ConMed Linvatec Anodized Aluminum Sterilization Trays is a family of containment devices for medical device sterilization. The Hip Arthroscopy Master Tray is intended for use only with the following instruments and/or other instruments of similar type and size: [List of specific instruments provided in document]. The PowerPro® Small System Sterilization Case (PRO5095) and the Hall® Sterilization Tray (PRO6000) are designed to hold a variety of powered ConMed Linvatec handpieces as well as attachments and/or accessories that may be used with the powered handpieces. Cannulated handpieces or attachments with an internal diameter not less than 0.031 inches (0.079cm) and a cannulated length no greater than 5.75 inches (14.61cm) and/or handpiece accessories such as pneumatic air supply hoses with a lumen not less than 0.187 inches (0.475 cm) in diameter and a cannulated length no greater than 15 feet (4.6 m) may be sterilized in this tray along with other non-cannulated, non-porous instruments.
Device Story
Anodized aluminum sterilization trays; perforated design to facilitate steam penetration, evacuation, and drying. Trays hold specific medical instruments (arthroscopic tools, powered handpieces, cannulated attachments) during steam sterilization cycles. Used in clinical settings; operated by hospital staff. Trays require FDA-cleared sterilization wrap to maintain sterility. Output is a sterilized instrument set ready for surgical use. Benefits include organized, secure containment of instruments during the sterilization process, ensuring effective steam exposure and preventing damage to delicate surgical tools.
Clinical Evidence
Bench testing only. Validation of sterilization parameters (pre-vacuum and gravity steam cycles) performed to ensure effective sterilization and drying. Testing confirmed that a minimum dry time of 35 minutes is required for gravity cycles to prevent wet packs.
Technological Characteristics
Constructed of anodized aluminum; perforated for steam penetration. Compatible with standard autoclaves. Reusable, non-repairable. Dimensions and configurations vary by model. Sterilization via steam (pre-vacuum or gravity).
Indications for Use
Indicated for use as containment devices for the sterilization of reusable medical instruments, including specific arthroscopic instruments and powered handpieces, in healthcare settings using steam sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for CONMED LINVATEC. The logo is black and white and includes the company name in a bold, sans-serif font. Below the name is the text "L I N V A T E C" in a smaller font. The image also includes the address "11311 Concept Boulevard • Largo, FL 33773-4908", the phone number "727-392-6464", and the website "www.linvatec.com".
### 510(k) Summary ConMed Linvatec Anodized Aluminum Sterilization Trays
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Linvatec is hereby submitting a 510(k) Summary of Safety and Effectiveness for 510(k) Number_ 1090560
| Submitter | ConMed Linvatec<br>11311 Concept Blvd.<br>Largo, Florida 33773-4908<br>Establishment Registration Number 1017294 | |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------|-------------|
| | | JUN 29 2009 |
| Company Contact | Rebecca Roberts, RAC<br>Regulatory Affairs Specialist<br>Telephone 727-399-5564<br>Fax 727-399-5264 | |
| Family Name | ConMed Linvatec Anodized Aluminum<br>Sterilization Trays | |
| Device Name | Hip Arthroscopy Master Tray, ST7900<br>PowerPro Small System Sterilization Case, PRO5095<br>Hall Sterilization Tray, PRO6000 | |
| Classification Name | Sterilization Wrap Containers, Cassettes and<br>other Accessories | |
| & Regulation number | 21 CFR 880.6850<br>Infection Control Devices Branch | |
| Proposed Class/Device | Class II | |
| Product Code | KCT | |
| Predicate Device | ConMed Linvatec Microfracture Instrument<br>Sterilization Tray (MFX-TRAY), K080531 | |
# Device Description
ConMed Linvatec Anodized Aluminum Sterilization Trays are constructed of metal with perforations to facilitate sterilant penetration, evacuation and drying, These trays are designed to fit any standard autoclave and are manufactured from durable, biocompatible materials that are corrosion resistant and compatible with the environment of repeated steam sterilization. Since the trays are perforated, an FDA cleared sterilization wrap must be used to maintain sterility of
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Conmed Linvatec. The logo consists of the word "CONMED" in large, bold letters, with the word "LINVATEC" underneath in smaller letters. Below the logo is the address "11311 Concept Boulevard, Largo, FL 33773-4908", followed by the phone number "727-392-6464" and the website "www.linvatec.com".
# 510(k) Summary ConMed Linvatec Anodized Aluminum Sterilization Trays
the contents. Although these trays are reusable they will not be serviced or repaired.
