TEVADAPTOR CLOSED DRUG RECONSTITUTION AND TRANSFER SYSTEM
Applicant
Migada Plant
Product Code
LHI · General Hospital
Decision Date
Sep 1, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The TEVADAPTOR™, Drug Reconstitution and Transfer System is a contained system to be used by pharmacists or other healthcare professionals to prepare drugs, including cytotoxic drugs, for intravenous infusion or injection.
Device Story
TEVADAPTOR is a closed system for drug reconstitution and transfer; used by pharmacists/healthcare professionals in clinical settings. Components include Vial Adaptor (with 13mm/20mm cap compatibility), Syringe Adaptor, Infusion Bag Adaptor, and L-Connector. Vial Adaptor features automatic sterile pressure equalization to prevent drug particle escape. Syringe Adaptor connects to standard luer lock syringes with non-latex elastomer needle protection. Infusion Bag Adaptor connects to infusion bag spike ports. System enables safe withdrawal of diluents and introduction of drugs. Benefits include containment of hazardous substances, such as cytotoxic drugs, during preparation and transfer, protecting the user and environment from exposure.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Components: Vial Adaptor, Syringe Adaptor, Infusion Bag Adaptor, L-Connector. Materials: Non-latex elastomers. Principle: Closed system with automatic sterile pressure equalization. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for use by pharmacists or healthcare professionals to prepare drugs, including cytotoxic agents, for intravenous infusion or injection.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
PhaSeal® closed system for the preparation and administration of parenteral drugs (K980381)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of four abstract human profiles facing to the right, with flowing lines suggesting movement or progress. The overall design is simple and conveys a sense of health, well-being, and human connection.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 2 2005
Ms. Yossi Shachar QA/RA Manager Migada Plant P.O. Box 888 Kiryat Shmona 10258 ISRAEL
Re: K051669
Trade/Device Name: TEVADAPTOR™, Drug Reconstitution and Transfer System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: June 14, 2005 Received: June 22, 2005
.Dear Ms. Shachar:
This letter corrects our substantially equivalent letter of September 1, 2005.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Shachar
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sylite y. Michie Davis.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K051669
page 1 of 3
Image /page/2/Picture/2 description: The image contains three lines of text. The first line contains the word "TEVA" in a stylized, blocky font, followed by the word "Medical" in a cursive font, and the abbreviation "ltd.". The second line contains the word "nigada" in a stylized, cursive font. Below the word "nigada" is a small, circular graphic.
Image /page/2/Picture/3 description: The image shows a logo with the text "CERTIFIED QMS ISO 13485: 2003". The logo is stamped with the date "SEP - 1 2005" at the top. The text "THE STANDARDS INSTITUTION OF ISRAEL" is at the bottom of the image. The logo appears to be a certification mark related to quality management systems.
מדיקל
דע שיות איוד
# 510(k) Summary: TEVADAPTOR", Drug Reconstitution and Transfer System
Company Name: Migada Plant
Contact Person:
Yosi Shahar QA/RA manager
Telephone: +972-4-690-8826 +972-9-892-1665 Fax: E-mail: yosi.shahar(@teva.co.il
#### US Agent:
George Svokos MIGADA, INC. 482 Hudson Terrace Englewood Cliffs, NJ 07632
Phone: 201-561-1010 Fax: 201-567-7994
Date prepared: May 10, 2005
Trade Name: TEVADAPTOR"M. Drug Reconstitution and Transfer System
Classification name: Set. IV Fluid Transfer
9-2
a MEMBER OF I GROUP
מְקֻבֶּצֶת
אזור התעשיה הצפוני ת.ד 888 קרית שמונה 10258 טלפון: 04-6908800 פקס: 09-8921665 NORTH IND. ZONE P.O.BOX 888 KIRYAT SHMONA 10258 ISBAEL TEL: (972)-4-6908800 FAX: (972)-9-8921665
{3}------------------------------------------------
K051669 page 2 of 3
Image /page/3/Picture/1 description: The image shows the logo for Teva Medical Ltd. The word "Teva" is in a bold, blocky font, and the words "Medical Ltd." are in a cursive font. Below that is the word "nigada" in a thin, stylized font with a small circle below the "g".
