TEVADAPTOR Closed Drug Reconstitution and Transfer System

K170680 · Teva Medical Ltd., Migada Plant · ONB · May 16, 2017 · General Hospital

Device Facts

Record IDK170680
Device NameTEVADAPTOR Closed Drug Reconstitution and Transfer System
ApplicantTeva Medical Ltd., Migada Plant
Product CodeONB · General Hospital
Decision DateMay 16, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

TEVADAPTOR is a Closed System Drug Transfer Device (CSTD) that mechanically prohibits the release of the drug in vapor, aerosol or liquid form during preparation and administration, and prevents the introduction of microbial and airborne contaminants into the drug or fluid path, allowing the system to minimize exposure of individuals, healthcare personnel, and the environment to hazardous drugs.

Device Story

System of components for reconstitution of liquid/pre-dissolved powder drugs into infusion bags, bottles, or syringes; uses elastomeric seals and active carbon filter to prevent drug release; hydrophobic acrylic copolymer membrane (0.2 micron) for pressure equalization; Syringe Adaptor Lock connects to standard Luer lock syringes; features two-component distal hub design to prevent syringe disconnection; stainless steel needle shielded during use/disposal; used in clinical settings by healthcare personnel; protects users/environment from hazardous drug exposure; bench testing confirms flow, leakage, and mechanical integrity.

Clinical Evidence

Bench testing only. Testing included bidirectional flow, Luer-lock disconnection, Luer fitting compliance (ISO 594/1 & 2), liquid/air leakage, and visual inspection for stress cracking. No clinical data provided.

Technological Characteristics

Components include elastomeric seals, active carbon filter, and 0.2 micron hydrophobic acrylic copolymer membrane. Syringe Adaptor Lock features stainless steel needle and two-component distal hub (SA Luer-lock and SA Hub-lock). Sterilized by Ethylene Oxide (SAL 10^-6). Mechanical connection via Luer lock. No software or electronics.