ConMed Linvatec Anodized Aluminum Sterilization Trays are available under various trade names and in various sizes and tray configurations. Interior structures of the trays have the ability to separately hold each individual instrument during the entire duration they are in contact with the trav.
### Intended Use:
ConMed Linvatec Anodized Aluminum Sterilization Trays is a family of containment devices for medical device sterilization.
The Hip Arthroscopy Master Tray is intended for use only with the following instruments and/or other instruments of similar type and size:
| REF | Description | REF | Description |
|---------|---------------------------------------|-------------|---------------------------------|
| S210GSP | Grasping Forceps, 210mm | 25.50014EL | Extended Length Liberator Knife |
| S210SRF | Suture Retrieval Forceps, 210mm | 25.50016EL | Extended Length Liberator Rasp |
| S210MSC | Micro-scissors, 210mm | C6105EL | Extended Length Crochet Hook |
| S210AFU | Aggressor Forcep, 15° Up, 210mm | 8748EL | Extended Length Knot Pusher |
| S210AFL | Aggressor Forcep, 15° Left, 210mm | 9837 | Extended Length Hip Obturator |
| S210AFR | Aggressor Forcep, 15° Right, 210mm | A55-012-022 | Bio Mini Revo® Blunt, Obturator |
| S210RFR | Rotary Forcep Right, 210mm | A55-012-023 | Bio Mini Revo Trocar, Obturator |
| S210RGF | Retrograde Forceps, 210mm | A55-012-113 | Bio Mini Revo Drill Bit |
| S210RFL | Rotary Forcep Left, 210mm | A55-012-130 | Bio Mini Revo Bone Punch, 3mm |
| C6171EL | Extended Length Drill Guide Fishmouth | A55-012-120 | Bio Mini Revo Tap, 3mm |
| C6172EL | Extended Length Drill Guide Serrated | GU1009EL | Extended Length Katana |
| SC7900 | Slotted Cannula, 6.7mm | GU1004EL | Extended Length Raven, Straight |
| SC6900 | Slotted Cannula, 5.0mm | 21.1001EL | XO Button Depth Probe |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Conmed Linvatec. The logo features a stylized "C" shape on the left, followed by the text "CONMED" in bold, sans-serif font. Below the company name is the word "LINVATEC" in a smaller font. The address "11311 Concept Boulevard, Largo, FL 33773-4908" and contact information "727-392-6464" and website "www.linvatec.com" are listed below the company name.
# 510(k) Summary ConMed Linvatec Anodized Aluminum Sterilization Trays
Sterilize the Hip Arthroscopy Master Tray using the following parameters:
| Method | Cycle | Minimum<br>Temperature | Minimum Exposure | Minimum Dry Cycle |
|-----------------|------------|------------------------|------------------|-------------------|
| Steam (wrapped) | Pre-vacuum | 270°F(132°C) | 4 minutes | 20 minutes |
| Steam (wrapped) | Gravity | 270°F(132°C) | 15 minutes | 25 minutes |
The PowerPro® Small System Sterilization Case (PRO5095) and the Hall® Sterilization Tray (PRO6000) are designed to hold a variety of powered ConMed Linvatec handpieces as well as attachments and/or accessories that may be used with the powered handpieces.
Cannulated handpieces or attachments with an internal diameter not less than 0.031 inches (0.079cm) and a cannulated length no greater than 5.75 inches (14.61cm) and/or handpiece accessories such as pneumatic air supply hoses with a lumen not less than 0.187 inches (0.475 cm) in diameter and a cannulated length no greater than 15 feet (4.6 m) may be sterilized in this tray along with other non-cannulated, non-porous instruments.
Sterilize the PowerPro® Small System Sterilization Case (PRO5095) and the Hall® Sterilization Tray (PRO6000) using the following parameters.
| Method | Cycle | Minimum<br>Temperature | Minimum Exposure | Minimum Dry Cycle |
|-----------------|------------|------------------------|------------------|-------------------|
| Steam (wrapped) | Pre-vacuum | 270°F(132°C) | 4 minutes | 25 minutes |
| Steam (wrapped) | Gravity | 270°F(132°C) | 15 minutes | 35 minutes |
CAUTION: Testing demonstrates that a minimum dry time of 35 minutes is required to prevent wet packs when using the gravity cycle.
