K052992 · Conmed Linvatec · FRG · Oct 16, 2006 · General Hospital
Device Facts
Record ID
K052992
Device Name
CONMED LINVATEC INSTRUMENT STERILIZATION TRAY
Applicant
Conmed Linvatec
Product Code
FRG · General Hospital
Decision Date
Oct 16, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The ConMed Linvatec® Instrument Sterilization Trays are for loading surgical instruments in order to conveniently organize, sterilize, transport and store the instrument between uses. ConMed Linvatec® Instrument Sterilization Trays can be used in a steam sterilization cycles. The validated sterilization cycles are provided in Table II. Important Notes The cycles were validated using Kimguard (Kimberly Clark) wrap. The sterilization exposure times are extended. Please refer to Table 2 for cycle times. Only sterilization wraps cleared by FDA for extended sterilization exposure times should be used with these trays. The sterilized contents are for immediate use and should not be stored for any duration. The trays are intended for sterilization of non-porous loads, e.g. hooks, grassers, scissors, knot pushers, probes, extractors, drivers, rasps, obdurators, etc. The travs are not intended to be stacked during the sterilization process
Device Story
ConMed Linvatec Instrument Sterilization Tray consists of rigid, multi-piece base trays, lids, and internal individualized insert trays with custom silicone brackets/mats. Designed to hold, protect, and organize non-porous surgical instruments during steam sterilization cycles in standard autoclaves. Operated by healthcare personnel in clinical settings. Tray lid locks to base via stainless steel latches. Perforated design allows steam passage. Reusable; not intended for repair. Benefits include organized transport and effective sterilization of surgical instruments.
Clinical Evidence
Bench testing only. Sterilization cycles validated using Kimguard (Kimberly Clark) wrap per AAMI TIR No.12-1994 standards. Validated for pre-vacuum (270°F, 5 min exposure, 10 min dry) and gravity (270°F, 30 min exposure, 10 min dry) steam cycles.
Technological Characteristics
Materials: Radel R-5000 polymer resin (lid/base/insert), 300 Series Stainless Steel (latches), biomedical grade silicone (brackets/mats meeting FDA CFR 177.2600). Design: Perforated plastic tray with locking lid. Energy: Steam sterilization. Connectivity: None. Sterilization: Steam (pre-vacuum and gravity).
Indications for Use
Indicated for the organization, sterilization, transport, and storage of non-porous surgical instruments (e.g., hooks, graspers, scissors, knot pushers, probes, extractors, drivers, rasps, obturators) via steam sterilization. For immediate use only; not for storage. Not intended for stacking during sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
# 0C7 1 6 2006
## 510(k) Submission
## Summary of Safety and Effectiveness
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Linvatec is hereby submitting a 510(k) Summary of Safety and Effectiveness for the ConMed Linvatec Instrument Sterilization Tray.
## 510(k) # K052992
#### Submitter A.
ConMed Linvatec 11311 Concept Boulevard Largo, Florida 33773-4908
### B. Company Contact
Elizabeth Paul Manager, Regulatory Affairs (727) 399-5234 Telephone FAX (727) 399-5264
## C. Device Name
Trade Name: Instrument Sterilization Tray
Common Name: Instrument Sterilization Tray
Classification Name: Accessory to Sterilization Wrap, CFR 880.6850
Proposed Class/Device: Class II
Product Code: FRG
{1}------------------------------------------------
Summary of Safety and Effectiveness Instrument Sterilization Tray 510(k) # K052992 Page 2 of 4
### D. Predicate/Legally Marketed Devices
K993535 - MetaPak Multi-Purpose Instrument Tray - Riley Medical, Inc. K023658 - Sterilization cases, trays and cassettes - Carr Medical Products K944025 - MetaPak Multi-Purpose Instrument Tray - Riley Medical, Inc.
### E. Device Description
The ConMed Linvatec sterilization trays consist of base trays, lids and internal individualized insert travs. Each trav is composed of multiple pieces and is designed to be integrated into a single unit, which will contain and protect the interior components during sterilization. The interior structures of the tray have the ability to separately hold each individual piece during the entire duration they are in contact with the tray.