Indications for Use

Indicated for use as a Closed System Drug Transfer Device (CSTD) to prevent hazardous drug exposure and microbial/airborne contamination during drug preparation and administration for individuals and healthcare personnel.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 16, 2017 Teva Medical Ltd., Mitgada Plant c/o Mr. Roger Gray Donawa Lifescience Consulting Piazza Albania 10 Rome 00153 ITALY Re: K170680 Trade/Device Name: TEVADAPTOR® Closed Drug Reconstitution and Transfer System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: ONB Dated: April 12, 2017 Received: April 17, 2017 Dear Mr. Roger Gray: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Tina Kiang Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K170680 Device Name TEVADAPTOR Closed Drug Reconstitution and Transfer System Indications for Use (Describe) TEVADAPTOR is a Closed System Drug Transfer Device (CSTD) that mechanically prohibits the release of the drug in vapor, aerosol or liquid form during preparation and administration, and prevents the introduction of microbial and airborne contaminants into the drug or fluid path, allowing the system to minimize exposure of individuals, healthcare personnel, and the environment to hazardous drugs. | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Type of Use (Select one or both, as applicable) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is divided into two sections. The left section is a light gray color and contains the word "TEVA" in a stylized font. The right section is green and contains the word "Medical" in a cursive font, with the words "Teva Medical Ltd." below it in a smaller, sans-serif font. The image appears to be a logo for a medical company. # K170680 # 510(k) Summary in accordance with 21 CFR 807.92(c) | Device Name: | TEVADAPTOR® Closed Drug Reconstitution and Transfer System | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Type of 510(k) submission: | Special | | Date of Submission: | May 12, 2017 | | Manufacturer: | Teva Medical Ltd., MIGADA Plant<br>North Industrial Zone<br>Kiryat Shmona 10258<br>ISRAEL | | FDA Registration Number: | 9611423 | | Owner/Operator Number: | 9001925 | | 510(k) Owner & Submitter: | Teva Medical Ltd., MIGADA Plant<br>North Industrial Zone<br>Kiryat Shmona 10258<br>ISRAEL | | Phone: | +972-4-6908826 | | Fax: | +972-9-892-1659 | | 510(k) Application Correspondent: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy | | | Phone: +39 06 578 2665<br>Fax: +30 06 574 3786<br>Email: rgray@donawa.com | | FDA Product Code: | ONB | | FDA Regulation Number: | 880.5440 | | FDA Regulation Name: | Set, Administration, Intravascular | | Classification Panel: | General Hospital | | Common Name: | Closed Drug Reconstitution and Transfer System | | FDA Classification: | Class II | | Indications for Use: | TEVADAPTOR® is a Closed System Drug Transfer Device (CSTD) that mechanically<br>prohibits the release of the drug in vapor, aerosol or liquid form, during<br>preparation and administration, and prevents the introduction of microbial and<br>airborne contaminants into the drug or fluid path, allowing the system to<br>minimize exposure of individuals, healthcare personnel, and the environment to<br>hazardous drugs. | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains two logos side by side. The first logo on the left has the word "TEVA" in a stylized font with a teal color. The second logo on the right has the word "Medical" in a white serif font on a green background, with the text "Teva Medical Ltd." below it in a smaller font. #### Predicate Device: The predicate device selected for comparison with the revised TEVADAPTOR® Closed Drug reconstitution and Transfer System is: Predicate Device Name: TEVADAPTOR® Closed Drug Reconstitution and Transfer System | 510(k) Sponsor: | Teva Medical | |--------------------|----------------------------------| | 510(k) Number: | K141448 | | Clearance Date: | 23 January 2015 | | FDA Product Code: | ONB | | Regulation Name: | Intravascular administration set | | Regulation Number: | 880.5440 | #### Reference Device: | | Reference Device Name: TEVADAPTOR® Closed Drug Reconstitution and Transfer System | |--------------------|-----------------------------------------------------------------------------------| | 510(k) Sponsor: | Teva Medical | | 510(k) Number: | K071741 | | Clearance Date: | 24 September 2007 | | FDA Product Code: | LHI | | Regulation Name: | Intravascular administration set | | Regulation Number: | 880.5440 | #### Device Description: The TEVADAPTOR® Closed Drug Reconstitution and Transfer System is a system of components that allows the reconstitution of liquid or pre-dissolved powder drugs into infusion bags, flexible bottles or syringes. Single, partial or multiple vials can be used for each infusion solution container. The TEVADAPTOR® Closed Drug Reconstitution and Transfer System prevents contamination of the user or the environment by the drug through the use of elastomeric seals and an active carbon filter. Sterility of the drug in the vial is maintained because any air entering the vial during pressure equalization enters through of a hydrophobic acrylic copolymer membrane with a pore size of 0.2 micron. The purpose of this Special 510(k) is to add an additional design of one of the components, the TEVADAPTOR® Syringe Adaptor Lock. The Syringe Adaptor (SA) Lock is one component of the system which is intended for connection to a standard Luer lock syringe. A non-latex elastomer protector covers the