COOK VASCULAR INC. VITAL-PORT VASCULAR ACCESS SYSTEM POWER INJECTABLE PORT
K081425 · Cook Vascular, Inc. · LJT · Sep 30, 2008 · General Hospital
Device Facts
Record ID
K081425
Device Name
COOK VASCULAR INC. VITAL-PORT VASCULAR ACCESS SYSTEM POWER INJECTABLE PORT
Applicant
Cook Vascular, Inc.
Product Code
LJT · General Hospital
Decision Date
Sep 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The devices are for use in patient therapy requiring repeated vascular access for infusion therapy, power injected diagnostic techniques using contrast media, and blood infusion/withdrawal. For the power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec using media with a maximum viscosity of 11.8 cP.
Device Story
Vital-Port Vascular Access System Power Injectable Port is a subcutaneous, implanted, intravascular infusion port and catheter. Device provides repeated vascular access for infusion therapy, blood withdrawal, and power injection of contrast media. Operated by clinicians in a clinical setting. During power injection, port is accessed via qualified power-injectable infusion sets; contrast media is delivered via power syringe. Device enables diagnostic imaging studies requiring high-flow contrast delivery. Benefits include reliable, repeated venous access for patients requiring frequent therapy or diagnostic imaging. Performance parameters for power injection are limited to 5mL/sec and 325psi to ensure system integrity and prevent catheter failure.
Clinical Evidence
Bench testing only. Tests included dynamic failure flow, instantaneous burst, static burst, puncture life, cyclic testing, and life cycle power injection testing. Testing confirmed performance under "worst-case" scenarios for contrast media viscosity (11.8cP) and catheter length (>70cm).
Technological Characteristics
Subcutaneous, implanted, intravascular infusion port and catheter. Materials and construction identical to predicate. System includes silicone catheter. Designed for power injection up to 5mL/sec and 325psi. No software or electronic components.
Indications for Use
Indicated for patients requiring repeated vascular access for infusion therapy, power-injected diagnostic contrast media administration, and blood infusion/withdrawal.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Related Devices
K140169 — VITAL-PORT(R) CT VASCULAR ACCESS SYSTEM · Cook Incorporated · May 21, 2014
K150733 — Vital-Port Vascular Access System Power Injectable Port · Cook Incorporated · May 21, 2015
K101017 — ANGIODYNAMICS, INC., SMART PORT CT SERIES PORT ACCESS SYSTEMS · AngioDynamics, Inc. · Jun 17, 2010
K090512 — POWERPORT DUO M.R.I. IMPLANTED PORT WITH 9.5 FR. DUAL LUMEN CHRONOFLEX POLYURETHANE CATHETER · C.R. Bard, Inc. · Mar 27, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
1081425
SEP 3 0 2008
510(k) Summary
Submitted by:
Jim Ferguson Quality Systems Manager Cook Vascular, Incorporated 1186 Montgomery Lane Vandergrift, Pa 15690 724-845-8621, XT 2227 April 10, 2006
Device:
| Trade name: | Cook Vascular, Inc. Vital-Port Vascular Access System<br>Power Injectable Port |
|--------------------------|--------------------------------------------------------------------------------|
| Proposed Classification: | Subcutaneous, implanted, intravascular infusion port and<br>catheter, 880.5965 |
#### Predicate Devices:
The Cook Vascular, Inc. Vital-Port Vascular Access System Power Injectable Port Set is similar in terms of intended use, and exactly the same as materials of construction and technology characteristics to the predicate devices that have been found substantially equivalent.
#### Device Description:
The proposed device is exactly the same as the predicate devices,
## Substantial Equivalence:
This device will be manufactured to specified process controls and a Quality Assurance program. This device will undergo packaging exactly the same as the devices currently manufactured by Cook Vascular. This device will undergo sterilization exactly the same as the devices currently marketed and distributed. Being the same as the predicate devices in respect to indication for use, materials and physical construction, this device meets the requirements for section 510(k) substantial equivalence. It is to be noted that this device has not changed in design, materials and/or methods.
It has come to Cook Vascular's attention that ports have been and are being used for power injection studies. Cook Vascular currently makes no claims and provides no instruction for use for this procedure. The purpose for this 510(k) is to establish a safe and effective way to conduct a power injection study through ports which are deemed appropriate for this type of use.
When used for power injection, the port is accessed, using qualified power injectable infusion sets, in the standard fashion, per the standard instructions. The supplemental instructions will then be used for power injection. Power injection involves the luse of a power syringe to inject contrast media in order to complete a study.
<sup>·</sup> Cook Vascular, Inc. • Special 510(k) Premarket Submission • Vital-Port Vascular Access System Power Injectable Port •
{1}------------------------------------------------
### Test Data:
The Cook Vascular, Inc. Vital-Port Vascular Access System Power Injectable Port was subjected to the flowing tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:
- Dynamic Failure Flow Test 1.
- 2. Instantaneous Burst Test
- റ്റ് Static Burst Test
- 4. Puncture Life Test
- 5. Cyclic Test
- 6. Life Cycle Power Injection Test
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a Vital-Port Vascular Access System Power Injectable Port.
Based on empirical evidence, Cook Vascular, Inc. recommends for power injection of 9000 series models of Vital-Ports® with a maximum limit of 5mL/sec and 325psi. We reasonably assume that our testing provides "worst-case" scenarios for contrast media viscosity, catheter length and septum access. Our tested catheter length is greater than 70cm, while we assume the implanted length to be within a 20cm - 50cm range. Furthermore, our tested contrast media viscosity was at 11.8cP, while suggested usage for vendor supplied contrast media calls for pre-heating of the media prior to injection, this pre-heating could reduce viscosity by nearly half as some heated viscous contrast media could be as low as 4.9cP. Applying these expected clinical parameters would reduce overall system pressure and thereby increase the operational safety factor. Lastly, sufficient warnings against over pressurization of the silicone catheter systems are highly recommended as empirical evidence suggests that a catheter could fail prior to the injection machine depressurizing the system.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
SEP 3 0 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Jim Ferguson Quality System Manager Cook Vascular, Incorporated 1186 Montgomery Lane Vandergrift, Pennsylvania 15690
Re: K081425
Trade/Device Name: Cook Vascular, Inc. Vital-Port Vascular Access System Power Injectable Port Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Codc: LJT Dated: September 19, 2008 Received: September 22, 2008
Dear Mr. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Ferguson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the elcctronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) promarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K081425
#### Cook Vascular Inc. Vital-Port Vascular Access System Device Name:_ Power Injectable Port
Indications For Use:
The devices are for use in patient therapy requiring repeated vascular access for infusion therapy, power injected diagnostic techniques using contrast media, and blood infusion/withdrawal. For the power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec using media with a maximum viscosity of 11.8 cP.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\$\mathscr{A}\$ for clu lin
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
KO 1425 510(k) Number:
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.