PICC Tip Positioning Aid

K152261 · Nostix, LLC · LJS · Oct 21, 2015 · General Hospital

Device Facts

Record IDK152261
Device NamePICC Tip Positioning Aid
ApplicantNostix, LLC
Product CodeLJS · General Hospital
Decision DateOct 21, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5970
Device ClassClass 2

Intended Use

The PICC Tip Positioning Aid is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.

Device Story

Device provides real-time PICC tip location during insertion. Inputs: patient cardiac electrical activity via ECG cables/electrodes. Operation: monitors P-wave changes as catheter approaches right atrium via superior vena cava. Output: real-time tip location information displayed on monitor. Used in clinical settings by clinicians. Replaces traditional chest x-ray or fluoroscopy for tip confirmation. Benefits: reduces radiation exposure and procedural time. System includes monitor, software, battery, ECG patient cable, remote control cable, probe cover, and ECG clip cable.

Clinical Evidence

Bench testing only. Device relies on simulated use/human factors testing previously conducted for the predicate Celerity System to validate the alternative method to chest x-ray/fluoroscopy. Compliance with IEC 60601-1 and IEC 60601-1-2 standards confirmed.

Technological Characteristics

Standalone monitor with software, battery, and power cord. Includes ECG patient cable, remote control cable, probe cover, and ECG clip cable. Complies with IEC 60601-1 (General safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for positioning Peripherally Inserted Central Catheters (PICC) in adult patients. Provides real-time tip location via cardiac electrical activity as an alternative to chest x-ray or fluoroscopy. Contraindicated/limited for patients with atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythm, or COPD; these patients require additional confirmation methods.

Regulatory Classification

Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and clean, and it is easily recognizable. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 21, 2015 Nostix LLC c/o Mr. Jim Lewis Salus Ventures LLC 5335 Holmes Place Boulder, Colorado, 80303 Re: K152261 Trade/Device Name: PICC Tip Positioning Aid Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: II Product Code: LJS Dated: August 7, 2015 Received: August 11, 2015 Dear Mr. Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152261 Device Name PICC Tip Positioning Aid Indications for Use (Describe) The PICC Tip Positioning Aid is indicated for the postioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location by using the patient's cardiac electrical activity. The PICC Tip Positioning Aid is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients. Note: In general, devices that utilize ECG technique to observe P-Wave arc limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave; including - Atrial fibrillation - Atrial flutter - Severe tachycardia - Pacemaker-driven rhythm - Chronic obstructive pulmonary disease (COPD) Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☑</span> | |----------------------------------------------|--------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <span>☐</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Nostix" in blue font with a gray shadow. The word is written in a serif font. The shadow is slightly offset from the word, giving it a 3D effect. # 510(k) Summary Prepared: 25 September 2015 ## Submitter | Company | Nostix LLC<br>5541 Central Av, Suite 170<br>Boulder, CO 80301 | |---------|-----------------------------------------------------------------| | Tel | 303 245 8895 | | Fax | 303 245 8909 | | Contact | Pete Nelson<br>Director of Engineering<br>petenelson@nostix.com | | Device | | | Trade Name | PICC Tip Positioning Aid | |--------------|-----------------------------------------------------------| | Common Name | PICC placement accessory | | Class Name | Percutaneous, implanted, long-term intravascular catheter | | Product Code | LJS | | Regulation | 21 CFR 880.5970 | | Class | 2 | # Predicate | Trade Name | Celerity System | |--------------|-----------------------------------------------------------| | Clearance | K142889, 27 January 2015 | | Common Name | PICC placement accessory | | Class Name | Percutaneous, implanted, long-term intravascular catheter | | Product Code | LJS | | Regulation | 21 CFR 880.5970 | | Class | 2 | {4}------------------------------------------------ ### Device Description The PICC Tip Positioning Aid includes a standalone Monitor containing software, battery and power cord accompanied by an ECG Patient Cable, a Remote Control Cable, probe cover and ECG Clip Cable. Other procedural accessories; including ECG Snap Leads, ECG Surface Electrodes, Cable Cover, Gloves and Prep Pads; may be provided as a convenience for the clinician but are not in the scope of this submission. #### Intended Use The PICC Tip Positioning Aid is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava. #### Indications for Use The PICC Tip Positioning Aid is indicated for the positioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. The PICC Tip Positioning Aid is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients. Note: In general, devices that utilize ECG technique to observe P-Wave arc limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave; including - Atrial fibrillation 트 - Atrial flutter - I Severe tachycardia - Pacemaker-driven rhythm - I Chronic obstructive pulmonary disease (COPD) Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location. #### Technological Characteristics Technological Characteristics of the subject PICC Tip Positioning Aid are identical to the predicate device. The name and address changes on the labels between the predicate and proposed devices do not raise new technological questions. #### Performance Data As the only differences between the device and its predicate are names, logos, and addresses in the labeling, the following recognized standards from the IEC 60601 (3ra Edition) series continue to be satisfied. - IEC 60601-1-1 Medical electrical equipment—Part 1-1: General requirements for safety-Collateral standard: Safety requirements for medical electrical systems {5}------------------------------------------------ - IEC 60601-1-2 Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility - Requirements and tests Human Factors Evaluation - Simulated Use Testing alternate to chest x-ray and fluoroscopy: Simulated Use / Human Factors Testing has been conducted to evaluate the application of the PICC Tip Placement Aid as embodied in the predicate Celerity System when used to confirm PICC tip position as an alternative method to chest X-ray or fluoroscopy confirmation. The use related events noted in the studies have been adequately reviewed and addressed in order to ensure the appropriate use of the device as an alternate to x-ray techniques for confirmation of tip location of PICC. Based on the content of the proposed PICC Tip Positioning Aid's Risk Analysis / Use and Design FMEAs, and the content of the Instructions for Use, the PICC Tip Positioning Aid has demonstrated its suitability for its intended purpose. ## Substantial Equivalence Conclusion As this device design and manufacturing are the same as the predicate except for name changes to the device and the manufacturer, the device is clearly equivalent to its predicate. The proposed device is clearly substantially equivalent to the predicate device based on identical: - 트 Intended Use - 트 Indications for Use - ' Design - 트 Production - I Operating principles, characteristics, and user interface - 트 Technology and specifications - I Labeling
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%