K170934 · Medical Components, Inc (Dba Medcomp_ · LJS · Dec 6, 2017 · General Hospital
Device Facts
Record ID
K170934
Device Name
C3 Wave System
Applicant
Medical Components, Inc (Dba Medcomp_
Product Code
LJS · General Hospital
Decision Date
Dec 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
C3 Wave System is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.
Device Story
C3 Wave System assists in PICC placement by monitoring cardiac electrical activity. System inputs include surface ECG signals from chest electrodes and intravascular ECG signals from a lead connected to the catheter stylet. A hub collects these signals and transmits them via Bluetooth to an iPad application. The application displays the ECG waveform, allowing the clinician to visualize P-wave amplitude changes as the catheter tip approaches the right atrium. Used in hospital/clinical settings by trained nurses or physicians. The real-time visual feedback allows clinicians to confirm catheter tip location without immediate reliance on X-ray or fluoroscopy, potentially reducing radiation exposure and procedure time.
Clinical Evidence
Clinical survey of 303 PICC placements across two facilities compared C3 Wave tip confirmation against portable chest X-ray. Clinicians used a 'Bundle Protocol' to assess tip location. Results demonstrated a 98% success rate when protocol parameters were met. Human factors testing with representative clinicians (nurses/physicians) showed no critical task failures.
Technological Characteristics
System components: iPad monitor, C3 hub, ECG leads, remote control. Connectivity: Bluetooth Low Energy. Energy: Mains power with battery backup. Sterilization: EO for accessory packs. Software: App version 2.0.3, hub firmware 39, remote firmware 16. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62133, ETSI EN 300 328, ETSI EN 301 489-17.
Indications for Use
Indicated for positioning of Peripherally Inserted Central Catheters (PICC) in adult patients as an alternative to chest X-ray or fluoroscopy for tip confirmation. Contraindicated/limiting conditions: atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythm, and COPD; these require additional confirmation methods.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
K112744 — SAPIENS TIP CONFIRMATION SYSTEM · C.R. Bard, Inc. · Oct 20, 2011
K152261 — PICC Tip Positioning Aid · Nostix, LLC · Oct 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2017
Medical Components, Inc (dba Medcomp) Courtney Nix Regulatory Affairs Manager, North American and Europe 1499 Delp Drive Harleysville, Pennsylvania 19438
Re: K170934
Trade/Device Name: C3 Wave System Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: Class II Product Code: LJS Dated: October 27, 2017 Received: November 6, 2017
Dear Courtney Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tina Kiang -
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170934
Device Name C3 Wave System
Indications for Use (Describe)
The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central Catheters (PICC). The C3 Wave provides real-ime cather tip location information by displaying changes in the patient's cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tin placement in adult patients.
Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave:
-Atrial fibrillation
-Atrial flutter
- Severe tachycardia
-Pacemaker-Driven Rhythm
-Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of additional method is necessary to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: K170934
B.
C.
#### A. Submitter Information:
| Submitter: | Medcomp®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201, x 2285<br>Fax: (215) 256-9191 |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Registration Number: | 2518902 |
| Contact: | Courtney Nix<br>Cnix@Medcompnet.com<br>Regulatory Affairs Manager: North America and EU |
| Date Prepared: | November 28, 2017 |
| Proposed or Subject Device Information: | |
| Trade Name: | C3 Wave System |
| Device: | Catheter, intravascular, Therapeutic Long-Term<br>Greater Than 30 Days |
| Product Code: | LJS |
| Regulation Description: | Percutaneous, implanted, long-term intravascular<br>catheter |
| C.F.R. Section: | 880.5970 |
| Class: | ll |
| Regulation Medical<br>Specialty and Review<br>Panel: | General Hospital |
| Predicate Device Information: | |
| 510(k) Number: | K143238 |
| 510(k) Holder: | Medcomp® |
| Trade Name: | C3 Wave System |
| Device: | Catheter, intravascular, Therapeutic Long-Term<br>Greater Than 30 Days |
| Product Code: | LJS |
{4}------------------------------------------------
| Regulation Description: | Percutaneous, implanted, long-term intravascular<br>catheter |
|------------------------------------------------------|--------------------------------------------------------------|
| C.F.R. Section: | 880.5970 |
| Class: | II |
| Regulation Medical<br>Specialty and Review<br>Panel: | General Hospital |
#### E. Purpose for Submission:
The primary purpose of this submission is to revise the indications for use for the C3 Wave System. The indications currently included, cleared by C3 Wave System K143238, "Confirmation of tip placement should be verified according to clinical judgement and established hospital protocol (e.q., Chest X-Ray, Fluoroscopy)." The proposed indications will revise this statement to read "for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients."