Image /page/3/Picture/2 description: The image shows a logo with the text "CERTIFIED QMS" in a circular arrangement above "ISO 13485: 2003". The logo is from "THE STANDARDS INSTITUTION OF ISRAEL". The logo is black and white and has a triangular shape.
עצים מודייקל
מיגדה תעשיות ציוד רפואי
Common/usual name: Drug reconstitution system
Product Code: LHI
Regulation No.: 880.5440
Class: II
Panel identification: General Hospital Panel
#### Predicate Device:
PhaSeal® closed system for the preparation and administration of parenteral drugs, Carmel Pharma AB. S-402 28 Gottingen, Sweden, cleared under 510(k) no. K980381.
### Description of the device:
The device comprises of the following components:
- Vial Adaptor System with 13 mm Vial converter -
- Syringe Adaptor System -
- Infusion Bag Adaptor -
- -I Connector
The Vial Adaptor System is intended to fit over vials with 20mm caps. A converter accessory enables the use of the Vial Adaptor System with a 13 mm cap vial. A special venting mechanism ensures automatic. sterile, pressure equalization within the vial, preventing drug particle escape into the environment. An elastomeric, non-latex septum is designed for connection of the Syringe Adaptor System for introduction or withdrawal of liquid.
9-3
A MEMBER OF = / I GROUP
מִקְבּוּצַת
אזור התעשיה הצפוני ת.ד 888 קרית שמונה 10258 טלפון: 04-6908800 פקם: 09-8921665 NORTH IND. ZONE P.O.BOX 888 KIRYAT SHMONA 10258 ISRAEL TEL: (972)-4-6908800 FAX: (972)-9-8921665
{4}------------------------------------------------
K051669 page 30.
Image /page/4/Picture/1 description: The image shows the logos for Teva Medical Ltd. and nigada. The Teva Medical Ltd. logo is on the top and consists of the word "TEVA" in a blocky font and the words "Medical Ltd." in a cursive font. The nigada logo is on the bottom and consists of the word "nigada" in a stylized font with a small circle below the "g".
Image /page/4/Picture/2 description: The image shows a logo for the Standards Institution of Israel. The logo is a triangle shape with the words "CERTIFIED QMS" in a circular pattern at the top. Inside the circle is the text "ISO 13485: 2003". Below the triangle is the text "THE STANDARDS INSTITUTION OF ISRAEL".
ת"ש מדוייקל
מיגדה תעשיות ציוד רפואי
The Syringe Adaptor System is intended for connection to a standard luer lock svringe. A special non-latex elastomer protector covers the liquid dispensing needle tip. A clamp mechanism reversibly connects the Syringe Adaptor System to the other components.
The Infusion Bag Adaptor is intended for connection to the spike port of an infusion bag. A non- latex septum enables connection to the Syringe Adaptor System for withdrawal of diluents or introduction of drug. A capped short tubing enables connection of the drug containing bag to a delivery set in the hospital ward.
An injection accessory enables the connection of the Syringe Adaptor System to an intravenous infusion line for direct drug injection.
#### Indications for Use:
The TEVADAPTOR™, Drug Reconstitution and Transfer System is a contained system to be used by pharmacists or other healthcare professionals to prepare drugs, including cytotoxic drugs, for intravenous infusion or injection.
## Substantial Equivalence:
The TEVADAPTORTM, Drug Reconstitution and Transfer System has the same intended use as the PhaSeal® closed system for the preparation and administration of parenteral drugs, cleared under 510(k) no. K980381 and has equivalent performance characteristics. It is therefore substantially equivalent to that device.
#### Conclusion -
The evaluation of the TEVADAPTORTM, Drug Reconstitution and Transfer System does not raise any additional concerns regarding safety and effectivity and may therefore be considered substantially equivalent to the predicate devices.
{5}------------------------------------------------
K051669
# Indications for Use
510(k) Number (if known): K051669
Device Name: TEVADAPTOR™, Drug Reconstitution and Transfer System
Indications for Use:
The TEVADAPTOR™, Drug Reconstitution and Transfer System is a contained system to be used by pharmacists or other healthcare professionals to prepare drugs, including cytotoxic drugs, for intravenous infusion or injection.
Prescription Use X (Part 21 CFR 801 Subpart D)
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Antone Vime
(Division Sign-Off) (Division Sign of Anesthesiology, General Hospital, Infection Control, Denta! Devices
Page 1 of 1
510(k) Number: K931669
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.