K152969 · Conmed Corporation · KCT · Mar 31, 2016 · General Hospital
Device Facts
Record ID
K152969
Device Name
ENHANCE ALLOGRAFT WEDGE INSTRUMENT TRAY
Applicant
Conmed Corporation
Product Code
KCT · General Hospital
Decision Date
Mar 31, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Enhance Allograft Wedge Instrument on Tray (7903) is a containment device intended for medical device sterilization. The Enhance Allograft Wedge Instrumentation Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses. This tray is intended for use with the following instruments: [List of instruments omitted for brevity]. The Enhance Allograft Wedge Instrumentation Tray (T903) is not intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to mainal instruments. Sterilize the instrument. Sterilize the instrument tray using the following sterilization parameters: [Table of parameters]. Double wrap tray using an FDA cleared wrap.
Device Story
Containment device for surgical instruments; aluminum tray with perforations; facilitates steam sterilization; organizes, transports, and stores instruments. Used in clinical settings; operated by sterile processing staff or surgical technicians. Input: surgical instruments loaded into tray; Output: sterilized instrument set. Process: tray loaded with instruments, double-wrapped in FDA-cleared sterilization wrap, subjected to validated steam sterilization cycles (pre-vacuum or gravity). Output allows healthcare providers to maintain sterile instruments for surgery; ensures organized, efficient surgical workflows; benefits patient by providing sterile, ready-to-use instrumentation.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included performance testing per ANSI/AAMI ST77:2013, handle strength, cleaning, biocompatibility, packaging, transportation, and non-clinical user validation.
Technological Characteristics
Materials: 5052-H32 anodized aluminum (tray/lid/base), 301/302/304/316/17-4 PH stainless steel (hardware), silicone (brackets). Perforated design for steam penetration. Dimensions: 21.29" x 10.03" x 3.30". Sterilization: Steam (pre-vacuum/gravity) with FDA-cleared wrap. No software/electronics.
Indications for Use
Indicated for use as a containment device for the organization, sterilization, transport, and storage of specific surgical instruments (Evans Trials, Cotton Trials, Mallet, Distractor, Osteotomes) between uses. No specific patient population or contraindications identified.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2016
ConMed Corporation Ms. Dionne Sanders, MS. COA. RAC Manager, Regulatory Affairs 11311 Concept Blvd. Largo, Florida 33773
Re: K152969
Trade/Device Name: Enhance Allograft Wedge Instrument Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: March 02, 2016 Received: March 03, 2016
Dear Ms. Sanders:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152969
### Device Name
Enhance Allograft Wedge Instrument Tray
### Indications for Use (Describe)
The Enhance Allograft Wedge Instrument on Tray (7903) is a containment device intended for medical device sterilization. The Enhance Allograft Wedge Instrumentation Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses. This tray is intended for use with the following instruments:
| REF | Description | REF | Description |
|--------|------------------|--------|------------------|
| 903106 | Evans Trial 6mm | 903206 | Cotton Trial 6mm |
| 903108 | Evans Trial 8mm | 903207 | Cotton Trial 7mm |
| 903110 | Evans Trial 10mm | 903208 | Cotton Trial 8mm |
| 903112 | Evans Trial 12mm | 903006 | Mallet |
| 903010 | Tamp Evans | 903008 | Distractor |
| 903012 | Tamp Cotton | 903002 | Osteotome 16mm |
| 903205 | Cotton Trial 5mm | 903004 | Osteotome 10mm |
The Enhance Allograft Wedge Instrumentation Tray (T903) is not intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to mainal instruments. Sterilize the instrument. Sterilize the instrument tray using the following sterilization parameters:
| Method | Cycle | Temperature | Exposure | Dry Cycle | Cool Down | Duration |
|--------------------|------------|---------------|------------|------------|--------------|------------|
| Steam<br>(wrapped) | Pre-vacuum | 270°F (132°C) | 4 minutes | 20 minutes | 118°F (48°C) | 25 minutes |
| Steam<br>(wrapped) | Gravity | 250°F (121°C) | 30 minutes | 15 minutes | 118°F (48°C) | 37 minutes |
Double wrap tray using an FDA cleared wrap.
] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for CONMED Corporation. The logo features a blue square with a white, stylized "C" shape inside. To the right of the square is the word "CONMED" in black, sans-serif font. Below the word "CONMED" is a thin blue line, and below that is the word "CORPORATION" in a smaller, sans-serif font.
# 510(k) SUMMARY
#### I. SUBMITTER
CONMED Corporation 11311 Concept Blvd. Largo, Florida 33773
Phone: 727-399-5564 Fax: 727-399-5264
Contact Person: Dionne Sanders, RAC Date Prepared: March 25, 2016 510k#: K152969
#### II. DEVICE
| Name of Device: | Enhance Allograft Wedge Instrument Tray |
|----------------------|-------------------------------------------|
| Common Name: | Sterilization Tray |
| Classification Name: | Sterilization Wrap (21 CFR Part 880.6850) |
| Regulatory Class: | Class II |
| Product Codes: | KCT |
#### III. PREDICATE DEVICE
| Device Name: | ConMed Linvatec Anodized Aluminum Sterilization Trays |
|---------------|-------------------------------------------------------|
| Company Name: | ConMed Linvatec |
| 510(k) #: | K090560 |
#### IV. DEVICE DESCRIPTION
The Enhance® Allograft Wedge Instrumentation Tray (T903) is constructed primarily of aluminum with perforations to facilitate sterilant penetration, evacuation and drying. The tray dimensions are approximately 21 x 10" and it is designed to fit any standard autoclave. The tray is manufactured from durable, biocompatible materials that are corrosion resistant and compatible with the environment of repeated steam sterilization. Since the tray is perforated, an FDA cleared sterilization wrap must be used to maintain sterility of the contents. Although the tray is reusable it will not be serviced or repaired. Interior structures of the trays have the ability to separately hold individual instruments during the entire duration they are in contact with the tray.
#### V. INTENDED USE / INDICATIONS FOR USE
The Enhance Allograft Wedge Instrumentation Tray (T903) is a containment device intended for medical device sterilization. The Enhance Allograft Wedge
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a blue square with a white, curved shape inside on the left. To the right of the square, the word "CONMED" is written in black, sans-serif font. Below the word "CONMED", the word "CORPORATION" is written in a smaller, sans-serif font, with a blue line above it.
Instrumentation Tray is used for loading surgical instruments in order to conveniently organize, sterilize, transport, and store the instruments between uses.
| REF | Description | REF | Description |
|--------|------------------|--------|------------------|
| 903106 | Evans Trial 6mm | 903206 | Cotton Trial 6mm |
| 903108 | Evans Trial 8mm | 903207 | Cotton Trial 7mm |
| 903110 | Evans Trial 10mm | 903208 | Cotton Trial 8mm |
| 903112 | Evans Trial 12mm | 903006 | Mallet |
| 903010 | Tamp Evans | 903008 | Distractor |
| 903012 | Tamp Cotton | 903002 | Osteotome 16mm |
| 903205 | Cotton Trial 5mm | 903004 | Osteotome 10mm |
This tray is intended for use with the following instruments:
The Enhance Allograft Wedge Instrumentation Tray (T903) is not intended to maintain sterility; it is intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed medical instruments. Sterilize the instrument tray using the following sterilization parameters:
| Method | Cycle | Temperature | Exposure | Dry Cycle | Cool Down | Duration |
|--------------------|----------------|------------------|------------|------------|-----------------|------------|
| Steam<br>(wrapped) | Pre-<br>vacuum | 270°F<br>(132°C) | 4 minutes | 20 minutes | 118°F<br>(48°C) | 25 minutes |
| Steam<br>(wrapped) | Gravity | 250°F<br>(121°C) | 30 minutes | 15 minutes | 118°F<br>(48°C) | 37 minutes |
Validation was conducted using an FDA cleared sterilization wrap with a maximum load 14 lbs. (6.4 kg).
Only single unstacked trays may be used during the sterilization process.
Double wrap tray using an FDA cleared wrap.