The proposed C3 Wave System is equivalent to the cleared C3 Wave System, K143238, in device design, materials and materials. The changes to the cleared device include the hardware to the hub portion of the system, software, and the indications for use.
#### E. Device Description:
C3 Wave is designed to provide a continuous display of electrocardiograph [ECG] waveform to be used as a quide in placement of peripherally-inserted central catheters [PICC] in the lower third of the Superior Vena Cava [SVC] of a patient. The principle for operation of this system uses three ECG leads placed on the patient's chest and generates a third ECG lead by switching from RA to PICC stylet. The ECG waveform is wirelessly transmitted to a tablet which allows the operator to view and record changes to the ECG waveform as the tip of the catheter approaches the heart. As the PICC catheter approaches the atrium of the heart, the P wave in the ECG waveform shows substantial changes. This system is designed to aid the visualization of changes in P wave amplitude.
The C3 Wave system must only be operated by a skilled nurse, physician, or trained medical professional who has been qualified in placement of PICC's and trained in the proper use of this device.
#### G. Indications for Use:
The C3 Wave System is indicated for use in the positioning of Peripherally Inserted Central Catheters (PICC). The C3 Wave provides real-time catheter tip location information by displaying changes in the patient's cardiac electrical activity. The C3 Wave is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients.
{5}------------------------------------------------
Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P-Wave:
- -Atrial fibrillation
- -Atrial flutter
- Severe tachycardia
- -Pacemaker-Driven Rhythm
- -Chronic obstructive pulmonary disease (COPD)
Such patients are easily identified prior to PICC insertion. Use of additional method is necessary to confirm catheter tip location.
#### H. Intended Use:
C3 Wave System is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.
#### l. Comparison to Predicate Device:
The C3 Wave system is substantially equivalent to the predicate device.
| Device | Proposed Device: | Predicate Device: | Substantially<br>Equivalent<br>Comparison | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The C3 Wave System | The C3 Wave System | | |
| | K170934 | (K143238) | | |
| Indications<br>for Use | The C3 Wave System is<br>indicated for use in the<br>positioning of Peripherally<br>Inserted Central Catheters<br>(PICC). The C3 Wave<br>provides real-time catheter<br>tip location information by<br>displaying changes in the<br>patient's cardiac electrical<br>activity. The C3 Wave is<br>indicated for use as an<br>alternative method to chest<br>X-ray or fluoroscopy<br>confirmation of PICC tip<br>placement in adult patients.<br>Note: Limiting, but not<br>contraindicated, situations<br>for this technique are<br>patients where cardiac<br>rhythms may change<br>presentation of the P-Wave:<br>-Atrial fibrillation<br>-Atrial flutter- Severe<br>tachycardia<br>-Pacemaker-Driven Rhythm | The C3 Wave System is<br>indicated for use as a<br>supplemental aid in<br>positioning for Peripherally<br>Inserted Central Catheters<br>(PICC) in adult patients. It<br>provides real-time catheter<br>tip location information by<br>using the patient's cardiac<br>electrical activity.<br>Confirmation of tip<br>placement should be<br>verified according to clinical<br>judgement and established<br>hospital protocol (e.g.,<br>Chest XRay, Fluoroscopy).<br>Note: Limiting, but not<br>contraindicated, situations<br>for this technique are<br>patients where cardiac<br>rhythms may change<br>presentation of the P wave:<br>atrial fibrillation, atrial<br>flutter, severe<br>tachycardia, pacemaker<br>driven rhythm, and chronic | The subject<br>device is<br>indicated as an<br>alternative<br>method to chest<br>x-ray or<br>fluoroscopy<br>confirmation of<br>PICC tip<br>placement in<br>adult patients,<br>subject to the<br>noted limitations.<br>The differences<br>between the<br>subject and<br>predicate device<br>indications for use<br>do not raise<br>different<br>questions of<br>safety and<br>effectiveness.<br>Performance data<br>was provided to<br>verify the design | |
| | | -Chronic obstructive<br>pulmonary disease (COPD)<br>Such patients are easily<br>identified prior to PICC<br>insertion. Use of additional<br>method is necessary to<br>confirm catheter tip<br>location. | obstructive pulmonary<br>disease (COPD).<br>Such patients are easily<br>identified prior to PICC<br>insertion. Use of additional<br>confirmation method is<br>necessary to confirm<br>catheter tip location. | and labeling<br>adequately<br>support the<br>revised<br>indications for<br>use. |