The rigid, multi piece tray holds medical device apparatus and accessories before, during and after the sterilization process. The tray set has a locking lid to contain products, which is held to the base by latches designed to fasten the lid to the base not allowing the two to separate. The trays are designed to fit any standard autoclave, which allows it to be effective for sterilization and be able to withstand the environment of repeated steam sterilization cycles in a normal operating room.
The tray is designed to allow repeated sterilization cycles. Though these travs are reusable they will not be serviced or repaired. Like the outer containing trays, the interior trays with individual containment custom brackets have been developed to withstand repeated use and will not be serviced or repaired. These custom brackets are made of materials that do not contaminate the tray or any of the parts or pieces contained within the tray.
The lid, base and insert for the proposed trays are made of Radel R-5000. This material is a polymer resin produced by Solvay Advanced Polymers, LLC. This material is identical to the Radel used in the predicate device cleared under K993535. Attachment I includes the material composition for the Radel R-5000.
The latches, which secure the lid to the base, are constructed of 300 Series Stainless Steel, identical to the stainless steel in the predicate devices cleared under K993535 and K944025.
{2}------------------------------------------------
Summary of Safety and Effectiveness Instrument Sterilization Tray 510(k) # K052992 Page 3 of 4
The instruments to be sterilized in the proposed trays are all non-porous devices and including instruments such as suture snares, various hooks, graspers, scissors, knot pushers, probes, extractors, drivers, rasps, obturators, etc.
Each tray includes brackets and containment mats which are made of a biomedical grade silicone. These are used to secure the instruments during transport, sterilization and storage. The drawings in Attachment III include the brackets and mats.
Additional device characteristics can be found in the chart included as Attachment A of this 510(k) Summary.
The brackets and containment mats are both constructed from biomedical grade silicone which has been manufactured to meet FDA CFR 177.2600. This material is identical to the predicate devices cleared under K993535 and K944025.
## Intended Use F. Intended Use
The ConMed Linvatec® Instrument Sterilization Trays are for loading surgical instruments in order to conveniently organize, sterilize, transport and store the instrument between uses.
ConMed Linvatec® Instrument Sterilization Trays can be used in a steam sterilization cycles. The validated sterilization cycles are provided in Table II. Important Notes
The cycles were validated using Kimguard (Kimberly Clark) wrap. The sterilization exposure times are extended. Please refer to Table 2 for cycle times. Only sterilization wraps cleared by FDA for extended sterilization exposure times should be used with these trays. The sterilized contents are for immediate use and should not be stored for any duration.
The trays are intended for sterilization of non-porous loads, e.g. hooks, grassers, scissors, knot pushers, probes, extractors, drivers, rasps, obdurators, etc.
The travs are not intended to be stacked during the sterilization process Substantial Equivalence ت
The ConMed Linvatec Instrument Sterilization Tray is substantially equivalent in materials and is similar in intended use and design to the below identified predicate devices. The proposed Instrument Sterilization Tray and the predicate devices are used for storage, transport, and sterilization of surgical instruments between uses. Both the proposed and predicate devices are suitable for use in steam sterilization processes.
The ConMed Linvatec Instrument Sterilization Tray will be offered in a combination of various metals and plastics.
{3}------------------------------------------------
Summary of Safety and Effectiveness Instrument Sterilization Tray 510(k) # K052992 Page 4 of 4
There are no significant differences between the proposed device and the predicate devices other than various tray sizes and layout of custom containment brackets. Also see Attachment B of this 510(k) summary for further details.
# Substantially Equivalent Predicate Marketed Device:
K993535 - MetaPak Multi-Purpose Instrument Tray - Riley Medical, Inc. K023658 - Sterilization cases, trays and cassettes - Carr Medical Products K944025 - MetaPak Multi-Purpose Instrument Tray - Riley Medical, Inc.