# Substantial Equivalence
The proposed and the predicate devices are substantially equivalent in design. materials and intended use. The proposed devices and the predicate devices are containment devices for sterilization of reusable medical devices using steam sterilization. The proposed devices present no new issues of safety and efficacy.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and a caduceus, a symbol of medicine, is visible in the lower right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the seal.
#### Public Health Service
JUN. 29 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Robecca Roberts Regulatory Affairs Specialist ConMed Linvatec 11311 Concept Boulevard Largo, Florida 33773-4908
. Re: K090560
Trade/Device Name: Hip Arthroscopy Master Tray (ST7900) PowerPro® Small System Sterilization Case, PRO5095, Hall Sterilization Tray, PRO6000 Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: June 16, 2009 Received: June 17, 2009
Dear Ms. Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. .
{4}------------------------------------------------
### Page 2- Mr. Roberts
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Susan Runner
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation. Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
| 510(k) Number (if known): | K090560/S001 |
|---------------------------|-----------------------------------------------------------------------------------------|
| Family Name: | ConMed Linvatec Anodized Aluminum Sterilization Trays |
| Trade Name: | PowerPro® Small System Sterilization Case, PRO5095<br>Hall® Sterilization Tray, PRO6000 |
### Indications for Use:
ConMed Linvatec Anodized Aluminum Sterilization Trays is a family of containment devices for medical device sterilization.
The PowerPro® Small System Sterilization Case (PRO5095) and the Hall® Sterilization Tray (PRO6000) are designed to hold a variety of powered ConMed Linvatec handpieces as well as attachments and/or accessories that may be used with the powered handpieces.
Cannulated handpieces or attachments with an internal diameter not less than 0.031 inches (0.079cm) and a cannulated length no greater than 5.75 inches (14.61cm) and/or handpiece accessories such as pneumatic air supply hoses with a lumen not less than 0.187 inches (0.475 cm) in diameter and a cannulated length no greater than 15 feet (4.6 m) may be sterilized in this tray along with other non-cannulated, non-porous instruments.
Sterilize the PowerPro® Small System Sterilization Case (PRO5095) and the Hall® Sterilization Tray (PRO6000) using the following parameters.
| Method | Cycle | Minimum<br>Temperature | Minimum Exposure | Minimum Dry Cycle |
|-----------------|------------|------------------------|------------------|-------------------|
| Steam (wrapped) | Pre-vacuum | 270°F(132°C) | 4 minutes | 25 minutes |
| Steam (wrapped) | Gravity | 270°F(132°C) | 15 minutes | 35 minutes |
CAUTION: Testing demonstrates that a minimum dry time of 35 minutes is required to prevent wet packs when using the gravity cycle.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-the-Counter Use × (21 CFR 801 Subpart C)
(PLÉASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE If NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <span style="text-decoration: underline;">Charles H. Murphree, M.D.</span> |
|--|----------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------|
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K091
{6}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): K090560
Family Name:
ConMed Linvatec Anodized Aluminum Sterilization Trays
Trade Name:
Hip Arthroscopy Master Tray (ST7900)
Indications for Use:
ConMed Linvatec Anodized Aluminum Sterilization Trays is a family of containment devices for medical device sterilization.
The Hip Arthroscopy Master Tray is intended for use only with the following instruments:
| REF | Description | REF | Description |
|---------|----------------------------|-------------|-----------------------|
| S210GSP | Grasping Forceps | 25.50014EL | EL Liberator |
| S210SRF | Suture Retrieval | 25.50016EL | EL Rasp |
| S210MSC | Micro-scissors | C6105EL | EL Crochet Hook |
| S210AFU | Aggressor Forcep Up | 8748EL | EL Knot Pusher |
| S210AFL | Aggressor Forcep Left | 9837 | EL Hip Obturator |
| S210AFR | Aggressor Forcep Right | A55-012-022 | BMR Obturator, Blunt |
| S210RFR | Rotary Forcep Right | A55-012-023 | BMR Obturator, Trocar |
| S210RGF | Retrograde Forceps | A55-012-113 | BMR Drill Bit |
| S210RFL | Rotary Forcep Left | A55-012-130 | BMR Bone Punch |
| C6171EL | EL Drill Guide Fishmouth | A55-012-120 | BMR Tap |
| C6172EL | EL Drill Guide Serrated | GU1009EL | EL Katana |
| SC7900 | Slotted Hip Cannula, 6.7mm | GU1004EL | EL Raven, Straight |
| SC6900 | Slotted Hip Cannula, 5.0mm | 21.1001EL | EL Probe |
Shale H. Maugher, D
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K90 562 C
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.