STAT-MAT MINOR

K110500 · Macpherson Medical, Inc. · KCT · Jun 23, 2011 · General Hospital

Device Facts

Record IDK110500
Device NameSTAT-MAT MINOR
ApplicantMacpherson Medical, Inc.
Product CodeKCT · General Hospital
Decision DateJun 23, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

The Stat-Mat® Minor multipurpose surgical instrument tray is used for loading an entire kit of surgical instruments for a specific case in order to conveniently organize, sterilize, transport and store the instruments between uses. The Stat-Mat® Minor is designed using plastic and metal materials that can be reused with prevacuum or gravity displacement steam sterilization methods. The Stat-Mat® Minor should be used only with FDA-cleared and legally marketed sterilization wraps that have been cleared for the sterilization cycles listed below. The Stat-Mat® Minor has been validated with the Dextex II Sterilization Wrap [K800123]. The base containing organized surgical instrumentation can be placed directly on the back table or onto a Mayo Tray. Instruments with lumens were not tested in the Stat-Mat® Minor validation process and are therefore not recommended for use with the Stat-Mat® Minor.

Device Story

Stat-Mat® Minor is a reusable surgical instrument tray system comprising a Radel plastic base and cover, and an aluminum instrument roll for securing ringed instruments. Used in clinical settings to organize, sterilize, transport, and store surgical instrument kits. Operates via standard steam sterilization (prevacuum or gravity displacement) when wrapped in FDA-cleared sterilization wraps. Designed for placement on back tables or Mayo trays. Does not contact patients. Benefits include improved organization and efficient sterilization processing of non-lumened surgical instruments.

Clinical Evidence

Bench testing only. Validation of gravity displacement (121°C) and pre-vacuum (132°C) steam sterilization cycles performed using Geobacillus stearothermophilus spore strips and thermocouples. Testing confirmed sterility assurance level of 10^-6 achieved at twice the exposure time. No clinical data.

Technological Characteristics

Materials: Radel plastic (base/cover), aluminum (instrument roll). Biocompatibility: ISO-10993 compliant. Design: Multipurpose tray with quick-release cover and ringed instrument separator; perforated for steam circulation. Sterilization: Steam (prevacuum/gravity). Connectivity: None.