liquid dispensing needle tip. A clamp mechanism reversibly connects the Syringe Adaptor Lock to ther components of the TEVADAPTOR® system. The TEVADAPTOR® Syringe Adaptor Lock includes a stainless steel needle that is shielded at all times during preparation, use and disposal. The Syringe Adaptor Lock is supplied with a сар. Disconnection of a syringe is not possible with the Syringe Adaptor Lock, because of a design change from the syringe adaptor that introduces a two-component distal hub. One of these components attaches, and rotates within the second component, together with the syringe, if disconnection is attempted. The components of the TEVADAPTOR® system available separately are: - Vial Adaptor 20 mm with 13 mm Vial Converter - Vial Adaptor 28 mm ● {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the Teva Medical logo. The logo consists of the word "TEVA" in a stylized font on the left side of the image. To the right of the word "TEVA" is the word "Medical" in a serif font, followed by the words "Teva Medical Ltd." in a smaller font. - Syringe Adaptor - Syringe Adaptor Lock - Spike Port Adaptor - Connecting Set - Luer Lock Adaptor The environment of use is unchanged from that of the predicate TEVADAPTOR® system cleared under K141448. ### Device Performance Testing: Bench testing has been carried out on the Syringe Adaptor Lock to demonstrate equivalence with the Syringe Adaptor, including: - Bidirectional flow - Needle to Syringe Adapter Luer-lock disconnection test - Breakage of SA Lock Luer retention teeth - Disconnection of male luer and SA Lock - Luer fitting compliance with ISO 594/1 & 2, including: - o Gauging test - O Resistance for overriding - Unscrewing torque O - O Ease of assembly - O Liquid leakage - O Air leakage - O Separation force of conical fitting assembly - O Stress cracking - Visual inspection for absence of cracks ● The packaging materials and process for the Syringe Adaptor Lock are unchanged from those utilized for the Syringe Adaptor and other TEVADAPTOR® components cleared under K141448. The subject devices are supplied sterile, with an SAL of 10 °. The devices are sterilized by Ethylene Oxide, with the same parameters as the remainder of the TEVADAPTOR® components, as described in K141448. No additional biocompatibility tests have been carried out on the components of the Syringe Adaptor Lock, because all materials within the fluid path of the item are unchanged from the Syringe Adaptor cleared under K141448 or its previous design, cleared under K071741, included within this submission as a reference device. The material of the hub component of the syringe adaptor lock is identical to the material of the hub component of the syringe adaptor cleared under K071741. A risk assessment of the changes in design introduced by the Syringe Adaptor Lock concluded that any new risks identified were adequately mitigated and verified by means of bench testing. ### Comparison with Predicate Device: The major difference between the Syringe Adaptor and Syringe Adaptor Lock is in the 'Luer Hub' component, which has been replaced by two components, the 'SA Luer-lock' and the 'SA Hub-lock'. The SA Hub-lock and End Cap for the Syringe Adaptor Lock have new colorants which are not used in the Syringe Adaptor components, for easy user-recognition purposes, however neither of these new components of the Syringe Adaptor Lock are in the fluid path, being external only components. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains two logos, one for TEVA and one for Teva Medical Ltd. The TEVA logo is on the left side of the image and is in a light blue color. The Teva Medical Ltd. logo is on the right side of the image and is in a green color. The text "Medical" is in a larger font than the text "Teva Medical Ltd." The separation of the 'Luer Hub' into two components has allowed the design to permanently attach a Luer syringe to the Syringe Adaptor Lock, whereas a Luer syringe can be disconnected from the predicate Syringe Adaptor. When attaching a Syringe to the Syringe Adaptor Lock, the SA Luer-lock will rotate with the Syringe until two teeth on the SA Luer-lock component engage with two of the four cam upstands on the inner circumference of the SA Hub-lock component. Once the SA Luer-lock teeth have engaged with the SA Hub-lock, no further rotation of the SA Luer-lock is possible. If disconnection of the Syringe Adaptor Lock is attempted, the SA Luer-lock will rotate with the Syringe. When a quarter rotation is reached, by means of a cam action, the SA Luer-lock teeth will slip over the upstands on the SA Hub-lock inner surface, and will thus continue to rotate with the Syringe, preventing disconnection of the Syringe from the Syringe Adaptor Lock. The only other difference is that 'Main Body' component of the Syringe Adaptor has been shortened by 5.2 mm to accommodate the above new components within the Syringe Adaptor Lock. All other components of the TEVADAPTOR® system remain unchanged from those cleared under K141448. #### Substantial Equivalence Conclusion: Based on the performance testing conducted on the subject device, the Syringe Adaptor Lock, does not raise new types of safety and effectiveness questions. It is concluded that the TEVADAPTOR® device that has been modified by the introduction of the Syringe Adaptor Lock is substantially equivalent to the identified predicate device cleared under K141448, which is already in interstate commerce within the USA.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...