Stryker ChestShield Sterilization Tray System

K183701 · Paragon Medical · KCT · May 7, 2019 · General Hospital

Device Facts

Record IDK183701
Device NameStryker ChestShield Sterilization Tray System
ApplicantParagon Medical
Product CodeKCT · General Hospital
Decision DateMay 7, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The Stryker ChestShield Sterilization Tray System is intended to organize, enclose, sterilize, transport, and store Stryker implantable devices and surgical instruments within a healthcare facility when used in conjunction with a validated figid sterliization container in order to maintain sterility of the enclosed devices. The Stryker ChestShield Sterilization Tray System has been validated for use in the following sterilization cycle: Pre-vacuum Steam Enclosure Rigid Container Temperature 132°C (270°F) Sterilization Time 4 minutes Minimum Dry Time 30 minutes Maximum Weight1 19.23 lb. / 8.72 kg

Device Story

Stryker ChestShield Sterilization Tray System; tray system for organizing, storing, and sterilizing surgical instruments and implants. Components: anodized aluminum implant module with inlay; stainless steel instrument trays with stainless steel/silicone brackets; stainless steel lids with slide latches; stainless steel storage rack. Used in healthcare facilities; operated by clinical staff. Device placed in rigid sterilization container for steam sterilization; does not maintain sterility independently. Laser-etched/silkscreened markings assist in component placement. Facilitates sterilization of implants and instruments via steam penetration through perforations. Benefits: organized, sterile transport and storage of surgical sets.

Clinical Evidence

Bench testing only. Sterilization and dry time validation performed per ANSI/AAMI/ISO 17665-1:2006 using Geobacillus stearothermophilus biological indicators. Leveraged predicate data for biocompatibility (ISO 10993-5), cleaning validation (AAMI TIR 30:2011), and design verification (AAMI/ANSI ST77:2013).

Technological Characteristics

Materials: Stainless steel, anodized aluminum, USP Class VI silicone, thermoplastic polymers. Design: Perforated tray system with slide-latch lids, stored in stainless steel racks. Sterilization: Pre-vacuum steam (132°C, 4 min, 30 min dry). Connectivity: None. Sterilization method: Steam autoclave. Form factor: Modular trays/racks.