| User/Setting | | Prescription Use –<br>Hospital/Clinical Setting | Prescription Use –<br>Hospital/Clinical Setting | Equivalent |
| Patient/User<br>Interface | | iPad | iPad | Equivalent |
| System | | iPad, C3 Wave Hub, ECG<br>Leads, C3 Wave Remote | iPad, C3 Wave Hub, ECG<br>Leads, C3 Wave Remote | Equivalent |
| Sterilization<br>Method | | Sterile Accessory Packs – EO | Sterile Accessory Packs – EO | Equivalent |
| Energy Source | | Mains power (battery<br>backed) | Mains power (battery<br>backed) | Equivalent |
| EMC & Electrical<br>Safety | | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | Equivalent |
| ATTRIBUTE | | C3 Wave (Proposed Device) | | C3 Wave System (K143238) |
| Operating<br>Principle/Technology: | | The C3 Wave system consists<br>of an iPad® monitor, a C3 hub,<br>and remote control.<br><br>First, the hub, which can be<br>placed on the patient's chest,<br>collects the ECG signal from<br>the surface ECG (via ECG<br>chest pad electrode leads)<br>and the intravascular ECG<br>(lead connected to the<br>catheter stylet).<br><br>Next, the waveform is sent<br>wirelessly by Bluetooth™<br>technology from the hub to the<br>mobile application.<br><br>Then, the C3 mobile<br>application, installed on an<br>iPad®, displays the ECG<br>waveform. Hence, the<br>medical professional can<br>monitor the location of the<br>PICC catheter through the<br>ECG waveform and,<br>specifically, monitor the P-<br>wave shape and amplitude<br>which indicates tip position<br>relative to the sinoatrial node.<br>Further, sterility can be | | The C3 Wave system consists<br>of an iPad® monitor, a C3 hub,<br>and remote control.<br><br>First, the hub, which can be<br>placed on the patient's chest,<br>collects the ECG signal from<br>the surface ECG (via ECG<br>chest pad electrode leads) and<br>the intravascular ECG (lead<br>connected to the catheter<br>stylet).<br><br>Next, the waveform is sent<br>wirelessly by Bluetooth™<br>technology from the hub to the<br>mobile application.<br><br>Then, the C3 mobile<br>application, installed on an<br>iPad®, displays the ECG<br>waveform. Hence, the medical<br>professional can monitor the<br>location of the PICC catheter<br>through the ECG waveform<br>and, specifically, monitor the P-<br>wave shape and amplitude<br>which indicates tip position<br>relative to the sinoatrial node.<br>Further, sterility can be |
| | | maintained during the entire | | maintained during the entire |
# 510(K) Summary: Comparison Table
{6}------------------------------------------------
{7}------------------------------------------------
| | process as the C3 Wave<br>system is controlled using a<br>remote control. | system is controlled using a<br>remote control. |
|--|---------------------------------------------------------------------------|-------------------------------------------------|
| | | |
The proposed C3 Wave System is equivalent to the cleared C3 Wave System, K143238, in device design, materials and materials. The changes to the proposed device from the predicate are in the hardware to the hub portion of the system, software and the expanded indications for use.
Through performance bench testing, the subject device has demonstrated that it is substantially equivalent to the predicate.
## J. Non-Clinical Performance Testing
| Standards | Performance<br>Testing |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2006+A12:2014 Medical electrical equipment -<br>Part 1: General requirements for basic safety and essential<br>performance | IEC 60601-1 Test<br>Report (E468297-D1-<br>IT-1) |
| IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests | IEC 60601-1-2 Test<br>Report (R-2642P) |
| IEC 62304:2006 Medical device software — Software life cycle<br>processes | IEC 62304 Test Report |
| | Equipment Interaction |
| | C3 Wave Remote –<br>Updated Firmware |
| | C3 Wave Application<br>Software - Equipment<br>Interaction “Software<br>Driven Warnings” |
| | C3 Wave – Equipment<br>Interaction (C3 Wave<br>Application iPad<br>Software) |
| | Lextech - Medcomp<br>functionality checklist<br>VPI - Hub firmware test |
| | VPI - Remote firmware<br>test |
| ISTA-2A Partial simulation performance test procedure | Non-Sterile C3 Wave<br>Accessory Pack- ECG<br>Electrodes Test<br>Summary |
| ISTA-1A Non-simulation integrity performance test procedure | C3 Wave Navigation<br>System w/ iPad – ISTA-<br>2A Shipping Test<br>Summary |
| IEC 62133:2013 Secondary cells and batteries containing<br>alkaline or other non-acid electrolytes - Safety requirements for<br>portable sealed secondary cells, and for batteries made from<br>them, for use in portable applications | C3 Wave Hub - ISTA-<br>1A Shipping Test<br>IEC 62133 CB Test<br>Certificate (4307556.50) |
