K140169 · Cook Incorporated · LJT · May 21, 2014 · General Hospital
Device Facts
Record ID
K140169
Device Name
VITAL-PORT(R) CT VASCULAR ACCESS SYSTEM
Applicant
Cook Incorporated
Product Code
LJT · General Hospital
Decision Date
May 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vital-Port® CT Vascular Access System is indicated for patient therapy requiring repeated vascular access for injection or infusion therapy and/or blood sampling. When used with a power injectable infusion set, Vital-Port® CT Vascular Access Systems are indicated for power injection of contrast media. The maximum pressure limit setting for power injectors used with the Vital-Port® CT may not exceed 325 psi, and the flow rate may not exceed the indicated maximum flow rate.
Device Story
Vital-Port® CT Vascular Access System is an implanted, subcutaneous, intravascular infusion port and catheter. Device facilitates repeated vascular access for infusion therapy, blood sampling, and power injection of contrast media. System comprises a plastic port body (Standard or Petite sizes) and a catheter (silicone or polyurethane). Used in clinical settings by healthcare professionals. Power injection requires compatible infusion sets; maximum pressure limit 325 psi; maximum flow rate 5 ml/sec. Clinicians access the port percutaneously to deliver fluids or withdraw blood. Device provides long-term vascular access, reducing need for repeated venipuncture, thereby improving patient comfort and clinical efficiency.
Clinical Evidence
Bench testing only. Verification testing included dynamic failure flow, instantaneous burst, static burst, puncture life, and cyclic testing. Biocompatibility testing performed per ISO 10993-1:2009 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, subacute toxicity, genotoxicity, implantation, and hemocompatibility). Device confirmed MR Conditional.
Technological Characteristics
Implantable infusion port and catheter. Materials: plastic port body, silicone or polyurethane catheters. Dimensions: 6.0 Fr, 7.5 Fr, or 9.5 Fr catheters, 50 cm length. Power injection rated up to 325 psi and 5 ml/sec. MR Conditional. Sterilization method not specified.
Indications for Use
Indicated for patients requiring repeated vascular access for injection, infusion therapy, or blood sampling. When used with power injectable infusion sets, indicated for power injection of contrast media at pressures up to 325 psi.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
Vital-Port® Vascular Access System Power Injectable Port (K081425)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters. There is an asterisk after the word "COOK".
750 DANIELS WAY, P.O W.COOKMEDIC.
MAY 2 1 2014
# 510(k) Summary - K140169 Vital-Port® CT Vascular Access System 21 CFR §807.92 Date Prepared: April 22, 2014
## 1. Submitter Information:
Applicant:
Cook Incorporated 750 Daniels Way P.O. Box 489 Bloomington, IN 47404 Phone: 724-845-8621 Ext. 2225 Fax: 724-845-2848
Contact:
Contact Address:
Thomas J. Kardos Email: Thomas.Kardos@cookmedical.com Cook Incorporated 1186 Montgomery Lane Vandergrift PA. 15690 Phone: 724-845-8621 Ext. 2225 Fax: 724-845-2848
## 2. Device Information:
| Trade name: | Vital-Port® CT Vascular Access System |
|-----------------|-------------------------------------------------------------------|
| Common name: | Port & Catheter, Implanted, Subcutaneous, Intravascular |
| Classification: | Class II |
| Regulation: | 21 CFR 880.5965 |
| | Subcutaneous, implanted, intravascular infusion port and catheter |
| Product Code: | LJT - Port & Catheter, Implanted, Subcutaneous, Intravascular |
## 3. Predicate Device:
The Vital-Port® CT Vascular Access System is substantially equivalent to the following 510(k) cleared devices manufactured by Cook Vascular Incorporated;
Vital-Port® Vascular Access System Power Injectable Port, K081425.
# 4. Device Description:
The Vital-Port® CT Vascular Access System consists of a plastic port body and catheter lock. There are two Vital-Port® CT body sizes, Standard and Petite.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on the top line, with a small trademark symbol to the right of the second "O". Below that, in smaller white letters, is the word "MEDICAL". The background is black.