Indications for Use

Indicated for organizing, enclosing, sterilizing, transporting, and storing Stryker implantable devices and surgical instruments within a healthcare facility. Used with a validated rigid sterilization container to maintain sterility. Validated for pre-vacuum steam sterilization at 132°C for 4 minutes with 30-minute dry time. Load includes single-use implants, single-use instruments, reusable surgical instruments, and lumened instruments.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 7, 2019 Paragon Medical Rebecca Walker Regulatory Affairs Manager 8 Matchett Dr Pierceton, Indiana 46562 Re: K183701 Trade/Device Name: Stryker ChestShield Sterilization Tray System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: April 3, 2019 Received: April 5, 2019 # Dear Rebecca Walker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For David Krause, PhD Acting Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183701 ### Device Name ## Stryker ChestShield Sterilization Tray System Indications for Use (Describe) The Stryker ChestShield Sterilization Tray System is intended to organize, enclose, sterilize, transport, and store Stryker implantable devices and surgical instruments within a healthcare facility when used in conjunction with a validated figid sterliization container in order to maintain sterility of the enclosed devices. The Stryker ChestShield Sterilization Tray System has been validated for use in the following sterilization cycle: | | Pre-vacuum Steam | |--------------------|---------------------| | Enclosure | Rigid Container | | Temperature | 132°C (270°F) | | Sterilization Time | 4 minutes | | Minimum Dry Time | 30 minutes | | Maximum Weight1 | 19.23 lb. / 8.72 kg | 'The validated load configuration for the Stryker Chestin Tray System included single-use implants (plates, bone screws, etc.), single-use instruments (drills), reusable surgical instruments (benders, etc.) and lumened instruments. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center. To the right of the pentagon are the words "PARAGON" and "MEDICAL" stacked on top of each other in black font. There are black lines above and below the words. ## 510(k) Summary for Stryker ChestShield Sterilization Tray System, K183701 In accordance with 21 CFR 807.92, the following 510(k) summary is provided: | 1. Applicant Information: | Paragon Medical<br>8 Matchett Drive<br>Pierceton, IN 46562 USA<br>Phone: 574-594-2140<br>Fax: 574-594-2154 | |---------------------------------------|--------------------------------------------------------------------------------------------------------------| | 2. Correspondent Contact Information: | Rebecca Walker<br>Regulatory Affairs Manager<br>Rebecca.walker@nninc.com<br>Phone: 574-594-2140 ext. 10384 | | 3. Date Prepared: | May 2, 2019 | | 4. Trade Name: | Stryker ChestShield Sterilization Tray System | | 5. Common Name: | Sterilization Tray | | 6. Classification Name: | Sterilization wrap containers, trays, cassettes and other accessories<br>(21 CFR 880.6850, Product Code KCT) | | 7. Predicate Device: | Stryker Universal Select Sterilization Tray System, K173615 | - 8. Device Description: The Stryker ChestShield Sterilization Tray System is a tray system designed to store implants and surgical instrumentation. The tray system consists of an anodized aluminum implant module with an inlay which stores the implantable components and stainless steel instrument trays, which have stainless steel and silicone brackets to contain the reusable and singleuse instruments. Both the implant module and instrument trays have laser-etched and silkscreened artwork to assist the end user in correct placement of the contents. The tray components have stainless steel lids which feature a slide latch. The trays and implant module are organized and stored in a stainless steel rack. The tray system does not maintain sterility; the rack is placed in a rigid sterilization container to act as a sterile barrier for sterilization. The tray system provided in a nonsterile condition and must be sterilized prior to and after each use. - 9. Indications for Use: The Stryker ChestShield Sterilization Tray System is intended to organize, enclose, sterilize, transport, and store Stryker implantable devices and surgical instruments within a {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center, and the words "PARAGON MEDICAL" in black text to the right of the pentagon. There are two black lines above and below the text. healthcare facility when used in conjunction with a validated, FDA cleared rigid sterilization container in order to maintain sterility of the enclosed devices. The Stryker ChestShield Sterilization Tray System has been validated for use in the following sterilization cycle: | | Pre-vacuum Steam | |--------------------|---------------------| | Enclosure | Rigid Container | | Temperature | 132°C (270°F) | | Sterilization Time | 4 minutes | | Minimum Dry Time | 30 minutes | | Maximum Weight1 | 19.23 lb. / 8.72 kg | ¹The validated load configuration for the Stryker ChestShield Sterilization Tray System included single-use implants (plates, bone screws, etc.), single-use instruments (drills), reusable surgical instruments (benders, etc.) and lumened instruments. ## 10. Technological Characteristics: Shown below is a comparison of the technological (Table 1) and performance (Table 2) characteristics of the subject and predicate device. | Table 1: Technological Characteristics | | | | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | | Predicate Device | Subject Device | Comparison | | Device Name | Stryker Universal Select Sterilization Tray System | Stryker ChestShield Sterilization Tray System | -- | | 510(k) Number | K173615 | K183701 | -- | | Product Code | KCT | KCT | Same | | | The Stryker Universal Select Sterilization Tray System<br>is intended to organize, enclose, sterilize, transport<br>and store Stryker implantable devices and surgical<br>instruments within a healthcare facility when used in<br>conjunction with a validated, FDA cleared rigid<br>sterilization container in order to maintain sterility of<br>the enclosed devices.