Indications for Use

Indicated for use as a multipurpose surgical instrument tray to organize, sterilize, transport, and store surgical instrument kits. Not indicated for instruments with lumens.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # REVISED 510(k) SUMMARY JUN 2 3 2011 KII0500 MacPherson Medical's Stat-Mat® Minor # Submitter: MacPherson Medical Inc. 930 Main Street, Suite 202 Acton, MA 01720 Phone: (978) 800-4641 Facsimile: (866) 280-2078 Contact Person: Brad MacPherson Date Prepared: June 21, 2011 ### Device Name: Stat-Mat® Minor # Sponsor: MacPherson Medical Inc. 930 Main Street, Suite 202 Acton, MA 01720 ### Common or Usual Name Sterilization Tray # Classification Name Accessory to sterilization wrap ### Predicate Devices PolyVac's Surgical Instrument Delivery Systems (K012105) Riley Medical, Inc.'s MetaPak Multi-Purpose Instrument Tray (K993535) - --------------------------------------------------------------------------------------------------------------------------------------------------------------------------- {1}------------------------------------------------ ### Device Description The Stat-Mat® Minor is a multipurpose surgical instrument tray that consists of three primary components: a base, cover, and instrument roll. The base and cover are composed of Radel plastic. which meets the requirements for biocompatibility pursuant to ISO-10993 and is compatible with standard steam sterilization methods. The instrument roll is composed of aluminum and is used to keep ringed instruments open during sterilization. The Stat-Mat® does not contact the patient. ### Intended Use / Indications for Use The Stat-Mat® Minor multipurpose surgical instrument tray is used for loading an entire kit of surgical instruments for a specific case in order to conveniently organize, sterilize, transport and store the instruments between uses. The Stat-Mat® Minor is designed using plastic and metal materials that can be reused with prevacuum or gravity displacement steam sterilization methods. The Stat-Mat® Minor should be used only with FDA-cleared and legally marketed sterilization wraps that have been cleared for the sterilization cycles listed below. The Stat-Mat® Minor has been validated with the Dextex II Sterilization Wrap [K800123]. The base containing organized surgical instrumentation can be placed directly on the back table or onto a Mayo Tray. Instruments with lumens were not tested in the Stat-Mat® Minor validation process and are therefore not recommended for use with the Stat-Mat® Minor. | Cycle Type | Temperature | Sterilization<br>Time | Pre-Vac Pulses | Dry Time | |------------|---------------|-----------------------|----------------|------------| | Gravity | 250°F (121°C) | 30 minutes | none | 30 minutes | | Pre-vacuum | 270°F (132°C) | 4 minutes | 4 | 30 minutes | #### Sterilization cycle parameters #### Technological Characteristics The Stat-Mat® Minor includes a quick release cover and ringed instrument separator. The Stat-Mat® Minor can be wrapped in CSR wrap with no sharp edges. It is made of durable materials that are appropriate for holding general surgical instruments and withstanding standard steam sterilization methods. It has ample holes to allow steam to circulate for sterilization. #### Performance Data (Nonclinical Testing) The company conducted testing to validate gravity displacement and pre-vacuum steam sterilization of the Stat-Mat® Minor. 10° Geobacillus stearothermophilus spore strips and thermocouples were placed in the test articles. The test articles were double-wrapped in a cleared sterilization wrap and placed in the sterilization chamber. The test articles were successfully sterilized in a pre-vacuum cycle at 132° for a 2-minute half-cycle and a gravity {2}------------------------------------------------ displacement cycle at 121° for a 15-minute half-cycle. The Stat-Mat® Minor was successfully sterilized for three consecutive and separate sterilization half cycles. The cycle conditions were considered adequate to achieve a sterility assurance level of 10 ° at twice the stated exposure time. The thermocouples passed the verification calibration, and all positive and negative controls were satisfactory. Conclusion: The results of the sterilization testing performed demonstrate that the Stat-Mat® Minor is as safe, as effective, and performs as well as the predicate devices. #### Substantial Equivalence The Stat-Mat® Minor is as safe and effective as Polyvac's Surgical Instrument Delivery Systems or Riley Medical, Inc.'s MetaPak Multi-purpose Instrument Tray. The Stat-Mat® Minor has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Stat-Mat® Minor and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Stat-Mat® Minor is as safe and effective as Polyvac's Surgical Instrument Delivery Systems or Riley Medical, Inc.'s MetaPak Multipurpose Instrument Tray. Thus, the Stat-Mat® Minor is substantially equivalent. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stripes forming its body and wings. The eagle appears to be in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MacPherson Medical, Incorporated C/O Mr. Howard Holstein Hogan Lovells US, LLP Columbia Square 555 13th Street, NW Washington, D.C. 20004 JUN 2 3 2011 Re: K110500 Trade/Device Name: Stat-Mat® Minor Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: June 16, 2011 Received: June 16, 2011 Dear Mr. Holstein: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Holstein Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Susan Punner Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use Statement | | 110500 | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number (if known): | 1 | | | CALLER SELECTRACTED COLLECTION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLEGION COLLECTION CONTRACTOR COLLEGION CONTRACTOR | Device Name: Stat-Mat® Minor Indications for Use: The Stat-Mat® Minor multipurpose surgical instrument tray is used for loading an entire kit of surgical instruments for a specific case in order to conveniently organize, sterilize, transport and store the instruments between uses. The Stat-Mat® Minor is designed using plastic and metal materials that can be reused with prevacuum or gravity displacement steam sterilization methods. The Stat-Mat® Minor should be used only with FDA-cleared and legally marketed sterilization wraps that have been cleared for the sterilization cycles listed below. The Stat-Mat® Minor has been validated with the Dextex II Sterilization Wrap [K800123]. The base containing organized surgical instrumentation can be placed directly on the back table or onto a Mayo Tray. Instruments with lumens were not tested in the Stat-Mat® Minor validation process and are therefore not recommended for use with the Stat-Mat® Minor. Sterilization cycle parameters: | Cycle Type | Temperature | Sterilization<br>Time | Pre-Vac Pulses | Dry Time | |------------|---------------|-----------------------|----------------|------------| | Gravity | 250°F (121°C) | 30 minutes | none | 30 minutes | | Pre-vacuum | 270°F (132°C) | 4 minutes | 4 | 30 minutes | Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH. Office of Dev E. Sindretta F. Clavijo-Will Division of Anesthesiology, General Hospital Infection Control, Dental Devices WDC - 038100/000001 - 3261988 v2 KI10500 510(k) Number:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...