The catheters offered are: 7.5 or 9.5 Fr. silicone catheters or 6.0 or 7.5 Fr. polyurethane catheters. All catheters are 50 cm in length and are supplied either pre-attached or detached.
| Device Name | Catheter Size and<br>Material | Maximum Flow<br>Rate* | Injection Pressure<br>Limit Setting |
|-----------------------|-------------------------------|-----------------------|-------------------------------------|
| Standard | 9.5 Fr silicone | 5 ml/sec | 325 psi |
| Vital-Port® CT | 7.5 Fr silicone | 5 ml/sec | 325 psi |
| Vital-Port® CT | 7.5 Fr polyurethane | 5 ml/sec | 325 psi |
| Vital-Port® CT | 6.0 Fr polyurethane | 5 ml/sec | 325 psi |
| Petite | 7.5 Fr silicone | 5 ml/sec | 325 psi |
| Petite Vital-Port® CT | 7.5 Fr polyurethane | 5 ml/sec | 325 psi |
| Petite Vital-Port® CT | 6.0 Fr polyurethane | 5 ml/sec | 325 psi |
| Configuration | Description |
|----------------------------|------------------------------------------------------------------------------------|
| Single Lumen | Single lumen catheter for contrast injection |
| Dual Lumen | Dual lumen catheter for contrast injection and saline flush |
| Power Injectable Rated | Catheter is power injectable rated to allow for high flow rates |
| Non-Power Injectable Rated | Catheter is not power injectable rated and should not be used with power injectors |
*Flow rates were achieved using a room temperature infusate equivalent to Omnipaque 300" and a Medrad Injector. Omnipaque 3000 has a viscosity of 11.8 centipoise at room temperature (20 degrees C). A change in the temperature or viscosity of the media will result in a change in achievable flow rate. Omnipaque 300% is a registered trademark of GE Healthcare.
## 5. Intended Use:
The Vital-Port® CT Vascular Access System is indicated for patient therapy requiring repeated vascular access for injection or infusion therapy and/or blood sampling. When used with a power injectable infusion set, Vital-Port® CT Vascular Access Systems are indicated for power injection of contrast media. The maximum pressure limit setting for power injectors used with the Vital-Port® CT may not exceed 325 psi, and the flow rate may not exceed the indicated maximum flow rate.
## 6. Comparison to Predicates:
It has been demonstrated that the Vital-Port® CT Vascular Access System is comparable to the predicate device in terms of intended use, duration of use, principles of operation, technological characteristics, insertion method, anatomical location, and method of sterilization.
## 7. Technological Characteristics:
The Vital-Port® CT Vascular Access System is a single lumen, long-term implantable infusion port catheter intended for in patient therapy requiring repeated vascular access, infusion therapy, power injected diagnostic techniques using contrast media and blood infusion/withdrawal.
The Vital-Port® CT Vascular Access System port body is available in two configurations, Standard and Petite, with either a pre-attached silicone or polyurethane catheter.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters. There is a small asterisk to the right of the word COOK.
To demonstrate reliable design and performance of the Vital-Port® CT Vascular Access System, the following verification testing was performed:
- Dynamic Failure Flow Test Testing demonstrated that the vital port did not fail during . simulated use.
- Instantaneous Burst Test Testing demonstrated that the burst test of the port system met � the acceptance criterion.
- Static Burst Test Testing demonstrated that the static failure pressure was at or above . the acceptance criterion.
- Puncture Life Test Testing demonstrated that the puncture life test met the acceptance . criterion.
- Cyclic Test - Testing demonstrated that the cyclic test met the acceptance criterion
- . MR Safety - Testing demonstrated that the Vital-Port® CT Vascular Access System is MR Conditional.
- Biocompatibility Tests of cytotoxicity, sensitization, intracutaneous reactivity, acute ◆ systemic toxicity, subacute toxicity, genotoxicity, implantation, and hemocompatibility demonstrated the Vital-Port® CT Vascular Access System to be biocompatible. In conformance with the applicable sections of ISO 10993-1:2009, the predetermined acceptance criteria were met.
The results of these tests support a conclusion that the Vital-Port® CT Vascular Access System is as safe and as effective as the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure with three arms or extensions reaching upwards.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
### May 21, 2014
Cook. Incorporated Mr. Thomas J. Kardos Vice President, Quality Assurance and Regulatory Affairs 750 Daniels Way Bloomington, IN 47404
Re: K140169
Trade/Device Name: Vital-Port® CT Vascular Access System Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port And Catheter Regulatory Class: II Product Code: LJT Dated: April 22, 2014 Received: April 23, 2014
Dear Mr. Kardos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Brackett
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note
the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K140169
#### Device Name
Vital-Port® CT Vascular Access System
#### Indications for Use (Describe)
The Vital-Port® CT Vascular Access System is indicated for patient therapy requiring repeated vascular access for injection or infusion therapy and/or blood sampling.
When used with a power injectable infusion set, Vital-Port® CT Vascular Access Systems are indication of contrast media. The maximum pressure limit setting for power injectors used with the Vital-Port® CT may not exceed 325 psi and the indicated maximum flow rate.
Type of Use (Select one or both, as applicable)
[] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY | | |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | | |
| | Digitally signed by<br>Richard C.<br>Chapman<br>Date: 2014.05.21<br>13:55:29 -04'00' | |
| FORM FDA 3881 (1/14) | Page 1 of 2 | PSC Publishing Services (301) 443-4740 |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.