<br>The Stryker Universal Select Sterilization Tray System<br>has been validated for use in the following<br>sterilization cycles: | The Stryker ChestShield Sterilization Tray System is<br>intended to organize, enclose, sterilize, transport, and<br>store Stryker implantable devices and surgical<br>instruments within a healthcare facility when used in<br>conjunction with a validated, FDA cleared rigid<br>sterilization container in order to maintain sterility of<br>the enclosed devices.<br>The Stryker ChestShield Sterilization Tray System has<br>been validated for use in the following sterilization<br>cycle: | Similar<br>Differences include<br>maximum weight<br>and validated load<br>contents | | Indications for<br>Use | Pre-vacuum Steam Cycles | | | | | Enclosure | Rigid Container | Rigid Container | | | Temperature | 132°C<br>(270°F) | 135°C<br>(275°F) | | | Sterilization Time | 4 min | 3 min | | | Minimum Dry Time | 30 min | 30 min | | | Maximum Weight1 | 25 lb./<br>11.36kg | 25 lb./<br>11.36kg | | | 1Contents in the validated Tray System included: Single-use<br>implants (plates, meshes, bone screws, etc.), single-use<br>instruments (drills), and reusable surgical instruments<br>(benders, forceps, handles, depth gauges, trocars, etc.) | | | | | | | | | | | Pre-Vacuum Steam Cycle | | | | Enclosure | Rigid Container | | | | Temperature | 132°C (270°F) | | | | Sterilization Time | 4 min | | | | Minimum Dry Time | 30 min | | | | Maximum Weight1 | 19.23 lb. / 8.72 kg | | | | 1The validated load configuration for the Stryker ChestShield<br>Sterilization Tray System included single-use implants (plates,<br>bone screws, etc.), single-use instruments (drills), reusable<br>surgical instruments (benders, forceps, drivers, etc.) and<br>lumened instruments. | | | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center. To the right of the pentagon are the words "PARAGON MEDICAL" stacked on top of each other, with a line above and below the words. | Table 1 (continued): Technological Characteristics | Predicate Device | Subject Device | Comparison | | | | | | | |----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------|-----------------------|-------------------|-------------------------------------------------------------|----------------------| | Intended Use | The Stryker Universal Select Sterilization Tray System<br>is intended to organize, enclose, sterilize, transport, and store medical devices between surgical uses in a healthcare facility. | The Stryker ChestShield Sterilization Tray System is intended to organize, enclose, sterilize, transport, and store medical devices between surgical uses in a healthcare facility. | Same | | | | | | | | Design | The Stryker Universal Select Sterilization Tray System platform is designed to be compatible with Stryker implantable devices and surgical instrumentation. Individual components were designed to slide in to two and three level storage racks. Racks have handles on the side to facilitate carrying. Module and tray lids have sliding latch mechanism to facilitate containment. | The design of the Stryker ChestShield Sterilization Tray System is the same as the predicate device, with specific instrument brackets and implant module layout for the intended Stryker surgical instrumentation and implantable devices. The individual tray components slide into the existing two-level storage rack part of the predicate device system. | Similar<br>Differences include specific bracket, implant module and inlay layouts | | | | | | | | Principles of Operation | Operates on the principles of allowing steam penetration through perforations and open area designed into the components to achieve sterilization of the contained devices. | Operates on the principles of allowing steam penetration through perforations and open area designed into the components to achieve sterilization of the contained devices. | Same | | | | | | | | Materials of Construction | Stainless Steel, Anodized Aluminum, USP Class VI Silicone, Thermoplastic Polymers | Same as predicate | Same | | | | | | | | Sterilization Parameters | Cycle<br>Pre-Vacuum<br><br><br> | Temp<br>132° C (270°F)<br>135°C (275°F) | Exposure<br>4 Min<br>3 Min | Min Dry Time<br>30 Min<br>30 Min | Cycle<br>Pre-Vacuum | Temp<br>132°C (270°F) | Exposure<br>4 Min | Min Dry Time<br>30 Min | Similar | | Perforations | Evenly distributed perforated steam hole pattern. | | | Evenly distributed perforated steam hole pattern. | | | Same | | | | Vent to Volume ratio | The predicate Stryker Universal Select Sterilization Tray System components have the following worst-case vent to volume ratios: | | | The subject Stryker ChestShield Sterilization Tray System components have the following worst-case vent to volume ratios: | | | Similar | | | | | Paragon Part Number | Stryker Part Number | Description | Vent to Volume Ratio | Paragon Part Number | Stryker Part Number | | Description | Vent to Volume Ratio | | | 3009-1013 | 29-13904 | Mandible Fracture Instrument Tray with Lid (3349-0006/29-13921) | 0.321 | 3009-1040 | 4700002 | | SternalPlate Instrument Tray 2 with Lid (3349-0009/4700101) | 0.330 | | | 2209-0109 | 29-23900 | 2.0/2.3 Mandible Screws Module with Lid (3349-0005/29-13941) | 0.054 | 2209-0115 | 4700201 | | SternalPlate Implant Module with Lid (3349-0008/4700102) | 0.082 | | | 2219-0036 | 29-17903 | 1.7 Orthognathic Plates Inlay | 0.265 | 2219-0002 | 4700202 | | SternalPlate Plates Inlay | 0.288 | | | 3009-1008 | 29-13900 | 2-Level Rack | 0.297 | | | | | | | Sterilization Method | Pre-vacuum Steam | | | Pre-vacuum Steam | | | Same | | | | Reusable | Yes | | | Yes | | | Same | | | | Patient Contact | No direct patient contact | | | No direct patient contact | | | Same | | | | Air Permeance | Yes | | | Yes | | | Same | | | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center, and the words "PARAGON MEDICAL" in black text to the right of the pentagon. There are two black lines above