device MicroAire, PAL® Sterilization Case, K243589 | Device Comparison | | --- | --- | --- | --- | | | | (30cm or less in length). The sterilization case is used to steam sterilize its contents via standard gravity or dynamic air removal autoclave processes. The silicone inserts hold the instruments in place. The case can be double wrapped using FDA-cleared sterilization wraps and loaded into the autoclave. | | | Device Description | The Sonicision Sterilization Tray is a reusable, optional accessory to the Sonicision System (cleared through K180149). This device is specially made to enclose up to one Sonicision Reusable Generator and up to two Sonicision Reusable Battery Insertion Guides in order to provide storage during sterilization, storage, and transportation within the hospital environment. | The PAL Sterilization Case is a reusable, optional accessory to the approved PAL System (cleared through K212024). This device is specially made to enclose up to one PAL Electric Handpiece, one PAL Cable and up to four PAL Multi-use Cannulas (30 cm or less in length) in order to provide storage during sterilization, storage, and transportation within the hospital environment. | Similar | | Sterilization methods | Steam (Pre-vacuum and Gravity) | Steam (Pre-vacuum and Gravity) | Same | | Steam sterilization parameters | **132 °C Pre-vacuum Steam Cycle** Exposure temperature: 270 °F (132 °C) Exposure time: 4 minutes Minimum dry time: 40 minutes | **132 °C Pre-vacuum Steam Cycle** Exposure temperature: 270 °F (132 °C) Exposure time: 4 minutes Minimum dry time: 20 minutes | Similar | | | **135 °C Pre-vacuum Steam Cycle** Exposure temperature: 275 °F (135 °C) Exposure time: 3 minutes Minimum dry time: 40 minutes | | | | | **132 °C Gravity Steam Cycle** Exposure temperature: 270 °F (132 °C) Exposure time: 15 minutes Minimum dry time: 40 minutes | **121 °C Gravity Steam Cycle** Exposure temperature: 250 °F (121 °C) Exposure time: 30 minutes Minimum dry time: 20 minutes | Similar | | | **135 °C Gravity Steam Cycle** Exposure temperature: 275 °F (135 °C) Exposure time: 10 minutes Minimum dry time: 30 minutes | | | | Stacking | Do not stack cases and trays in the sterilization chamber. | Do not stack cases and trays in the sterilization chamber. | Same | | Maximum Load Capacity | < 1800 grams (3.97 pounds, including tray) | 5.582 lbs ± 0.026 lbs (including case) | Different | {10} Page 7 of 8 K243589 # 9. Non-Clinical Performance Data Non-clinical testing was performed to verify and validate the design of the PAL Sterilization Case against specified design requirements and to assure conformance with applicable recognized national consensus standards and FDA Guidance documents. The PAL Sterilization Case passed all non-clinical testing. The PAL Sterilization Case complies with all applicable voluntary consensus standards and FDA Guidance documents for Sterilization Cases (per Product Code - KCT), shown in the table below. Table 11-B Summary of Non-Clinical Testing | Test Description | Applicable Standards | Acceptance Criteria | Test Results | | --- | --- | --- | --- | | PAL Sterilization Case Gravity Steam Sterilization Validation | • ANSI/AAMI ST77:2013/(R)2018 • AAMI TIR12:2020 • ISO 17665:2024 • FDA Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling | A full cycle of 30 minutes at 121°C using gravity air displacement for the PAL Sterilization Case in a double-wrapped configuration. | Passed The results of the study found that a 121°C, 30-minute gravity sterilization and dry time of 20 minutes is sufficient to sterilize (and dry) the PAL product family (not including cannulas over 30 cm in length) to a Sterility Assurance Level (SAL) of 10^{-6} in the PAL Sterilization Case (PAL-CASE) in a double wrapped configuration. | | PAL Sterilization Case Pre-Vacuum Steam Sterilization Validation | • ANSI/AAMI ST77:2013/(R)2018 • AAMI TIR12:2020 • ISO 17665:2024 • FDA Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling | A full cycle of four (4) minutes at 132°C using pre-vacuum air removal for the PAL Sterilization Case in a double wrapped configuration. | Passed The results of the study found that a 132°C, 4-minute pre-vacuum sterilization and dry time of 20 minutes is sufficient to sterilize (and dry) the PAL product family (not including cannulas over 30 cm in length) to a Sterility Assurance Level (SAL) of 10^{-6} in the PAL Sterilization Case (PAL-CASE) in a double wrapped configuration. | | Validation of Automated and Manual Cleaning Methods for PAL Sterilization Case | • AAMI ST98:2022 • FDA Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling | In accordance with ST98 requirements. There shall be no visible evidence of test soil remaining on any surface of the device post cleaning from a distance of 18 inches with 20/20 vision. | Passed The results confirm that MicroAire’s draft automated and manual cleaning instructions for the PAL Sterilization Case (PAL-CASE) are effective in removing soil from the devices to the point that they are considered cleaned. | | Design Verification - PAL Sterilization Case Weight | • ANSI/AAMI ST77:2013/(R)2018 (for sample size) | There shall be 90% confidence that the mean weight of the PAL Sterilization Case population is less than 25lb while containing the following: one PAL Electric Handpiece, one PAL Instrument Cable, and four PAL Cannula. | Passed There is a 90% confidence that the fully loaded PAL Sterilization case population has a mean weight of 5.582 lb ± 0.026 lb, which is within the required upper bound of 25 lb. | {11} Page 8 of 8 K243589 ## 10. Clinical Performance Data The PAL Sterilization Case has not been the subject of clinical testing. ## 11. Conclusion The conclusions drawn from nonclinical testing demonstrate that the subject device, K243589, Microaire PAL Sterilization Case, is as safe, as effective and performs as well or better than the legally marketed predicate device, K180915 Covidien Sonicision™.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...