## H. Technological Characteristics
The design and materials used in of the ConMed Linvatec Instrument Sterilization Trays, a perforated plastic tray with a stainless steel latch and silicone containment brackets. that allows free steam passage is the substantially equivalent to the predicate devices. All trays have various dimensions such that the proposed trays and the predicates devices are substantially equivalent in size and volume. The sterilization validation of all trays has met the requirements of the same performance standard, AAMI TIR No.12-1994. The associated procedures for the use of the devices , i.e. the manufacturer's recommended sterilization cycles, are all similar.
### I. Summary and Conclusion
The Conmed Linvatec Instrument Sterilization Tray has the same intended use, technological characteristics and performance characteristics as the predicate devices. Thus, the proposed tray does not raise any new issues of safety and efficacy and per the definition under 21 CFR 807.100 is substantially equivalent to the referenced predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The caduceus is depicted in black, and the text is also in black, set against a white background. The overall design is simple and recognizable, representing the department's role in public health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 16 2006 Manager, Regulatory Affairs
Re: K052992
Ms. Elizabeth M. Paul
11311 Concept Boulevard Largo, Florida 33773-4908
ConMed Linvatec
Trade/Device Name: ConMed Linvatec Instrument Sterilization Tray Regulation Number: 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: September 19, 2006 Received: September 21, 2006
Dear Ms. Paul:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Ms. Paul
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K052992
# Device Name: ConMed Linvatec Instrument Sterilization Tray
# Indications for Use:
The ConMed Linvatec Instrument Sterilization Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport and store the instruments between uses.
The ConMed Linvatec Instrument Sterilization Tray is for use with steam sterilization.
The validated sterilization cycles are provided in Table II.
The cycles were validated using Kimguard (Kimberly Clark).
The sterilization exposure times are extended. Please refer to Table 2 for cycle tines. Only sterilization wraps cleared by FDA for extended sterilization exposure times should be used with these trays.
The sterilized contents are for immediate use and should not be stored for any duration.
The trays are intended for sterilization of non-porous loads, e.g. hooks, graspers, scissors, knot pushers, probes, extractors, drivers, rasps, obdurators, etc.
| Table 1 Tray Characteristics | | | | | |
|------------------------------|---------------------------------------|------------------------------------|--------------------|-------------------|-------------------|
| Catalog<br>Number | Description | Recommended<br>Max load<br>(lbs.)* | Length<br>(Inches) | Width<br>(Inches) | Depth<br>(Inches) |
| 9895 | ROTATOR CUFF STERILATION CASE<br>TRAY | 2.54 | 14.60 | 9.52 | 0.82 |
| | ROTATOR CUFF STERILE CASE (BASE) | | 14.45 | 9.4 | 0.4 |
| 8217 | OSTEOPREP STERILE TRAY (LID) | 4.08 | 20.88 | 10.13 | 0.82 |
| | OSTEOPREP STERILE TRAY (BASE) | | 20.63 | 9.81 | 1.19 |
| C6355 | SPECTRUM II (LID) | 11.90 | 20.75 | 9.93 | 1.116 |
| | SPECTRUM II (UPPER TRAY) | | 19.75 | 8.94 | 1.97 |
| | SPECTRUM II (BASE) | | 20.46 | 9.67 | 4.76 |
Table 1 lists the trays included in this submission.
*Includes the weight of the tray.
{7}------------------------------------------------
| Table 2 | | | | |
|--------------------------------|------------|------------------------|---------------------|----------------------|
| Validated Sterilization Cycles | | | | |
| Method | Cycle | Minimum<br>Temperature | Minimum<br>Exposure | Minimum<br>Dry Cycle |
| Steam | Pre-vacuum | 270°F (132°C) | 5 minutes | 10 minutes |
| Steam | Gravity | 270°F (132°C) | 30 minutes | 10 minutes |
Prescription Use______________________________________________________________________________________________________________________________________________________________ OR Over-the-Counter Use __X
(Part 21 CFR 801 subpart D)
(Part 21 CFR 807 subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE if NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elaine S. Maglull fus. Murphy
(Sign-Off)
n of Anesthesiology, General Hospital.
nesthesiology, General Hospital, tal Devices
K052992
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.