and below the text. | Configuration<br>and<br>Dimensions of<br>Device Models | | | | | | |--------------------------------------------------------|---------------------|-----------------------------------|--------------------------------------------|-----------------------------------------------------------------|--| | Two-level Rack | 9.80 x 9.70 x 5.27" | Two-level Rack¹ | 9.80 x 9.70 x 5.27" | Similar | | | Three level Rack | 9.80 x 9.70 x 7.57" | | | | | | ¼ DIN Inst/Accessory Tray | 9.31 x 4.09 x 1.54" | ¼ DIN Instrument Tray | 9.31 x 4.09 x 1.54" | Differences include<br>slight dimensional<br>differences in the | | | ¼ DIN Inst/Accessory Tray Lid | 9.29 x 4.30 x 0.49" | ¼ DIN Instrument Tray Lid | 9.21 x 4.30 x 0.49" | | | | ½ DIN Inst/Accessory Tray | 9.31 x 8.55 x 1.54" | | | inlay of the subject<br>device to | | | ½ DIN Inst/Accessory Tray Lid | 9.29 x 8.71 x 0.55" | | | | | | ¼ DIN Implant Module | 9.31 x 4.25 x 1.05" | | | accommodate the | | | ¼ DIN Screw Module | 9.31 x 4.25 x 1.75" | ¼ DIN Implant/Screw Module | 9.31 x 4.25 x 1.75" | | | | ¼ DIN Implant Module Lid | 9.33 x 4.22 x 0.44" | ¼ DIN Implant Module Lid<br>Inlay | 9.33 x 4.22 x 0.44"<br>4.09 x 3.31 x 0.63" | intended implants. | | | Inlay | 4.34 x 3.85 x 0.53" | | | | | | Inlay, large | 8.91 x 3.85 x 0.34" | | | | | | Drill Caddy | 3.30 x 1.33 x 1.47" | | | | | | ¼ DIN Silicone Mat | 8.95 x 3.67 x 0.63" | ¹Two-level Rack is identical to | | | | | ½ DIN Silicone Mat | 8.95 x 8.14 x 0.63" | the predicate. | | | | | Table 2: Performance Characteristics | | | | |------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | | Predicate Device | Subject Device | Comparison | | Material<br>Compatibility<br>with Sterilization<br>Process | Predicate device is compatible with steam<br>sterilization. | Subject device is compatible with steam sterilization. | Same | | Biocompatibility | Cytotoxicity (ISO 10993-5) testing results<br>demonstrated that the predicate device is non-<br>cytotoxic | Subject device is identical in formulation and<br>processing to the predicate device; therefore,<br>cytotoxicity results of the predicate are applicable to<br>the subject device. | Same | | Performance<br>Testing | The following non-clinical testing was performed<br>on the Stryker Universal Select Sterilization Tray<br>System:<br>• Sterilization Validation<br>• Dry Time Validation<br>• Design Verification<br>• Life Cycle Testing<br>• Cleaning Validation<br>• Biocompatibility | The following non-clinical testing was performed on<br>the Stryker ChestShield Sterilization Tray System:<br>• Sterilization Validation<br>• Dry Time Validation<br>The following non-clinical testing was leveraged from<br>the predicate to support the performance of the<br>subject device:<br>• Design Verification<br>• Life Cycle Testing<br>• Cleaning Validation<br>• Biocompatibility | Same | # 11. Non-Clinical Performance Testing: The following non-clinical testing was performed on the Stryker ChestShield Sterilization Tray System: ### . Sterilization and Dry Time Validation Sterilization validations were performed to verify the effectiveness of steam sterilization of a fully loaded Stryker ChestShield Sterilization Tray System sterilized in a rigid sterilization container using an autoclave cycle with pre-vacuum air removal. The study evaluated the resistance of biological indicators (Bls) with 10° Geobacillus stearothermophilus spores to 132°C pre-vacuum steam autoclave half cycle exposures. Dry time was evaluated after sterilization. The Styker ChestShield Sterilization Tray System was validated per ANSI/AAMI/ISO 17665-1:2006. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for Paragon Medical. The logo consists of a red pentagon with a white star in the center. To the right of the pentagon are the words "PARAGON MEDICAL" stacked on top of each other, with a black line above and below the words. The subject device is identical in formulation and processing to the predicate device; therefore, the following non-clinical performance testing performed on the predicate Stryker Universal Select Sterilization Tray System is applicable to the subject Stryker ChestShield Sterilization Tray System: ### Lifecycle Testing . Lifecycle testing was performed on the predicate device to verify that the tray system maintained functional quality requirements, material compatibility, and traceability after exposure to repeated pre-vacuum steam sterilization cycles and automated washing cycles and simulated functional use of the components. ### . Biocompatibility (ISO 10993-5) Cytotoxicity testing was performed on the predicate device per ISO 10993-5 using the MEM Elution method. Test article extracts met the criteria of grade 2 or less cell lysis and reactivity, demonstrating the device is non-cytotoxic. The following non-clinical testing was leveraged from the predicate device to support the performance of the subject device: ### . Design Verification Design verification was performed on the predicate device to demonstrate the device complies with AAMI/ANSI ST77:2013 for handle strength testing. Additional design verification included containment verification during transport and simulated use, stacking verification and edge sharpness testing. ### . Cleaning Validation Manual and automated cleaning validations were performed on the predicate device to validate the cleaning instructions using the worst case configured tray system per AAMI TIR 30:2011. The results met the acceptance criteria of residual protein of less than 6.4 µg/cm² and residual hemoglobin of less than 2.2 µg/cm², indicating that the recommended cleaning methods were effective in removing soil from all designated surfaces of the tray system that might be accessible to the end user. - 12. Conclusion: Based on the intended use, technological characteristics, performance data and nonclinical tests performed, the subject Stryker ChestShield Sterilization Tray System is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K173615.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...