### VI. COMPARISION OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
ConMed's Enhance Allograft Wedge Instrument Tray is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the ConMed Linvatec Anodized Aluminum Sterilization Trays, specifically the Hip Arthroscopy Master Tray, and raises no new issues of safety or effectiveness.
The similarities and differences between the predicate and proposed sterilization trays are the following-
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Conmed Corporation. The logo consists of a blue square with a white curved shape inside, followed by the word "CONMED" in black, bold letters. Below the word "CONMED" is the word "CORPORATION" in smaller, thinner letters, with a blue line above it.
| | Proposed Device<br>Enhance Allograft Wedge<br>Instrument Tray | Predicate Device<br>ConMed Linvatec Anodized Aluminum<br>Sterilization Tray (K090560) |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same | A family of containment devices for<br>medical device sterilization. |
| Indications for<br>Use | Same | The trays are intended for use only with<br>the instruments sets listed in the IFU. |
| Contraindications | None Known | None known |
| Materials of<br>Construction | Tray lid, base and internal tray:<br>5052-H32 grade aluminum,<br>anodized<br>Rivets, Handle Wire, Handle<br>Clip, Latch Component, Channel:<br>301SS, 302SS, 304SS, Silicone<br>Latches, Screws, Nuts, Washers<br>and Brackets: 17-4 PH SS, 304-<br>2B SS, 316SS, 18-8 SS<br>Brackets, Internal: Silicone<br>Silk Screen Ink: Sericoil 301 TP<br>Black | Tray lid, base and internal tray: 5052-H32<br>grade aluminum, anodized<br>Handles: 304SS<br>Latches: 304SS<br>Brackets and protective mat: Silicone<br>(21 CFR177.2600) |
| Design | Same | Perforated tray and lid used to organize,<br>sterilize, transport, and store instruments<br>between uses |
| Overall<br>Dimensions | 21.29" x 10.03" x 3.30" | 20.77" x 10.02" x 6.00" |
| Filled Weight | 11.70lbs | 18.20lbs |
| Gaskets, Filters,<br>Valves, Seals | Same | None |
| Latches | Same | Yes, to secure the tray lid to the base |
| Sterilization<br>Method | Same | Steam sterilization using a<br>FDA-cleared sterilization wrap |
| External Stacking | Same | Do not stack |
| | Same w/added cool down time | Min temperature: 270° F (132° C) |
| Pre-vacuum cycle<br>Steam (wrapped) | Cool down: 48° C for 25 minutes | Min exposure: 4 min<br>Min dry cycle: 20 min |
| | Method: Steam (wrapped) | Method: Steam (wrapped) |
| Gravity cycle<br>Steam (wrapped) | Temperature: 250° F (121° C)<br>Exposure: 30 min<br>Dry Cycle: 15 min<br>Cool Down: 48° C for 37 min | Min temperature: 270° F (132° C)<br>Min exposure: 15 min<br>Min dry cycle: 25 min |
| Sterility<br>Assurance Level<br>(SAL) | SAL 10-6 | SAL 10-6 |
| Tamper Evident | Same | Secured with autoclave tape |
| Instrumentation | Same | For use with specific instrumentation |
| Packaging | Shipped empty | Shipped loaded with instrumentation |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside on the left. To the right of the square is the company name, "CONMED," in bold, black letters. Below the name is the word "CORPORATION" in smaller, gray letters.
#### VII. PERFORMANCE DATA
# Non-Clinical Performance
ConMed conducted verification and validation testing including performance testing in accordance with ANSI/AAMI ST77:2013 Containment Devices for Reusable Medical Device Sterilization as well as: handle strength verification, cleaning, biocompatibility, packaging and transportation and a non-clinical user validation. Based on this testing, it has been determined that the subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate device cleared under (K090560).
# Clinical Performance
This submission does not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
#### VIII. CONCLUSION
The Enhance Allograft Wedge Instrument Tray is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the predicate device.
#### IX. SUBSTANTIAL EQUIVALENCE STATEMENT
The performance testing data for the subject device Enhance Allograft Wedge Instrument Tray (T903) demonstrates the subject device is as safe, as effective, and performs as well as the predicate device (K090560).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.