| 47 CFR Part 15, Subpart C Intentional Radiators | Bluetooth Low Energy<br>Certification Test<br>Report (10463456A) |
| | Bluetooth Low Energy<br>Certification Test<br>Report (10463456B) |
| | EN 300 328 V1.8.1<br>Certification Test<br>Report (14L18625-<br>E1)(Remote) |
| | EN 300 328 V1.8.1<br>Certification Test<br>Report (14L18625-<br>E2)(Hub) |
| RSS-210 License-Exempt Radio Apparatus: Category I<br>Equipment | Bluetooth Low Energy<br>Certification Test<br>Report (10463456A) |
| | Bluetooth Low Energy<br>Certification Test<br>Report (10463456B) |
| | EN 300 328 V1.8.1<br>Certification Test<br>Report (14L18625-E1) |
| | EN 300 328 V1.8.1<br>Certification Test<br>Report (14L18625-E2) |
| ETSI EN 300 328 V1.8.1 Electromagnetic compatibility and<br>Radio spectrum Matters (ERM); Wideband transmission<br>systems; Data transmission equipment operating in the 2,4 GHz<br>ISM band and using wide band modulation techniques;<br>Harmonized EN covering the essential requirements of article<br>3.2 of the R&TTE Directive | EN 300 328 V1.8.1<br>Certification Test<br>Report (14L18625-E1) |
| | EN 300 328 V1.8.1<br>Certification Test<br>Report (14L18625-E2)) |
| ETSI EN 301 489-17 V2.2.1 Electromagnetic compatibility and<br>Radio spectrum Matters (ERM); ElectroMagnetic Compatibility<br>(EMC) standard for radio equipment; Part 17: Specific<br>conditions for Broadband Data Transmission Systems | Bluetooth Low Energy<br>Certification Test<br>Report (10463456A) |
| | Bluetooth Low Energy<br>Certification Test<br>Report (10463456B) |
| Bluetooth 4.1 | Low Energy RF PHY<br>Test Specification (TS):<br>Test Suite Structure<br>(TSS) and Test<br>Purposes (TP) RF-<br>PHY.TS.4.1.0 (UL-<br>BQT-<br>RP10522414JD01A<br>V2.0) (Hub) |
| | Low Energy RF PHY<br>Test Specification (TS):<br>Test Suite Structure<br>(TSS) and Test<br>Purposes (TP) RF-<br>PHY.TS.4.1.0 UL-BQT-<br>RP10522414JD02A<br>V2.0 (Remote) |
{8}------------------------------------------------
{9}------------------------------------------------
In addition to demonstrating performance in accordance with these standards, the submission contained information recommended in FDA quidance documents for software and cybersecurity:
- Content of Premarket Submissions for Software Contained in Medical Devices. ●
- . Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
The current released version of the C3 Wave app is 2.0.3. The current released remote firmware is version 16. The current released hub firmware is version 39.
## Clinical and Human Factors Testing
Clinical and human factors studies were collected/performed to evaluate the C3 Wave System.
### Clinical Study
Clinical Surveys were completed to demonstrate that various PICC line nurses of different years of experience can successfully insert the PICC Line based on their ability to read the ECG display on the iPad
- . Objective: To demonstrate clinical acceptance between the MedComp C3 Wave System ECG Tip Confirmation System positioning results and traditional portable chest xray in determining proper distal tip location during a PICC procedure.
- Methods: A total of 303 PICCs were to be placed at two healthcare facilities using the C3 Wave System ECG Tip Confirmation System in accordance with the instructions for use. A subsequent portable chest x-ray was to be obtained, with the patient position similar of PICC placement, to ascertain acceptable tip location of PICC. The clinician was to complete an electronic survey document to capture specific data points within each procedure including insertion related
{10}------------------------------------------------
assessments and final catheter functionality questions referred to as the Bundle Protocol.
Outcome: The C3 Wave System has demonstrated a 98% success rate when the Bundle Protocol Parameters were met.
### Human Factors
Human Factors Testing has been conducted to evaluate the application of the C3 Wave System when used to confirm PICC tip position as an alternative method to chest X-ray or fluoroscopy confirmation. Participants in the study were representative of the population who place PICC lines in the clinical environment where the product is intended to be used, e.g., hospital wards, acute care centers, extended care centers, alternate care centers, and outpatient clinics. Clinicians with experience in the placement of PICC lines were recruited for the study. No critical task failures were observed in the study.
#### K. Summary/Conclusion of Substantial Equivalence:
In conclusion, the proposed device, C3 Wave System, is equivalent to the predicate device, C3 Wave System (K143238).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.